Immunoassay Methods and their Applications in Pharmaceutical Analysis: Basic Methodology and Recent Advances

Department of Pharmaceutical Analytical Chemistry, Faculty of Pharmacy, Assiut University, Egypt, Ibrahim A. Darwish

International Journal of Biomedical Science · 2006 · Review

doi:10.59566/ijbs.2006.2217

Abstract

Immunoassays are bioanalytical methods in which the quantitation of the analyte depends on the reaction of an antigen (analyte) and an antibody. Immunoassays have been widely used in many important areas of pharmaceutical analysis such as diagnosis of diseases, therapeutic drug monitoring, clinical pharmacokinetic and bioequivalence studies in drug discovery and pharmaceutical industries. The importance and widespread of immunoassay methods in pharmaceutical analysis are attributed to their inherent specificity, high-throughput, and high sensitivity for the analysis of wide range of analytes in biological samples. Recently, marked improvements were achieved in the field of immunoassay development for the purposes of pharmaceutical analysis. These improvements involved the preparation of the unique immunoanalytical reagents, analysis of new categories of compounds, methodology, and instrumentation. The basic methodologies and recent advances in immunoassay methods applied in different fields of pharmaceutical analysis have been reviewed.

Shown for reading. Its publisher allows quoting it, not republishing it.

Full text

From PubMed Central, U.S. National Library of Medicine, as current on Oct 24, 2026. Paragraph numbers are ours, so an answer can point at one.

INTRODUCTION

Immunoassays are bioanalytical methods in which the quantitation of the analyte depends on the reaction of an antigen (analyte) and an antibody. Principally, these methods are based on a competitive binding reaction between a fixed amount of labelled form of an analyte and a variable amount of unlabelled sample analyte for a limited amount of binding sites on a highly specific anti-analyte antibody. When these immunoanalytical reagents are mixed and incubated, the analyte is bound to the antibody forming an immune complex. This complex is separated from the unbound reagent fraction by physical or chemical separation technique. Analysis is achieved by measuring the label activity (e.g. radiation, fluorescence, or enzyme) in either of the bound or free fraction. A standard curve, which represents the measured signal as a function of the concentration of the unlabelled analyte in the sample is constructed. Unknown analyte concentration is determined from this calibration curve.

Immunoassay methods have been widely used in many important areas of pharmaceutical analysis such as diagnosis of diseases, therapeutic drug monitoring, clinical pharmacokinetic and bioequivalence studies in drug discovery and pharmaceutical industries. The analysis in these areas usually involves measurement of very low concentrations of low molecular weight drugs, macromolecular biomolecules of pharmaceutical interest, metabolites, and/or biomarkers which indicate disease diagnosis or prognosis. The importance and widespread of immunoassay methods in pharmaceutical analysis are attributed to their inherent specificity, high-throughput, and high sensitivity for the analysis of wide range of analytes in biological samples. The detection system in immunoassays depends on readily detectable labels (e.g. radioisotopes or enzymes) coupled to one of the immunoanalytical reagents (i.e. analyte or antibody). The use of these labels in immunoassays results in assay methods with extremely high sensitivity and low limits of detection. In circumstances whereas the specific measurements of large molecules at the femtomole to attomole level in complex biological matrices is required, no doubt that immunoassays are the methods of choice because of their high specificity and sensitivity.

In the early stages of drug discovery and development, particularly during the clinical pharmacokinetic studies for the new drug candidate, screening of large number of samples is required. This can be achieved only by using an analytical method of high throughput. The analysis of complex biological matrices (e.g. blood or urine) by immunoassay methods, being based on a specific binding reaction, can be achieved without pretreatment for the sample. Although the developing of a new immunoassay method for an analyte may take months (due to the time needed for generating the desired antibody), however, once suitable immuoanalytical reagents become available, the immunoassay method can be established in a time frame that is competitive with chromatographic methods. Furthermore, novel techniques were developed to enable the rapid production of specific antibodies. These techniques resulted in dramatic shortening of the time required for developing of immunoassay methods. These potential advantages of immunoassay methods, in addition to the relatively low cost of the instruments, tools, or the reagents made immunoassays the methods of choice in many areas of pharmaceutical analysis.

REAGENTSTS REQUIRED FOR IMMUNOASSAY DEVELOPMENT

These reagents are the antibodies, signal-generating labels, and separation matrices. Antibodies are the key reagents on which the success of any immunoassay depends. The antibodies can be either polyclonal or monoclonal. However, for immunoassay development for pharmaceutical analysis purposes, monoclonal antibodies are more advantageous than polyclonal ones. This is attributed to their higher degree of affinity and specificity towards the analyte. Even that, many successful immunoassays were developed using polyclonal antibodies because it was possible to generate the antibodies with high affinity to the analyte.

The signal generating labels in immunoassays include radioactive atoms (mostly 125I, 3H, and 14C). The use of radioactive labels offers extremely sensitive and quite precise assays, however, they have some drawbacks (e.g. health hazards, special attention for handling of the reagents, training of staff, short half-life time of the isotope, and expensive instrumentation for the counting of radioactivity. Therefore, alternative non-radioactive labels such as enzymes, fluorescent probes, chemiluminescent substances, metals and metal chelates, and liposomes were introduced. On the basis of number of publications, enzymes are the most common labels employed in immunoassay methods for pharmaceutical compounds. A potential advantage in the use of enzyme labels for immunoassay is the possibility of the amplification of the signal, and subsequently the potential increasing in the sensitivity of the method. This is beneficial when the original signal is not sufficient to get the desirable sensitivity for the analysis.

The matrices used for separation of the immune complexes that formed as a result of immunoanalytical reactions include charcoal, polyethylene glycol, second antibody, microbeads or the most useful 96-well microwell plates; each well of the plate serves as a separate reaction tube. One component of the reaction (analyte or antibody) is coated onto the surface of the bottom of the plate wells, and the immune complex is formed on the surface of the wells. The use of these plates facilitates the washing steps, and reagents pipetting, and thus leads to semi-automation of the method.

BASIC METHODOLOGY INVOLVED IN PHARMACEUTICAL ANALYSIS

Immunoassay methods that have been applied in pharmaceutical analysis, based on whether the separation step is or is not required, can be classified into heterogeneous or homogeneous assay, respectively. These methods can be performed in either competitive or non-competitive designs. The choice from these designs is based on nature of the analyte, labeling chemistry available and the analytical parameter required from the assay (e.g. sensitivity, dynamic range, and precision). Competitive design of immunoassays can be carried out in an antigen-capture or antibody-capture format, depending on whether the solid phase is coated with antibody or antigen (analyte), respectively. The features of these formats are shown in Figure 1. In the antigen-capture format (Figure 1A), the competition reaction occurs between the analyte (in sample) and a labelled analyte for the binding to a limited amount of anti-analyte antibody coated onto a solid support. After equilibration and separation, the label activity on the solid phase is measured, and the measured signal is inversely correlated to the concentrations of analyte in the sample. In antibody-capture format (Figure 1B), the analyte (or its protein conjugate) is coated onto a solid support. The competition occurs between the analyte (in sample) and the immobilized analyte for the binding to a limited amount of labelled anti-analyte antibody. After equilibration and separation, the activity of the label bound to the solid support is measured, and the signal is inversely correlated to the concentration of the analyte. The non-competitive design (usually called “two-site” or “sandwich” assay) is used for large analytes possessing more than one recognition epitopes on the molecule. It requires two antibodies that bind to non-overlapping epitopes on the analyte molecules. One of the two antibodies is bound to the solid phase, and the second one is labelled and used for detection. Figure 2 illustrates the features of this assay. The sample analyte is allowed to bind to an immobilized antibody. After washing, the solid support (contains the formed analyte-antibody complex) is incubated with an excess of the second labelled antibody, which binds to the remaining epitope on the analyte molecule. After washing, the activity of the label bound to the solid support is measured.

Radioimmunoassay

The history and development of radioimmunoassay (RIA) were reviewed by Najjar and Weintraub. RIA methods have been used successfully for the determination of limitless number of pharmaceutically important compounds in biological fluids. Most of these methods are now automated with separation assisted by the use of antibody bound to a solid phase matrices. Table 1 shows list of the compounds that have been analyzed by RIA in biological fluids. The most important advantage of RIA in the measurement of compounds in biological fluids is the quite precision and extreme sensitivity, which can not be achieved by other analytical techniques (except GC-MS).

Enzyme Immunoassay

Enzyme immunoassay (EIA) is analogous to RIA except that the label is an enzyme rather than a radioisotope. The basic approach for use of an enzyme as an immunoassay label is appreciated by coupling an enzyme molecule into one of the immunoanalytical reagents (analyte or antibody), by appropriate chemical technique, and then carrying out the immunoanalytical reaction in the normal way. Following the separation of bound and free fractions, the enzyme activity is monitored in either of the two fractions. This is achieved by adding substrate, and subsequent monitoring the turnover of the substrate to product. The product must possess measurable physical or chemical differences from the substrate. For example, colourless chromogenic substrates which are converted into coloured products by the action of the enzyme label. The coloured products can be simply measured by a spectrophotometer. The measured signal is then correlated to the analyte concentration. Although enzyme/chromogenic substrate system is frequently used, however, many other systems can be used. A list for compounds that have been analyzed by competitive assays; antigen-capture and antibody-capture EIA are given in Tables 2 and 3, respectively. As well, sensitive non-competitive EIA methods were developed for various compounds of pharmaceutical interest; these methods are shown in Table 4. Homogenous EIA, exemplified by the enzyme multiplied immunoassay (EMIA) is the major breakthrough in immunoassay technology. Homogenous EIA was successfully applied to therapeutic drug monitoring of some pharmaceutical compounds, e.g. aminoglycoside antibiotics. Reagent kits are available for the analysis of amikacin, gentamicin, netilmicin and tobramycin. These kits are used with the Viva automated analyser or with automated analysers from different manufacturers. Other homogenous EMIA were developed for measuring the concentration of thyroxine, triiodothyronine, theophylline, and cyclosporine in biological fluids (Table 4).

Fluoroimmunoassay

Fluoroimmunoassay (FIA) is analogous to RIA except that the label is a fluorophore rather than a radioisotope. As in other immunoassays, FIA can be categorized into heterogeneous and homogeneous assays, depending on whether the separation step is or is not needed, respectively. Either of heterogeneous or homogeneous assays can be performed in a competitive or non-competitive format. Heterogeneous competitive FIA methods are currently available for analysis of various compounds of pharmaceutical importance in biological fluids: aminoglycoside antibiotics, morphine-3-glucuronide, the major urinary metabolite of heroin and morphine, benzoylecogonine in urine, and endocrine disruptTing chemicals (estrone, estradiol, and ethinylestradiol) in a complex synthetic aqueous matrix. These FIA offered sensitivities ranged from 0.01-2 ng/ml. A heterogeneous competitive FIA approach was developed for the measurement of the fluorescent signal directly on the solid supports, wells of microwell plates. This approach was used for the determination of the total thyroxine in human serum. In a further work, the fluoTrescence signal measured directly onto the solid phase was considerably stabilized by pretreatment of the wells with a glycerin solution. This treatment caused in an increase in the sensitivity of the assay. In homogeneous FIA, the antibody-bound analyte is not needed to be separated from the free analyte before the fluorescence measurement. Almost all the homogeneous FIA applied in pharmaceutical analysis are performed as competitive. In these assays, the antibody binding causes some changes in the fluorescence properties (e.g. polarization) of the labelled analyte. The analyte concentration in a sample can be monitored directly from the reaction mixture. The most common type of these assays is the fluorescence polarization fluoroimmunoassay (FPFIA). This technique is based on the following: when a fluorescent analyte conjugate is excited with polarized light, the polarization of the resulting emission depends inversely on the decay constant of the probe (4-5 ns for fluorescein isothiocyanate) and on the rotational motion of the conjugate. With small molecules (e.g. drugs), random rotation decreases the polarization signal; when bound to specific antibodies, their rotation slows, and the polarization signal increases. The increase in the polarization signal is related to the concentration of the analyte. Originally, this technique was developed for single tube analytical instruments, but the technology was rapidly converted Immunoassays into high-throughput screening assays when automated analyzers became commercially available. The FPFIA methods utilizing automated analyzer are simple, precise, and easy to perform. Therefore, these methods are widely applied in drug discovery studies as well as in therapeutic monitoring of a wide range of analytes, shown in Table 5. Recently, an improved one-step FPFIA was developed for analysis of progesterone hormone using an immunocomplex single reagent (a pre-equilibrated mixture of antibody and tracer). This assay is a speedy and does not need incubation period before the measurement of the fluorescence polarization signal (the total assay time is about 7 min for 10 samples). The limit of detection of this assay was 2.7 ng/ml with 50 ml samples.

Time-resolved fluoroimmunoassay (TRFIA) is a separate group of FIA because its principles can be adapted to both heterogenous and homogenous assay format. In TRFIA, chelates of lanthanides and palladium ions are used as labels. Highly sensitive TRFIA methods were developed for analysis of many pharmaceutical compounds and hormones in biological fluids (Table 6). The advantage in these assays was the possibility to resolve the background fluorescence (due to biological fluids) from the assay fluorescence. Therefore, the non-specific endogenous fluorescence was eliminated.

Chemiluminescence Immunoassay

Chemiluminescence immunoassay (CLIA) involves a chemiluminescent substance as a label. The growing success of this technique in pharmaceutical analysis due to its high performance, low detection limits, and good precision. The principles, methodology, instrumentation, and applications of chemiluminescence immunoassay are the subject of some books and reviews. The CLIA was used for analysis of many compounds of pharmaceutical importance; these compounds are shown in Table 7.

Liposome Immunoassay

Liposome immunoassay (LIA) is the assays involving a liposome-encapsulating marker. In LIA, liposomes are prepared and then coupled to either analyte or antibody by a suitable procedure, and then carrying out the assay in normal way. Detection in LIA relies on the lysis of the liposome and releasing the encapsulated marker, which is then measured and related to the analyte concentration. LIA methods are applied as homogenous assays often rely on the lytic activity of complement on antibody-associated liposomes, or the ability of free mellitin (a bee venom protein) conjugate with an analyte to lyse the liposomes, and release markers. Heterogeneous LIA are performed competitively or non-competitively in a similar way to stanTdard EIA, except replacement of the enzyme label with a liposome. Washing steps are used to separate bound liposome- labelled reagent from unbound reagents. All bound liposomes are lysed by addition of a detergent releasing the signal-generating marker (Figure 3). Different LIA methods were developed for analysis of various pharmaceutical compounds (Table 8).

IMMUNOASSAY FOR SIMULTANEOUS ANALYSIS OF MULTI-ANALYTES

Immunoassay methods can be applied for analysis of two or even more analytes in the same sample, employing different approaches. A dual-label assay was developed for the simultaneous analysis of thyroxine and triiodothyronine. Thyroxine was conjugated with alkaline phosphatase enzyme, and the triiodothyronine was conjugated with β-galactosidase enzyme. Competitive heterogeneous assays were performed for each hormone, in presence of each other. A double antibody precipitation technique was used for separating the immune complexes, followed by measuring the activity of the enzyme conjugates. The alkaline phosphatase conjugate was measured at 540 nm by the formation of phenolphthalein from its monophosphate salt. The β-galactosidase conjugate was measured at 420 nm by the formation of o-nitrophenol from o-nitrophenyl-β-galactoside substrate. The principles of this dual-label approach were used for the development of a similar system for the simultaneous analysis of phenytoin and phenobarbital in presence of each other.

A new heterogeneous FIA system was developed and automated for the simultaneous analysis of serum albumin and transferring. This system employed dual-fluorescent molecules as labels, and incorporated thiophilic gel solid phase reactor to separate antibody-bound and unbound molecules. Antibody elution was achieved by changes in ionic strength. Detection of the two separate fluorophores was achieved by high speed synchronous fluorescence scanning at two different wavelength intervals, one for each fluorophore.

A sensitive dual-label TRFIA was developed for the simultaneous analysis of methamphetamine and chloropromazine in serum. The assay is based on a novel coating strategy and enhanced lanthanide-based FIA technology. The coating strategy involved the coating of microwell plate wells with high concentrations of a mix Antibody Analyte Marker Signal Concture of anti-methamphetamine and anti-chloropromazine antibodies. Methamphetamine was labelled with Eu (III), and chloropromazine was labelled with Sm (III). In the assay, the labelled compounds were mixed with their non-labelled standard solutions or samples, and the mixture was then added to the wells coated with the antibodies. Methamphetamine was analyzed by measuring the fluorescence intensity of Eu (III) at 615 nm. Chloropromazine was analyzed by measuring the fluorescence intensity of Sm (III) at 643 nm. The assay showed rapid kinetics, and high sensitivities; the lower limits of detections were 1 and 10 ng/ml for methamphetamine and chloropromazine, respectively. A similar system was developed for the simultaneous analysis of thyroxine and thyroid-stimulating hormone in serum. The limits of detection for the assay were 4.1 nmol for thyroxine and 0.028 mIU/ml for thyroid-stimulating hormone with 20 μl sample volume.

A capillary-based FIA system was developed for the simultaneous determination of penicillin-G, ampicillin, amoxicillin, cloxacillin, cephapirin, and ceftiofur. The assay system consists of an assay cartridge containing 4 glass capillaries, a reagent tray with 4 wells of dried reagents, and a processor, which processes the assay, reads the fluorescent output, and finally reports the results. The assay system was specific, and sensitive; the limits of detection were 3.2, 2.9, 3.6, 7.4, 16.3, and 33.7 ng/ml for penicillin-G, ampicillin, amoxicillin, cloxacillin, cephapirin, and ceftiofur, respectively. The capacity of this system was subsequently increased to be able to the simultaneously analyze 9 compounds in the same samples. These compounds are tetracycines (tetracycline, oxytetracycline, chlorotetracycline, and doxycycline), cephalosporins (cephapirin, ceftiour and cefquinome), and b-lactam antibiotics (penicillin-G, ampicillin, and amoxicillin).

Microsphere-based competitive immunoassay was developed for the simultaneous analysis of digoxin and theophylline. This assay was performed using microspheres coated with both compounds, and antibodies labelled with horseradish peroxidase. Two flurogenic substrates were used to differentiate analytical signals from each compound. An epifluorescence microscope and a camera interfaced with a computer were used to measure the individual fluorescence signals. The measured signals were then related to the concentration of the compounds.

Carbonylmetalloimmunoassay method was developed for the simultaneous analysis of carbamazepine, phenobarbital, and diphenylhydantoin. This assay used various transition element-carbonyl complexes as labels followed by the sensitive quantitation of these complexes by Fourier transform infrared spectroscopy. The measurements utilized the particular spectral features of the carbonyl complexes, which show very strong absorption bands in the 1800-2200 cm-1 region. The signals were individually assignable and the intensities were used for quantitation.

ADVANCES IN IMMUNOASSAYS FOR PHARMACEUTICAL ANALYSIS

Advances in immunoassay continue in all areas of the technology; preparation of the unique reagents, analysis of new categories of compounds, and improvement in the methodology and instrumentation:

Advances in Preparation of the Immunoanalytical Rreagents

Immunoassay methods, particularily the monoclonal antibody-based assays, are of potential advantages in parmaceutical analysis because of their high-throughput and high-sensitivity for the analysis in biological samples. However, the protocol for preparation of a specific monoclonal antibody for small drug molecules (Figure 4) is difficult because the low incidence of positive hybridomas from which selection of the most specific one is performed. Furthermore, the production time is excessive (typically 3-9 months). Recent studies were performed to solve these problems. Molecular modeling of the hapten structure was emerged as a useful tool in predicting the immunogenicity of the immunogen, and subsequently the specificity of the generated antibody. By this technique, it was possible to speculate if the antigenic epitope of the hapten molecule is projected and exposed to the immune system, or enveloped into the molecular structure, and thus losses its immunogenicity. This approach was used for synthesis of a sulphacetamide-protein immunogen aimed to the generation of broad specificity antibodies for sulphonamide drugs.

A novel immunization technique was developed for the rapid production of monoclonal antibodies. This technique employed a repetitive, multiple sites, and short regime (8-13 day) immunization. This immunization was followed by fusion of popiteal and bracial lymph node cells from the immunized mouse with myeloma cells. The antibody production time, including immunizations, fusion, screening and cloning takes ~30 days to be completed. This technique was used for generation of monoclonal antibody specific to ranitidine; the generated antibody was used in the development of an immunoassay method for ranitidine. An efficient shorter immunization regime was developed by Darwish et al.. This technique employed only a single immunization of rats, rather than mice. Two weeks later, cell fusion is performed between the medial iliac lymph nodes and rat myeloma cells. By this technique, high number of positive hybridomas was obtained, which offered the chance for selection of more specific hybridoma cells. This technique was employed for generating of highly specific monoclonal antibody for 2`-deoxycytidine; marker for the prognosis of breast cancer to chemotherapy. The generated antibody was subsequenly used for development of specific immunoassay method for this marker in plasma. The use of only single injection in this technique saves not only the time and effort given to the purification of the immunogens (usually by time-consuming chromatographic procedures), but also the amount of the immunogen which is often very valuable or difficult to obtain in an amount enough for multiple injections, or difficult to purify.

Advances to Involve New Categories of Compounds

Most of immunoassay methods applied in pharmaceutical analysis are directed toward organic compounds, however this technique is theoretically applicable to any analyte, even inorganic metal ions, if a suitable antibody can be generated. The ability of generating monoclonal antibodies that recognize metal ions has been demonstrated. The immunogens were prepared by chelating the metal ions with a bifunctional chelator (e.g. p-isothiocyanate-benzyl EDTA), followed by coupling the metal-chelator complex to a carrier protein via the isothiocyanate group of the chelator. The screening assay was designed to select hybridomas that secrete antibodies that show higher affinity to the metal-chelator complex, rather than to the metal-free chelator. By this approach, specific monoclonal antibodies were generated for uranium, cadmium, lead, cobalt, and mercuric. These antibodies were used in the development of competitive immunoassays for the accurate measurement of these metal ions in water samples as well as in human serum. The basic protocol for the competitive immunoassay for metal ions is illustrated in Figure 5. Briefly, the sample containing metal ions is treated with an excess of metal-free chelator to form metal chelate complexes. The resulting solution, which contains the excess chelator and the metal chelate complexes, is mixed with the antibody in microwell plates containing immobilized metal chelate-protein conjugate. The soluble metal chelate complexes compete with the immobilized metal chelate-protein conjugate for the binding sites of a specific antibody. After attaining the equilibrium, the unbound reagents are removed by washing, and an enzyme labelled secondary antibody is added. The unbound secondary antibody is reomved by a second wash step. The signal is generated by addition of a chromogenic enzyme substrate. The measured signal is then related to the concentration of metal ions present in the original sample. These assays were highly sensitive, remarkably quick, easily performed, reasonably portable to the analysis site, require minimum sample pretreatment, and inexpensive.

More recently, FPFIA methods were developed for lead (II) and cadmium (II). The antibodies were raised against protein conjugates of the metal ionsplexes of polyaminopolycarboxylate chelating agent. Fluorophore-labelled analogues of the metal chelate complexes were used as tracer in performing the competitive assay. This method was very sensitive; the limits of detection were 20, and 100 ppt for lead (II) and cadmium (II), respectively.

Although immunoassays for metal ions are still in their infancy, very promising results have been obtained. Recent advances in the genetic manipulation of recombinant single-chain antibodies provided a new opportunity to optimize the binding properties of the parent antibody to provide new highly specific recombinant reagents. The new immunoassays for heavy metals are considered as a promising alternative approach for the existing technologies for analysis of metal ions (atomic absorption spectroscopy, inductively coupled plasma emission spectroscopy, etc.), which are expensive and the analysis needs excessive pretreatment of the sample.

Cloned Enzyme Donor Immunoassay

Cloned enzyme donor immunoassay (CEDIA) methodology is a novel approach which uses the DNA technology to produce homogenous enzyme immunoassays for drugs. The principle of this method is illustrated in Figure 6. Enzyme donor units combine with enzyme acceptor units to form a complete and fully active tetrameric enzyme molecule, which reacts with a colourless substrate to produce coloured product. An enzyme donor-drug conjugate is prepared by linking the drug molecule to the enzyme donor fragment. Competitive binding reaction results in the formation of active enzyme and consequently coloured product, which is directly proportional to the concentration of the drug present. This assay is rapid and has high throughput. Successful cloned enzyme donor immunoassay methods were developed for analysis of amphetamine, methamphetamine, barbiturates, opiates, phencycline, phenytoin, and benzodiazepines. The CEDIA methods were validated, in terms of the sensitivity and precision, in reference to gas chromatography- mass spectrometry, and their validity was proved for application in routine drug screening.

Flow-Injection Immunoassay

Flow-injection immunoassay (FIIA) methods were recently introduced to enhance the efficiency of immunochemical reaction, as well as to increase the performance of the analysis. Most of the developed FIIA systems are based on a principle of displacement (Figure 7). In this case, a column packed with immobilized solid phase antibodies is incorporated into the system. The column is saturated with a solution containing labelled analyte. Injection of the sample (contains the free unlabelled analyte) into the column results in displacement of the labelled analyte (due to the affinity of antibody for labelled analyte is usually significantly lower than its affinity for unlabelled free analyte). The displaced labelled analyte is eluted and detected at the outlet of the column. The generated signal is directly proportional to the concentration of the analyte in the sample (Figure 7A). A similar system was constructed using a column packed with immobilized analyte, and presaturated with labelled antibody. The injection of the sample into this system results in displacement of the analyte-labelled antibody complexes, which is the detected at the outlet of the column (Figure 7B).

FIIA systems based on displacement were developed for analysis of gentamicin, phenobarbital, and digoxigenin. The limits of detections of these flow-injection systems were in the range of 10-100 nM. The prototype of a flow-injection based instrument has been constructed by United States Drug Testing to screen urine for the presence of trace amounts of drugs of abuse. For analysis by displacement-based FIIA, definite conditions such as column, flow rate, and buffer solution should be optimized for each particular analyte, and should be met in the analysis apparatus. To overcome this diasadvantage, an alternative system was developed for theophylline. The general assay process involved in this system is shown in Figure 8. This system consists of a column containing immobilized protein-A, which possess the property of capturing most antibodies with high affinity. The preliminary step in the analysis is a saturation of the column with anti-theophylline antibody. The analysis is conducted by injection of a mixture of free theophylline and theophylline labelled with alkaline phosphatase enzyme. The labelled and free theophylline compete to interact with the antibodies bound to protein-A. The amount of labelled theophylline bound to antibodies is inversely proportional to the concentration of free theophylline. This amount is determined by injecting p-aminophenylphosphate substrate and amperometric detection of the produced p-aminophenol. After the assay procedure, the column is regenerated by removing complexes by acid elution through the column. Immobilized protein-A is then saturated again with new antibodies, and the assay is repeated. This system is time consuming since four steps are involved for each measurement, i.e. incubation of reagents, separation on the column, injection of substrate, and finally regeneration of the protein-A column. The applications of this approach of protein-A column regeneration was applied in the analysis of various compounds.

An improved flow system has been developed by Darwish for analysis of aminoglycosides (e.g. tobramycin) in human serum. A schematic diagram for this assay system is shown in Figure 9. This system is based on off-line competitive reaction between tobramycin and tobramycin labelled with β-galactosidase enzyme for the anti-tobramycin antibody. After the equilibrium was reached, the sample was introduced into the flow system. The antibody complexes of both free tobramycin and labelled tobramycin were trapped in a protein G column while the unbound labelled tobramycin was eluted and monitored by colorimetric detection down-stream after reaction with chlorophenolic red-β-D-galactopyranoside as a substrate. The signal was directly proportional to the concentration of tobramycin in the original sample. Protein-G column used in this system showed a good stability and capacity, thus allowing analysis of large number of samples before the need for regeneration of the column. Furthermore, the continuous detection of the eluting free fraction of the labelled tobramycin, rather than the bound fraction, allowed a higher sample throughput.

Capillary Electrophoresis Immunoassay

Capillary electrophoresis immunoassay (CEIA) has been recently introduced as a sensitive analytical technique, particularly when combined with a sensitive detection method. This technique is a combined use of the principles of the capillary electrophoresis separation and the heterogeneous immunoassay. In this technique, the antibody is attached covalently to the modified interior surface of a microcapillary, which is used as a solid phase immunoreactor. The detection is based on the reaction process of the immunoassay. The analyte and labelled analyte (tracers) are mixed and the mixture is then injected into the system. Tracers are detected with on-column detector. The concentration dependent peaks produced serve as the basis for quantitation. These principles were utilized in the development of an assay system for determination of phenobarbital in serum. Alkaline phosphatase labelled phenobarbital was used as tracer, and p-aminophenyl phosphate was used as a substrate. The detection was carried out amperometrically for the produced p-aminophenol. The capillary electrophoresis with competitive assay design was capable to achieve detection limits in the picomole levels for less than 5 ml sample volume.

A CEIA based on homogeneous fluorescence polarization was developed. The imunoanalytical reagents (analyte, fluorescein-labelled analyte, and antibody) are mixed off-line, and the mixture is then injected into the electrophoresis capillary. The fluorescein-labelled analyte and the antibody-labelled analyte complexes are separated under applied potential, and detected by on-column laser-induced fluorescence detector. This system was applied to the analysis of various drugs in biological fluids, e.g. methotrexate, methamphetamine, theophylline, ethosuximide, paracetamol, salicylate, quinidine, hirudin, codeine, dihydrocodeine, and their glucuronides.

A sensitive CEIA method with electrochemical detection was developed for the analysis of thyroxine in human serum. The method is a combined use of capillary electrophoresis separation and homogeneous enzyme immunoassay principles. The assay was performed in competitive configuration between sample thyroxine and horseradish peroxidase-labelled thyroxine for anti-thyroxine antibody. The free enzyme-labelled thyroxine and the antibody-bound labelled thyroxine complexes were separated by capillary electrophoresis in a separation capillary. Then, they catalyzed the oxidation of tetramethylbenzidine substrate with hydrogen peroxide in a reaction capillary connected to the system after the separation capillary. The reaction product of the substrate was amperometrically determined using an electrode at the outlet of the reaction capillary. The assay was very sensitive; the limit of detection was 23.2 attomole. A similar system was developed for cortisol, with limit of detection of 7.8 attomole. The same principle was used in designing a sensitive assay for tumor markers.

The CEIA with fluorescence detection was employed for the simultaneous analysis of urinary methadone, opiates, benzoyl-ecgonine (cocaine metabolite), and amphetamine. After incubation of the sample with the imunoreactants, a small aliquot of the mixture is applied onto a fused silica capillary and the unbound fluorescein labelled drugs were monitored by capillary electrophoresis with on-column laser induced fluorescence detector. This multi-analyte system was also applied for analysis of digoxin and gentamicin.

Immunosensors

Immunosensors represent the most technological progress in the field of immuoassay development. These sensors are analytical devices composed of an immunochemical recognition element directly interfaced to a signal transducer, which together relate the concentration of an analyte to a measurable response. Wide range of signal transducers can be used, e.g. optical, electrochemical, and piezoelectric transducers. The principles underlying immunosensors technology, and their analytical applications are the subject of many reviews. Immunosensors were widely applied in the analysis of many compounds of pharmaceutical importance; examples are given in Table 9. These sensors usually employ either reusable/regenerable or disposable assay format. Sensors using disposable formats are potential for multi-analyte analysis, format versatility, system cost, the assay time, sensitivity, and reproducibility. In contrast to the disposable formats, the multi-use immunosensors (which can recharged or regenerated) offer certain advantages, particularily for use as detector for chromatographic and flow injection systems. For example, continous flow and electrochemical immunosensors developed for analysis of cephalexin, paclitaxel, progesterone, and thyroid hormones.

Although many immunosensors were developed for analysis of wide range of drugs and hormones. However, it was not possible to perform the analysis in complex biological fluid (e.g. serum), in a one-step procedure, due to the interference resulting from the non-specific binding to the sensor surface. To control this interference, an optical chip interferometer has been developed. This interferometer configuration employed a reference sensing region that can be functionalized separately from the measuring sensing region, and additional reduction in signal drift in serum was achieved by controlling the surface chemistry of the optical chip. This functionalized procedure was utilized in the development of one-step assay for human chorionic gonadotropin in human serum with a detection limit of 0.1 ng/ml for a 35 min assay.

CONCLUSIONS

Immunoassay methods are bioanalytical methods in which quantitation of analyte depends on its reaction with specific antibody. The response signal is generated from a label attached to either the analyte or antibody. The property of highly specific recognition of analytes by antibodies leads to the high selectivity of these assays. The extreme affinity of analyte-antibody interaction results in greater sensitivity of immunoassay methods. The instrumentation technology leads to automation of immunoassays and consequently increasing their throughput. Immunoassay methods are capable of quantifying wide variety of compounds such as low molecular weight drugs, macromolecular biomolecules, metabolites, and/or biomarkers which indicate disease diagnosis or prognosis. Therefore, these methods found wide applicability in many important areas of pharmaceutical analysis such as diagnosis of diseases, therapeutic drug monitoring, clinical pharmacokinetic and bioequivalence studies in drug discovery and pharmaceutical industries.

Inspite of the numerous advantages of immunoassays, they have some limitations. Immunoassays depend mainly on reaction between analyte and a biological antibody, they may have more inherent imprecision than other methods employed in pharmaceutical analysis (e.g. chromatography). The specificity of immunoassays depend mainly on the antibody directed to the analyte, however some immunoassays are not highly selective, and they may respond to a group of compounds (e.g. aminoglycosides, pesticides,...etc.) rather than individual compounds. Moreover, lack of specificity may be observed due to non-specific binding of the antibody to matrix component(s). In these circumstances, care must be taken to ensure the absence of interfering- substance in the analyte sample and/or matrix.

Recently, a marked improvement was achieved in the field of immunoassay development for the purposes of pharmaceutical analysis. This improvement involved the preparation of the unique immunoanalytical reagents, analysis of new categories of compounds, methodology, and instrumentation. The most important examples in this field are the development of continuous-flow immunoassays and immunosensors. These technologies resulted in improvement of the analysis performance by increasing the sensitivity, decreasing the analysis time, simplification of the assay procedure, automation of the method, and miniaturization of the analytical equipment.

References 250

  1. 1.

    Kellner R, Mermet JM, Otto M, Widmer HM. Analytical Chemistry. 1998;405-429.

    Not in our library

  2. 2.

    Findlay JW, Smith WC, Lee JW, Nordblom GD, et al. J. Pharm. Biomed. Anal. 2000;21:1249-1273.

    Not in our library10.1016/s0731-7085(99)00244-7

  3. 3.

    Chuanlai X, Cifang P, Kai H, Zhengyu J, et al. Luminescence. 2006;21:126-128.

    Not in our library10.1002/bio.892

  4. 4.

    Samsonova ZhV, Shchelokova OS, Ivanova NL, Rubtsova MU, et al. Prikl. Biokhim. Mikrobiol. 2005;41:668-675.

    Not in our library

  5. 5.

    Lachenmeier K, Musshoff F, Madea B. Forensic. Sci. Intl. 2006;159:189-199.

    Not in our library10.1016/j.forsciint.2005.08.009

  6. 6.

    Benito-Pena E, Moreno-Bondi MC, Orellana G, Maquieira A, et al. J. Agric. Food Chem. 2005;53:6635-6642.

    Not in our library10.1021/jf0511502

  7. 7.

    Koszegi T. J. Biochem. Biophys. Methods. 2002;53:157-164.

    Not in our library10.1016/s0165-022x(02)00104-5

  8. 8.

    Fraser AD, Worth D. Ther. Drug. Monit. 2002;24:746-750.

    Not in our library10.1097/00007691-200212000-00011

  9. 9.

    Kobayashi N, Kubota K, Oiwa H, Goto J, et al. J. Immunol. Methods. 2003;272:1-10.

    Not in our library10.1016/s0022-1759(02)00115-1

  10. 10.

    Karakulova IV, Shutov AA. Klin. Lab. Diagn. 2006;1:9-10.

    Not in our library

  11. 11.

    Desruisseau S, Palmari J, Giusti C, Romain S, et al. Br. J. Cancer. 2006;94:239-246.

    Not in our library10.1038/sj.bjc.6602920

  12. 12.

    Ball L, Jones A, Boogaard P, Will W, et al. Biomarkers. 2005;10:127-137.

    Not in our library10.1080/13547500500158938

  13. 13.

    Signo P, Barassi A, Novario R, Melzi d’Eril GV. Clin. Chem. Lab. Med. 2005;43:883-885.

    Not in our library10.1515/cclm.2005.149

  14. 14.

    Fiets WE, Blankenstein MA, Struikmans H, Ruitenberg HM, et al. Int. J. Biol. Markers. 2002;17:24-32.

    Not in our library10.1177/172460080201700103

  15. 15.

    Pieniaszek HJ, Davidson AF, Walton HL, Pinto DJ, et al. J. Pharm. Biomed. Anal. 2003;30:1441-1449.

    Not in our library10.1016/s0731-7085(02)00481-8

  16. 16.

    Suzuki Y, Arakawa H, Maeda M. Anal. Sci. 2003;19:111-115.

    Not in our library10.2116/analsci.19.111

  17. 17.

    Kobayashi N, Shibusawa K, Kubota K, Hasegawa N, et al. J. Immunol. Methods. 2003;274:63-75.

    Not in our library10.1016/s0022-1759(02)00501-x

  18. 18.

    Ito K, Nakagawa K, Murakami S, Arakawa H, et al. Anal. Sci. 2003;19:105-109.

    Not in our library10.2116/analsci.19.105

  19. 19.

    Maeda M. J. Pharm. Biomed. Anal. 2003;30:1725-1734.

    Not in our library10.1016/s0731-7085(02)00514-9

  20. 20.

    Gomez-Henz A, Aguilar-Caballos MP. Comb. Chem. High Throughput Screen. 2003;6:177-182.

    Not in our library10.2174/138620703106298356

  21. 21.

    Burke TJ, Loniello KR, Beebe JA, Ervin KM. Comb. Chem. High Throughput Screen. 2003;6:183-194.

    Not in our library10.2174/138620703106298365

  22. 22.

    Giese RW. J. Chromatogr. A. 2003;1000:401-412.

    Not in our library10.1016/s0021-9673(03)00306-6

  23. 23.

    Hermida J, Boveda MD, Vadillo FJ, Tutor JC. Clin. Biochem. 2002;35:251-254.

    Not in our library10.1016/s0009-9120(02)00299-0

  24. 24.

    Zhao X, Kottegoda S, Shippy S. Analyst. 2003;128:357-362.

    Not in our library10.1039/b210782a

  25. 25.

    Draisci R, Quandri F, Achene L, Volpe G, et al. Analyst. 2001;126:1942-1946.

    Not in our library10.1039/b104939a

  26. 26.

    Darwish I, Akizawa T, Hirose K, Omura K, et al. Anal. Chim. Acta. 1998;365:121-128.

    Not in our library

  27. 27.

    Wring SA, Kilpatrick KE, Hutchins JT, Witherspoon SM, et al. J. Pharm. Biomed. Anal. 1999;19:695-707.

    Not in our library10.1016/s0731-7085(98)00296-9

  28. 28.

    Harlow E, Lane D. Antibodies: a Laboratory Manual. 1988;92-103.

    Not in our library

  29. 29.

    Terrana B, Giuliani MM, Cosco E, Mannucci D, Masseyeff RF, Alber WH, Staines NA. Methods of Immunological Analysis. 1993;2:492-510.

    Not in our library

  30. 30.

    Köhler G, Milstein C. Nature. 1975;256:495-497.

    Not in our library10.1038/256495a0

  31. 31.

    Freshney RI. Culture of Animal Cells. 1987;336-338.

    Not in our library

  32. 32.

    Majima K, Fukui T, Yunan J, Wang G, et al. Anal. Sci. 2002;18:869-874.

    Not in our library10.2116/analsci.18.869

  33. 33.

    Putalun M, Tanaka H, Muranaka T, Shoyama Y. Analyst. 2002;127:328-1332.

    Not in our library10.1039/b203351h

  34. 34.

    Micheli L, Di Stefano S, Moscone D, Palleschi G, et al. Anal. Bioanal. Chem. 2002;373:678-684.

    Not in our library10.1007/s00216-002-1399-3

  35. 35.

    Okuda K, Kotake Y, Ohta S. Life Sci. 2002;70:2871-2883.

    Not in our library10.1016/s0024-3205(02)01543-6

  36. 36.

    Fiet J, Giton F, Boudi A, Boudou P, et al. Steroids. 2001;66:81-86.

    Not in our library10.1016/s0039-128x(00)00207-5

  37. 37.

    Yeung JM, Newsome WH, Abbott MA. J. AOAC Int. 2000;83:139-143.

    Not in our library

  38. 38.

    Weeks I, Sevehla G. Chemiluminescence Immunoassay. 1992;24:18-118.

    Not in our library

  39. 39.

    Simmonet F, Guilloteau D, Masseyeff RF, Albert WH, Staines NA. Method of Immunological Analysis. 1993;1:270-282.

    Not in our library

  40. 40.

    Kobayashi N, Katumata H, Uto Y, Goto J, et al. Steroids. 2002;67:827-833.

    Not in our library10.1016/s0039-128x(02)00036-3

  41. 41.

    Arakawa H, Shiokama M, Imamura O, Aeda M. Anal. Biochem. 2003;314:206-211.

    Not in our library10.1016/s0003-2697(02)00657-7

  42. 42.

    Abad A, Manclus JJ, Moreno MJ, Montoya A. J. AOAC Int. 2001;84:156-161.

    Not in our library

  43. 43.

    Darwish IA, Blake DA. Anal. Chem. 2002;74:52-58.

    Not in our library10.1021/ac010510r

  44. 44.

    He Z, Jin W. Anal. Biochem. 2003;313:34-40.

    Not in our library10.1016/s0003-2697(02)00508-0

  45. 45.

    Porstsmann B, Portsmann T, Ngo TT. Non-isotopic Immunoassay. 1988;60.

    Not in our library

  46. 46.

    Burestedt E, Nistor C, Schagerlöf U, Emnéus J. Anal. Chem. 2000;72:4171-4177.

    Not in our library10.1021/ac000128z

  47. 47.

    Seto Y, Iba T, Abe K. Luminescence. 2001;16:285-290.

    Not in our library10.1002/bio.654

  48. 48.

    Wilson R, Clavering C, Hutchinson A. Analyst. 2003;128:480-485.

    Not in our library10.1039/b301942j

  49. 49.

    Inouye K, Ueno I, Yokoyama S, Sakaki T. J. Biochem. 2002;13:97-105.

    Not in our library10.1093/oxfordjournals.jbchem.a003082

  50. 50.

    Hemmilä I. Clin. Chem. 1985;31:359-370.

    Not in our library

  51. 51.

    Adamezyk M, Mattingly PG, Moore JA, Pan Y, et al. Bioconjug. Chem. 2001;12:329-331.

    Not in our library10.1021/bc000152j

  52. 52.

    Baeyens WR, Schulman SG, Calokerinos AC, Zhao Y, et al. J. Pharm. Biomed. Anal. 1998;17:941-953.

    Not in our library10.1016/s0731-7085(98)00062-4

  53. 53.

    Garcia AM, Baeyens WR, Zhang XR, Smet E, et al. Calokerinos, Biomed. Chromatogr. 2000;14:166-172.

    Not in our library10.1002/1099-0801(200005)14:3<166::aid-bmc954>3.0.co;2-p

  54. 54.

    Ankelo M, Westerlund A, Ilonen J, Knip M, et al. Clin. Chem. 2003;49:908-915.

    Not in our library10.1373/49.6.908

  55. 55.

    Kimura H, Mukaida M, Wang G, Yuan J, et al. Forensic Sci. Int. 2000;113:345-351.

    Not in our library10.1016/s0379-0738(00)00257-7

  56. 56.

    Yunan J, Wang G, Majima K, Matsumoto K. Anal. Chem. 2001;73:869-1876.

    Not in our library10.1021/ac0013305

  57. 57.

    Wheeler MJ. Ann. Clin. Biochem. 2001;38:217-229.

    Not in our library10.1258/0004563011900641

  58. 58.

    Rongen HA, Bult A, Van Bennekom WP. J. Immunol. Methods. 1997;204:105-133.

    Not in our library10.1016/s0022-1759(97)00041-0

  59. 59.

    Agasan AL, Stewart BJ, Watson TG. J. Immunol. Methods. 1994;177:251-260.

    Not in our library10.1016/0022-1759(94)90163-5

  60. 60.

    Bowsher RR, Compton JA, Kirkwood JA, Place GD, et al. Antimicrob. Agents Chemother. 1994;38:2134-2142.

    Not in our library10.1128/aac.38.9.2134

  61. 61.

    Laurent V, Lindberg I. Anal. Biochem. 2002;309:143-149.

    Not in our library10.1016/s0003-2697(02)00252-x

  62. 62.

    Hage DS. Anal. Chem. 1995;67:455R-462R.

    Not in our library10.1021/ac00108a030

  63. 63.

    Cracium AM, Vermeer C, Eisenwiener HG, Drees N, et al. Clin. Chem. 2000;46:252-257.

    Not in our library

  64. 64.

    Lavery P, Brown MJ, Pope AJ. J. Biomol. Screen. 2001;6:3-9.

    Not in our library10.1177/108705710100600102

  65. 65.

    Asthagiri AR, Horwitz AF, Lauffenburger DA. Anal. Biochem. 1999;269:342-347.

    Not in our library10.1006/abio.1999.4055

  66. 66.

    Cirimele V, Kentez P, Lohner S, Ludes B. J. Anal. Toxicol. 2003;27:103-105.

    Not in our library10.1093/jat/27.2.103

  67. 67.

    Najjar SM, Weintraub BD, De Pablo F, Scanes CG, Weintraub BD. Handbook of Endocrine Research Techniques. 1993;3-23.

    Not in our library

  68. 68.

    Chang YC, Li CM, Li LA, Jong SB, et al. Analyst. 2003;128:363-368.

    Not in our library10.1039/b210111b

  69. 69.

    Lynas L, Currie D, Elliott CT, McEvoy JD, et al. Analyst. 1998;123:2773-2777.

    Not in our library10.1039/a804904a

  70. 70.

    Blackwell LF, Brown JB, Vigil P, Gross B, et al. Steroids. 2003;68:465-476.

    Not in our library10.1016/s0039-128x(03)00049-7

  71. 71.

    Ikeda Y, Araki T, Takimoto H, Fujii Y. Biol. Pharm. Bull. 2002;25:422-425.

    Not in our library10.1248/bpb.25.422

  72. 72.

    Ikeda Y, Fujii M, Yamazaki M, Fujii Y. Clin. Chim. Acta. 2001;314:245-247.

    Not in our library10.1016/s0009-8981(01)00635-0

  73. 73.

    Vakkuri O, Arnason SS, Pouta A, Vuolteenaho O, et al. J. Endocrinol. 2000;165:669-677.

    Not in our library10.1677/joe.0.1650669

  74. 74.

    Meyer MT, Bumgarner JE, Varns JL, Daughtridge JV, et al. Sci. Total Environ. 2000;248:181-187.

    Not in our library10.1016/s0048-9697(99)00541-0

  75. 75.

    Lin CC, Veals J, Korduba C, Hilbert MJ, et al. Ther. Drug Monit. 1997;19:675-681.

    Not in our library10.1097/00007691-199712000-00012

  76. 76.

    Tadepalli SM, Quinn RP. J. Pharm. Biomed. Anal. 1996;15:157-163.

    Not in our library10.1016/0731-7085(96)01835-3

  77. 77.

    Bayliss MA, Preston JR, Wilkinson D, Wilkinson DJ, et al. J. Pharm. Biomed. Anal. 1996;15:21-31.

    Not in our library10.1016/0731-7085(96)01816-x

  78. 78.

    Mannaert E, Tytgat J, Daenens P. Clin. Chim. Acta. 1996;253:103-115.

    Not in our library10.1016/0009-8981(96)06350-4

  79. 79.

    Yamanaka K, Morikawa S, Murata K, Banno K, et al. J. Pharm. Biomed. Anal. 1996;14:281-287.

    Not in our library10.1016/0731-7085(95)01595-7

  80. 80.

    Aravagiri M, Marder SR, Van Putten T, Hawes EM, et al. Ther. Drug Monit. 1994;16:21-29.

    Not in our library10.1097/00007691-199402000-00003

  81. 81.

    He H, McKay G, Midha KK. J. Pharm. Sci. 1993;82:1027-1232.

    Not in our library

  82. 82.

    He H, McKay G, Wirshing B, Midha KK. J. Pharm. Sci. 1995;84:561-567.

    Not in our library10.1002/jps.2600840509

  83. 83.

    Mei JV, Hannon WH, Dobbs TL, Bell CJ, et al. Clin. Chem. 1998;44:281-286.

    Not in our library

  84. 84.

    De Clerck I, Daenens P. Analyst. 1997;122:1119-1124.

    Not in our library10.1039/a702869e

  85. 85.

    Narongchai P, Sribanditmonkol P, Thampithug S, Narongchai S, et al. J. Med. Assoc. Thai. 2002;85:82-86.

    Not in our library

  86. 86.

    Tagliaro F, Valentini R, Manetto G, Crivellente F, et al. Forensic Sci. Int. 2000;107:121-128.

    Not in our library10.1016/s0379-0738(99)00157-7

  87. 87.

    Marsh A, Evans MB, Strang J. J. Pharm. Biomed. Anal. 1995;84:829-839.

    Not in our library10.1016/0731-7085(95)01517-o

  88. 88.

    Tagliaro F, Camilot M, Valentini R, Mengarda F, et al. J. Chromatogr. B Biomed. Sci. Appl. 1998;16:77-82.

    Not in our library10.1016/s0378-4347(98)00196-0

  89. 89.

    Deberg M, Houssa P, Frank BH, Sodoyez-Goffaux F, et al. Clin. Chem. 1998;44:1504-1513.

    Not in our library

  90. 90.

    Shirtcliff EA, Granger DA, Schwartz EB, Curran MJ, et al. Horm. Behav. 2000;38:137-147.

    Not in our library10.1006/hbeh.2000.1614

  91. 91.

    Giton F, Valleix A, Boudou P, Villette JM, et al. J. Steroid. Biochem. Mol. Biol. 2002;81:85-94.

    Not in our library10.1016/s0960-0760(02)00051-1

  92. 92.

    Lo MS, Ng ML, Wu LL, Khalid BA. Malays. J. Pathol. 1996;18:53-58.

    Not in our library

  93. 93.

    Lu YC, Chatterton RT, Vogelsong KM, May LK. J. Immunoassay. 1997;18:149-163.

    Not in our library10.1080/01971529708005810

  94. 94.

    Riepe FG, Wonka S, Partsch CJ, Sippell WG. J. Chromatogr. B Biomed. Sci. Appl. 2001;763:99-106.

    Not in our library10.1016/s0378-4347(01)00373-5

  95. 95.

    Clive DR, Sudhaker D, Giacherio D, Gupta M, et al. Clin. Biochem. 2002;35:517-521.

    Not in our library10.1016/s0009-9120(02)00378-8

  96. 96.

    Abe F, Omata K, Yamada M, Tsunoda K, et al. Immunopharmacol. 1999;44:199-204.

    Not in our library10.1016/s0162-3109(99)00090-9

  97. 97.

    Kominami G, Nakamura M, Mizobuchi M, Ueki K, et al. J. Pharm. Biomed. Anal. 1996;15:175-182.

    Not in our library10.1016/0731-7085(96)01849-3

  98. 98.

    Sommer RG, Belchak TL, Blockzynski ML, Oguslawski SJ, et al. Clin. Chem. 1990;36:201-206.

    Not in our library

  99. 99.

    Kerriagen S, Phillips WH. Clin. Chem. 2001;47:540-547.

    Not in our library

  100. 100.

    Wolthers BG, Kema IP, Volmer M, Wesemann R, et al. Clin. Chem. 1997;43:114-120.

    Not in our library

  101. 101.

    Kaibe K, Nishimura S, Ishil H, Sunahara N, et al. Clin. Chem. 1990;36:24-27.

    Not in our library

  102. 102.

    Adam A, Gervais N, Panoyan A, Ong H. Analyst. 1994;119:2663-2666.

    Not in our library10.1039/an9941902663

  103. 103.

    Haasnoot W, Stouten P, Cazemier G, Lommenn A, et al. Analyst. 1999;124:301-305.

    Not in our library10.1039/a807846g

  104. 104.

    Lind C, Chen J, Byrjalsen I. Clin. Chem. 1997;43:943-949.

    Not in our library

  105. 105.

    Kim JK, Adam A, Loo JC, Ong H. J. Pharm. Biomed. Anal. 1995;13:885-891.

    Not in our library10.1016/0731-7085(95)01503-d

  106. 106.

    Draisci R, Volpe G, Compagnone D, Purificato I, et al. Analyst. 2000;125:1419-1423.

    Not in our library10.1039/b003176n

  107. 107.

    Marnet PG, Volland H, Pradelles P, Grassi J, et al. J. Immunoassay. 1994;15:35-53.

    Not in our library10.1080/15321819408009570

  108. 108.

    Craciun AM, Vermeer C, Eisenwiener HG, Drees N, et al. Clin. Chem. 2000;46:252-257.

    Not in our library

  109. 109.

    Hammer P, Kirchhoff H, Hahn G. Anal. Chim. Acta. 1993;275:313-316.

    Not in our library

  110. 110.

    Thacker JD, Casale TS, Tucker CM. J. Agric. Food Chem. 1996;44:2680-2685.

    Not in our library

  111. 111.

    Tanaka T, Ikebuchi H, Sawada JI, Okada M, et al. J. AOAC Int. 1996;79:426-430.

    Not in our library

  112. 112.

    Kolosova AY, Blintsov AN, Samsonova JV, Egorov AM. Fresenius J. Anal. Chem. 1998;361:329-330.

    Not in our library

  113. 113.

    Usleber E, Lorber M, Straka M, Terplan G, et al. Analyst. 1994;119:2765-2768.

    Not in our library10.1039/an9941902765

  114. 114.

    Hagemann K, Piek K, Stockigt J, Weiler EW. Planta Medica. 1992;58:68-72.

    Not in our library10.1055/s-2006-961392

  115. 115.

    Haasnoot W, Stouten P, Lommen A, Cazemier G, et al. Analyst. 1994;119:2675-2680.

    Not in our library10.1039/an9941902675

  116. 116.

    Hassan SS, Rowell FG, Hambleton P, Jackson LS. Anal. Commun. 1998;35:249-252.

    Not in our library

  117. 117.

    Iosefsohn M, Soldin SJ, Hicks JM. Ther. Drug. Monit. 1990;12:201-205.

    Not in our library10.1097/00007691-199003000-00016

  118. 118.

    Laurie D, Mason AJ, Piggott NH, Rowell FJ, et al. Analyst. 1996;121:951-954.

    Not in our library10.1039/an9962100951

  119. 119.

    Sachetelli S, Beaulac C, Lagacé J. Biochim. Biophys. Acta. 1998;1379:35-41.

    Not in our library10.1016/s0304-4165(97)00079-2

  120. 120.

    Westermann J, Hubl W, Kaiser N, Salewski L. Clin. Lab. 2002;48:61-71.

    Not in our library

  121. 121.

    Kohno T, Murasugi N, Sakurai H, Watabe K, et al. Clin. Lab. Anal. 1998;12:268-275.

    Not in our library10.1002/(sici)1098-2825(1998)12:5<268::aid-jcla3>3.0.co;2-2

  122. 122.

    Choi MJ, Song EY, Chung TW. Arch. Pharm. Res. 1998;21:231-235.

    Not in our library10.1007/bf02975280

  123. 123.

    Cao Y, Smith WC, Bowsher RR. J. Pharm. Biomed. Anal. 2001;26:53-61.

    Not in our library10.1016/s0731-7085(01)00345-4

  124. 124.

    Kohno H, Sueshige N, Oguri K, Izumidate H, et al. Clin Biochem. 2000;33:243-253.

    Not in our library10.1016/s0009-9120(00)00065-5

  125. 125.

    White LO. Ther. Drug Monit. 1998;20:464-468.

    Not in our library10.1097/00007691-199810000-00003

  126. 126.

    Karapitta CD, Xenakis A, Papadimitriou A, Sotiroudis TG. Clin. Chem. Acta. 2001;308:99-106.

    Not in our library10.1016/s0009-8981(01)00469-7

  127. 127.

    Karapitta CD, Sotiroudis TG, Papadimitriou A, Xenakis A. Clin. Chem. 2001;47:569-574.

    Not in our library

  128. 128.

    Yao H, Jenkins SH, Pesce AJ, Halsall HB, et al. Clin. Chem. 1993;39:1432-1434.

    Not in our library

  129. 129.

    Kimura S, Lyama S, Yamaguchi Y, Kanakura Y. J. Clin. Lab. Anal. 2001;15:319-323.

    Not in our library10.1002/jcla.1044

  130. 130.

    Stead DA. J. Chromatogr. B. 2000;747:69-93.

    Not in our library10.1016/s0378-4347(00)00133-x

  131. 131.

    Eldefrawi ME, Azer NL, Nath N, Anis NA, et al. Appl. Biochem. Biotechnol. 2000;87:25-35.

    Not in our library10.1385/abab:87:1:25

  132. 132.

    O`Connell KP, Valdes JJ, Azer NL, Schwartz RP, et al. J. Immunol. Methods. 1999;225:157-169.

    Not in our library10.1016/s0022-1759(99)00041-1

  133. 133.

    Coille I, Reder S, Bucher S, Gauglitz G. Biomol. Eng. 2002;18:273-280.

    Not in our library10.1016/s1389-0344(01)00111-3

  134. 134.

    Kakabakos SE, Georgiou S, Petrou PS, Christofidis I. J. Immunol. Methods. 1999;222:183-187.

    Not in our library10.1016/s0022-1759(98)00197-5

  135. 135.

    Petrou PS, Georgiou S, Christofidis I, Kakabakos SE. J. Immunol. Methods. 2002;266:175-179.

    Not in our library10.1016/s0022-1759(02)00144-8

  136. 136.

    Hackett LP, Dusci LJ, Ilett KF. Ther. Drug Monit. 1998;20:30-34.

    Not in our library10.1097/00007691-199802000-00006

  137. 137.

    Chetty M, Wilson JF. Ther. Drug Monit. 1999;21:215-218.

    Not in our library10.1097/00007691-199904000-00012

  138. 138.

    Berry DJ, Patsalos PN. Ther. Drug. Monit. 2000;22:460-464.

    Not in our library10.1097/00007691-200008000-00016

  139. 139.

    Banerjee SK, Wells A, Dasgupa A. Ther. Drug Monit. 1999;21:540-543.

    Not in our library10.1097/00007691-199910000-00009

  140. 140.

    Sym D, Smith C, Meenan G, Lehrer M. Ther. Drug Monit. 2001;23:441-444.

    Not in our library10.1097/00007691-200108000-00020

  141. 141.

    Morigawa N, Mori T, Kawashima H, Takeyama M, et al. Biol. Pharm. Bull. 2000;23:774-777.

    Not in our library10.1248/bpb.23.774

  142. 142.

    Morigawa N, Mori T, Abe T, Kawashima H, et al. Biol. Pharm. Bull. 2000;23:784-787.

    Not in our library10.1248/bpb.23.784

  143. 143.

    Felscher D, Schulz K. J. Forensic Sci. 2000;45:1327-1331.

    Not in our library

  144. 144.

    Schwenzer KS, Pearlman R, Tsilimidos M, Salamone SJ, et al. J. Anal. Toxicol. 2000;24:726-732.

    Not in our library10.1093/jat/24.8.726

  145. 145.

    Agbuya PG, Li L, Miles MV, Zaritsky AL, et al. Ther. Drug Monit. 1996;18:194-199.

    Not in our library10.1097/00007691-199604000-00014

  146. 146.

    Fraser AD, Bryan W. J. Anal. Toxicol. 1995;19:281-284.

    Not in our library10.1093/jat/19.5.281

  147. 147.

    Keller T, Schneider A, Dirnhofer R, Jungo R, et al. Med. Sci. Law. 2000;40:258-262.

    Not in our library10.1177/002580240004000312

  148. 148.

    Fraser AD, Worth D. Ther. Drug Monit. 2002;4:746-750.

    Not in our library10.1097/00007691-200212000-00011

  149. 149.

    Hong JY, Choi MJ. Biol. Pharm. Bull. 2002;25:1258-1262.

    Not in our library10.1248/bpb.25.1258

  150. 150.

    Tuomola M, Cooper KM, Lahdenpera S, Baxter GA, et al. Analyst. 2002;127:83-86.

    Not in our library10.1039/b108671p

  151. 151.

    Wu FB, Han SQ, Zhang C, He YF. Anal. Chem. 2002;74:5882-5889.

    Not in our library10.1021/ac025727f

  152. 152.

    Tscop M, Behre HM, Nieschlag E, Dressendorfer RA, et al. Clin. Chem. Lab. Med. 1998;36:223-230.

    Not in our library10.1515/cclm.1998.038

  153. 153.

    Vihma V, Tiitinen A, Ylikorkala O, Tikanen MJ. J. Clin. Endocrinol. Metab. 2003;88:2552-2555.

    Not in our library10.1210/jc.2002-021898

  154. 154.

    Castillo B, Alvarez-Builla J, Lerner DA. In: Luminescence Techniques in Chemical and Biochemical Analysis, Practical Spectroscopy Series. 1991;12:73-139.

    Not in our library

  155. 155.

    Frank EL, Schwarz EL, Juenke J, Annesley TM, et al. Am. J. Clin. Pathol. 2002;118:124-131.

    Not in our library10.1309/4502-68h7-1k1b-1jvr

  156. 156.

    Pou K, Ong H, Adam A, Lamothe P, et al. Analyst. 1994;119:2659-2662.

    Not in our library10.1039/an9941902659

  157. 157.

    Sato K, Muramatsu K, Aano S. Anal. Biochem. 2002;308:1-4.

    Not in our library10.1016/s0003-2697(02)00216-6

  158. 158.

    Tanaka T, Matsunaga T. Anal. Chem. 2000;72:3518-3522.

    Not in our library10.1021/ac9912505

  159. 159.

    Christofides ND, Sheehan CP. Clin. Chem. 1995;41:17-23.

    Not in our library

  160. 160.

    De Boever J, Kohen F, Bouve J, Leyseele D, et al. Clin. Chem. 1990;36:2036-2041.

    Not in our library

  161. 161.

    Luppa P, Hauck S, Schwab I, Birkmayer C, et al. Clin. Chem. 1995;41:564-570.

    Not in our library

  162. 162.

    Luppa P, Bruckner C, Schwab I, Hauck S, et al. Clin. Chem. 1997;43:2345-2352.

    Not in our library

  163. 163.

    Roda A, Manetta AC, Portanti O, Mirasoli M, et al. Luminescence. 2003;18:72-78.

    Not in our library10.1002/bio.708

  164. 164.

    Kim JK, Adam A, Loo JC, Ong H. J. Phram. Biomed. Anal. 1995;13:885-891.

    Not in our library10.1016/0731-7085(95)01503-d

  165. 165.

    Bacigalupo MA, Ius A, Simoni P, Piazza F, et al. Fresenius J. Anal. Chem. 2001;370:82-87.

    Not in our library10.1007/s002160100717

  166. 166.

    Xu XN, Jeffers RB, Gao J, Logan B. Analyst. 2001;126:1285-1292.

    Not in our library10.1039/b104180k

  167. 167.

    Haga M, Hoshino S, Okada H, Hazemoto N, et al. Chem. Pharm. Bull. 1990;38:252-254.

    Not in our library10.1248/cpb.38.252

  168. 168.

    Kim CK, Park KM. J. Immunol. Methods. 1994;170:225-231.

    Not in our library10.1016/0022-1759(94)90397-2

  169. 169.

    Litchfield WJ, Freytag JW, Adamich M. Clin. Chem. 1984;30:1441-1445.

    Not in our library

  170. 170.

    Kubotsu K, Goto S, Fujita M, Tuchiya H, et al. Clin. Chem. 1992;38:808-812.

    Not in our library

  171. 171.

    Canona-Davis E, Redemann CT, Vollmer YP, Kung VT. Clin. Chem. 1986;32:1687-1691.

    Not in our library

  172. 172.

    Rongen HA, Van der Horst HM, Hugenholtz GW, et al. Anal. Chem. Acta. 1994;287:191-199.

    Not in our library

  173. 173.

    Blake C, El-Bassem MN, Gould BJ, Marks V, et al. Clin. Chem. 1982;28:1469-1473.

    Not in our library

  174. 174.

    Dean KJ, Thompson SG, Burd JF, Buckler RT. Clin. Chem. 1983;29:1051-1056.

    Not in our library

  175. 175.

    Guo JC, Miller JM, Evans M, Palmer DA. Analyst. 2000;125:1707-1708.

    Not in our library10.1039/b005575l

  176. 176.

    Wu FB, Han SQ, Xu T, He YF. Anal. Biochem. 2003;314:87-96.

    Not in our library10.1016/s0003-2697(02)00624-3

  177. 177.

    Huth SP, Warholic PS, Devou JM, Chaney LK, et al. J. AOAC Int. 2002;85:355-364.

    Not in our library

  178. 178.

    Okerman L, De Wasch K, Van Hoof J, Smedts W. J. AOAC. 2003;86:236-240.

    Not in our library

  179. 179.

    Szurdoki F, Michael KL, Walt DR. Anal. Biochem. 2001;291:219-228.

    Not in our library10.1006/abio.2001.5041

  180. 180.

    Varenne A, Vessieres A, Salmain M, Durand S, et al. Anal. Biochem. 1996;242:172-179.

    Not in our library10.1006/abio.1996.0450

  181. 181.

    Salmain M, Vessieres A, Varenne A, Brossier B. J. Organomet. Chem. 1999;589:192-197.

    Not in our library

  182. 182.

    Blake DA, Pavlov AR, Yu H, Khosraviani M, et al. Anal. Chem. Acta. 2001;444:3-11.

    Not in our library

  183. 183.

    Spinks CA, Wyatt GM, Lee HA, Morgan MR. Bioconjug. Chem. 1999;10:583-588.

    Not in our library10.1021/bc980054m

  184. 184.

    Kilpatrick KE, Wring SA, Walker DH, Macklin MD, et al. Hybridoma. 1997;16:381-389.

    Not in our library10.1089/hyb.1997.16.381

  185. 185.

    Yoshioka M, Abu-Zeid M, Kubo T, El-Merzebani M. Biol. Pharm. Bull. 1994;17:169-172.

    Not in our library10.1248/bpb.17.169

  186. 186.

    Darwish I, Emara S, Askal H, El-Rabbat N, et al. Anal. Chim. Acta. 2000;404:179-186.

    Not in our library

  187. 187.

    Blake DA, Chakrabarti P, Khosraviani M, Hatcher FM, et al. J. Biol. Chem. 1996;271:27677-27685.

    Not in our library10.1074/jbc.271.44.27677

  188. 188.

    Khosraviani M, Blake RC II, Pavlov AR, Lorbach SC, et al. Bioconj. Chem. 2000;11:267-277.

    Not in our library10.1021/bc9901548

  189. 189.

    Blake DA, Blake RC II, Khosraviani M, Pavlov AR. Anal. Chem. Acta. 1998;376:13-19.

    Not in our library

  190. 190.

    Marx A, Hock B. Anal. Lett. 1998;31:1633-1650.

    Not in our library

  191. 191.

    Marx A, Hock B. Methods. 2000;22:49-52.

    Not in our library10.1006/meth.2000.1035

  192. 192.

    Khosraviani M, Pavlov AR, Flowers GC, Blake DA. Environ. Sci. Technol. 1998;32:137-142.

    Not in our library

  193. 193.

    Blake DA, Pavlov AR, Khosraviani M, Flowers GC. Current Protocols in Field Analytical Chemistry. 1998.

    Not in our library

  194. 194.

    Blake DA, Jones RM, Blake RC II, Pavlov AR, et al. Biosens. Bioelectron. 2001;16:799-809.

    Not in our library10.1016/s0956-5663(01)00223-8

  195. 195.

    Darwish IA, Blake DA. Anal. Chem. 2001;73:1889-1895.

    Not in our library10.1021/ac0012905

  196. 196.

    Darwish IA, Blake DA. Anal. Chem. 2002;74:52-58.

    Not in our library10.1021/ac010510r

  197. 197.

    Johnson DK. Comb. Chem. High Througput Screen. 2003;6:245-255.

    Not in our library10.2174/138620703106298293

  198. 198.

    Jones RM, Yu H, Delehanty JB, Blake DA. Bioconjug. Chem. 2002;13:408-415.

    Not in our library10.1021/bc0155418

  199. 199.

    Loor R, Lingenfelter C, Wason PP, Tang K, et al. Anal. Toxicol. 2002;26:267-273.

    Not in our library10.1093/jat/26.5.267

  200. 200.

    Shindelman J, Mahal J, Hemphill G, Pizzo P, et al. J. Anal. Toxicol. 1999;23:506-510.

    Not in our library10.1093/jat/23.6.506

  201. 201.

    Armbruster DA, Hubster EC, Kaufman MS, Ramon MK. Clin. Chem. 1995;41:92-98.

    Not in our library

  202. 202.

    Weide J, Luiting HJ, Veefkind AH. Ther. Drug Monit. 1993;15:344-348.

    Not in our library10.1097/00007691-199308000-00014

  203. 203.

    Meatherall RC, Faser AD. Ther. Drug Monit. 1998;20:673-675.

    Not in our library10.1097/00007691-199812000-00016

  204. 204.

    Meatherall RC, Fraser AD. J. Anal. Toxicol. 1998;22:270-273.

    Not in our library10.1093/jat/22.4.270

  205. 205.

    Fitzgerald RL, Herold DA. J. Anal. Toxicol. 1997;21:32-35.

    Not in our library10.1093/jat/21.1.32

  206. 206.

    Verstraete AG, Steyaert S. Anal. Toxicol. 1998;22:601-604.

    Not in our library10.1093/jat/22.7.601

  207. 207.

    Way BA, Walton KG, Koenig JW, Eveland BJ, et al. Clin. Chem. Acta. 1998;27:1-9.

    Not in our library10.1016/s0009-8981(97)00227-1

  208. 208.

    George C, George S, Parmer S. J. Anal. Toxicol. 2002;26:233-235.

    Not in our library10.1093/jat/26.4.233

  209. 209.

    Yang HH, Zhu QZ, Qu HY, Chen XL, et al. Anal. Biochem. 2002;308:71-76.

    Not in our library10.1016/s0003-2697(02)00222-1

  210. 210.

    Berecki A, Horvath V, Horvai G. J. Chromatogr. B: Biomed. Sci. Appl. 2000;749:215-223.

    Not in our library10.1016/s0378-4347(00)00409-6

  211. 211.

    Lovgren U, Kronkvist K, Backstrom B, Edholm LE, et al. J. Immunol. Methods. 1997;208:159-168.

    Not in our library10.1016/s0022-1759(97)00144-0

  212. 212.

    Yu H, Kusterbeck AW, Hale MJ, Ligler FS, et al. Biosens. Bioelectron. 1996;11:725-734.

    Not in our library10.1016/0956-5663(96)85923-9

  213. 213.

    Palmer DA, Edmonds TE, Seare NJ. Anal. Lett. 1993;26:1425-1439.

    Not in our library

  214. 214.

    Ghindilis AL, Atanasov P, Wilkins M, Wilkins E. Biosens. Bioelectron. 1998;13:113-131.

    Not in our library10.1016/s0956-5663(97)00031-6

  215. 215.

    Darwish IA. J. Pharm. Biomed. Anal. 2003;30:1539-1548.

    Not in our library10.1016/s0731-7085(02)00543-5

  216. 216.

    Burestedt E, Nistor C, Schagerlöf U. J. Emnéus, Anal. Chem. 2000;72:1707-1708.

    Not in our library10.1021/ac000128z

  217. 217.

    Zhang J, Heineman WR, Halsall HB. J. Pharm. Biomed. Anal. 1999;19:145-152.

    Not in our library10.1016/s0731-7085(98)00053-3

  218. 218.

    Taylor J, Picelli G, Harrison DJ. Electrophoresis. 2001;22:3699-3708.

    Not in our library10.1002/1522-2683(200109)22:17<3699::aid-elps3699>3.0.co;2-a

  219. 219.

    Choi J, Kim C, Choi MJ. Electrophoresis. 1998;19:2950-2955.

    Not in our library10.1002/elps.1150191626

  220. 220.

    Choi J, Kim C, Choi MJ. J. Chromatogr. B Biomed. Sci. Appl. 1998;705:277-282.

    Not in our library10.1016/s0378-4347(97)00527-6

  221. 221.

    Steinmann L, Thormann W. Electrophoresis. 1996;17:1348-1356.

    Not in our library10.1002/elps.1150170812

  222. 222.

    Ban E, Nam HS, Yoo YS. J. Chromatogr. A. 2001;924:337-344.

    Not in our library10.1016/s0021-9673(01)00829-9

  223. 223.

    Wey AB, Caslavska J, Thormann W. J. Chromatogr. A. 2000;895:133-146.

    Not in our library10.1016/s0021-9673(00)00636-1

  224. 224.

    Jia M, He Z, Jin W. J. Chromatogr. A. 2002;966:187-194.

    Not in our library10.1016/s0021-9673(02)00701-x

  225. 225.

    He Z, Gao N, Jin W. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 2003;784:343-350.

    Not in our library10.1016/s1570-0232(02)00823-1

  226. 226.

    Caslavska J, Allemann D, Thormann W. J. Chromatogr. A. 1999;838:197-211.

    Not in our library10.1016/s0021-9673(99)00115-6

  227. 227.

    Wan QH, Le XC. J. Chromatogr. B Biomed. Sci. Appl. 1999;734:31-38.

    Not in our library10.1016/s0378-4347(99)00327-8

  228. 228.

    Michkova M, Feng J, Sanchez-Baeza F, Marco MP, et al. Anal. Chem. 2003;75:83-90.

    Not in our library10.1021/ac025933n

  229. 229.

    Zeravik J, Ruzgas T, Franek M. Biosens. Bioelectron. 2003;18:1321-1327.

    Not in our library10.1016/s0956-5663(03)00076-9

  230. 230.

    Halamek J, Hepel M, Skladal P. Biosens. Bioelectron. 2001;16:253-260.

    Not in our library10.1016/s0956-5663(01)00132-4

  231. 231.

    Hatch A, Kamholz AE, Hawkins KR, Unson MS, et al. Nat. Biotechnol. 2001;19:461-465.

    Not in our library10.1038/88135

  232. 232.

    Sand TT, Zielinski JE, Arthur C, Bradley D, et al. Biosens. Bioelectron. 2003;18:797-804.

    Not in our library10.1016/s0956-5663(03)00048-4

  233. 233.

    Ikeda M, Toda K. Anal. Sci. 2003;19:155-158.

    Not in our library10.2116/analsci.19.155

  234. 234.

    Morgan CL, Newman DJ, Price CP. Clin. Chem. 1996;42:193-209.

    Not in our library

  235. 235.

    Stefan RI, Staden JF, Aboul-Enein HY. Fresenius J. Anal. Chem. 2000;366:659-668.

    Not in our library10.1007/s002160051560

  236. 236.

    Stefan RI, Staden JF, Aboul-Enein HY. Comb. Chem. High Throughput Screen. 2000;3:445-454.

    Not in our library10.2174/1386207003331382

  237. 237.

    Luppa PB, Skoll LJ, Chan DW. Clin. Chim. Acta. 2001;314:1-26.

    Not in our library10.1016/s0009-8981(01)00629-5

  238. 238.

    Lu B, Iwuoha EI, Smyth MR, ÓKennedy R. Anal. Chim. Acta. 1997;345:59-66.

    Not in our library

  239. 239.

    Suleiman AA, Xu YH. Electroanal. 1998;10:240-243.

    Not in our library

  240. 240.

    Xu YH, Suleiman AA. Anal. Lett. 1997;30:2675-2689.

    Not in our library

  241. 241.

    Ivnitski D, Rishpon J. Biosens. Bioelectron. 1996;11:409-417.

    Not in our library10.1016/0956-5663(96)82736-9

  242. 242.

    Attili BS, Suleiman AA. Microchem. J. 1996;54:174-179.

    Not in our library

  243. 243.

    Attili BS, Suleiman AA. Anal. Lett. 1995;28:2149-215.

    Not in our library10.1080/00032719508000035

  244. 244.

    Hanbury CM, Miller WG, Harris RB. Biosens. Bioelectron. 1996;11:1129-1138.

    Not in our library10.1016/0956-5663(96)82335-9

  245. 245.

    Yu H, Kusterback AW, Hale MJ, Ligler FS, et al. Biosens. Bioelectron. 1996;11:725-734.

    Not in our library10.1016/0956-5663(96)85923-9

  246. 246.

    Meyer UJ, Zhi Z, Loomans E, Spener E, et al. Analyst. 1999;124:1605-1610.

    Not in our library10.1039/a907121k

  247. 247.

    Sheikh SH, Mulchandani A. Biosens. Bioelectron. 2001;16:647-652.

    Not in our library10.1016/s0956-5663(01)00193-2

  248. 248.

    Pemberton RM, Hart JP, Ottram TT. Biosens. Bioelectron. 2001;16:715-723.

    Not in our library10.1016/s0956-5663(01)00212-3

  249. 249.

    Stefan RL, Aboul-Enein HY. J. Immunoassay Immunochem. 2002;23:429-437.

    Not in our library10.1081/ias-120015474

  250. 250.

    Schneider BH, Dickinson EL, Vach MD, Hoijer JV, et al. Biosens. Bioelectron. 2000;15:13-22.

    Not in our library10.1016/s0956-5663(00)00056-7