Diagnostic Reference Levels (DRLs) and Image Quality Evaluation for Digital Mammography in a Nigerian Facility

A. E. Anasthesia, U. Ibrahim, S. D. Yusuf and 8 more

Journal of Nigerian Society of Physical Sciences · 2022

doi:10.46481/jnsps.2022.734 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Diagnostic Reference Levels (DRLs) for digital mammography and image quality evaluation are important optimization tools in medical imaging. High quality mammograms are essential to the successful early detection of breast cancer. The objective of the study is to establish DRLs for digital mammography and to assess image quality of the mammograms for optimization. DRLs were established using thermoluminescent dosimeter (TLD) chips to estimate the mean glandular dose for both cranio-caudal and medio-lateral oblique projections. The TLD chips were calibrated. The DRLs were set at the 75th percentile of the distribution of the median value of mean glandular dose. Image quality was assessed using European Commission guideline for mammographic image quality assessment. Results for DRLs were 0.53 mGy for cranio-caudal and also 0.53 mGy for medio-lateral oblique. Image quality evaluation showed criteria scores for cranio-caudal and medio-lateral oblique projections as 76 % and 61.2 % respectively. The mammograms scored the highest and lowest score of 100 % and 44 % on criteria 2 and criteria 6 (absence of skin fold) respectively for cranio-caudal projections while for the mediolateral oblique projections, criteria 1 (all breast tissue clearly shown) and criteria 5 (inframammary angle clearly demonstrated) have the highest and lowest score of 96 % and 8 % respectively. The study showed that the DRLs in this study was lower than the established values in other regions of Nigeria and international established values. Image quality was within acceptable level. DRLs for digital mammography and image quality evaluation are important optimization tool that should be adopted by every radiology department with mammography unit.

Abstract licence: CC BY

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.