Skip to content
naulon
Not checkedNot checked for notices yet.

Investigation and Analysis of Hemoglobin A1c Measurement Systems' Performance for 135 Laboratories in China

Hai-Jian Zhao, Tian-Jiao Zhang, Jie Zeng and 3 more

Chinese Medical Journal · 2017

doi:10.4103/0366-6999.204932 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Background: Hemoglobin A1c (HbA1c) measurement is of great value for the diagnosis and monitoring of diabetes. Many manufacturers have developed various experiments to determine the HbA1c concentration. However, the longitudinal use of these tests requires strict quality management. This study aimed to analyze the quality of HbA1c measurement systems in China using six sigma techniques to help improve their performances. Methods: A total of 135 laboratories were involved in this investigation in 2015. Bias values and coefficients of variation were collected from an HbA1c trueness verification external quality assessment program and an internal quality control program organized by the National Center of Clinical Laboratories in China. The sigma (σ) values and the quality goal index (QGI) were used to evaluate the performances of different groups, which were divided according to principles and instruments. Results: The majority of participants (88, 65.2%) were scored as “improvement needed (σ < 3)”, suggesting that the laboratories needed to improve their measurement performance. Only 8.2% (11/135) of the laboratories were scored as “world class (σ ≥ 6)”. Among all the 88 laboratories whose σ values were below 3, 52 (59.1%) and 23 (26.1%) laboratories needed to improve measurement precision (QGI <8.0) and trueness (QGI >1.2), respectively; the remaining laboratories (13, 14.8%) needed to improve both measurement precision and trueness. In addition, 16.1% (5/31) and 15.0% (3/20) of the laboratories in “TOSOH” and “ARKRAY” groups, respectively, were scored as “world class”, whereas none of the laboratories in “BIO-RAD” group were scored as “world class”. Conclusions: This study indicated that, although participating laboratories were laboratories with better performance in China, the performances were still unsatisfactory. Actions should be taken to improve HbA1c measurement performance before we can include HbA1c assays in diabetes diagnosis in China.

Abstract licence: CC BY-NC-ND

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.