A case of delayed cardiac perforation of active ventricular lead
Hangyuan Guo, Yangbo Xing, Fukang Xu
Clinics and Practice · 2011
doi:10.4081/cp.2011.e104 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
A 65-year-old man was admitted as for one month of repetitive dizziness and one episode of syncope. Electrocardiogram showed sinus bradycardia and his Holter monitoring also showed sinus bradycardia with sinus arrest, sino-atrial block and a longest pause of 4.3 s. Then sick sinus syndrome and Adam-Stokes syndrome were diagnosed. Then a dual chamber pacemaker (Medtronic SDR303) was implanted and the parameters were normal by detection. The patient was discharged 1 week later with suture removed. Then 1.5 month late the patient was presented to hospital once again for sudden onset of chest pain with exacerbation after taking deep breath. Pacemaker programming showed both pacing and sensing abnormality with threshold of?5.0V and resistance of 1200?. Lead perforation was revealed by chest X-ray and confirmed by echocardiogram. Considering the fact that there was high risk to remove ventricular lead, spiral tip of previous ventricular lead was withdrew followed by implantation of a new ventricular active lead to the septum. Previous ventricular lead was maintained. As we know that the complications of lead perforation in the clinic was rare. Here we discuss the clinical management and the possible reasons for cardiac perforation of active ventricular lead.
Abstract licence: CC BY
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.