Optoacoustic Imaging and Tomography: Reconstruction Approaches and Outstanding Challenges in Image Performance and Quantification

Christian Lutzweiler, Daniel Razansky

Sensors · 2013 · Review · CC BY

doi:10.3390/s130607345

Abstract

This paper comprehensively reviews the emerging topic of optoacoustic imaging from the image reconstruction and quantification perspective. Optoacoustic imaging combines highly attractive features, including rich contrast and high versatility in sensing diverse biological targets, excellent spatial resolution not compromised by light scattering, and relatively low cost of implementation. Yet, living objects present a complex target for optoacoustic imaging due to the presence of a highly heterogeneous tissue background in the form of strong spatial variations of scattering and absorption. Extracting quantified information on the actual distribution of tissue chromophores and other biomarkers constitutes therefore a challenging problem. Image quantification is further compromised by some frequently-used approximated inversion formulae. In this review, the currently available optoacoustic image reconstruction and quantification approaches are assessed, including back-projection and model-based inversion algorithms, sparse signal representation, wavelet-based approaches, methods for reduction of acoustic artifacts as well as multi-spectral methods for visualization of tissue bio-markers. Applicability of the different methodologies is further analyzed in the context of real-life performance in small animal and clinical in-vivo imaging scenarios.

Full text

From PubMed Central, U.S. National Library of Medicine, as current on Oct 27, 2026. Paragraph numbers are ours, so an answer can point at one.

Introduction

Nowadays, the terms optoacoustic and photoacoustic are equally used to describe the effect of acoustic wave generation by transient light absorption. Optoacoustic sensing and imaging draws its roots from the discovery of the photophone by Bell and his assistant Tainter in 1880, which the inventors used as the first practical wireless telephony or, in fact, optical communication device. Bell also named the effect the photophonic phenomenon and later suggested the so-called spectrophone. Following its discovery in solid and liquid media, the effect had been further confirmed by Tyndall and Rontgen to occur in gases. However, despite multiple early attempts to put the photophone into practical use, mainly for military communications, it was not until 1938 that Veingerov developed the first widely accepted implementation of the phenomenon, which he called method for gas analysis based on Tyndall-Rontgen optic-acoustic effect. The term photoacoustic was later adopted in 1970s when sensitive spectroscopy of solids was first demonstrated. Rosencwaig was the first to suggest the phenomenon for biological sensing. However, due to the lack of appropriate laser sources and ultrasound detection technologies, theoretical understanding of the underlying phenomena as well as backlogs with the algorithmic and data processing capacities, utilization of optoacoustics for imaging of real biological tissues and organisms has only evolved in the last decade.

The field of biological optoacoustic imaging is developing tremendously with new technical approaches and applications continuously emerging. By combining now commercially available pulsed laser technology in the nano-second range and sensitive acoustic detectors, it was shown possible to generate opto-acoustic responses from tissue that carry significant spatially-resolved biochemical information. Due to its hybrid nature, i.e., optical excitation and acoustic detection, the optoacoustic imaging technology benefits both from the rich and versatile optical contrast and high (diffraction-limited) spatial resolution associated with low-scattering nature of ultrasonic wave propagation as compared to photon propagation.

In vascularized tissues, highly absorbing hemoglobin manifests good optoacoustic contrast. It is therefore natural for optoacoustics to attain high fidelity images of vascular anatomy, dynamic microcirculation, tumor neovascularization, as well as blood oxygenation levels deep within highly diffuse tissues without introduction of contrast agents. Besides blood-related contrast, optoacoustics is sensitive to other intrinsic tissue contrast. Nevertheless, imaging of extrinsic photo-absorbing agents may offer important contrast advantage and imaging specificity but would typically require differentiation of these agents on top of spectrally varying background tissue absorption. In response, the so-called multi-spectral optoacoustic tomography (MSOT) technique relies on the spectral identification of chromophoric molecules and particles distributed in tissue over background tissue absorption. Pulses of different wavelengths are used, in a time-shared fashion, whereas the wavelengths are selected to sample a certain spectral characteristic in the absorption spectrum of intrinsic bio-markers and reporter agents of interest. Essentially, when operated at multiple wavelengths, optoacoustics is capable of resolving spectral signatures of chromophores in three dimensions (volumetrically), i.e., it can sense color, or more precisely, listen to color.

Despite its great promise, like ultrasound, optoacoustic imaging possesses several limitations related to imaging through acoustically mismatched areas, such as lungs and bones. Furthermore, even though spatial resolution performance here does not directly suffer from light scattering in tissue, it is still affected by penetration limits and signal degradation due to light attenuation. Therefore, whole-body imaging with optoacoustics is currently only possible in small animals whereas clinical applications are limited to relatively superficial, low attenuating or otherwise conveniently accessible areas of the body. Optoacoustic imaging is indeed a rapidly evolving area in biomedical imaging sciences. Its in-vivo use puts forth a number of challenging problems demanding intensive investigations, from imaging instrumentation, quantified reconstruction algorithms, spectral processing schemes, detection sensitivity, and other technical issues, to biology-related topics, such as effectiveness of imaging contrast approaches or animal handling. We refer an interested reader to some of the recently published review articles summarizing on progress of optoacoustic imaging applications. On the technical side, while some focused reviews address the mathematical inverse acoustic problem and quantification challenges of multispectral optoacoustic reconstruction methods, the current paper attempts to comprehensively cover the most recent experimentally-driven algorithmic developments in the optoacoustic field and also review some newly introduced image acquisition methodologies and selected practical in vivo imaging approaches.

The Optoacoustic (OA) Effect

Optoacoustic (OA) imaging is based on absorption of light radiation in tissue and conversion of the deposited energy into heat (Figure 1(a)). This results in an expansion of the tissue and mechanical stress which propagates in the form of pressure waves. Typically, light in the visible or near-infrared spectrum (400 nm–1,200 nm) is used for excitation of OA responses due to the relatively weak absorption of biological tissues in this spectral region, also known as the optical window. Alternatively, tissue can be excited with energy in the radiofrequency and microwave spectra, which is however not included in the scope of this paper. A large variety of chromophores absorb at the optical wavelengths of interest, which further leads to a high contrast between different tissues with varying chromophore concentration. For biomedical applications, examples of usable contrast include intrinsic chromophores (e.g., hemoglobin in its oxygenated and deoxygenated form, melanin, fat), extrinsically administered agents (e.g., nanoparticles, fluorophores) or genetically encoded markers, such as fluorescent proteins. Vast majority of the modern OA imaging setups utilize lasers that emit ultra-short pulses, mainly due to a better signal to noise performance and parallelization capacity. In this paper, we therefore limit our discussion to the theory and applications of pulsed OA sources, although the interested reader is indeed encouraged to explore more about OA imaging using modulated continuous-wave sources.

To simplify the model of the induced pressure variations, two assumptions are usually made on the duration x of the laser pulse: the first is the so-called thermal confinement assuming that the heat conductance during a laser pulse in one image voxel does not affect the neighboring voxels (τ ≪ Δl/4Dt, where Δl is the dimension of the voxels and Dt is the thermal diffusivity). The second is the so-called stress confinement assuming the volume expansion during the laser pulse to be negligible (τ ≪ Δl/c, where c is the speed of sound). With these simplifications, the initially induced pressure p0(r) is generally proportional to the total absorbed optical energy density H(r): where Γ is the dimensionless Grüneisen parameter, β is the isobaric thermal expansion coefficient, and Cp is the isobaric specific heat capacity. In a more general sense, Γ represents the combined OA efficiency of conversion from heat into pressure. In contrast to significant spatial variations in the absorption coefficient, Γ does not normally exhibit dramatic variations among different soft tissues, thus it is most conveniently assumed to be constant. Nevertheless, accounting for spatial variations in Γ may still benefit the efforts on quantification of optoacoustic images. Moreover, some studies attempted utilizing the strong dependence of the Griineisen parameter on temperature in order to sense temperature optoacoustically.

For biomedical applications, it is the distribution of the absorption coefficient μa(r) that is usually directly related to the concentration ci(r) of the chromophores rather than the totally absorbed energy H(r). Both are related via the expression: where Φ(r,μa,μs,g) is the light fluence (or photon density), μs is the scattering coefficient and g is the anisotropy factor. Although it might seem that H(r) is a simple product, it depends nonlinearly on the absorption coefficient μa because the light fluence generally depends on the underlying optical properties, among them μa itself. Thus, it is important for image quantification purposes to account for the light fluence since it may considerably vary as a function of depth. Figure 1(b) summarizes the main contributions to the OA signal generation, which need to be accounted for when developing accurate image reconstruction algorithms.

The Optoacoustic Wave Equation

The initial goal of optoacoustic imaging is to retrieve the initial pressure distribution inside the object due to the absorbed laser energy. However, the generated pressure fields can normally only be measured outside the object. Propagation of pressure toward the detection point is described by the optoacoustic wave equation, which, for a non-absorbing homogeneous medium, is written as:

For tomographic data acquisition, the detectors are placed on a (closed) surface S surrounding the volume of interest. The most common detection patterns are a spherical, a cylindrical and a planar surface or their 2-D counterparts (Figure 2(a)). On the surface the temporal pressure profile pd(rd, t), rd ∈ S, t > 0, is considered known. To reconstruct an OA image from pd(rd, t), the initial pressure p0(r) satisfying: has to be found. To solve this partial differential equation, one usually takes advantage of the free-space Green's function G(r, r′, t, t′). It relates the pressure at two different spatio-temporal points, i.e.,

Consequently, the pressure profile pd(rd, t) on the detection surface S can be directly related to the initial pressure p0(r) via a Poisson-type integral:

This type of equation is known as spherical Radon transform or spherical mean transform. This is because contributions of all sources that are located on the surface of a sphere with radius R = c · t = |rd − r| from the detector are integrated (Figure 2(b)).

Imaging Instrumentation

A typical optoacoustic setup consists of several key components, an example shown in Figure 1(c). In the pulsed excitation mode, the tissue is illuminated by a laser emitting monochromatic pulses of light with a typical duration of some nanoseconds. For deep tissue imaging applications, optical parametric oscillators are often used to provide a tunable wavelength in the spectrum of interest with pulse repetition rate in the order of a few tens of Hertz and per-pulse energies in the millijoule range. In optoacoustic microscopy and other superficial applications, where such high per-pulse energies are not required, other types of sources in the microjoule and nanojoule range are considered as well, including high repetition passive Q-switched and dye lasers, laser diodes and fiber lasers. For tomographic imaging, the optoacoustically-generated pressure profiles are subsequently captured with detectors surrounding the object whereas the acoustic coupling between object and detector is usually ensured by water or gel medium. In contrast to ultrasonic (US) imaging, OA signal amplitudes are relatively low while their spectral content is broad, spanning the frequencies from several tens of kHz up to a hundred MHz for micron-scale structures. Thus, high sensitivity, ultrawide bandwidth as well as good tomographic coverage are key for ensuring image quality. In addition, for acquisition of spatially resolved data with any type of detector, either a single detector is scanned around the object or, alternatively, multiple detection elements acquire data in parallel. The latter allows fast data acquisition, e.g., for reconstruction of images from a single laser shot with commercially available or tailored detection arrays. Figure 3 illustrates the broad variety of OA setups presented in the literature including OA raster scanning (Figure 3(a,c)), tomography (Figure 3(b,e,f)) and endoscopy systems Figure 3(d). The signals were detected optically with an interferometer for the setups in Figure 3(c,e), whereas in the other systems conventional piezoelectric detection elements were used.

Two main categories of sensors are used to detect OA signals: The most commonly used type is based on piezoelectric elements, such as piezocomposite- or polymer-based films, which are also widely utilized in US imaging. This well developed technology possesses relatively low cost of implementation and can be readily applied for highly sensitive measurements. However their sensitivity scales with detector's size, making miniaturization challenging. Of the second type are optical resonators and interferometry detectors like microring resonators, Fabry-Perot interferometers, Mach-Zehnder interferometers, or fiber Bragg gratings, which are sensitive to changes in the length of the optical path induced by pressure waves. Interferometric detectors usually possess ultrawideband detection characteristics and further allow all-optical optoacoustic imaging, in some cases also greatly simplifying delivery of the excitation light to the imaged object through transparent optical components. Sensitivity of interferometric detectors is not directly dependent upon the physical size of the detection element, which simplifies miniaturized designs suitable for small scale imaging, such as intravascular or endoscopic applications. On the other hand, interferometric detection approaches are usually difficult to parallelize in order to achieve faster tomography while it is also not always possible to increase detection sensitivity by integration of signals over large areas. Recently, other detection technologies have been introduced and successfully applied for optoacoustic imaging, such as capacitive micromachined ultrasonic transducers (CMUT). The advantages and disadvantages of the various optoacoustic sensor approaches are summarized in Table 1.

Following their detection, the optoacoustic waveforms are pre-amplified and digitized by a data acquisition system (DAQ). For tomographic data acquisition, a motion stage can be translated or rotated in order to alter the position of the detector/s with respect to the imaged object or vice versa in order to increase the amount of the available data points (projections) around the imaged area. A sufficiently high number of projections and a suitable detection pattern are crucial for good image quality. A PC controls the acquisition process, stores the data and is used to reconstruct and render the images. Prior to image formation, an additional signal processing step normally takes place that may include some of the following: averaging over multiple pulses to enhance signal-to-noise ratio (SNR), denoising the signals, deconvolving the signals from electrical frequency response of the detector, band pass filtering to remove noise and/or enhance visibility of certain spatial frequency components in the images.

Inversion of the Optoacoustic Wave Equation

The first step to obtain OA images consists in reconstructing the initial pressure distribution, i.e., solving Equation (4) or inverting Equation (6). Since the early reconstruction approaches were introduced, plenty of research has been done on the problem in the last decade, driven from both theoretical interest and the need to improve images acquired from specific experimental setups.

Types of Inversion Algorithms

The inversion schemes are usually based on idealized assumptions of infinitely small point detectors located on a closed measurement surface and a non-absorbing homogeneous acoustic medium. Modifications for deviations from these idealized assumptions will be further presented below.

An inversion formula was initially presented by Norton et al. for ultrasonic imaging. The basic idea is to represent the solution p0(r) as a convergent series: where Ψk(r) are eigenfuctions of the negative Laplacian −Δ in the domain of the measurement geometry and αk are the so-called Fourier coefficients. The coefficients αk can be calculated from pd(rd, t). Although it has been shown that series solutions exist for arbitrary closed detection surfaces, only the geometries with known eigenfunctions (sphere, cylinder, plane) are usually of practical experimental interest.

For a planar detection geometry, this approach results in a fairly simple reconstruction algorithm. First, the signals pd(x, y, t) are transformed to Fourier space in x, y and t coordinates, resulting in pd(kx, ky, ω). Second, the temporal component is mapped to its corresponding spatial coordinate with the dispersion relation, resulting in p0(kx, ky, kz). Finally, the reconstruction is obtained by applying the inverse Fourier transform in all three spatial coordinates. In this way, fast reconstructions of 3D volumes can be obtained using an FFT algorithm. A similar FFT-based approach has been presented by Kunyansky for a finite cube used as the detection surface instead of an infinite planar surface which has to be truncated in practical cases. For a spherical measurement geometry, Wang et al. have recently derived a simple FFT-based algorithm in the frequency domain.

Another approach is based on closed-form analytical filtered back-projection (BP) formulae in the time domain. This type of formula is commonly used in computed x-ray tomography to invert the classical Radon transform. Similar to Equation (7), the initial pressure is obtained by integrating the detected signals over spherical surfaces. Prior to or following the integration, a filter function is applied. The first exact back-projection formulae were obtained by Finch et al. for a sphere in 3-D and later also in 2-D. Xu et al. presented an approximate BP formula for a spherical detection geometry, which was later generalized to the widely used universal back-projection (UBP) formula:

The latter scheme can be efficiently applied to spherical, cylindrical or planar (with 4π replaced by 2π) geometries. In the far-field approximation with detectors located far from the object, only the derivative term in Equation (8) contributes significantly.

Time reversal inverse methods are based instead on the Huygens' principle: In three dimensions, the waves from an initial pressure distribution p0(r), confined inside a finite volume V, will leave the volume after a finite duration T while the pressure p(r, t) inside the volume will vanish for t > T. Although Huygens' Principle does not hold in two dimensions, the pressure inside the volume usually decays fast enough to result in good reconstructions for large T. Time reversal methods aim in solving Equation (4) backwards in time: Starting from zero initial condition at time t = T, the pressure distribution is propagated backwards in time step-by-step, reemitting the pressure profile pd(rd, t) on the detection surface for each time point (Figure 2(c)). Arriving at t = 0, the pressure distribution p(r, t) equals the sought-after initial pressure p0(r). Time reversal methods can be implemented by finite differences methods or by k-space methods, which accelerate inversion by using larger time steps, and furthermore offer greater flexibility as they generally work for arbitrary closed measurement surfaces.

Whereas series solution or the back-projection formulae perform an approximated inversion of Equation (6) analytically, the so-called model-based (MB) algorithms seek instead for an accurate numerical solution. Rosenthal et al. presented an algorithm that is based on a discretized semi-analytical forward model M̃ of the wave propagation that linearly maps the discretized initial pressure p̃0 to the corresponding discretized signals p̃d. Numerical models can also be obtained by finite element (FE) calculations of Equation (4), which can be implemented in the time or frequency domain. The image is then reconstructed by finding a solution that minimizes the difference between the measured data p̃d and the pressure corresponding to the forward-modeled image p̃0:

In Equation (9), P is an optional regularization term, such as Tikhonov or total variation regularization, which is intended for stabilization of the numerical inversion. The solution can be calculated via the Moore-Penrose pseudo-inverse M̃†. Although calculation of M̃† may become burdensome in terms of CPU time and memory consumption, it needs to be calculated only once for a given experimental system. Then the reconstruction process reduces to a simple and fast matrix-vector multiplication:

Alternatively, the solutions can be calculated iteratively, for example with the LSQR algorithm that takes advantage of the sparse nature of the model matrix. Benefiting from the advances in computation technology, MB algorithms have drawn considerable interest in the recent years because of their flexibility to account for a broad variety of experimental imperfections in the model. However, their drawbacks are often associated with longer calculation times or a considerable consumption of memory needed to store the model (or its pseudo-inverse), especially when considering the full three dimensional problem.

Computational Efficiency

Besides image quality and technical feasibility, computational efficiency is yet another important criterion determining applicability of the different methodologies, especially when dealing with large 3D datasets or in applications involving real-time rendering of images. For instance, different approaches can be characterized by their algorithmic complexity in reconstructing 3-D datasets-O(n4) for series solutions, O(n5) for BP algorithms, O(n4) for time reversal methods and O(n9) for calculating the pseudo-inverse in MB algorithms. The series solutions can result in particularly fast algorithms whenever fast methods for summation of the eigenfunctions are applicable (e.g., FFT), resulting in low complexity in the order of O(n3 · logn). In addition to the algorithmic complexity, additional implementation aspects include the different memory consumption requirements between algorithms that only involve standard linear algebra operations for inversion, such as BP, versus the extensive memory size needed to store matrices for some MB schemes. The graphic processing units (GPUs), which are increasingly used in medical imaging in the recent years due to their vast computational power, are expected to mitigate or significantly reduce complications concerned with computational complexities in optoacoustic imaging.

As mentioned above, MB algorithms offer better reconstruction accuracy and flexibility in accounting various experimental imperfections; however they demand large computational resources. Consequently, methods to overcome this problem have been presented. Wang et al. proposed a method that circumvents storing the large model matrix by approximating voxels in the image by means of spheres, whose corresponding pressure profiles are known analytically. For the GPU implementation of their algorithm they reported a runtime of 313 min per iteration for a 3-D high resolution data set, whereas the corresponding BP reconstruction took only 46 s. Rosenthal et al. developed an efficient framework for model-based inversions based on wavelet packets. In this case, both the image and the signals are decomposed into a two-level wavelet-packet basis. The model matrix is approximately block-diagonal in this basis and thus an approximate inverse matrix can be calculated (and stored) easily. From the obtained inverse, the image is directly calculated in a more efficient and faster manner and can be then further improved iteratively. The authors reported a runtime of 0.1 s for a 2-D data set by direct reconstruction from the approximate inverse, which constituted a 40-fold increase in the reconstruction speed as compared to the regular iterative MB algorithm.

Reconstructions with Finite-Sized Detection Elements

Due to mathematical and numerical simplicity, most inversion schemes assume infinitely small point detectors. Conversely, in experimental reality, larger detector sizes are often preferable as they lead to a better detection SNR, which in many cases scales with detector size. Furthermore, large area detectors are also required for focused detection geometries. For instance, if the object is placed in the focal zone of a cylindrically focused detector (Figure 4(a)), the signals acquired by the detector originate only from a small region close to the imaging plane. The reconstruction is thus effectively reduced to a 2-D problem. So data acquisition and image rendering can be performed faster or even in real-time. OA microscopy setups often use spherically focused detectors which are in first approximation only collecting signals from a line, in which case images can be formed without complicated reconstruction algorithms. However, simplified assumptions readily lead to out-of-plane (out-of-focus) artifacts in the images, loss of quantification, highly anisotropic resolution and degradation of the overall image quality.

Li et al. introduced the virtual detector concept to OA in order to improve the reconstructions from focused detectors outside the focal point (Figure 4(b)). The detected signals are assumed to originate from a virtual point detector located in the focal spot, delayed by the time corresponding to the focal distance. These signals can then be used to form an image, e.g., by a simple delay-and-sum algorithm for OA microscopy or a back-projection algorithm for tomographic reconstructions.

A very straightforward reconstruction approach can be applied for flat detectors with dimensions much larger than the object (in contrary to point detectors with dimensions much smaller than the object). The so-called integrating line detectors (Figure 4(c)) simultaneously record signals along a line thus effectively the pressure field of the object integrated along a certain axis. A complete tomographic data set is subsequently obtained by rotating the detectors' orientation. The reconstruction process is then performed in two steps: First, the integrated 2-D images are reconstructed for a given detector orientation with an arbitrary 2-D method. In a second step, the 3-D reconstruction is obtained from the integrated 2-D reconstructions for all detector orientations with the inverse Radon transform. Alternatively to line detectors, detection elements, that are integrating in two dimensions or integrating in one and focused in the second dimension, have been suggested as well.

An interesting approach, which is also based on detection of integrated pressure fields, was presented by Nuster et al.. Instead of detecting a time-resolved integrated field along a certain line, the approach consists of recording the full field at a given time instant td using a CCD camera. In this way, the integrated acoustic pressure field is detected in a direction perpendicular to the imaging plane. td is chosen in such a way that essential parts of the waves have already left the reconstruction domain. A volumetric image can be then reconstructed from a few simple steps: (1) The 2-D pressure field is integrated (Radon transform) along a fixed direction in the imaging plane. (2) The integrated 1-D signals are back-propagated in time from t = td to t = 0 using D'Alembert's formula, which results an integrated 1-D image. (3) An integrated 2-D image is obtained by the inverse Radon transform from the integrated 1-D images for all in-plane directions. (4) A final 3-D image is reconstructed by the inverse Radon transform from the 2-D images for all orientations of the imaging plane.

Rosenthal et al. presented a generalized method that allows accounting for arbitrary-shape detectors. It is based on including the effects of the exact detector shape in the forward model of a model-based inversion approach by temporal convolution. As a result, image artifacts related to finite detector size can be readily eliminated using this universal approach, while also improving spatial resolution of the images.

Effects of Heterogeneous Acoustic Properties

In real tissues, acoustic properties may significantly deviate from the idealized homogeneous properties initially assumed in the governing equations, which may lead to deterioration of the image quality. Indeed, speed of sound c may vary considerably for different types of tissue within a range of 1,400–1,600 m/s and be also spatially dependent (c = c(r)). Information on the speed of sound within the object can be obtained using various approaches, e.g., from a priori knowledge on the object, ultrasound or hybrid measurements, auto-focusing approaches or can be reconstructed from the signals by a combined FE approach. Although simplified series solutions exist in principle for optoacoustic reconstructions with variable speed of sound, iterative and time reversal methods were further shown efficient for this purpose.

An additional effect taking place is attenuation of the optoacoustic waves as they propagate through the tissue. Often these effects are also dispersive, i.e., acoustic attenuation increases with frequency, which is usually modeled by exponential attenuation with a power-law of ωn, 1 ≤ n ≤ 2. The penetration depth from which acoustic waves can be detected also limits the imaging depth. Since the maximum OA resolution is determined by the frequency content of the signals, the resolution thus scales with depth. Time-gain compensation applied in US imaging cannot be used for OA signals because they are generally broadband. Instead, the attenuation is accounted for by adding a dissipation term in the time domain for time reversal methods or by a complex frequency-dependent dispersion relation in an iterative approach.

In media with heterogeneous acoustic properties, scattering and reflection of the acoustic waves may occur at boundaries of highly mismatching media. Examples include bones or air cavities such as the lung. Wang et al. studied the effects of reflection at soft and hard boundaries and were able to considerably improve image quality in the presence of reflections at planar boundaries, which might however not always be applicable to realistic imaging scenarios. Anastasio et al. showed that tomographic OA signals, acquired from multiple view angles (projections), contain complementary information thus images can be equally reconstructed from truncated signals, which may improve image quality in heterogeneous acoustic media due to data redundancy principles. Dean-Ben et al. presented a statistically weighted BP algorithm that efficiently accounted for strong acoustic reflections. The algorithm weights each detected signal by a factor that represents the probability of the signal being undisturbed during propagation. In this way, reconstruction artifacts arising from scattering can be suppressed (Figure 5).

Tomographic Reconstructions in Limited View Geometries

Ideal tomographic imaging scenarios assume that the object is fully surrounded by detectors on a closed surface, creating the so-called full-view scenario. Full view detection is favorable for the reconstruction process but cannot always be realized in experimental conditions, e.g., because the imaged region is simply not accessible for efficient detection from all directions or cannot be fully immersed in water for efficient acoustic coupling to the detectors. Such scenarios are known as limited view reconstructions and the inversion has to be performed with an incomplete set of data.

Xu et al. investigated the effects of limited view in an OAT setup with detectors on a truncated circle (Figure 6(a)). It has been found that regions enclosed by the detection surface can be recovered stably. Sharp boundaries of the object are thus visible in the reconstructions if they are facing the detection surface but turn invisible if they are perpendicular to the detection surface.

Buehler et al. showed that limited view reconstructions suffer from stripe artifacts (Figure 6(b)). The artifacts were reduced using a MB approach with a regularization term that suppresses ripples in the direction of the artifacts. Alternatively, total variation regularization terms can be applied in limited view scenarios to stabilize the inversion and reduce the artifacts. Closed form reconstruction formulae can be also optimized for limited view scenarios. Paltauf et al. presented a BP approach that weights the signals with a smooth function before back-projecting. The weighting function needs to consider whether complementary data from an opposite direction are available or these projections are missing. This approach may reduce artifacts resulting from the limited view while also preserving the simple BP implementation. Kunyansky presented an approach for non-closed detection surfaces based on single-layer potentials which, however, necessitates a non-trivial step of finding and calculating such potential. Haltmeier et al. presented an efficient algorithm for a truncated planar detection geometry based on a nonuniform FFT that reduces the limited view artifacts while keeping computational efficiency of the fast Fourier methods.

Compressed Sensing Reconstruction Approaches

Recent reconstruction algorithms for OA have been also driven by the rapidly growing field of compressed sensing (CS), which had been so far successfully applied to biomedical imaging in the field of MRI imaging. Provost et al. were the first to suggest a CS scheme in OA imaging, followed by contributions from other groups. The basic idea of CS is that images p̃0 can be represented by only a few large coefficients in a suitable basis instead of many small coefficients. CS exploits sparsity of the images based on a certain model M̃ in that basis. Instead of minimizing the residual for an image in L2-norm (‖·‖2, Euclidian norm) only, CS also enforces the sparsity of the retrieved image by concurrently minimizing the L1-norm (‖·‖1, Taxicab norm) of the image. Thus, instead of Equation (9), one has to minimize: where Ψp̃0 is the representation of the image in the CS basis and λ is a scalar regularization parameter. Equation (11) can be solved by nonlinear minimization methods, which typically require longer calculation times as compared to just solving Equation (9). In comparison to standard image reconstruction methods, CS approaches have further shown reduction of artifacts in images when reconstructing from only a few tomographic projections (Figure 7). Thus, by reducing the number of acquired projections, these approaches hold great promise for dramatic reduction of image acquisition and reconstruction times as well as instrumentation costs.

Accounting for the Transport of Light

The last section has been entirely devoted to the acoustic part of the optoacoustic phenomena, i.e., reconstruction of the initial pressure distribution and the deposited heat H(r) in the imaged object from the detected signals pd(rd, t). Nevertheless, understanding processes behind the optical counterpart governing propagation of the excitation light is equally important, mainly for two reasons. First, since photons are absorbed as they propagate in tissue, the depth at which a sufficient number of photons is left to excite a detectable signal limits the effective penetration depth of the imaging modality. Second, as stated earlier, the eventual goal of optoacoustic tomography is retrieving distribution of the optical absorption coefficient μa(r) in the imaged object, which can be further related to distribution of the various chromophores and bio-markers in tissues. But the inversion process, described in the previous section, only retrieves maps of the deposited energy H(r), not μa(r). Both distributions are related by the light fluence Φ(r, μa, μs, g) (see Equation (2)), which generally depends on the optical properties of the medium. In order to accurately quantify the values of the optical absorption coefficients and the corresponding concentrations of tissue chromophores, one needs to correct for the light fluence, which may vary considerably versus imaging depth. The reader is referred to a review article on that issue by Cox et al., which has been published recently and covers the subject of image quantification in more detail.

Models of Light Transport

Several approaches exist to describe transport of light in diffusive tissues, which account for both absorption-which is mainly responsible for the contrast-and scattering processes. For visible and near-infrared light propagating in biological tissues, scattering can be typically considered as the dominant process. As depicted in Figure 8(a), one distinguishes between two main regimes of light transport: ballistic photons that have not been scattered and diffuse photons that have undergone multiple scattering events before being absorbed. At depths relevant for deep tissue OA tomography, typically between several millimeters to several centimeters, diffuse photons are dominant.

One frequently applied approach to model the transport of diffuse light is the Monte Carlo (MC) method, which simulates a random walk of a great number of photons through the scattering tissue. The ensemble of all simulated paths results in the light fluence. Although MC calculations can be easily parallelized and optimized, they generally require an enormous calculation effort, which cannot be performed in real time.

An alternative description is given by the time-independent radiative transfer equation (RTE) derived from the Boltzmann equation (see Equation (9) in). It models the light transport as integro-differential equation and holds for both ballistic and diffuse photons.

Because of the dominating scattering process, light transport in OA is most often modeled by a simpler approximation to the RTE, the so-called light diffusion equation (LDE): where q0(r) is a source term and is the diffusion coefficient (with the reduced scattering coefficient expressed via ). For light incident upon a homogeneous scattering half-space, expression for the light fluence can be readily simplified as Φ(z) = Φ0 · exp (−μeff · z) -well known as Beer's law-with an effective absorption coefficient. For instance, if μeff = 1 cm−1, the light fluence will drop to 1/e of its maximal value after 1 cm (Figure 8 (b)). In this case, light penetration depth in tissue will be limited to a few centimeters.

If collimated light is incident on the object, the LDE is not accurate for depths close to the surface, where the light has not fully diffused. The δ-Eddington approximation is a higher order approximation to the RTE providing better results in the so-called mesoscopic region, typically up to a few millimeters in most tissues.

Model-Based Correction Schemes

Given a certain light transport model and H(r), Equation (2) needs to be solved for retrieving μa. In general, this means solving a non-linear problem because μa also non-linearly and implicitly depends on H(r) via the light fluence Φ(r, μa, μs, g). Various methods aim at obtaining both μa(r) and μs(r) simultaneously or only μa(r) for a given μs(r).

The solution is trivial for superficial imaging applications where only ballistic photons are of interest, e.g., as in case of optical resolution OA microscopy. Alternatively, simple correction schemes, accounting for the drop of light fluence with depth, can be applied, e.g., analytical solutions to the LDE considering only the average optical properties of the medium.

A more complex light transport model can be also solved by an iterative fixed point calculation. Here, the unknown μa(r) is first written as an explicit function:

A better approximation of the unknown μa(r) is then obtained with each additional iteration and is subsequently used as an improved estimate for the right-hand term in the next iteration. Similarly, the problem can be treated in a linearized way by expanding μa and Φ as a Taylor series (μa = μa,0 + δμa, Φ = Φ0 + δΦ0). The problem can be linearly solved then for small perturbations δμa and δΦ from their known solutions μa,0 and Φ0. The fluence deviation δΦ0 is assumed either as unchanged or as being linearly related to δμa in the Born approximation. The solutions can be then obtained by a standard FE method.

Instead of separately treating the acoustic and the optical inverse problem, the solution can also be obtained in a hybrid manner. Laufer et al. presented a combined algorithm for modeling distributions of both the light (FE method) and the generated pressure field (k-space method). The method directly recovers the optical parameters quantitatively by iterative non-linear minimization of the difference between the computed and the measured pressure profiles.

Yet, optoacoustics is a high resolution imaging modality, which makes accurate modeling of the light transport and quantification of OA images challenging. If the various models and equations used in the reconstruction process are not too far away from the experimental reality, the estimate for μa(r) is expected to converge to a stable solution and obtain a good approximate after only a few iterations. In some other cases, solution for the absorption coefficient was shown to be divergent or otherwise the solution was found to be non-unique and ill-posed when trying to recover μa and μs simultaneously.

Other Approaches

The importance of knowing the light fluence in real experiments and the challenges with modeling light propagation in optically heterogeneous tissues have led to some alternative approaches attempting to correct for light attenuation.

One method to increase the amount of available information regarding optical properties of the imaged tissues is to use multi-illumination patterns. The object is illuminated from different angles with light of the same wavelength. Bal et al. presented an approach that can—under some assumptions—recover the optical properties from a multi-illumination pattern in a non-iterative way. Alternatively to multi-illumination patterns, the inversion process can also be done simultaneously for multi-wavelength illumination with knowledge on the spectral dependence of the optical parameters (see next section).

Rosenthal et al. presented a blind separation approach which does not consider any particular light transport model. Instead, it is based on a fairly realistic assumption that the light fluence Φ(r) is a slowly varying function of space whereas the absorption coefficient μa(r) has mainly high spatial frequency components. The idea is to sparsely represent a logarithmic representation of the reconstructed optoacoustic image log H(r) in a set of two bases, a wavelet basis Λn(r) and a Fourier basis Ψm(r), i.e., where the coefficients αn are attributed to the absorption coefficient μa(r) and βm to the light fluence Φ(r). In this way, both the absorption and the light fluence can be reconstructed from the OA image blindly (Figure 9), without modeling the light propagation in tissues.

Cox et al. suggested a method that directly ‘measures’ the local light fluence. Its key component is a chromophore whose absorption has a highly non-linear dependence on the light fluence at a certain threshold intensity. With increase of the illumination intensity, the chromophores are ‘switched’ on (or off) at positions where the local light fluence crosses that threshold. The applicability of this approach however requires the availability of a contrast agent with such highly non-linear absorbance.

Another approach is to experimentally determine the light fluence using a hybrid imaging approach that also accommodates diffuse optical tomography (DOT). Because of their purely optical nature, DOT images are heavily affected by light diffusion thus cannot provide the same spatial resolution as OA. But they can be instead used to improve quantification of OA images. Maslov et al. implanted an absorber with known optical parameter in tissue at depth to investigate the effects of light fluence. Although being an interesting experimental exercise, this invasive method seems impractical for in-vivo imaging applications.

Multispectral Processing in Optoacoustic Imaging

The contrast in OA stems from differential light absorption and can be of either intrinsic or extrinsic nature. Figure 10 shows the absorption spectra of the most dominant intrinsic absorbers in tissue in the visible and near-infrared range.

For many applications, the blood-related absorption is of particular interest, providing valuable physiological or functional information. The main challenge arises from the fact that, at any given wavelength, more than one chromophore contributes to the total absorption coefficient μa(r). For instance, at 800 nm, the oxygenated (HbO2) and the deoxygenated (Hb) forms of hemoglobin contribute equally to absorption and thus are indistinguishable using single wavelength measurements. Therefore, it is the concentrations of the different chromophores ci(r), not the total absorption μa(r), which is mainly of interest from the biological point of view. This relation may be expressed via linear superposition of the different chromophores: where ϵi(λ) are their wavelength-dependent molar extinction coefficients and μback(r, λ) is the residual (background) absorption, which might also include noise. To differentiate between contributions of different chromophores, their distinct spectral dependence on wavelength can be assessed. This multi-wavelength approach is known as multispectral optoacoustic tomography (MSOT) or spectroscopic imaging. The process of recovering ci(r) from multispectral measurements is known as unmixing. It can be combined with calculation of the light fluence or treated as a separate image processing step.

Due to versatility and wide availability of optical molecular agents, sensitive and accurate spectral processing to recover concentration of extrinsically-administered agents may enable longitudinal molecular imaging studies. This can be done by resolving accumulation of agents with specific spectral signatures, such as targeted and activatable fluorescent molecular agents, nanoparticles or genetic markers.

Unmixing with Known Spectra

If spectra of the different components (chromophores) are known, their concentrations can be directly calculated from the OA absorption images. In some special cases this step is rather trivial, e.g., if the certain chromophore of interest is entirely dominant over other absorbers at a given wavelength or if one is only interested in the total blood concentration measured at the isobestic point, for which Hb and HbO2 are equally absorbing.

The simplest multispectral OA imaging approach was presented by Kruger et al. and consisted of subtracting the images obtained at two different wavelengths. It was used to recover distribution of an optical agent, having sharp variations in its absorption spectrum, over wavelength-independent absorbing background. A similar approach proved efficient for detection of dynamic contrast based on a pump-probe excitation. Here, a phosphorescent chromophore was pumped to an excited state with a laser pulse at the pump wavelength. The transient absorption of the excited chromophore was then probed with a slightly delayed second laser pulse with a different wavelength. The dynamic contrast can be extracted by subtracting the pump-only absorption from the combined pump-probe absorption. Also, it is well known that blood oxygenation level saO2 is independent of the total concentration of Hb and HbO2 but only depends on their ratio, i.e.,

Consequently, the oxygenation level can in principle be obtained from the ratio of OA images acquired at two different wavelengths.

Despite its relative simplicity, the image subtraction method cannot provide quantitative measurements of distribution of several chromophores. The approach can be thus generalized to an arbitrary number of chromophores with unknown concentrations and known spectra. In this way, for a given set of wavelengths, the concentrations are linearly fitted to Equation (15) on a per-voxel basis using a linear regression method. This can be efficiently realized with a matrix relation: where ci and μλ are respectively the concentrations of the components and the absorption coefficient in a certain reconstructed image voxel. is the Moore-Penrose pseudo-inverse of the spectra matrix S derived from the components' molar extinction coefficients. In general, the number of measured wavelength should be greater than the number of chromophores.

Blind Unmixing Methods

In many cases, the absorption spectra of all the absorbing components present in the imaged object may not be exactly known. As a result, the unmixing method needs to simultaneously retrieve both the chromophore concentrations and their spectra. In contrary to the previously presented methods, the so-called blind unmixing methods do not operate on a per-voxel basis but exploit the statistical properties of a set of images without requiring prior knowledge of the spectra.

Glatz et al. investigated the performance of blind spectral unmixing on OA images using two multivariate methods. Principal component analysis (PCA) is based on the assumption that the different chromophore distributions are statistically uncorrected and yields an orthonormal transformation into a new base, in which the largest data variance is projected onto the first principal component, the largest remaining variance onto the second one, and so on. In this way, the spectrally correlated measurement data are unmixed by being transformed to the uncorrected components. Similarly, independent component analysis (ICA) is based on a more general assumption that the source components are statistically independent. Consequently, the spectrally mixed measurement data are transformed to statistically independent source components. Both methods are able to recover the spectra of the different components via the transformation matrix and their concentrations via the transformed images. Figure 11 summarizes the results obtained with the linear fitting, PCA and ICA algorithms.

Selected Biomedical Applications

In this section, we shortly highlight several representative examples of the emerging optoacoustic imaging applications while a more comprehensive review on biomedical applications of optoacoustics can be gained from other recently published review articles.

Naturally, the best intrinsic tissue contrast arises from highly absorbing hemoglobin, thus blood is clearly visible in the images. Functional OA microscopy is thus geared towards investigation of vascular structures, providing an excellent intrinsic contrast and a high spatial resolution in the order of some tens of μm and penetration of several millimeters into scattering tissues. In order to obtain high resolution 3-D images, a single spherically-focused high frequency detection element is mechanically raster-scanned along a plane. Significantly higher spatial resolution is further achieved by using a tightly focused optical illumination spot, resulting in the so-called optical resolution microscopy. Despite providing much better spatial resolution in the sub-micron range, similarly to all the other optical microscopy techniques, the latter method suffers from limited penetration depth of only several hundreds of microns due to intense photon scattering in biological tissues. An example of a typical acoustic resolution in vivo functional optoacoustic microscopy system is shown in Figure 3(a). A tunable laser generated pulses of 6-ns duration which passed through a fiber and formed a ring-shaped illumination pattern to suppress OA signals from the surface. A coaxially aligned focused 50-MHz central frequency detector acquired time-resolved one-dimensional signals. The transducer was scanned in 50 μm steps in the x-y plane to form a 3-D image without an inversion algorithm needed. With this setup, subcutaneous vasculature of a Sprague-Dawley rat was demonstrated. Figure 12(a) shows the structural image obtained at the isobestic wavelength of 584 nm in grayscale and the functional saO2 image obtained by a linear fitting to four different wavelengths in color. The saO2 levels obtained were 0.97 and 0.77 for arterial and venous blood, respectively. Changes in saO2 for hypoxic and hyperoxic conditions of the animal were successfully tracked, offering interesting prospects for non-invasive functional brain imaging.

Preclinical whole-body imaging of small animals with multispectral optoacoustic tomography (MSOT) systems is yet another key application of OA imaging. These systems can provide tomographic images with a resolution in the order of hundred μm for depths between several millimeters to several centimeters. Possible applications include monitoring of tumor hypoxia, drug response or molecular targets in biological model organisms. The ability of MSOT to visualize deep-seated fluorescent proteins with high resolution has also been demonstrated. The setup utilized a selective-plane light illumination and a confocal ultrasound detection pattern (Figure 3(b)). An OPO laser provided 8-ns laser pulses of tunable wavelength and the signals were captured by rotating a 3.5-MHz central frequency transducer around the object in 3° steps for multiple slices. The images for multiple wavelengths were reconstructed with the UBP algorithm and unmixed with a linear fitting on a per-pixel basis. In this way, a transgenic three-month old zebra fish expressing mCherry fluorescent proteins in the vertebral column was visualized in vivo. Figure 12(b) shows the structural OA image of the hindbrain at 585 nm and its corresponding histological slice (top row). The multispectral OA imaging has accurately attained the mCherry expression, which has also well corresponded with the findings from epi-fluorescent histology (bottom row).

Buehler et al. presented the capabilities of MSOT for targeted molecular imaging in real time. In the in vivo mouse experiments, a ring-shaped illumination was provided by a tunable OPO (680–950 nm, 10-ns, 10 Hz) and delivered through a fiber bundle. A 64 element transducer array (5-MHz central frequency) covering 172° solid angle detected the signals in parallel without multiplexing. The CD-1 nude mice containing 4T1 tumor allografts were wrapped in a transparent membrane to prevent direct contact of the mouse with water. A near-infrared apoptosis targeting probe (PSS-794, 50-100 nanomoles) was injected in three mice at different time points. The images were reconstructed using a model-based algorithm and unmixed with an ICA spectral unmixing algorithm. Figure 13(a) shows the MSOT image containing both high resolution anatomical and functional information for a tumor-bearing mouse with PSS-794 probe. From the MSOT images (Figure 13(a)), it was clearly determined that the PSS-794 probe mainly accumulated in the blood vessels surrounding the tumors and did not infiltrate into the tumor mass. On the other hand, the use of three-dimensional optical tomography (FMT) has not attained the sufficient spatial resolution that would determine the precise location of the probe (Figure 13(c)), mainly due to its poor spatial resolution resulting from photon scattering.

At present, the great potential of optoacoustic imaging that was showcased in preclinical research has encouraged translation of this technology into clinical practice. The clinical promise of OA is greatly supported by its non-ionizing radiation, rich contrast, real-time operation, and relatively low cost. It is likely that OA will increasingly enter the clinical imaging segments in areas such as skin, breast, vasculature, and ocular imaging, visualization of tumor and inflammation-related pathology etc.. Indeed, early detection of breast cancer will capture a significant spot in the future developments of OA imaging applications. One example of such clinical system for OA breast angiography without the need for a contrast agent was recently presented. The system employed 128 detection elements (5-MHz central frequency) which were placed on a 100 mm radius hemisphere in a spiral pattern and acquired the signals in parallel (Figure 3(f)). The hemisphere was rotated for higher projection sampling and one complete scan took between 6 to 24 s. The volunteer's breast was immersed into a water-filled bowl from the top and immobilized by an optically and acoustically transparent cup. 3-D images were reconstructed from a filtered BP algorithm with prior frequency filtering and transducers' impulse response deconvolution. The system was used to visualize the vasculature of a 57 year-old volunteer without a contrast agent and vessels from superficial regions up to a considerable depth of several centimeters (Figure 14(a)).

Naturally, clinical application of OA imaging is limited to regions that can be efficiently illuminated with the excitation light. When the imaged region of interest is not reachable from the outside, certain areas of the human body can also be imaged with minimally invasive catheter- or endoscopy-based techniques. To this end, such systems have been realized for other imaging modalities, such as intra-vascular ultrasound (IVUS) or fluorescent imaging (NIRF). In the recent years, similar approaches have been also attempted with OA for intravascular, esophageal or colonoscopic imaging. One example is shown in Figure 3(d), where a system for simultaneous functional OA and US endoscopy in vivo is presented. The endoscope was able to acquire co-registered pulse-echo US and OA images. A-scans were obtained from an ultrasonic pulse-echo system while two OA pulses, having different wavelengths for spectroscopic imaging, were delivered through a fiber and detected by the focused transducer. 3-D images could be obtained by constant rotation of a scanning mirror and a mechanical pullback of the catheter. Using this system, rabbit esophagus was imaged in vivo, as shown in Figure 14(b).

Limitations and Future Challenges of Optoacoustic Imaging

Indeed, optoacoustic imaging offers multiple advantages over other imaging modalities. Most favorable is the combination of US-diffraction limited spatial resolution and excellent optical absorption contrast without any ionizing radiation involved. Clearly, much like all other imaging modalities, optoacoustics is limited in certain aspects. As presented in the current review, due to the (non-ideal) highly heterogeneous and wavelength-dependent nature of biological tissues, images reconstructed from experimental data usually contain artifacts, such as negative values in the images which otherwise have no physical meaning. In addition, high noise levels, limited bandwidth of the detection system, inaccurate assumptions on the speed of sound, acoustic attenuation, as well as limited view in detection, can all alter the signals in a way that the reconstructed image accuracy is highly compromised. While some simple algorithms, such as Hilbert transform for envelope detection or thresholding of negative values, can be used to keep image values in the positive region, these non-linear operations in fact only conceal the underlying problem while not assisting with achieving quantitative results. Instead, reliable detection and modeling of signal propagation in tissues is required.

Imaging at depth brings further challenges and limitations. First, imaging depth is compromised by attenuation of the light fluence in optically opaque tissues. Some of it can be compensated by increasing the amount of the deposited laser energies. However, for in vivo applications in the near-infrared, illumination on the skin surface is limited by the maximum permissible exposure (MPE,) to about 20 mJ/cm2. As a result, imaging depth is usually restricted to regions where the light fluence is sufficiently high to generate detectable pressure variations, typically up to a few centimeters in most soft tissues. In addition, even though spatial resolution of optoacoustic imaging is not affected by light scattering, resolution may still deteriorate with increasing imaging depth due to its dependence on the frequency content of the detected ultrasonic responses. This is because of the dispersive nature of ultrasound waves with high frequency components being rapidly attenuated as they propagate through tissues.

In multispectral imaging applications, one also faces the problem of so-called ‘spectral coloring’. Due to the non-local and non-linear dependence of light transport on the object's optical properties, spectra of various tissue chromophores and agents, extracted by means of optoacoustics, might be corrupted. For improving quantitative determination of chromophore concentrations using their known spectra, an accurate correction for the light distribution needs to be employed, which, as discussed before, constitutes a challenging problem and an open area of research. Another important aspect influenced by depth is the limits of the imaging system with respect to detection of minimal concentrations of certain intrinsic tissue chromophores and extrinsically-administered contrast agents. To this end, several publications have reported sensitivity in the femtomole to picomole range, depending on the molecular weight of the probe employed, while a recent study has also systematically addressed the question of how the detection limits are affected by imaging depth. In reality, sensitivity limits are affected by multiple additional factors, such as the total volume, spectrum and absorption coefficient of the imaged chromophore, the noise equivalent pressure (NEP) of the detectors, level and spectral dependence of background tissue absorption.

Instrumentation-related limitations are an additional factor in translating optoacoustic imaging methods into routine application. One of the important advantages, which helps reducing motion-related image artifacts but may also enable powerful applications in in-vivo tracking of dynamic processes, is the possibility for reliable measurements and image rendering in real-time. This requires introducing high repetition pulsed laser technology and fast wavelength tuning capabilities, especially for multi-spectral imaging applications involving real-time visualization of distributions of spectrally distinct contrast agents. On the other hand, the detection technology should become increasingly sensitive but also parallelized to enable real-time image acquisition. Often a trade-off between acquisition speed and quality on the one hand and costs of instrumentation on the other hand has to be found. In this respect, optoacoustic imaging has greatly benefited from the advances in both laser and detection technology over the last years and significant performance enhancements are expected as the technological frontiers advance. Providing information from hybrid US imaging will be of added value for future clinical systems. Besides fast data acquisition, fast image rendering is also of great importance to readily provide findings for efficient real-time guidance and optimization of diagnostic or therapeutic procedures. Recently, extensive research is devoted to the subject of real-time visualization, benefiting both from algorithmic developments and advances in parallel computing technology.

Quantitative rendering of chromophore concentration is perhaps among the most important but also among the most challenging tasks of the optoacoustic methods. Hemoglobin in its oxygenated and deoxygenated form is omnipresent in tissue and is the most dominant intrinsic absorber in both the visible and most of the near-infrared range of the optical spectrum. Thus, quantitative determination of blood parameters is of great interest as it is related to a variety of physiological parameters. Prominent examples are monitoring angiogenesis or hypoxic states in tumors, functional imaging in the brain or responses to environmental changes. In most cases, the oxygen saturation saO2 (Equation (16)) is of main interest, whose quantitative determination is non-trivial, especially in the presence of strong wavelength dependence of the excitation light fluence in tissues. First attempts to reliably determine saO2 were conducted in vitro whereas accuracy of ±2.5% saO2 and a minimum detectable change of ±1% saO2 were reported. Furthermore, oxygenation saturation of single blood vessels, embedded deep in tissue mimicking phantom, was determined using numerical forward light transport model, reporting an accuracy of ±7% saO2. Most in vivo results were so far obtained from multispectral optical resolution optoacoustic microscopy where light transport issues can be effectively omitted. Zhang et al. reported 4% saO2 systematic error for their ex vivo study with bovine blood, while accuracy for in vivo determination of arterial and venous blood under hypoxic, normal and hyperoxic conditions was not reported. Applications of efficient and practical saO2-determination schemes for experimental data in 3-D in an accurate, robust and computationally feasible manner presents therefore one of the key challenges for the current algorithmic developments. Finally, when imaging other chromophores or extrinsically-administered contrast agents, the blood background needs to be accounted for to enable quantified extraction of biomarker concentrations.

Conclusions

Owing to its hybrid nature, i.e., optical excitation and ultrasonic detection, optoacoustics benefits from both rich and versatile optical contrast and high (diffraction-limited) spatial resolution, associated with relatively low scattering of ultrasonic waves. Optoacoustic biosensing and imaging provides an excellent platform for multi-scale investigations using the same contrast, from microscopic observations at the single capillary and cell level to whole body imaging of small animals and deep tissue imaging of humans. Much like other optical imaging modalities, optoacoustics is safe for both small animals and clinical use as it utilizes non-ionizing radiation at the visible and near-infrared wavelengths.

In the last decade, OA imaging is considered to be the fastest growing biomedical imaging modality with multiple already implemented and envisioned applications in biomedical research and clinical practice, from diagnostic applications in cancer research and brain imaging to drug development and treatment monitoring. Clearly, progress of the image reconstruction and quantification methods has been central to those developments and has significantly contributed to creation of new exciting applications of the OA imaging technologies. As discussed in this review, multiple frontiers are still open in the algorithmic areas, where many challenges related to removal of image artifacts, image quantification, reconstruction strategies in the presence of acoustically mismatched areas, real-time operation, and multi-spectral data processing need to be further addressed.

References 177

  1. 1.

    Tainter C.S., Bell A.G. Selenium and the photophone. Nature. 1880;22:500-503.

    Not in our library

  2. 2.

    Bell A.G. Production of sound by radiant energy. Manuf. Build. 1881;13:156-158.

    Not in our library10.1126/science.os-2.49.242

  3. 3.

    Bell A.G. The spectrophone. Bull. Phil. Soc. 1881;4:143-162.

    Not in our library

  4. 4.

    Tyndal J. Action of an intermittent beam of radiant heat upon gaseous matter. Proc. R. Soc. Lond. 1880;31:307-317.

    Not in our library10.1126/science.os-2.38.110-a

  5. 5.

    Röntgen W.C. On tones produced by the intermittent irradiation of a gas. Philos. Mag. 1881;11:308-311.

    Not in our library

  6. 6.

    Groth M. Photophones revisted. Aust. Melb. 1987;55:14.

    Not in our library

  7. 7.

    Veingerov M.L. New method of gas analysis based on Tyndall-Röntgen optic-acoustic effect. Dokl. Akad. Nauk. USSR. 1938;19:687-688.

    Not in our library

  8. 8.

    Delany M.E. The optic-acoustic effect in gases. Sci. Prog. 1959;47:459-467.

    Not in our library

  9. 9.

    Bobylev B.A., Kravchenko A.F. Photoacoustic spectra of high-resistivity gallium arsenide. Fizika i Tekhnika Poluprovodnikov. 1969;3:1352-1356.

    Not in our library

  10. 10.

    Rosencwaig A. Photoacoustic spectroscopy of solids. Opt. Commun. 1973;7:305-308.

    Not in our library

  11. 11.

    Rosencwaig A. Photoacoustic spectroscopy of biological materials. Science. 1973;181:657-658.

    Not in our library10.1126/science.181.4100.657

  12. 12.

    Oraevsky A.A., Jacques S.L., Esenaliev R.O., Tittel F.K. Laser based optoacoustic imaging in biological tissues. Proc. SPIE. 1994;2134A:122-128.

    Not in our library

  13. 13.

    Kruger R.A., Kiser W.L., Reinecke D.R., Kruger G.A., Miller K.D. Thermoacoustic molecular imaging of small animals. Mol. Imaging. 2003;2:113-123.

    Not in our library10.1162/15353500200303109

  14. 14.

    Wang X.D., Pang Y.J., Ku G., Xie X.Y., Stoica G., Wang L.H.V. Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain. Nat. Biotechnol. 2003;21:803-806.

    Not in our library10.1038/nbt839

  15. 15.

    Ntziachristos V., Razansky D. Molecular imaging by means of multispectral optoacoustic tomography (MSOT). Chem. Rev. 2010;110:2783-2794.

    Not in our library10.1021/cr9002566

  16. 16.

    Ntziachristos V., Ripoll J., Wang L.H.V., Weissleder R. Looking and listening to light: The evolution of whole-body photonic imaging. Nat. Biotechnol. 2005;23:313-320.

    Not in our library10.1038/nbt1074

  17. 17.

    Zhang E.Z., Laufer J.G., Pedley R.B., Beard P.C. In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy. Phys. Med. Biol. 2009;54:1035-1046.

    Not in our library10.1088/0031-9155/54/4/014

  18. 18.

    Song L., Maslov K., Wang L.V. Section-illumination photoacoustic microscopy for dynamic 3D imaging of microcirculation. in vivo. Opt. Lett. 2010;35:1482-1484.

    Not in our library10.1364/ol.35.001482

  19. 19.

    Razansky D., Vinegoni C., Ntziachristos V. Imaging of mesoscopic-scale organisms using selective-plane optoacoustic tomography. Phys. Med. Biol. 2009;54:2769-2777.

    Not in our library10.1088/0031-9155/54/9/012

  20. 20.

    Razansky D., Vinegoni C., Ntziachristos V. Multispectral photoacoustic imaging of fluorochromes in small animals. Opt. Lett. 2007;32:2891-2893.

    Not in our library10.1364/ol.32.002891

  21. 21.

    Beard P. Biomedical photoacoustic imaging. Interface Focus. 2011;1:602-631.

    Not in our library10.1098/rsfs.2011.0028

  22. 22.

    Wang L.H.V., Hu S. Photoacoustic tomography: In vivo imaging from organelles to organs. Science. 2012;335:1458-1462.

    Not in our library10.1126/science.1216210

  23. 23.

    Kuchment P., Kunyansky L. A survey in mathematics for industry mathematics of thermoacoustic tomography. Eur. J. Appl. Math. 2008;19:191-224.

    Not in our library

  24. 24.

    Cox B., Laufer J.G., Arridge S.R., Beard P.C. Quantitative spectroscopic photoacoustic imaging: A review. J. Biomed. Opt. 2012;17.

    Not in our library10.1117/1.jbo.17.6.06120

  25. 25.

    Kruger R.A., Miller K.D., Reynolds H.E., Kiser W.L., Reinecke D.R., Kruger G.A. Breast cancer in vivo: Contrast enhancement with thermoacoustic CT at 434 MHz-Feasibility study. Radiology. 2000;216:279-283.

    Not in our library10.1148/radiology.216.1.r00jl30279

  26. 26.

    Razansky D., Kellnberger S., Ntziachristos V. Near-field radiofrequency thermoacoustic tomography with impulse excitation. Med. Phys. 2010;37:4602-4607.

    Not in our library10.1118/1.3467756

  27. 27.

    Amiot C.L., Xu S.P., Liang S., Pan L.Y., Zhao J.X.J. Near-infrared fluorescent materials for sensing of biological targets. Sensors. 2008;8:3082-3105.

    Not in our library10.3390/s8053082

  28. 28.

    Ku G., Wang L.H.V. Deeply penetrating photoacoustic tomography in biological tissues enhanced with an optical contrast agent. Opt. Lett. 2005;30:507-509.

    Not in our library10.1364/ol.30.000507

  29. 29.

    Tong L., Wei Q.S., Wei A., Cheng J.X. Gold nanorods as contrast agents for biological imaging: Optical properties, surface conjugation and photothermal effects. Photochem. Photobiol. 2009;85:21-32.

    Not in our library10.1111/j.1751-1097.2008.00507.x

  30. 30.

    De La Z.A., Zavaleta C., Keren S., Vaithilingam S., Bodapati S., Liu Z., Levi J., Smith B.R., Ma T.J., Oralkan O., et al. Carbon nanotubes as photoacoustic molecular imaging agents in living mice. Nat. Nanotechnol. 2008;3:557-562.

    Not in our library10.1038/nnano.2008.231

  31. 31.

    Zimmer M. Green fluorescent protein (GFP): Applications, structure, and related photophysical behavior. Chem. Rev. 2002;102:759-781.

    Not in our library10.1021/cr010142r

  32. 32.

    Maslov K., Wang L.V. Photoacoustic imaging of biological tissue with intensity-modulated continuous-wave laser. J. Biomed. Opt. 2008;13.

    Not in our library10.1117/1.290496

  33. 33.

    Telenkov S.A., Mandelis A. Fourier-domain methodology for depth-selective photothermoacoustic imaging of tissue chromophores. Eur. Phys. J. Spec. Top. 2008;153:443-448.

    Not in our library

  34. 34.

    Kellnberger S., Deliolanis N.C., Queiros D., Sergiadis G., Ntziachristos V. In vivo frequency domain optoacoustic tomography. Opt. Lett. 2012;37:3423-3425.

    Not in our library10.1364/ol.37.003423

  35. 35.

    Bal G., Ren K. Multi-source quantitative photoacoustic tomography in a diffusive regime. Inverse Probl. 2011;27.

    Not in our library10.1088/0266-5611/27/7/07500

  36. 36.

    Bal G., Ren K. On multi-spectral quantitative photoacoustic tomography in diffusive regime. Inverse Probl. 2012;28.

    Not in our library10.1088/0266-5611/28/2/02501

  37. 37.

    Larina I.V., Larin K.V., Esenaliev R.O. Real-time optoacoustic monitoring of temperature in tissues. J. Phys. D Appl. Phys. 2005;38:2633.

    Not in our library10.1088/0022-3727/38/15/015

  38. 38.

    Pramanik M., Wang L.V. Thermoacoustic and photoacoustic sensing of temperature. J. Biomed. Opt. 2009;14.

    Not in our library10.1117/1.3247155

  39. 39.

    Kruger R.A., Liu P., Fang Y.R., Appledorn C.R. Photoacoustic ultrasound (PAUS)— reconstruction tomography. Med. Phys. 1995;22:1605-1609.

    Not in our library10.1118/1.597429

  40. 40.

    Ma R., Söntges S., Shoham S., Ntziachristos V., Razansky D. Fast scanning coaxial optoacoustic microscopy. Biomed. Opt. Express. 2012;3:1724-1731.

    Not in our library10.1364/boe.3.001724

  41. 41.

    Zemp R.J., Song L.A., Bitton R., Shung K.K., Wang L.H.V. Realtime photoacoustic microscopy in vivo with a 30-MHz ultrasound array transducer. Opt. Express. 2008;16:7915-7928.

    Not in our library10.1364/oe.16.007915

  42. 42.

    Allen T.J., Beard P.C. Pulsed near-infrared laser diode excitation system for biomedical photoacoustic imaging. Opt. Lett. 2006;31:3462-3464.

    Not in our library10.1364/ol.31.003462

  43. 43.

    Wang Y., Maslov K., Zhang Y., Hu S., Yang L.M., Xia Y.N., Liu J.A., Wang L.H.V. Fiber-laser-based photoacoustic microscopy and melanoma cell detection. J. Biomed. Opt. 2011;16.

    Not in our library10.1117/1.3525643

  44. 44.

    Zhang H.F., Maslov K., Stoica G., Wang L.H.V. Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging. Nat. Biotechnol. 2006;24:848-851.

    Not in our library10.1038/nbt1220

  45. 45.

    Razansky D., Distel M., Vinegoni C., Ma R., Perrimon N., Koster R.W., Ntziachristos V. Multispectral opto-acoustic tomography of deep-seated fluorescent proteins. in vivo. Nat. Photonics. 2009;3:412-417.

    Not in our library

  46. 46.

    Zhang E., Laufer J., Beard P. Backward-mode multiwavelength photoacoustic scanner using a planar Fabry-Perot polymer film ultrasound sensor for high-resolution three-dimensional imaging of biological tissues. Appl. Opt. 2008;47:561-577.

    Not in our library10.1364/ao.47.000561

  47. 47.

    Yang J.M., Favazza C., Chen R.M., Yao J.J., Cai X., Maslov K., Zhou Q.F., Shung K.K., Wang L.H.V. Simultaneous functional photoacoustic and ultrasonic endoscopy of internal organs. in vivo. Nat. Med. 2012;18:1297-1302.

    Not in our library10.1038/nm.2823

  48. 48.

    Paltauf G., Nuster R., Haltmeier M., Burgholzer P. Photoacoustic tomography using a Mach-Zehnder interferometer as an acoustic line detector. Appl. Opt. 2007;46:3352-3358.

    Not in our library10.1364/ao.46.003352

  49. 49.

    Kruger R.A., Lam R.B., Reinecke D.R., del Rio S.P., Doyle R.P. Photoacoustic angiography of the breast. Med. Phys. 2010;37:6096-6100.

    Not in our library10.1118/1.3497677

  50. 50.

    Manohar S., Kharine A., van Hespen J.C.G., Steenbergen W., van Leeuwen T.G. The twente photoacoustic mammoscope: System overview and performance. Phys. Med. Biol. 2005;50:2543-2557.

    Not in our library10.1088/0031-9155/50/11/007

  51. 51.

    Xia W.F., Piras D., van Hespen J.C.G., van Veldhoven S., Prins C., van Leeuwen T.G., Steenbergen W., Manohar S. An optimized ultrasound detector for photoacoustic breast tomography. Med. Phys. 2013;40.

    Not in our library10.1118/1.4792462

  52. 52.

    Huang S.-W., Chen S.-L., Ling T., Maxwell A., O'Donnell M., Guo L.J., Ashkenazi S. Low-noise wideband ultrasound detection using polymer microring resonators. Appl. Phys. Lett. 2008;92.

    Not in our library10.1063/1.2929379

  53. 53.

    Chen S.-L., Huang S.-W., Ling T., Ashkenazi S., Guo L. Polymer microring resonators for high-sensitivity and wideband photoacoustic imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2009;56:2482-2491.

    Not in our library10.1109/tuffc.2009.1335

  54. 54.

    Rosenthal A., Razansky D., Ntziachristos V. High-sensitivity compact ultrasonic detector based on a pi-phase-shifted fiber Bragg grating. Opt. Lett. 2011;36:1833-1835.

    Not in our library10.1364/ol.36.001833

  55. 55.

    Oralkan O., Ergun A.S., Johnson J.A., Karaman M., Demirci U., Kaviani K., Lee T.H., Khuri-Yakub B.T. Capacitive micromachined ultrasonic transducers: Next-generation arrays for acoustic imaging? IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2002;49:1596-1610.

    Not in our library10.1109/tuffc.2002.1049742

  56. 56.

    Caliano G., Carotenuto R., Cianci E., Foglietti V., Caronti A., Iula A., Pappalardo M. Design, fabrication and characterization of a capacitive micromachined ultrasonic probe for medical imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2005;52:2259-2269.

    Not in our library10.1109/tuffc.2005.1563268

  57. 57.

    Vaithilingam S., Ma T.-J., Furukawa Y., Wygant I.O., Xuefeng Z., de la Zerda A., Oralkan O., Kamaya A., Jeffrey R.B., Khuri-Yakub B.T. Three-dimensional photoacoustic imaging using a two-dimensional CMUT array. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2009;56:2411-2419.

    Not in our library10.1109/tuffc.2009.1329

  58. 58.

    Viator J.A., Choi B., Ambrose M., Spanier J., Nelson J.S. In vivo port-wine stain depth determination with a photoacoustic probe. Appl. Opt. 2003;42:3215-3224.

    Not in our library10.1364/ao.42.003215

  59. 59.

    Holan S.H., Viator J.A. Automated wavelet denoising of photoacoustic signals for circulating melanoma cell detection and burn image reconstruction. Phys. Med. Biol. 2008;53.

    Not in our library10.1088/0031-9155/53/12/n01

  60. 60.

    Wang Y., Xing D., Zeng Y.G., Chen Q. Photoacoustic imaging with deconvolution algorithm. Phys. Med. Biol. 2004;49:3117-3124.

    Not in our library10.1088/0031-9155/49/14/006

  61. 61.

    Rosenthal A., Ntziachristos V., Razansky D. Optoacoustic methods for frequency calibration of ultrasonic sensors. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2011;58:316-326.

    Not in our library10.1109/tuffc.2011.1809

  62. 62.

    Norton S.J., Linzer M. Ultrasonic reflectivity imaging in three dimensions: Exact inverse scattering solutions for plane, cylindrical, and spherical apertures. IEEE Trans. Biomed. Eng. 1981;28:202-220.

    Not in our library10.1109/tbme.1981.324791

  63. 63.

    Kostli K.P., Frenz M., Bebie H., Weber H.P. Temporal backward projection of optoacoustic pressure transients using Fourier transform methods. Phys. Med. Biol. 2001;46:1863-1872.

    Not in our library10.1088/0031-9155/46/7/309

  64. 64.

    Kunyansky L.A. A series solution and a fast algorithm for the inversion of the spherical mean Radon transform. Inverse Probl. 2007;23:S11-S20.

    Not in our library

  65. 65.

    Xu Y., Feng D.Z., Wang L.H.V. Exact frequency-domain reconstruction for thermoacoustic tomography—I: Planar geometry. IEEE Trans. Med. Imaging. 2002;21:823-828.

    Not in our library10.1109/tmi.2002.801172

  66. 66.

    Xu Y., Xu M.H., Wang L.H.V. Exact frequency-domain reconstruction for thermoacoustic tomography—II: Cylindrical geometry. IEEE Trans. Med. Imaging. 2002;21:829-833.

    Not in our library10.1109/tmi.2002.801171

  67. 67.

    Kunyansky L. Fast reconstruction algorithms for the thermoacoustic tomography in certain domains with cylindrical or spherical symmetries. Inverse Probl. Imaging. 2012;6:111-131.

    Not in our library

  68. 68.

    Wang K., Anastasio M.A. A simple Fourier transform-based reconstruction formula for photoacoustic computed tomography with a circular or spherical measurement geometry. Phys. Med. Biol. 2012;57:N493-N499.

    Not in our library10.1088/0031-9155/57/23/n493

  69. 69.

    Finch D., Rakesh P.S.K. Determining a function from its mean values over a family of spheres. Siam J. Math. Anal. 2004;35:1213-1240.

    Not in our library

  70. 70.

    Finch D., Rakesh H.M. Inversion of spherical means and the wave equation in even dimensions. Siam J. Appl. Math. 2007;68:392-412.

    Not in our library

  71. 71.

    Kunyansky L.A. Explicit inversion formulae for the spherical mean Radon transform. Inverse Probl. 2007;23:373-383.

    Not in our library

  72. 72.

    Xu M.H., Wang L.H.V. Time-domain reconstruction for thermoacoustic tomography in a spherical geometry. IEEE Trans. Med. Imaging. 2002;21:814-822.

    Not in our library10.1109/tmi.2002.801176

  73. 73.

    Xu M.H., Wang L.H.V. Universal back-projection algorithm for photoacoustic computed tomography. Phys. Rev. E. 2005;71.

    Not in our library10.1103/physreve.71.016706

  74. 74.

    Xu Y., Wang L.H.V. Time reversal and its application to tomography with diffracting sources. Phys. Rev. Lett. 2004;92.

    Not in our library10.1103/physrevlett.92.033902

  75. 75.

    Hristova Y., Kuchment P., Nguyen L. Reconstruction and time reversal in thermoacoustic tomography in acoustically homogeneous and inhomogeneous media. Inverse Probl. 2008;24.

    Not in our library10.1088/0266-5611/24/5/055006

  76. 76.

    Treeby B.E., Zhang E.Z., Cox B.T. Photoacoustic tomography in absorbing acoustic media using time reversal. Inverse Probl. 2010;26.

    Not in our library10.1088/0266-5611/26/11/115003

  77. 77.

    Cox B.T., Kara S., Arridge S.R., Beard P.C. k-space propagation models for acoustically heterogeneous media: Application to biomedical photoacoustics. J. Acoust. Soc. Am. 2007;121:3453-3464.

    Not in our library10.1121/1.2717409

  78. 78.

    Paltauf G., Viator J.A., Prahl S.A., Jacques S.L. Iterative reconstruction algorithm for optoacoustic imaging. J. Acoust. Soc. Am. 2002;112:1536-1544.

    Not in our library10.1121/1.1501898

  79. 79.

    Rosenthal A., Razansky D., Ntziachristos V. Fast semi-analytical model-based acoustic inversion for quantitative optoacoustic tomography. IEEE Trans. Med. Imaging. 2010;29:1275-1285.

    Not in our library10.1109/tmi.2010.2044584

  80. 80.

    Wang K., Ermilov S.A., Su R., Brecht H.P., Oraevsky A.A., Anastasio M.A. An imaging model incorporating ultrasonic transducer properties for three-dimensional optoacoustic tomography. IEEE Trans. Med. Imaging. 2011;30:203-214.

    Not in our library10.1109/tmi.2010.2072514

  81. 81.

    Yuan Z., Jiang H.B. Quantitative photoacoustic tomography: Recovery of optical absorption coefficient maps of heterogeneous media. Appl. Phys. Lett. 2006;88.

    Not in our library10.1063/1.2209883

  82. 82.

    Dean-Ben X., Buehler A., Ntziachristos V., Razansky D. Accurate model-based reconstruction algorithm for three-dimensional optoacoustic tomography. IEEE Trans. Med. Imaging. 2012;31:1922-1928.

    Not in our library10.1109/tmi.2012.2208471

  83. 83.

    Yao L., Jiang H.B. Finite-element-based photoacoustic tomography in time domain. J. Opt. A Pure Appl. Opt. 2009;11.

    Not in our library10.1088/1464-4258/11/8/085301

  84. 84.

    Golub G.H., van Loan C.F. Matrix Computations. 1996.

    Not in our library

  85. 85.

    Paige C.C., Saunders M.A. LSQR: An algorithm for sparse linear equations and sparse least squares. ACM Trans. Math. Softw. 1982;8:43-71.

    Not in our library

  86. 86.

    Buehler A., Deán-Ben X., Claussen J., Ntziachristos V., Razansky D. Three-dimensional optoacoustic tomography at video rate. Opt. Express. 2012;20:22712-22719.

    Not in our library10.1364/oe.20.022712

  87. 87.

    Wang K., Huang C., Kao Y.-J., Chou C.-Y., Oraevsky A.A., Anastasio M.A. Accelerating image reconstruction in three-dimensional optoacoustic tomography on graphics processing units. Med. Phys. 2013;40.

    Not in our library10.1118/1.4774361

  88. 88.

    Pratx G., Xing L. GPU computing in medical physics: A review. Med. Phys. 2011;38.

    Not in our library10.1118/1.3578605

  89. 89.

    Rosenthal A., Jetzfellner T., Razansky D., Ntziachristos V. Efficient framework for model-based tomographic image reconstruction using wavelet packets. IEEE Trans. Med. Imaging. 2012;31:1346-1357.

    Not in our library10.1109/tmi.2012.2187917

  90. 90.

    Razansky D., Buehler A., Ntziachristos V. Volumetric real-time multispectral optoacoustic tomography of biomarkers. Nat. Protoc. 2011;6:1121-1129.

    Not in our library10.1038/nprot.2011.351

  91. 91.

    Gamelin J., Maurudis A., Aguirre A., Huang F., Guo P.Y., Wang L.V., Zhu Q. A real-time photoacoustic tomography system for small animals. Opt. Express. 2009;17:10489-10498.

    Not in our library10.1364/oe.17.010489

  92. 92.

    Taruttis A., Herzog E., Razansky D., Ntziachristos V. Real-time imaging of cardiovascular dynamics and circulating gold nanorods with multispectral optoacoustic tomography. Opt. Express. 2010;18:19592-19602.

    Not in our library10.1364/oe.18.019592

  93. 93.

    Li M.L., Zhang H.F., Maslov K., Stoica G., Wang L.H.V. Improved in vivo photoacoustic microscopy based on a virtual-detector concept. Opt. Lett. 2006;31:474-476.

    Not in our library10.1364/ol.31.000474

  94. 94.

    Yang X.M., Li M.L., Wang L.H.V. Ring-based ultrasonic virtual point detector with applications to photoacoustic tomography. Appl. Phys. Lett. 2007;90.

    Not in our library10.1063/1.2749856

  95. 95.

    Li C.H., Wang L.V. High-numerical-aperture-based virtual point detectors for photoacoustic tomography. Appl. Phys. Lett. 2008;93.

    Not in our library10.1063/1.2963365

  96. 96.

    Burgholzer P., Hofer C., Paltauf G., Haltmeier M., Scherzer O. Thermoacoustic tomography with integrating area and line detectors. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2005;52:1577-1583.

    Not in our library10.1109/tuffc.2005.1516030

  97. 97.

    Burgholzer P., Bauer-Marschallinger J., Grun H., Haltmeier M., Paltauf G. Temporal back-projection algorithms for photoacoustic tomography with integrating line detectors. Inverse Probl. 2007;23:S65-S80.

    Not in our library

  98. 98.

    Haltmeier M., Scherzer O., Burgholzer P., Paltauf G. Thermoacoustic computed tomography with large planar receivers. Inverse Probl. 2004;20:1663-1673.

    Not in our library

  99. 99.

    Gratt S., Passler K., Nuster R., Paltauf G. Photoacoustic section imaging with an integrating cylindrical detector. Biomed. Opt. Express. 2011;2:2973-2981.

    Not in our library10.1364/boe.2.002973

  100. 100.

    Nuster R., Zangerl G., Haltmeier M., Paltauf G. Full field detection in photoacoustic tomography. Opt. Express. 2010;18:6288-6299.

    Not in our library10.1364/oe.18.006288

  101. 101.

    Rosenthal A., Ntziachristos V., Razansky D. Model-based optoacoustic inversion with arbitrary-shape detectors. Med. Phys. 2011;38:4285-4295.

    Not in our library10.1118/1.3589141

  102. 102.

    Duck F.A. Physical Properties of Tissue: A Comprehensive Reference Book. 1990.

    Not in our library

  103. 103.

    Manohar S., Willemink R.G.H., van der Heijden F., Slump C.H., van Leeuwen T.G. Concomitant speed-of-sound tomography in photoacoustic imaging. Appl. Phys. Lett. 2007;91.

    Not in our library10.1063/1.2789689

  104. 104.

    Jose J., Willemink R.G.H., Resink S., Piras D., van Hespen J.C.G., Slump C.H., Steenbergen W., van Leeuwen T.G., Manohar S. Passive element enriched photoacoustic computed tomography (PER PACT) for simultaneous imaging of acoustic propagation properties and light absorption. Opt. Express. 2011;19:2093-2104.

    Not in our library10.1364/oe.19.002093

  105. 105.

    Treeby B.E., Varslot T.K., Zhang E.Z., Laufer J.G., Beard P.C. Automatic sound speed selection in photoacoustic image reconstruction using an autofocus approach. J. Biomed. Opt. 2011;16.

    Not in our library10.1117/1.3619139

  106. 106.

    Yuan Z., Jiang H.B. Three-dimensional finite-element-based photoacoustic tomography: Reconstruction algorithm and simulations. Med. Phys. 2007;34:538-546.

    Not in our library10.1118/1.2409234

  107. 107.

    Agranovsky M., Kuchment P. Uniqueness of reconstruction and an inversion procedure for thermoacoustic and photoacoustic tomography with variable sound speed. Inverse Probl. 2007;23:2089-2102.

    Not in our library

  108. 108.

    Modgil D., Anastasio M.A., La Riviere P.J. Image reconstruction in photoacoustic tomography with variable speed of sound using a higher-order geometrical acoustics approximation. J. Biomed. Opt. 2010;15.

    Not in our library10.1117/1.3333550

  109. 109.

    Xu M.H., Wang L.V. Analytic explanation of spatial resolution related to bandwidth and detector aperture size in thermoacoustic or photoacoustic reconstruction. Phys. Rev. E. 2003;67.

    Not in our library10.1103/physreve.67.056605

  110. 110.

    La Riviere P.J., Zhang J., Anastasio M.A. Image reconstruction in optoacoustic tomography for dispersive acoustic media. Opt. Lett. 2006;31:781-783.

    Not in our library10.1364/ol.31.000781

  111. 111.

    Wang L.H.V., Yang X.M. Boundary conditions in photoacoustic tomography and image reconstruction. J. Biomed. Opt. 2007;12.

    Not in our library10.1117/1.2709861

  112. 112.

    Anastasio M.A., Zhang J., Pan X.C., Zou Y., Ku G., Wang L.H.V. Half-time image reconstruction in thermoacoustic tomography. IEEE Trans. Med. Imaging. 2005;24:199-210.

    Not in our library10.1109/tmi.2004.839682

  113. 113.

    Dean-Ben X.L., Ma R., Razansky D., Ntziachristos V. Statistical approach for optoacoustic image reconstruction in the presence of strong acoustic heterogeneities. IEEE Trans. Med. Imaging. 2011;30:401-408.

    Not in our library10.1109/tmi.2010.2081683

  114. 114.

    Dean-Ben X.L., Ntziachristos V., Razansky D. Artefact reduction in optoacoustic tomographic imaging by estimating the distribution of acoustic scatterers. J. Biomed. Opt. 2012;17.

    Not in our library10.1117/1.jbo.17.11.110504

  115. 115.

    Xu Y., Wang L.V., Ambartsoumian G., Kuchment P. Reconstructions in limited-view thermoacoustic tomography. Med. Phys. 2004;31:724-733.

    Not in our library10.1118/1.1644531

  116. 116.

    Buehler A., Rosenthal A., Jetzfellner T., Dima A., Razansky D., Ntziachristos V. Model-based optoacoustic inversions with incomplete projection data. Med. Phys. 2011;38:1694-1704.

    Not in our library10.1118/1.3556916

  117. 117.

    Yao L., Jiang H.B. Photoacoustic image reconstruction from few-detector and limited-angle data. Biomed. Opt. Express. 2011;2:2649-2654.

    Not in our library10.1364/boe.2.002649

  118. 118.

    Paltauf G., Nuster R., Burgholzer P. Weight factors for limited angle photoacoustic tomography. Phys. Med. Biol. 2009;54:3303-3314.

    Not in our library10.1088/0031-9155/54/11/002

  119. 119.

    Kunyansky L.A. Thermoacoustic tomography with detectors on an open curve: An efficient reconstruction algorithm. Inverse Probl. 2008;24.

    Not in our library10.1088/0266-5611/24/5/055021

  120. 120.

    Haltmeier M., Scherzer O., Zangerl G. A reconstruction algorithm for photoacoustic imaging based on the nonuniform FFT. IEEE Trans. Med. Imaging. 2009;28:1727-1735.

    Not in our library10.1109/tmi.2009.2022623

  121. 121.

    Donoho D.L. Compressed sensing. IEEE Trans. Inf. Theory. 2006;52:1289-1306.

    Not in our library

  122. 122.

    Candes E.J., Wakin M.B. An introduction to compressive sampling. IEEE Signal Process. Mag. 2008;25:21-30.

    Not in our library

  123. 123.

    Lustig M., Donoho D., Pauly J.M. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn. Reson. Med. 2007;58:1182-1195.

    Not in our library10.1002/mrm.21391

  124. 124.

    Provost J., Lesage F. The application of compressed sensing for photo-acoustic tomography. IEEE Trans. Med. Imaging. 2009;28:585-594.

    Not in our library10.1109/tmi.2008.2007825

  125. 125.

    Guo Z.J., Li C.H., Song L.A., Wang L.H.V. Compressed sensing in photoacoustic tomography in vivo. J. Biomed. Opt. 2010;15.

    Not in our library10.1117/1.3381187

  126. 126.

    Meng J., Wang L.H.V., Liang D., Song L. In vivo optical-resolution photoacoustic computed tomography with compressed sensing. Opt. Lett. 2012;37:4573-4575.

    Not in our library10.1364/ol.37.004573

  127. 127.

    Wang L., Jacques S.L., Zheng L. MCML-Monte Carlo modeling of light transport in multi-layered tissues. Comput. Methods Programs Biomed. 1995;47.

    Not in our library10.1016/0169-2607(95)01640-f

  128. 128.

    Bu S.H., Liu Z.B., Shiina T., Kondo K., Yamakawa M., Fukutani K., Someda Y., Asao Y. Model-based reconstruction integrated with fluence compensation for photoacoustic tomography. IEEE Trans. Biomed. Eng. 2012;59:1354-1363.

    Not in our library10.1109/tbme.2012.2187649

  129. 129.

    Zhu C., Liu Q. Review of monte carlo modeling of light transport in tissues. J. Biomed. Opt. 2013;18.

    Not in our library10.1117/1.jbo.18.5.050902

  130. 130.

    Arridge S.R., Schweiger M., Hiraoka M., Delpy D.T. A finite element approach for modeling photon transport in tissue. Med. Phys. 1993;20:299-309.

    Not in our library10.1118/1.597069

  131. 131.

    Arridge S.R. Optical tomography in medical imaging. Inverse Probl. 1999;15:R41-R93.

    Not in our library

  132. 132.

    Tarvainen T., Vauhkonen M., Kolehmainen V., Arridge S.R., Kaipio J.P. Coupled radiative transfer equation and diffusion approximation model for photon migration in turbid medium with low-scattering and non-scattering regions. Phys. Med. Biol. 2005;50:4913-4930.

    Not in our library10.1088/0031-9155/50/20/011

  133. 133.

    Paulsen K.D., Jiang H. Spatially varying optical property reconstruction using a finite element diffusion equation approximation. Med. Phys. 1995;22:691-701.

    Not in our library10.1118/1.597488

  134. 134.

    Kienle A., Patterson M.S. Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium. J. Opt. Soc. Am. A Opt. Image Sci. Vis. 1997;14:246-254.

    Not in our library10.1364/josaa.14.000246

  135. 135.

    Cox B.T., Arridge S.R., Kostli K.P., Beard P.C. Quantitative Photoacoustic imaging: Fitting a model of light transport to the initial pressure distribution. Proc. SPIE. 2005;5697:49-55.

    Not in our library

  136. 136.

    Cox B.T., Arridge S.R., Kostli K.P., Beard P.C. Two-dimensional quantitative photoacoustic image reconstruction of absorption distributions in scattering media by use of a simple iterative method. Appl. Opt. 2006;45:1866-1875.

    Not in our library10.1364/ao.45.001866

  137. 137.

    Maslov K., Zhang H.F., Hu S., Wang L.V. Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries. Opt. Lett. 2008;33:929-931.

    Not in our library10.1364/ol.33.000929

  138. 138.

    Jetzfellner T., Razansky D., Rosenthal A., Schulz R., Englmeier K.H., Ntziachristos V. Performance of iterative optoacoustic tomography with experimental data. Appl. Phys. Lett. 2009;95.

    Not in our library10.1063/1.3167280

  139. 139.

    Shao P., Cox B., Zemp R.J. Estimating optical absorption, scattering, and Grueneisen distributions with multiple-illumination photoacoustic tomography. Appl. Opt. 2011;50:3145-3154.

    Not in our library10.1364/ao.50.003145

  140. 140.

    Laufer J., Cox B., Zhang E., Beard P. Quantitative determination of chromophore concentrations from 2D photoacoustic images using a nonlinear model-based inversion scheme. Appl. Opt. 2010;49:1219-1233.

    Not in our library10.1364/ao.49.001219

  141. 141.

    Bal G., Uhlmann G. Inverse diffusion theory of photoacoustics. Inverse Probl. 2010;26.

    Not in our library10.1088/0266-5611/26/8/085010

  142. 142.

    Shao P., Harrison T., Zemp R.J. Iterative algorithm for multiple illumination photoacoustic tomography (MIPAT) using ultrasound channel data. Biomed. Opt. Express. 2012;3:3240-3249.

    Not in our library10.1364/boe.3.003240

  143. 143.

    Rosenthal A., Razansky D., Ntziachristos V. Quantitative optoacoustic signal extraction using sparse signal representation. IEEE Trans. Med. Imaging. 2009;28:1997-2006.

    Not in our library10.1109/tmi.2009.2027116

  144. 144.

    Cox B.T., Laufer J.G., Beard P.C. Quantitative photoacoustic image reconstruction using fluence dependent chromophores. Biomed. Opt. Express. 2010;1:201-208.

    Not in our library10.1364/boe.1.000201

  145. 145.

    Yin L., Wang Q., Zhang Q.Z., Jiang H.B. Tomographic imaging of absolute optical absorption coefficient in turbid media using combined photoacoustic and diffusing light measurements. Opt. Lett. 2007;32:2556-2558.

    Not in our library10.1364/ol.32.002556

  146. 146.

    Li X.Q., Xi L., Jiang R.X., Yao L., Jiang H.B. Integrated diffuse optical tomography and photoacoustic tomography: Phantom validations. Biomed. Opt. Express. 2011;2:2348-2353.

    Not in our library10.1364/boe.2.002348

  147. 147.

    Bauer A.Q., Nothdurft R.E., Erpelding T.N., Wang L.H.V., Culver J.P. Quantitative photoacoustic imaging: Correcting for heterogeneous light fluence distributions using diffuse optical tomography. J. Biomed. Opt. 2011;16.

    Not in our library10.1117/1.3626212

  148. 148.

    Maslov K., Zhang H.F., Wang L.V. Effects of wavelength-dependent fluence attenuation on the noninvasive photoacoustic imaging of hemoglobin oxygen saturation in subcutaneous vasculature. in vivo. Inverse Probl. 2007;23:S113-S122.

    Not in our library

  149. 149.

    Kostli K.P., Frenz M., Weber H.P., Paltauf G., Schmidt-Kloiber H. Optoacoustic infrared spectroscopy of soft tissue. J. Appl. Phys. 2000;88:1632-1637.

    Not in our library

  150. 150.

    Laufer J., Delpy D., Elwell C., Beard P. Quantitative spatially resolved measurement of tissue chromophore concentrations using photoacoustic spectroscopy: Application to the measurement of blood oxygenation and haemoglobin concentration. Phys. Med. Biol. 2007;52:141-168.

    Not in our library10.1088/0031-9155/52/1/010

  151. 151.

    Sethuraman S., Amirian J.H., Litovsky S.H., Smalling R.W., Emelianov S.Y. Spectroscopic intravascular photoacoustic imaging to differentiate atherosclerotic plaques. Opt. Express. 2008;16:3362-3367.

    Not in our library10.1364/oe.16.003362

  152. 152.

    Razansky D. Multispectral optoacoustic tomography-volumetric color hearing in real time. IEEE J. Sel. Top. Quantum Electron. 2012;18:1234-1243.

    Not in our library

  153. 153.

    Van Veen R.L.P., Sterenborg H.J., Pifferi A., Torricelli A., Cubeddu R. Determination of VIS-NIR Absorption Coefficients of Mammalian Fat, with Time- and Spatially Resolved Diffuse Reflectance and Transmission Spectroscopy. 2004.

    Not in our library10.1117/1.2085149

  154. 154.

    Ashkenazi S., Huang S.W., Horvath T., Koo Y.E.L., Kopelman R., Kopelman R. Photoacoustic probing of fluorophore excited state lifetime with application to oxygen sensing. J. Biomed. Opt. 2008;13.

    Not in our library10.1117/1.2927466

  155. 155.

    Huang S.W., Eary J.F., Jia C.X., Huang L.Y., Ashkenazi S., O'Donnell M. Differential-absorption photoacoustic imaging. Opt. Lett. 2009;34:2393-2395.

    Not in our library10.1364/ol.34.002393

  156. 156.

    Glatz J., Deliolanis N.C., Buehler A., Razansky D., Ntziachristos V. Blind source unmixing in multi-spectral optoacoustic tomography. Opt. Express. 2011;19:3175-3184.

    Not in our library10.1364/oe.19.003175

  157. 157.

    Joliffe I.T. Principal Component Analysis. 2002.

    Not in our library

  158. 158.

    Hyvarinen A., Oja E. Independent component analysis: Algorithms and applications. Neural Netw. 2000;13:411-430.

    Not in our library10.1016/s0893-6080(00)00026-5

  159. 159.

    Maslov K., Stoica G., Wang L.H.V. In vivo dark-field reflection-mode photoacoustic microscopy. Opt. Lett. 2005;30:625-627.

    Not in our library10.1364/ol.30.000625

  160. 160.

    Taruttis A., Wildgruber M., Kosanke K., Beziere N., Licha K., Haag R., Aichler M., Walch A., Rummeny E., Ntziachristos V. Multispectral optoacoustic tomography of myocardial infarction. Photoacoustics. 2013;1:3-8.

    Not in our library10.1016/j.pacs.2012.11.001

  161. 161.

    Buehler A., Herzog E., Ale A., Smith B.D., Ntziachristos V., Razansky D. High resolution tumor targeting in living mice by means of multispectral optoacoustic tomography. EJNMMI Res. 2012;2.

    Not in our library10.1186/2191-219x-2-14

  162. 162.

    Smith B.A., Akers W.J., Leevy W.M., Lampkins A.J., Xiao S.Z., Wolter W., Suckow M.A., Achilefu S., Smith B.D. Optical imaging of mammary and prostate tumors in living animals using a synthetic near infrared Zinc(II)-dipicolylamine probe for anionic cell surfaces. J. Am. Chem. Soc. 2010;132:67-69.

    Not in our library10.1021/ja908467y

  163. 163.

    Xiao J.Y., Yao L., Sun Y., Sobel E.S., He J.S., Jiang H.B. Quantitative two-dimensional photoacoustic tomography of osteoarthritis in the finger joints. Opt. Express. 2010;18:14359-14365.

    Not in our library10.1364/oe.18.014359

  164. 164.

    Razansky D., Harlaar N.J., Hillebrands J.L., Taruttis A., Herzog E., Zeebregts C.J., van Dam G.M., Ntziachristos V. Multispectral optoacoustic tomography of matrix metalloproteinase activity in a vulnerable human carotid plaque. Mol. Imaging Biol. 2012;14:277-285.

    Not in our library10.1007/s11307-011-0502-6

  165. 165.

    Nissen S.E., Yock P. Intravascular ultrasound-novel pathophysiological insights and current clinical applications. Circulation. 2001;103:604-616.

    Not in our library10.1161/01.cir.103.4.604

  166. 166.

    Razansky R.N., Rosenthal A., Mallas G., Razansky D., Jaffer F.A., Ntziachristos V. Near-infrared fluorescence catheter system for two-dimensional intravascular imaging. in vivo. Opt. Express. 2010;18:11372-11381.

    Not in our library10.1364/oe.18.011372

  167. 167.

    Sethuraman S., Aglyamov S.R., Amirian J.H., Smalling R.W., Emelianov S.Y. Intravascular photoacoustic imaging using an IVUS imaging catheter. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2007;54:978-986.

    Not in our library10.1109/tuffc.2007.343

  168. 168.

    Yang J.M., Maslov K., Yang H.C., Zhou Q.F., Shung K.K., Wang L.H.V. Photoacoustic endoscopy. Opt. Lett. 2009;34:1591-1593.

    Not in our library10.1364/ol.34.001591

  169. 169.

    Jansen K., van der Steen A.F.W., van Beusekom H.M.M., Oosterhuis J.W., van Soest G. Intravascular photoacoustic imaging of human coronary atherosclerosis. Opt. Lett. 2011;36:597-599.

    Not in our library10.1364/ol.36.000597

  170. 170.

    ANSI Standard Z136. 1-2000: For Safe Use of Lasers. 2000.

    Not in our library

  171. 171.

    Deán-Ben X.L., Razansky D., Ntziachristos V. The effects of acoustic attenuation in optoacoustic signals. Phys. Med. Biol. 2011;56.

    Not in our library10.1088/0031-9155/56/18/021

  172. 172.

    Razansky D., Baeten J., Ntziachristos V. Sensitivity of molecular target detection by multispectral optoacoustic tomography (MSOT). Med. Phys. 2009;36.

    Not in our library10.1118/1.3077120

  173. 173.

    Ku G., Wang X., Xie X., Stoica G., Wang L.V. Imaging of tumor angiogenesis in rat brains in vivo by photoacoustic tomography. Appl. Opt. 2005;44:770-775.

    Not in our library10.1364/ao.44.000770

  174. 174.

    Stein E.W., Maslov K., Wang L.V. Noninvasive, in vivo imaging of blood-oxygenation dynamics within the mouse brain using photoacoustic microscopy. J. Biomed. Opt. 2009;14.

    Not in our library10.1117/1.3095799

  175. 175.

    Fainchtein R., Stoyanov B.J., Murphy J.C., Wilson D.A., Hanley D.F. Local Determination of Hemoglobin Concentration and Degree of Oxygenation in Tissue by Pulsed Photoacoustic Spectroscopy. Proc. SPIE. 2000;3916:19-33.

    Not in our library

  176. 176.

    Laufer J., Elwell C., Delpy D., Beard P. In vitro measurements of absolute blood oxygen saturation using pulsed near-infrared photoacoustic spectroscopy: Accuracy and resolution. Phys. Med. Biol. 2005;50:4409-4428.

    Not in our library10.1088/0031-9155/50/18/011

  177. 177.

    Zhang H.F., Maslov K., Sivaramakrishnan M., Stoica G., Wang L.V. Imaging of hemoglobin oxygen saturation variations in single vessels in vivo using photoacoustic microscopy. Appl. Phys. Lett. 2007;90.

    Not in our library10.1063/1.2435697

Cited by papers we hold 2

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.