HPLC-DAD Determination of Nitrite and Nitrate in Human Saliva Utilizing a Phosphatidylcholine Column
Małgorzata Tatarczak-Michalewska, Jolanta Flieger, Justyna Kawka and 4 more
Molecules · 2019
doi:10.3390/molecules24091754 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
The aim of this research was to optimize the separation and quantitative determination of nitrites and nitrates in human saliva. HPLC with UV absorption (HPLC/DAD) using a phosphatidylcholine column (IAM.PC.DD2 Regis HPLC) was applied in this assay. Nitrates were detected directly by their absorbance at 210 nm, whereas nitrites were detected after oxidation to nitrates by potassium permanganate at acidic conditions. The kinetics of the permanganate−nitrite reaction was measured chromatographically. The calibration graph for nitrates was linear in the range of 0.5−35 µg mL<sup>−1</sup> with a correlation coefficient of 0.9999. The limit of detection was 4.56 ng mL<sup>−1</sup>. The calibration graph for nitrites (after oxidation to nitrates) was linear in the range of 0.5−15 µg mL<sup>−1</sup> with a correlation coefficient of 0.9972. The limit of detection was 4.21 ng mL<sup>−1</sup>. The nitrate concentrations in the saliva samples were found in the range of 8.98−18.52 μg mL<sup>−1</sup>, whereas nitrite was in the range of 3.50−5.34 μg mL<sup>−1</sup>.
Abstract licence: CC BY
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.