Amperometric Biosensor for Oxalate Determination in Urine Using Sequential Injection Analysis
Enrique Barrado, Yolanda Castrillejo, Veronica Salazar and 2 more
Molecules · 2012
doi:10.3390/molecules17088859 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
An amperometric flow biosensor for oxalate determination in urine samples after enzymatic reaction with oxalate oxidase immobilized on a modified magnetic solid is described. The solid was magnetically retained on the electrode surface of an electrode modified with Fe (III)-tris-(2-thiopyridone) borate placed into a sequential injection system preceding the amperometric detector. The variables involved in the system such as flow rate, aspired volumes (modified magnetic suspension and sample) and reaction coil length were evaluated using a Taguchi parameter design. Under optimal conditions, the calibration curve of oxalate was linear between 3.0–50.0 mg·L<sup>−1</sup>, with a limit of detection of 1.0 mg·L<sup>−1</sup>. The repeatability for a 30.0 mg·L<sup>−1</sup> oxalate solution was 0.7%. The method was validated by comparing the obtained results to those provided by the spectrophotometric method; no significant differences were observed.
Abstract licence: CC BY
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.