Detection of Human Plasma Glucose Using a Self-Powered Glucose Biosensor
Gymama Slaughter, Tanmay Kulkarni
Energies · 2019
doi:10.3390/en12050825 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
This work presents the characterization of a self-powered glucose biosensor using individual sequential assays of human plasma glucose obtained from diabetic patients. The self-powered glucose biosensor is exploited to optimize the assay parameters for sensing plasma glucose levels. In particular, the biofuel cell component of the system at pH 7.4, 37 °C generates a power density directly proportional to plasma glucose and exhibited a maximum power density of 0.462 mW·cm−2 at a cell voltage of 0.213 V in 5 mM plasma glucose. Plasma glucose is further sensed by monitoring the charge/discharge frequency (Hz) of the integrated capacitor functioning as the transducer. With this method, the plasma glucose is quantitatively detected in 100 microliters of human plasma with unprecedented sensitivity, as high as 104.51 ± 0.7 Hz·mM−1·cm−2 and a detection limit of 2.31 ± 0.3 mM. The results suggest the possibility to sense human plasma glucose at clinically relevant concentrations without the use of an external power source.
Abstract licence: CC BY