The Possibility for Iodine Concentration Determination in a Phantom with Known Titers of an Iodine-Containing Contrast Agent, by Using Dual-Energy Computed Tomography

A. S. Chaban, V. E. Sinitsyn

Вестник рентгенологии и радиологии · 2020

doi:10.20862/0042-4676-2019-100-6-335-338 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Objective : to study the capabilities of single-source dual-energy computed tomography (DECT) in quantifying the concentration of iodine in solutions. Material and methods . Single-source DECT was performed using a phantom containing a set of 5 tubes with a different titer of the iodine-containing contrast agent Iopamidol. Further, the obtained images were used to construct iodine maps; and the concentration of iodine was measured within the volume of the titrated contrast agent. Results. Despite a high correlation between the measured iodine concentration in solution with the true concentration (Pearson's correlation coefficient r = 0.98; p < 0.01), there is a measurement error that was 4.8 to 23% at different dilutions. Conclusion. Signal-source rapid voltage switching DECT does not allow precise measurements of the true concentration of iodine in solution. To eliminate measurement errors in further in vivo studies using singlesource DECT, it may be that attention must be paid to the measurement of normalized iodine concentration.

Abstract licence: CC BY-NC

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.