Skip to content
naulon
Not checkedNot checked for notices yet.

Introducing a Dedicated Cardiac SPECT Imaging System: ProSPECT

Behnoosh Teimourian Fard, Kiarash Amirmozaffari Sabet, Mohmmadreza Ay

Frontiers in Biomedical Technologies · 2019

doi:10.18502/fbt.v6i3.1700 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Purpose: Cardiac SPECT imaging is widely used for many clinical practices such as the diagnostic of coronary artery diseases (CAD) and Myocardial Perfusion Imaging (MPI). In this work, we introduced the ProSPECT system, a dedicated cardiac SPECT system with open-gantry design. Materials and Methods: In this study we assessed the performance of the system based on NEMA-NU1-2007 standards. The ProSPECT system was characterized by measurement of planar, tomographic and clinical parameters. Results: Planar measurements showed 7.6 mm and 8.9% of spatial resolution and energy resolution at 140 keV, respectively. The tomographic resolution at 250 mm radius of rotation was 12.3 mm. Also, the maximum LOR and COR error for the tomographic imaging are obtained 1.1 mm and 1.8 mm, respectively. The clinical images that obtained using the system were confirmed by nuclear medicine physician expert. Conclusion: It concluded that the ProSPECT imaging is qualified for cardiac SPECT imaging.

Abstract licence: CC BY-NC

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.