Tracking blood pressure changes by means of non-invasive intermittent blood pressure measurements in clinical application

Huhle Robert, Richter Thorsten, Gama de Abreu Marcelo

Current Directions in Biomedical Engineering · 2021

doi:10.1515/cdbme-2021-2208 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Considering accuracy/precision cut-offs of 5 ± 8 mmHg and cut-off values for inter-class correlation coefficients (ICC=0.37...1, from DIN EN ISO 81060-2), absolute and relative errors in time independent measurement of blood pressure changes with non-invasive intermittent devices (NiBP) are derived mathematically for mean arterial blood pressure range of 40-180 mmHg. As a clinically relevant value for change of arterial blood pressure 20% of the baseline blood pressure is considered. The mean ratio between the change of BP measured by the NiBP and measured by the invasive reference device (TE%) were proposed as quality measure for the evaluation of NiBP device tracking capability. The proposed measure TE%is theoretically independent of absolute accuracy but depends on precision and ICC of a device. NiBP devices show considerable maximum TE% of 41% in tracking mean blood pressure changes respectively. In 10% of the measurements in the low blood pressure range TE% exceeding 100%. The mean 50th/90th TE% percentile over the whole blood pressure range were 25/61%, respectively. Furthermore, TE% was relatively insensitive to assumed blood

Abstract licence: CC BY-NC-ND

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.