Skip to content
naulon
Not checkedNot checked for notices yet.

Radiopacity assessment of neurovascular implants

Hoffmann Thomas, Gugel Sebastian, Beuing Oliver and 1 more

Current Directions in Biomedical Engineering · 2016

doi:10.1515/cdbme-2016-0118 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

An experimental method for radiopacity evaluation of neurovascular implants is presented. State of the art methods are described for cardiovasular implants. To extend the methods to neurovascular devices, imaging parameters were determined from clinical radiography data sets and protocols. To reduce background noise, a thresholding method is introduced. This method is evaluated on a setup with four different laser cut and braided stents. The results are compared to current evaluation methods without filtering. It is shown that unfiltered methods underestimate the radiopacity by 0.89 mm ± 0.21 mm (aluminum equivalent). The introduced method is highly repeatable and uses modern clinical imaging technology. Thus an image evaluation under realisitc conditions is possible. Slightly visible vascular devices and implants can be compared with high accuracy.

Abstract licence: CC BY-NC-ND

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.