Determination of intralumenal individual bile acids by HPLC with charged aerosol detection
Maria Vertzoni, Helen Archontaki, Christos Reppas
Journal of Lipid Research · 2008
doi:10.1194/jlr.d800039-jlr200 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
An isocratic HPLC charged aerosol detector (CAD) method was developed, validated, and applied for the determination of individual bile acids in human gastric and duodenal aspirates. The method requires a low volume of aspirates (50–100 μl) and minimal sample pretreatment. A Hypersil BDS RP-C18 column (250 × 4.6 mm, 5 μm particle size) was equilibrated with a mobile phase composed of methanol-[ammoniun formate 20 mM, formic acid 0.5%, triethylamine 0.2% (pH 3)] 67:33 v/v. Its flow rate was 1 ml/min. The elution times for taurocholate, glycocholate, taurochenodeoxycholate, ursodeoxycholate, glycochenodeoxycholate, cholate, and glycodeoxycholate were approximately 9.9, 16.2, 18.2, 21.3, 31.6, 34.5, and 38.5 min, respectively. Calibration curves in the mobile phase were constructed in the concentration range of 0.5–500 μM. Limits of detection and quantification were in the range of 0.07–0.60 μM and 0.20–1.80 μM, respectively. This method was applied first, in gastric aspirates collected in the fasted state, in which bile acid presence is minimal and, second, in duodenal aspirates collected in the fed state, in which a large number of potentially interfering compounds exists. Intra-day relative standard deviation in fasted gastric aspirates and in fed duodenal aspirates was less than 2.2% and 6.0%, respectively.
Abstract licence: CC BY-NC-ND
Cited by papers we hold1
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.