Skip to content
naulon

Not checkedNot checked for notices yet.

Bearing failure in a mobile bearing unicompartmental knee arthroplasty: an uncommon presentation of an implant-specific complication

Sravya P. Vajapey, Paul M. Alvarez, Douglas Chonko

Arthroplasty · 2021

doi:10.1186/s42836-021-00073-9 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Background We present two cases of unicompartmental knee arthroplasty (UKA) bearing failure in this report—one case of bearing dislocation and one case of bearing fracture. The causes of failure in both cases are evaluated in depth and recommendations are provided regarding intraoperative technique to reduce risk of bearing failure in mobile bearing UKAs. Case presentation In the first case, intraoperative evidence of metallosis and chronic pain preceding the traumatic event may indicate that the patient had attenuation of her collateral ligaments that precipitated the instability event. In the second case, the relatively atraumatic nature of the bearing fracture-dislocation and intraoperative evidence of extensive poly wear suggest that the bearing fracture was likely due to a 3-mm bearing selection in the initial surgery. Conclusions This case report shows that late bearing in mobile bearing unicompartmental knee arthroplasty can often be a multifactorial event and treatment must address all the risk factors that led to bearing dislocation. Bearing fracture is a very rare complication associated with mobile bearing UKA and patients with thin polyethylene inserts are at risk for bearing fracture even in the absence of poly wear.

Abstract licence: CC BY

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.