Not checkedNot checked for notices yet.
Clinical application and technical details of cook zenith devices modification to treat urgent and elective complex aortic aneurysms
Jesse Manunga, Lia Jordano, Aleem K. Mirza and 3 more
CVIR Endovascular · 2021
doi:10.1186/s42155-021-00233-7 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
Purpose To describe technical details of modifying four different Cook Zenith devices to treat complex aortic aneurysms. Material In the first three cases, the modification process involved complete stent graft deployment on a sterile back table. Fenestrations were created using an ophthalmologic cautery and reinforced with a radiopaque snare using a double-armed 4–0 Ethibond locking suture based on measurements obtained on centerline of flow. In each instance, a nitinol wire was withdrawn and redirected through and through the fabric and used as a constraining wire. In the fourth patient, modification involved partial stent graft deployment and creation of additional two fenestrations to accommodate renal arteries. The devices are resheathed and implanted in the standard fashion. Results Four patients underwent exclusion of their aneurysms, including thoracoabdominal aneurysms ( n = 2), a contained ruptured juxtarenal aneurysm ( n = 1), and a ruptured failed previous endovascular repair ( n = 1). Fifteen fenestrations were successfully bridged with Atrium iCAST stent grafts. Average graft modification time, operative time, contrast volume, radiation dose, estimated blood loss, and hospital length of stay were 89 min, 155.25 min, 58.8 mL, 2451 mGy, 175 mL, and 4.3 days, respectively. One patient required a secondary intervention to treat a type Ib endoleak. During an average follow-up of 25 months, aneurysm sacs progressively shrank without additional intervention. Conclusion Physician-modified fenestrated/branched endografts are a safe alternative to custom made devices, especially in urgent cases and should be part of the armamentarium of any complex aortic program.
Abstract licence: CC BY
References11
11 more references, whose wording the paper’s licence keeps private.
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.