Spurious electroencephalographic activity due to pulsation artifact in the depth of anesthesia monitor

Kotoe Kamata, Tarmo Lipping, Arvi Yli-Hankala and 2 more

JA Clinical Reports · 2021

doi:10.1186/s40981-021-00441-z (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Abstract Background The depth of anesthesia (DOA) is estimated based on the anesthesia-induced electroencephalogram (EEG) changes. However, the surgical environment, as well as the patient him/herself, generates electrical interferences that cause EEG waveform distortion. Case presentation A 52-year-old patient required general anesthesia due to the right femur necrotizing fasciitis. He had no history of epilepsy or head injury. His cardiovascular status was stable without arrhythmia under propofol and remifentanil anesthesia. The DOA was evaluated with Root® with SedLine® Brain Function Monitoring (Masimo Inc, Irvine, CA). The EEG showed a rhythmic, heart rate time-locked pulsation artifact, which diminished after electrode repositioning. Offline analysis revealed that the pulse wave-like interference in EEG was observed at the heart rate frequency. Conclusions We experienced an anesthesia case that involves a pulsation artifact generated by the superficial temporal artery contaminating the EEG signal. Numerous clinical conditions, including pulsation artifact, disturb anesthesia EEG.

Abstract licence: CC BY

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.