Malenka M. Bissell, Francesca Raimondi, Lamia Ait Ali and 38 more (opens in a new tab)

Journal of Cardiovascular Magnetic Resonance · 2023 · Review · CC BY

doi:10.1186/s12968-023-00942-z (opens in a new tab)

Abstract

Hemodynamic assessment is an integral part of the diagnosis and management of cardiovascular disease. Four-dimensional cardiovascular magnetic resonance flow imaging (4D Flow CMR) allows comprehensive and accurate assessment of flow in a single acquisition. This consensus paper is an update from the 2015 '4D Flow CMR Consensus Statement'. We elaborate on 4D Flow CMR sequence options and imaging considerations. The document aims to assist centers starting out with 4D Flow CMR of the heart and great vessels with advice on acquisition parameters, post-processing workflows and integration into clinical practice. Furthermore, we define minimum quality assurance and validation standards for clinical centers. We also address the challenges faced in quality assurance and validation in the research setting. We also include a checklist for recommended publication standards, specifically for 4D Flow CMR. Finally, we discuss the current limitations and the future of 4D Flow CMR. This updated consensus paper will further facilitate widespread adoption of 4D Flow CMR in the clinical workflow across the globe and aid consistently high-quality publication standards.

Full text

From PubMed Central, U.S. National Library of Medicine, as current on Oct 27, 2026. Paragraph numbers are ours, so an answer can point at one.

Introduction

This is an update to the 4D Flow CMR Consensus Statement published in 2015.

Hemodynamics evaluation is crucial for the assessment of cardiovascular diseases, and is essential for understanding pathophysiology and explaining clinical manifestations. Four-dimensional cardiovascular magnetic resonance flow imaging (4D Flow CMR) uniquely provides comprehensive, in vivo characterization of cardiovascular blood flow. With this approach, the blood flow velocity is measured through motion encoding in all three spatial directions and resolved relative to all three dimensions of space and to the dimension of time along the cardiac cycle (3D + time = 4D).

4D Flow CMR is an extension of 2D Flow CMR which is currently the most used clinical flow application. Visualization of flow direction and magnitude are also valuable in clinical practice. More advanced quantification parameters are to date still largely confined to the research arena.

Several review papers are now available on 4D Flow CMR, detailing its advantages over 2D Flow as well as descriptions of useful clinical applications, especially for aortic disease, but also congenital heart diseases, particularly in the neonatal population, and other cardiovascular conditions. We therefore will not be covering detailed benefits of 4D flow CMR and its clinical application in this consensus statement.

The previously published consensus statement covers background information, clinical and scientific significance, and potential utility. Its recommendations regarding patient preparation, 4D Flow CMR data acquisition, data pre-processing, and flow visualization remain valid.

Since the publication of the original consensus statement, the field of 4D Flow CMR and the size of its user base has grown, supported by advances in CMR scanner hardware and coils, data acquisition and reconstruction strategies, vendor support, and availability of commercial post-processing solutions. Key advances in the last five years include further acceleration and diversification in acquisition methods. However, the most important development has been that 4D Flow CMR is now clinically available and supported by the major vendors of CMR scanners. Additionally, post-processing tools are commercially available, United States Food and Drug Administration (FDA) approved, European (CE-) marked for clinical use and in some countries, approved for reimbursement. These developments have enlarged the user base and paved the way for the more widespread clinical application of 4D Flow CMR, which is now used in the clinical routine at multiple centers worldwide. This is prompting large cohort, longitudinal and multi-center clinical studies. As the variety of acquisition and analysis platforms grows, standardized imaging acquisition, analysis, and publication approaches will simplify pooling data for meta-analysis studies and increase the validity of study results.

It is important to note that clinical and research 4D Flow CMR applications have differing priorities. Clinical acquisitions need to be fast with reliable flow and velocity quantification. In research, scan duration is less important whereas comprehensiveness of data is prioritized. Conversely, validation is more complex in the research setting as there often is no predefined gold standard for comparison for advanced measures beyond velocity and flow.

This update statement builds on the previously published consensus statement and focuses on:Recommended acquisition parameters for clinical use—with a growing number of clinical centers starting out in 4D Flow CMR we have summarized updated clinical parameter recommendations based on consensus from centers clinically using 4D Flow CMR.Clinical post-processing workflow—this section describes key elements to consider and follow when choosing a clinical post-processing platform and setting up a clinical workflow.Quality assurance and validation advice—with a growing number of different sequences and post-processing platforms commercially available, we have further extended the advice for clinical quality assurance and validation when starting out with 4D Flow CMR. Furthermore, we address the challenges faced in quality assurance and validation in the research setting.Integration into clinical practice—this section covers a selection of advice from centers that have integrated 4D Flow CMR into the clinical workflow.Recommended publication standards—this section provides a checklist specific for 4D Flow CMR.Overcoming limitations and future considerations—focusing on what is on the horizon for 4D Flow CMR.Appendix: 4D Flow CMR sequence options—since the last consensus statement, the options of available 4D Flow CMR sequences have considerably increased. This appendix, therefore, summarizes some aspects to consider when choosing a sequence prior to setting up a research study or clinical service.

Recommended acquisition parameters for clinical use—with a growing number of clinical centers starting out in 4D Flow CMR we have summarized updated clinical parameter recommendations based on consensus from centers clinically using 4D Flow CMR.

Clinical post-processing workflow—this section describes key elements to consider and follow when choosing a clinical post-processing platform and setting up a clinical workflow.

Quality assurance and validation advice—with a growing number of different sequences and post-processing platforms commercially available, we have further extended the advice for clinical quality assurance and validation when starting out with 4D Flow CMR. Furthermore, we address the challenges faced in quality assurance and validation in the research setting.

Integration into clinical practice—this section covers a selection of advice from centers that have integrated 4D Flow CMR into the clinical workflow.

Recommended publication standards—this section provides a checklist specific for 4D Flow CMR.

Overcoming limitations and future considerations—focusing on what is on the horizon for 4D Flow CMR.

Appendix: 4D Flow CMR sequence options—since the last consensus statement, the options of available 4D Flow CMR sequences have considerably increased. This appendix, therefore, summarizes some aspects to consider when choosing a sequence prior to setting up a research study or clinical service.

This consensus update is based on published data, where available, and consensus experience. It aims to cover a large audience, including clinicians and scientists interested in starting out in 4D Flow CMR as well as bringing together established groups in the area.

Advised acquisition parameters for clinical use

The choice of 4D Flow CMR acquisition parameters requires careful consideration of a balance between accuracy and scan time. For clinical use, it is advisable to keep the 4D Flow CMR acquisition to 5–10 min. Thereby it can easily be added to a clinical workflow, such as in the waiting period between gadolinium administration late enhancement enhancement (LGE) imaging, without interfering with or extensively prolonging the established protocols in the institution.

Equipment and set-up

The improved signal to noise ratio (SNR) at higher field strength such as 3T can be beneficial in the younger pediatric setting given the higher spatial resolution needed due to the small body size anatomy but is less important in older children and adults where body size is sufficient for good SNR at lower field strengths.

Coil selection largely depends on local protocol and availability. The routine number of coil elements used in standard cardiac imaging is sufficient for good quality 4D Flow CMR acquisition. The number of coil elements needs to be balanced against the ability of the scanner to reconstruct the data in a timely manner.

Volume coverage ideally includes at least the valves and aortic and pulmonary sinuses (even if focusing on intra-cardiac anatomy) for data quality assurance purposes (see section “Quality assurance and validation advice for clinical use”). Appropriate field-of-view that fully covers the anatomy of interest (plus a couple of additional slices) can be confirmed using anatomical scout images on the scanner.

For ease of use, it is best to aim for standardized protocols which might need to be adjusted for individual pathologies. Congenital heart disease centers in particular might have a variety of protocols for different age groups and/or pathologies. These should all be individually validated (see section “Quality assurance and validation advice for clinical use”).

Scan parameters

Accuracy and precision in flow imaging are influenced by several physiological patient parameters. Different body sizes and heart rates influence spatial and temporal resolutions, SNR, and therefore velocity-to-noise ratio (VNR). Hence, we recommend adjusting the spatial resolution for different age groups. Voxels should be isotropic and at least 6 voxels should cover a vessel diameter. Specific resolution guidance based on the most common resolutions used in clinical centers are detailed in Table 1.

It is important to note here the difference between acquired resolution, based on the field-of-view and k-space matrix size, and the reconstructed resolution, which is often higher than the acquired due to the use of spatial interpolation during image reconstruction. This is an important distinction, since sequence performance is primarily defined by the acquired rather than the reconstructed resolution. For completeness and easier comparison, both acquired and reconstructed resolutions should be stated in scientific publications.

4D Flow CMR requires the selection of an upper velocity encoding limit (VENC) during scan prescription to avoid velocity aliasing. This setting will adjust the motion encoding gradients accordingly to the desired motion sensitivity. In non-contrast acquisitions, the VENC should be as low as possible to keep adequate VNR and improve accuracy while avoiding aliasing.

Choice of VENC should be close to the maximum velocity (< 25% above) and can be guided by a previous imaging examination when available, such as a recently acquired echocardiogram or a previous CMR study. Otherwise, a 2D phase-contrast acquisition or rapid velocity scout sequence can be used at the aortic valve or area of interest. If 4D Flow CMR is acquired without a previous 2D phase-contrast acquisition and stenosis is suspected, consider an initial VENC at 250 cm/s. If no stenosis is suspected, the following VENC can be used as guidance:Large vessels (pulmonary artery and aorta): 150 cm/sDissection false lumen: 50–150 cm/sVenous blood flow (including extracardiac conduit and pulmonary arteries in Fontan patients): 50–80 cm/sIntra-cardiac: 100–150 cm/s

Large vessels (pulmonary artery and aorta): 150 cm/s

Dissection false lumen: 50–150 cm/s

Venous blood flow (including extracardiac conduit and pulmonary arteries in Fontan patients): 50–80 cm/s

Intra-cardiac: 100–150 cm/s

Electrocardiographic (ECG) gating should be retrospective whenever possible to capture hemodynamics throughout the complete cardiac cycle. Operators should monitor the ECG signal and acquisition time estimates to determine if electrodes require repositioning, as poor ECG signals can lead to prolonged scans and reduced image quality and accuracy. Irregular heartbeats can be a challenge, but 4D Flow CMR acquisition can still be accurate in patients with atrial fibrillation but is not always reliable. Prospective gating and arrhythmia rejection might be useful in these cases.

Respiratory motion suppression can improve image quality and does not automatically increase scan time, and guidance should be available from vendors or sequence developers whether respiratory motion suppression is advised for the specific sequence. In practice many clinical centres do not use respiratory motion suppression as it is not available from all vendors for all sequences. Furthermore, in patients with fast heart rates (HR; such as neonates or during stress or exercise with HR > 120 bpm) diaphragmatic respiratory navigators are often not feasible. If respiratory suppression is desirable, other respiratory gating methods such as self-gating or respiratory bellows can be considered. When self-gating becomes readily available, this would be the recommended gating method.

Contrast agent and flip angle

Advised scan parameters are summarized in Table 1.

As many acquisition parameters are likely to influence 4D Flow CMR acquisition, it is important to have consistent protocols which have undergone local quality assurance testing.

The spoiled gradient-echo sequence with short repetition time (TR) generates phase-contrast angiograms without the need for contrast agents. SNR and VNR improve with T1 shortening achieved by administering gadolinium-based contrast or superparamagnetic iron oxide agent (ferumoxytol). Therefore, contrast administration to enhance image quality can be useful (but is not essential), especially in challenging cases such as neonates and dissections to enhance image quality, but good image quality can be achieved without contrast administration especially when scanning at 3T. In adults, if contrast is given for other reasons, it is useful to perform the 4D Flow CMR acquisition after gadolinium-based contrast administration. It is important to note, that contrast administration for 4D Flow CMR acquisition alone is not required especially as the wider CMR community is moving more towards non-contrast acquisitions.

The standard flip angle for non-contrast 4D Flow CMR acquisition should be set around the Ernst angle which equates to a flip angle of around 7° for non-contrast 4D Flow CMR with repetition time and echo time chosen as short as possible. After contrast administration a higher flip angle is often beneficial, but this depends on contrast agent used and time past since administration. Several clinical centres use the following guidance: If acquiring 4D Flow CMR directly after gadolinium administration, it is advisable to increase the flip angle to 15–25° 1.5T and 12° at 3T.. If 4D Flow CMR is acquired after LGE a lower flip angle (similar to non-contrast values) is likely needed. If using ferumoxytol a higher flip angle of around 15–25° is often required. In neonates, limits in specific absorption rate (SAR) often necessitate dropping the flip angle with ferumoxytol to 12° in this patient cohort.

Clinical post-processing workflow

Data pre-processing steps are described in detail by the previous 4D Flow CMR consensus statement and this remains valid. Key elements are summarized in Fig. 1.

Post-processing of 4D Flow CMR data includes the following steps: (1) background phase offset correction, (2) anti-aliasing if required, (3) segmentation, (4) visualization (optional), and (5) quantification including an internal consistency check (Fig. 1).

There are several commercially available software packages for post-processing and analysis of 4D Flow CMR. Most software packages have regulatory approval for basic flow quantification in clinical routine. In addition, they allow visualization of blood flow and the analysis of various advanced research parameters such as flow eccentricity, vortices, kinetic energy (KE), flow components as well as relative pressure distribution for research purposes. Many software packages now also include valve tracking, circumventing the issue of through-plane motion and permitting valve motion to be factored in when computing flow, which improves accuracy especially in mitral and tricuspid valve assessment.

Step 1: Background phase offset correction

For accurate flow measurements, 4D Flow CMR requires correction for phase offset errors associated with eddy currents and concomitant gradient fields if not corrected during image reconstruction. While offset errors can be corrected by repeating the exam with a stationary phantom and subtracting the flow measurements of the static tissue from the patient’s data, this is too time-consuming for clinical practice. Static-tissue interpolation offset correction can be applied during post-processing with equivalent performance. All software should have this capability using linear or polynomial fits to static tissue. Particular attention should be given to large fields-of-view where regions-of-interest may reside far away from the magnet isocenter as offset errors increase with distance from the magnet isocenter.

Step 2: Velocity anti-aliasing

In cases where maximum blood flow velocity surpasses the chosen VENC, velocity aliasing can result in corrupted velocity measurements. In these cases, phase unwrapping can improve the accuracy of the flow and velocity measurements. Most software can detect a large shift in adjacent voxel velocity values and perform automatic correction. However, visual inspection of the peak systolic and diastolic cardiac phases is required to check all three primary velocity encoding directions for un-correctable velocity aliasing. Image regions affected by incorrigible aliasing, should not be considered for flow analysis.

Step 3: Segmentation

Depending on the software solution, evaluation of flow data starts either with 3D segmentation of the vessel or direct placement of regions of interest in the imaging volume delineating the vessel contour in 2D cross-sectional planes. Appropriate regions of interest selection, orientation, and segmentation are important parts of the flow and velocity quantification process. Care must be taken to select regions unaffected by artefacts, e.g., caused by partial volumes, metal implants, or motion. Vascular flow values should be measured in 2D planes that are orthogonal to the vessel. Regions of interest need to be propagated and adjusted throughout the cardiac cycle to account for vessel motion. Centerline-based plane positioning and registration-based contour propagation can support this process.

Valve tracking

Using retrospective valve tracking, a dynamic reformatted 2D plane of through-plane velocity is created from the time-resolved 3D velocity data (Fig. 2). Two orthogonal cine views per valve (for instance, left ventricular (LV) two-chamber and four-chamber views for the mitral valve) should be used to track the valve annulus over the whole cardiac cycle. Misalignment between the cine views and the 4D Flow CMR data should be resolved by manual or automatic image registration. It is advised to quantify regurgitant jets separately, by defining a reformatted plane perpendicular to the regurgitant jet. Aliasing in the regurgitant jet is common, as regurgitant flow is usually characterized by high blood velocity, turbulence, and incoherent flow and in these cases indirect quantification is advised (see section “Step 5: Quantification” below). The regurgitant jet regions of interest should be segmented and propagated in the reformatted 2D plane as described above (see Fig. 2).

Step 4: Visualization

Visualization can be performed using multiple tools such as velocity-based color coding, maximum velocity projections (“velocity MIP”), instantaneous streamlines, and time-resolved pathlines. The differences between streamlines and pathlines are described in detail in the first 4D Flow CMR Consensus Statement. We emphasize that streamlines do not represent flow pathways in pulsatile blood flow; keeping streamlines short minimizes the risk for misinterpretation. Visualization should include dynamic visualization of the complete 3D volume as well as localized visualization tools for the particular region of interest. Visualization can serve as a quick quality assessment in cases where velocity values are inverted. Visualization facilitates the detection and understanding of blood flow alterations in different pathologies, such as shunts or valve insufficiencies. Further detailed background information of 4D flow CMR visualization can be found in the 2015 consensus statement.

Step 5: Quantification

Guided by the visualization of anatomy and blood flow, 2D planes can be placed to measure flow parameters at anatomical landmarks or in areas of pathological flow. The most relevant clinical 4D Flow CMR-derived parameters are flow volumes and flow velocities that should be provided in the clinical report. Quantitative data should always be validated for internal consistency (see section “Quality assurance and validation advice for clinical use”).

The accuracy of blood flow can be compromised in certain flow geometries, and readers should be cautious of flow measurements in areas of high velocity flow jets, regions with substantial dephasing due to turbulence, and highly vortical blood flow, especially in ascending aortic aneurysm or aneurysmal pulmonary arteries. In these circumstances, alternative flow measurements outside of regions of abnormal flow or combined use of ventricular volumetry may be necessary to guide clinical management. Examples include using superior vena cava and descending aortic flow as net forward flow in the aorta. For evaluation of valvular regurgitant volume both direct and indirect jet quantification methods are used.

Direct jet quantification

The direct jet tracking method should be used in regurgitant lesions with only one central jet such as aortic and pulmonary regurgitation, in functional mitral regurgitation or atrioventricular valve regurgitation after atrioventricular septal defect correction. The advantage of this direct measurement approach is that no assumptions are made with respect to regurgitant jet morphology or mass conservation through other valves or over the atrial or ventricular septum, and that flow quantification over all four valves is performed from the same dataset from the same average cardiac cycle.

Indirect quantification method

The standard CMR method for mitral regurgitation quantification, here called indirect quantification, involves the subtraction of the aortic net LV ventricular stroke volume (SV) determined by LV cine short-axis volumetric assessment. Valve tracking has led to the improved indirect method involving subtraction of aortic net flow from the mitral forward flow. In cases where there are multiple jets with different directions or the regurgitation jet has uncorrectable aliasing, we recommend using the indirect method with valve tracking through the mitral and aortic valves as this has been shown to be more accurate in these cases.

While using any 4D Flow CMR method for assessment of valvular regurgitation, it is recommended to cross-check the quantification against standard methods. If there is a significant discrepancy in the quantification of regurgitation volume between methods (> 15 ml or > 10%), it is recommended to revisit the analysis and investigate the cause of the discrepancy using the conservation of mass principle (i.e. flow into and out of a chamber should be balanced). The following equations can be used to check the consistency of flow data:

LV stroke volume (short-axis cine segmentation) = mitral forward flow + aortic backward flow = aortic forward flow + mitral backward flow.

Retrospective valve tracking can be applied in patients with atrial fibrillation. However, a cautious approach should be used as there is a possibility of underestimating flows. In these cases, relative flow quantification, for example, regurgitation fraction, is possibly more reliable than absolute numbers of regurgitation volume.

Quality assurance and validation advice for clinical use

Both initial validation and ongoing quality assurance are important aspects of clinical 4D Flow CMR. This section builds and expands on the 2015 consensus statement.

For incorporation into standard clinical practice, 4D Flow CMR acquisitions must meet quality thresholds that provide the interpreting clinician with confidence in both the qualitative and quantitative accuracy of the data. Accuracy in 4D Flow CMR can be influenced by the choice of vendor sequences, acquisition parameters and postprocessing software. Therefore, local validation and ongoing quality assurance are an important part of the clinical 4D Flow CMR workflow.

Initial validation is advised to be undertaken when using a new sequence, updating a sequence (such as a significant sequence change with system updates), gradient servicing, applying significant changes in acquisition parameters or using a new post-processing platform:

We advise acquiring 10 datasets (healthy subjects and/or patients without any intra- or extra-cardiac shunts) with both the institution’s standard 2D Flow and 4D Flow CMR including at least: ascending aorta, pulmonary trunk, left branch pulmonary artery, right branch pulmonary artery, superior vena cava, descending aorta and pulmonary veins. If possible we also advise to rescan the volunteers/patients ideally after exiting and then re-entering the scanner either on the same day or a defined short recall period (< 1 month).

Initial visual assessment of the velocity and magnitude images should include assessment of motion artefacts, wrap around artefacts and any aliasing in systole.

We advise to include 3 steps in the quantitative assessment: (1) comparison to 2D Flow CMR, (2) within dataset validation, (3) inter- and intra-reader comparison (when changing/updating post-processing platform). Ideally, differences in flow assessment should be ≤ 5%. Scan-rescan differences up to 10% are acceptable due to minor physiological differences between scans.

Comparison to 2D Flow CMR: We suggest comparing forward flow and peak velocity for at least ascending aorta, pulmonary trunk, left branch pulmonary artery, right branch pulmonary artery, superior vena cava and descending aorta between 2D Flow and 4D Flow CMR in the 10 validation datasets. This captures arterial and venous flow with different flow velocities as well as a variety of vessel diameters (see Fig. 3).

Within dataset validation: This makes use of the conservation of mass principle. Mass is neither created nor destroyed and so flow volumes should stay equal. Therefore, the following forward flow comparisons can be made in all 10 datasets and should show equal flow:Aortic flow (add 5% for coronary flow if measuring above sinuses) = pulmonary flowRight + left pulmonary artery flow = main pulmonary artery flowBranch pulmonary artery flow = pulmonary vein flow (If not equal check for pulmonary vein anomalies)Superior vena cava + descending aortic flow = ascending aortic flow

Aortic flow (add 5% for coronary flow if measuring above sinuses) = pulmonary flow

Right + left pulmonary artery flow = main pulmonary artery flow

Branch pulmonary artery flow = pulmonary vein flow (If not equal check for pulmonary vein anomalies)

Superior vena cava + descending aortic flow = ascending aortic flow

This allows the assessment of measurement planes in a variety of directions within the 4D Flow CMR dataset.

Furthermore, we advise placing 2–4 measurement planes in the ascending aorta between the sinuses of Valsalva and the 1st branching vessel. Again, using the conservation of mass principle, the flow volume should match in all planes (< 5% difference).

Inter- and intra-reader comparison: We advise each reader involved in the clinical service to complete the above flow validations in all 10 datasets twice at least 1 week apart to evaluate any inter or intra-reader bias. Differences in flow assessment should be ≤ 5%.

Everyday quality assurance in every dataset acquired should at least include:

Initial visual assessment of the phase contrast and magnitude datasets should include assessment of motion artefacts, wrap around artefacts and any aliasing during systole.

Quantitative assessment using within dataset validation using the conservation of mass principle. At least one of the above ‘within dataset’ forward flow comparisons can be completed. The choice of these depends on the underlying anatomy and physiology and is determined by the reading physician.

Considerations for integration into clinical practice

When embarking on integrating 4D Flow CMR into the clinical workflow, several considerations are important. Initial validation of the chosen 4D Flow sequence on the local CMR scanner is paramount (see section "Quality assurance and validation advice for clinical use"). 4D Flow CMR datasets are large and may require additional space on the hospital’s image storage solutions. Stored datasets need to be accessible by 4D Flow CMR analysis software which often requires integrated graphics processing units (GPUs) and higher processor powers than standard hospital computers. An alternative is cloud-based 4D Flow CMR offered by some software vendors.

As with any new imaging technique, it takes a while for the teams involved to get confident with image acquisition and analysis. Only when this is achieved should the data be used for clinical reporting. Initially, both 2D and 4D Flow CMR should be acquired and analyzed in parallel. All published prognostic values are based on 2D Flow CMR and the clinical team will need time to evaluate whether 4D Flow CMR assessment can be used interchangeably with 2D Flow CMR in all or some of their patient cohorts.

Where does 4D Flow CMR fit in a clinical protocol?

Historically, 4D Flow CMR has been considered a research technique and thus placed at the end of clinical exams after diagnostic sequences; however, with the emergence of clinically validated applications and postprocessing software, some centers are now adding 4D Flow CMR to routine clinical CMR protocols. When deciding where to place 4D Flow CMR within a clinical CMR protocol there are several considerations, which take into account scan time and pathology-specific considerations:Non-contrast 4D Flow CMR is sufficient in many scenarios. If administering gadolinium-based contrast agents for other clinical questions, 4D Flow CMR should be placed after the CMR angiogram or during the delay before myocardial LGE. It is important to note that technical factors such as respiratory compensation, multi-VENC and large field-of-view increase scan time and may prohibit 4D Flow CMR acquisition during the 10-min post-gadolinium delay window.If the clinical indication requires flow quantification (e.g., shunt evaluation, quantification of valvular regurgitation) then 4D Flow CMR may take higher priority and be acquired earlier in the imaging protocol, especially if 4D Flow CMR is used in place of standard 2D Flow imaging.In pathologies where 4D Flow CMR plays an adjunctive role (e.g., aortic aneurysm), ensuring that all diagnostic sequences are completed before the acquisition of 4D Flow CMR is a common approach

Non-contrast 4D Flow CMR is sufficient in many scenarios. If administering gadolinium-based contrast agents for other clinical questions, 4D Flow CMR should be placed after the CMR angiogram or during the delay before myocardial LGE. It is important to note that technical factors such as respiratory compensation, multi-VENC and large field-of-view increase scan time and may prohibit 4D Flow CMR acquisition during the 10-min post-gadolinium delay window.

If the clinical indication requires flow quantification (e.g., shunt evaluation, quantification of valvular regurgitation) then 4D Flow CMR may take higher priority and be acquired earlier in the imaging protocol, especially if 4D Flow CMR is used in place of standard 2D Flow imaging.

In pathologies where 4D Flow CMR plays an adjunctive role (e.g., aortic aneurysm), ensuring that all diagnostic sequences are completed before the acquisition of 4D Flow CMR is a common approach

Quantitative analysis

Quantitative analysis of flow volumes and peak velocity can easily be integrated into standard CMR reporting templates. We advise reporting both 2D and 4D Flow CMR concomitantly until both imaging and clinical cardiology teams feel confident in basing clinical decisions on 4D Flow CMR results alone.

More advanced parameters can be derived from 4D Flow CMR velocity maps using dedicated software, but these may not be formally approved yet for clinical use.

Qualitative analysis

To take full advantage of the comprehensive nature of 4D Flow CMR, users should interact with the 4D Flow CMR datasets that can be reformatted into any plane to derive the optimum qualitative patient data for display. This is not possible on standard viewing and analysis platforms accessible to clinicians using images for decision making. We, therefore, recommend developing a workflow to save useful 4D Flow CMR images/videos in DICOM format which can then be loaded onto standard viewing platforms. This makes 4D Flow CMR more accessible to clinicians and can be used at multi-disciplinary team meetings even without a 4D Flow CMR specialist present. Especially in congenital heart disease qualitative analysis can be useful in delineating stenosed vessels in more detail, providing information on the exact position and length of flow acceleration. It also easily identifies any flow reversal.

Quality assurance and validation advice in the research setting

Several useful options exist for the validation of different aspects of 4D Flow CMR, such as sequence development, reconstruction algorithms and post-processing workflows. These options can broadly be categorized into (1) in vivo studies, (2) phantom studies, and (3) computer simulations (summarized in Table 2). We emphasize that there is no single evaluation or validation methodology that can target all aspects of 4D Flow CMR. Instead, evaluation and validation need to be tailored for the specific sequence, parameter, or application in question.

In vivo studies

In vivo studies are used to evaluate and validate new 4D Flow CMR methods in comparison to other modalities such as echocardiography, 2D Flow CMR, and other 4D Flow CMR methods. Furthermore, it is possible to use consistency criteria such as conservation of mass principles, given the fact that flow into and out of a closed system (e.g., the heart, or the aorta) must be the same. The main advantage of using in vivo studies for evaluation and validation is that it represents the final utility of the method. A challenging aspect of in vivo studies is that reference data is often not available when advanced hemodynamic parameters such as wall shear stress, turbulence stresses, intracardiac flow component, kinetic energy and vorticity are evaluated. Furthermore, we emphasize that in vivo validation of the capability of a sequence to measure basic parameters such as flow volume cannot be interpreted as evidence that the sequence permits accurate estimation of advanced hemodynamic parameters, and these require separate and targeted validation.

In vitro studies

In vitro phantom studies in idealized or anatomically accurate vascular and cardiac models, so-called flow phantoms, permit evaluation and validation in well-known and repeatable flow conditions. An advantage of in-vitro phantoms is that long sessions with scans using many different parameter settings can be performed on the same flow setup. Another advantage is the possibility to validate post-processing software and compare the results against flow meter, “timer and beaker” and pressure probe measurements, as well as other experimental fluid dynamics techniques, such as particle image velocimetry and direct pressure measurements. A disadvantage is that in vitro phantoms typically do not have realistic surrounding tissue. Furthermore, we note that while advanced flow phantoms with realistic geometry and pulsating flow are highly valuable, even simplified phantom experiments can provide valuable insight. Examples include a large container of stationary water or agarose gel for the evaluation of background phase offsets and rotating phantoms consisting of gel-filled wheels or rings. Finally, we encourage the continuing development of standardized 4D Flow CMR phantoms and pump setups that facilitates reproducible in-vitro flow experiments across multiple sites.

Computer simulations

Simulated 4D Flow CMR measurments in numerical velocity data, also referred to as synthetic phantoms and digital reference objects, permit detailed studies of the impact of sequence design and parameter settings in a fully known flow environment. Another advantage of this approach is that synthetic phantoms can be created with a model of more realistic surrounding tissue. This is relevant for reconstruction algorithms and processing tools. Hybrid in vitro/synthetic phantoms in which in vitro data are embedded into synthetic backgrounds may also be considered. The main disadvantage of synthetic phantoms is the question of simulation fidelity, i.e., how well the simulation results represent reality. Increased fidelity usually requires more computational resources. It can therefore be relevant to consider how complete the simulation needs to be and if the aim of the study permits any trade-offs between simulation time and completeness of the 4D Flow CMR simulations. We note that the generation of accurate numerical velocity data using computational fluid dynamics is a separate field of research. However, when used as input and reference in 4D Flow CMR simulations, the physical accuracy of the numerical velocity data is of secondary importance.

In summary, the development of 4D Flow CMR methods can be guided by in vivo studies, phantom studies, computer simulations, or a combination thereof. We recommend that all these approaches be considered in the evaluation and validation of new developments in 4D Flow CMR.

Recommended publication standards

In this section, we describe the essential and recommended standards that should be adhered to for any scientific publication containing 4D Flow CMR. Recommendations differ in parts between technical publications (often aiming to propose and evaluate a new technique) and clinical studies (applying 4D Flow CMR to clinical questions). High-quality and standardized publications will enable easier replication of proposed sequence protocols for clinical use and facilitate easier and higher quality meta-analysis to move the field forward.

Where possible, sharing of published datasets, code and materials to replicate, verify and extend the research presented in the manuscript is encouraged.

Below are 4D flow CMR specific considerations:

Introduction

For studies based on a priori stated hypotheses, all hypotheses should be clearly stated when describing the aim(s) of the study.

Methods-acquisition

All data processing methods that can affect the quality of the 4D flow CMR data should be described, including correction methods for eddy currents, distortions resulting from gradient field non-uniformity, intravoxel dephasing and concomitant gradient fields, velocity aliasing, as well as noise filtering (if not using commercially available sequences and data processing software). Many parameters, including hardware specifics, acquisition parameters, and post-processing software, affect the image quality and properties of 4D Flow CMR data and should be reported as such. The essential and recommended standards are listed in Table 3 and elucidated below. We specifically highlight that both acquired and reconstructed resolutions should be given, both for temporal and spatial resolution.

Method-data processing

Data processing can be performed by several commercial CE- and FDA approved software packages, while in-house developed tools enable techniques for research. Open-source software solutions facilitate reproducible research, and the availability of such tools should be clarified in publications. For commercially available as well as open-source software, the software release version should be detailed.

Method-quantification

After processing the data, hemodynamic parameters can be extracted from the velocity fields. A range of metrics can be derived, each with relevance depending on the specific application. Thus, a detailed description of the analysis methodology should be provided so that a similar analysis could be performed at other centers.

Methods-statistics

Clinical diagnosis and/or outcome studies need to be designed with an adequate sample size resulting from a power analysis. The methods section should contain a statistics paragraph describing all used statistical methods appropriate for the study size.

Results

The results section should include the number of included and excluded subjects, as well as the reason for exclusion, the results of the data quality assurance assessment, at least including a within-dataset validation (see section on "Quality assurance and validation advice in the research setting" and for clinical use), and inter- and intra-reader agreement for a subset of data (or referenced to previous publication with the same technique in the same setting) should be included. Given the variety in algorithms and their performance, it is important to mention whether velocity aliasing was present and have an estimate of how well the velocity fields have been corrected.

Besides the quantitative results, preferably shown in tables, well-made illustrations should be added of typical as well as extreme findings or participants. Movies should be included as supplementary material, if allowed by the publisher, to illustrate the behavior over the cardiac cycle. The anatomy should be well-annotated, preferably in combination with segmentation. Color bars should be included for quantitative parameters. If particle traces were used in the visualization, it should be clarified which particle traces were used (e.g., pathlines, streamlines).

Discussion

If acquisition settings are significantly different from these and previous recommendations or vary between participants, this should be mentioned in the limitation section, and it should be discussed how this might affect the results.

References

References to methods and techniques should reference the original work.

These recommendation standards will contribute to consistently high-quality publications, will improve the review process, and allow for easier comparison of different publications. We would therefore like to stress the importance of following these standards.

Overcoming limitations and future considerations

4D Flow CMR is becoming more widely used in medical centers with the technical and clinical capabilities to incorporate its use into standard-of-care protocols for heart valve, aortic, pulmonary, and congenital heart disease. However, several challenges remain to achieve widespread adoption and application of 4D Flow CMR remain. This includes limited velocity dynamic range due to a single user-selected VENC, long and unpredictable scan times, data storage of large datasets especially in clinical work flows as well as manual and time-consuming data processing (which can affect user confidence). Below we discuss several promising new developments which are ongoing to overcome these limitations.

Velocity dynamic range

Acceleration techniques enable reductions in acquisition time or acquisition of additional data within the same total acquisition time. In particular, acceleration enables the acquisition of additional data to reduce the dynamic range issues associated with velocity encoding. Evaluation of altered cardiovascular hemodynamics often requires measurement of flow across a wide range of velocities, e.g., high-velocity flow jets (up to 400–600 cm/s) with adjacent regions of low circulating or venous flows (as low as 10 cm/s). Commonly available 4D Flow CMR techniques measure blood flow velocity based on a single pre-defined VENC, but acceleration techniques have enabled 4D Flow CMR with dual- or multi-VENC velocity encoding, i.e., acquisition of both low- and high-VENC data within a single scan. Multi-VENC reconstruction can generate 4D Flow CMR data with the favorable VNR of a low-VENC acquisition but without velocity aliasing. In addition to multi-VENC acquisitions, initial deep learning-based studies have demonstrated the potential of using physics-informed neural networks to reduce noise, enhance resolution and automatically unwrap aliased velocity values in 4D Flow CMR velocity data.

Respiratory and cardiac self-gating

4D Flow CMR techniques are also being developed that permit respiratory and cardiac self-gating and thereby simplify and streamline acquisition for optimized clinical workflows. Respiratory self-gating eliminates the need for respiratory navigators and can be achieved by repeatedly acquiring a central k-space line that corresponds to a projection of the image volume in the feet-to-head direction. This has recently been incorporated in so-called five-dimensional “5D” and extra-dimensional “XD” Flow CMR which also permit the analysis of respiratory-driven changes in cardiovascular hemodynamics. Cardiac self-gating can be achieved with similar principles as respiratory self-gating and has recently been incorporated in 4D Flow CMR. Fully self-gated 5D free-running approaches that exploit compressed sensing reconstruction remove the need for respiratory navigators, have constant scan time, and are independent of the patient’s breathing pattern or heart-rate, which makes it particularly well-suited to be integrated as part of a clinical protocol while scan planning is much facilitated. However, at this juncture, reconstruction times are prohibitive for clinical use.

Accelerated data processing workflows

Current 4D Flow CMR data analysis workflows are often non-standardized and time-consuming, thus limiting reproducibility and clinical translation. Addressing these limitations will require the development of efficient image analysis strategies with minimal user dependence. This is becoming feasible with advances in image processing techniques. For example, automated segmentation of the aorta and pulmonary artery has been demonstrated with atlas-based as well as deep learning-based methods. In addition to segmentation, machine learning has the potential to speed up and automate image processing tasks such as background phase offset correction, although work in this area is still in an early stage and unpublished. Another time-consuming task for which machine learning has demonstrated impressive results is the reconstruction of highly undersampled 4D Flow CMR images from raw data in less than 1 min.

Summary/conclusion

4D Flow CMR has moved from “pretty pictures” to providing robust flow quantification in clinical practice. 4D Flow CMR has greatly benefitted from the advances in CMR acceleration making it feasible for clinical use. The worldwide 4D Flow CMR community has grown exponentially since the last consensus statement. 4D Flow CMR is no longer just a tool for researchers but for clinicians. This consensus statement aims to help clinicians initiate a 4D Flow CMR program in their institutions. Furthermore, it aims to set standards for both clinical and research settings to assure consistent high-quality 4D Flow CMR output.

References 169

  1. 1.
  2. 2.

    Doyle CM, Orr J, Greenwood JP, Plein S, Tsoumpas C, Bissell MM. Four-dimensional flow magnetic resonance imaging in the assessment of blood flow in the heart and great vessels: a systematic review. J Magn Reson Imaging. 2021.

    Not in our library10.1002/jmri.27874 (opens in a new tab)

  3. 3.

    Kamphuis VP, van der Palen RLF, de Koning PJH, Elbaz MSM, van der Geest RJ, de Roos A, et al. In-scan and scan-rescan assessment of LV in- and outflow volumes by 4D flow MRI versus 2D planimetry. J Magn Reson Imaging. 2018;47(2):511-522.

    Not in our library10.1002/jmri.25792 (opens in a new tab)

  4. 4.

    Kamphuis VP, Westenberg JJM, van der Palen RLF, van den Boogaard PJ, van der Geest RJ, de Roos A, et al. Scan-rescan reproducibility of diastolic left ventricular kinetic energy, viscous energy loss and vorticity assessment using 4D flow MRI: analysis in healthy subjects. Int J Cardiovasc Imaging. 2018.

    Not in our library10.1007/s10554-017-1291-z (opens in a new tab)

  5. 5.

    Juffermans JA-O, Westenberg JA-O, van den Boogaard PJ, Roest AAW, van Assen HC, van der Palen RLF, et al. Reproducibility of aorta segmentation on 4D flow MRI in healthy volunteers. J Magn Reson Imaging. 2021.

    Not in our library10.1002/jmri.27431 (opens in a new tab)

  6. 6.

    van der Palen RLF, Roest AAW, van den Boogaard PJ, de Roos A, Blom NA, Westenberg JJM. Scan-rescan reproducibility of segmental aortic wall shear stress as assessed by phase-specific segmentation with 4D flow MRI in healthy volunteers. MAGMA. 2018.

    Not in our library10.1007/s10334-018-0688-6 (opens in a new tab)

  7. 7.

    Stankovic Z, Allen BD, Garcia J, Jarvis KB, Markl M. 4D flow imaging with MRI. Cardiovasc Diagn Ther. 2014;4(2):173-192.

    Not in our library10.3978/j.issn.2223-3652.2014.01.02 (opens in a new tab)

  8. 8.

    Markl M, Frydrychowicz A, Kozerke S, Hope M, Wieben O. 4D flow MRI. J Magn Reson Imaging. 2012;36(5):1015-1036.

    Not in our library10.1002/jmri.23632 (opens in a new tab)

  9. 9.

    Takehara Y. 4D flow when and how? Radiol Med. 2020;125(9):838-850.

    Not in our library10.1007/s11547-020-01249-0 (opens in a new tab)

  10. 10.

    Calkoen EE, Roest AA, van der Geest RJ, de Roos A, Westenberg JJ. Cardiovascular function and flow by 4-dimensional magnetic resonance imaging techniques: new applications. J Thorac Imaging. 2014;29(3):185-196.

    Not in our library10.1097/rti.0000000000000068 (opens in a new tab)

  11. 11.

    Kamphuis VP, Westenberg JJM, van der Palen RLF, Blom NA, de Roos A, van der Geest R, et al. Unravelling cardiovascular disease using four dimensional flow cardiovascular magnetic resonance. Int J Cardiovasc Imaging. 2017;33(7):1069-1081.

    Not in our library10.1007/s10554-016-1031-9 (opens in a new tab)

  12. 12.

    Zhong LA-O, Schrauben EA-O, Garcia JA-O, Uribe S, Grieve SA-OX, Elbaz MA-O, et al. Intracardiac 4D flow MRI in congenital heart disease: recommendations on behalf of the ISMRM flow and motion study group. J Magn Reson Imaging. 2019.

    Not in our library10.1002/jmri.26893 (opens in a new tab)

  13. 13.

    Markl M, Schnell S, Barker AJ. 4D flow imaging: current status to future clinical applications. Curr Cardiol Rep. 2014;16(5):481.

    Not in our library10.1007/s11886-014-0481-8 (opens in a new tab)

  14. 14.

    Soulat G, McCarthy P, Markl M. 4D flow with MRI. Annu Rev Biomed Eng. 2020;22:103-126.

    Not in our library10.1146/annurev-bioeng-100219-110055 (opens in a new tab)

  15. 15.

    Burris NS, Hope MD. 4D flow MRI applications for aortic disease. Magn Reson Imaging Clin N Am. 2015;23(1):15-23.

    Not in our library10.1016/j.mric.2014.08.006 (opens in a new tab)

  16. 16.

    Garcia J, Barker AJ, Markl M. The role of imaging of flow patterns by 4D Flow MRI in aortic stenosis. JACC Cardiovasc Imaging. 2019;12(2):252-266.

    Not in our library10.1016/j.jcmg.2018.10.034 (opens in a new tab)

  17. 17.

    Takahashi K, Sekine T, Ando T, Ishii Y, Kumita S. Utility of 4D flow MRI in thoracic aortic diseases: a literature review of clinical applications and current evidence. Magn Reson Med Sci. 2022;21(2):327-339.

    Not in our library10.2463/mrms.rev.2021-0046 (opens in a new tab)

  18. 18.

    Catapano F, Pambianchi G, Cundari G, Rebelo J, Cilia F, Carbone I, et al. 4D flow imaging of the thoracic aorta: is there an added clinical value? Cardiovasc Diagn Ther. 2020;10(4):1068-1089.

    Not in our library10.21037/cdt-20-452 (opens in a new tab)

  19. 19.

    Cave DGW, Panayiotou H, Bissell MM. Hemodynamic profiles before and after surgery in bicuspid aortic valve disease: a systematic review of the literature. Front Cardiovasc Med. 2021;8:629227.

    Not in our library10.3389/fcvm.2021.629227 (opens in a new tab)

  20. 20.

    Oyama-Manabe N, Aikawa T, Tsuneta S, Manabe O. Clinical applications of 4D flow MR imaging in aortic valvular and congenital heart disease. Magn Reson Med Sci. 2022;21(2):319-326.

    Not in our library10.2463/mrms.rev.2021-0030 (opens in a new tab)

  21. 21.

    Ruiz-Munoz A, Guala A, Rodriguez-Palomares J, Dux-Santoy L, Servato L, Lopez-Sainz A, et al. Aortic flow dynamics and stiffness in Loeys-Dietz syndrome patients: a comparison with healthy volunteers and Marfan syndrome patients. Eur Heart J Cardiovasc Imaging. 2022;23(5):641-649.

    Not in our library10.1093/ehjci/jeab069 (opens in a new tab)

  22. 22.

    Lawley CM, Broadhouse KM, Callaghan FM, Winlaw DS, Figtree GA, Grieve SM. 4D flow magnetic resonance imaging: role in pediatric congenital heart disease. Asian Cardiovasc Thorac Ann. 2018;26(1):28-37.

    Not in our library10.1177/0218492317694248 (opens in a new tab)

  23. 23.

    Rizk J. 4D flow MRI applications in congenital heart disease. Eur Radiol. 2021;31(2):1160-1174.

    Not in our library10.1007/s00330-020-07210-z (opens in a new tab)

  24. 24.

    Azarine A, Garçon P, Stansal A, Canepa N, Angelopoulos G, Silvera S, et al. Four-dimensional flow MRI: principles and cardiovascular applications. Radiographics. 2019;39(3):632-648.

    Not in our library10.1148/rg.2019180091 (opens in a new tab)

  25. 25.

    Jacobs K, Hahn L, Horowitz M, Kligerman S, Vasanawala S, Hsiao A. Hemodynamic assessment of structural heart disease using 4D flow MRI: how we do it. AJR Am J Roentgenol. 2021;217(6):1322-1332.

    Not in our library10.2214/ajr.21.25978 (opens in a new tab)

  26. 26.

    Vasanawala SS, Hanneman K, Alley MT, Hsiao A. Congenital heart disease assessment with 4D flow MRI. J Magn Reson Imaging. 2015;42(4):870-886.

    Not in our library10.1002/jmri.24856 (opens in a new tab)

  27. 27.

    Elsayed A, Gilbert K, Scadeng M, Cowan BR, Pushparajah K, Young AA. Four-dimensional flow cardiovascular magnetic resonance in tetralogy of Fallot: a systematic review. J Cardiovasc Magn Reson. 2021;23(1):59.

    Not in our library10.1186/s12968-021-00745-0 (opens in a new tab)

  28. 28.

    Warmerdam E, Krings GJ, Leiner T, Grotenhuis HB. Three-dimensional and four-dimensional flow assessment in congenital heart disease. Heart. 2020;106(6):421-426.

    Not in our library10.1136/heartjnl-2019-315797 (opens in a new tab)

  29. 29.

    Geiger J, Callaghan FM, Burkhardt BEU, Valsangiacomo Buechel ER, Kellenberger CJ. Additional value and new insights by four-dimensional flow magnetic resonance imaging in congenital heart disease: application in neonates and young children. Pediatr Radiol. 2021;51(8):1503-1517.

    Not in our library10.1007/s00247-020-04885-w (opens in a new tab)

  30. 30.

    Zhuang B, Sirajuddin A, Zhao S, Lu M. The role of 4D flow MRI for clinical applications in cardiovascular disease: current status and future perspectives. Quant Imaging Med Surg. 2021;11(9):4193-4210.

    Not in our library10.21037/qims-20-1234 (opens in a new tab)

  31. 31.

    Markl M, Schnell S, Wu C, Bollache E, Jarvis K, Barker AJ, et al. Advanced flow MRI: emerging techniques and applications. Clin Radiol. 2016;71(8):779-795.

    Not in our library10.1016/j.crad.2016.01.011 (opens in a new tab)

  32. 32.

    Crandon S, Elbaz MSM, Westenberg JJM, van der Geest RJ, Plein S, Garg P. Clinical applications of intra-cardiac four-dimensional flow cardiovascular magnetic resonance: a systematic review. Int J Cardiol. 2017;249:486-93.

    Not in our library10.1016/j.ijcard.2017.07.023 (opens in a new tab)

  33. 33.

    Kaur H, Assadi H, Alabed S, Cameron D, Vassiliou VS, Westenberg JJM, et al. Left ventricular blood flow kinetic energy assessment by 4D flow cardiovascular magnetic resonance: a systematic review of the clinical relevance. J Cardiovasc Dev Dis. 2020;7(3):37.

    Not in our library10.3390/jcdd7030037 (opens in a new tab)

  34. 34.

    Oechtering TH, Roberts GS, Panagiotopoulos N, Wieben O, Roldan-Alzate A, Reeder SB. Abdominal applications of quantitative 4D flow MRI. Abdom Radiol (NY). 2022;47(9):3229-3250.

    Not in our library10.1007/s00261-021-03352-w (opens in a new tab)

  35. 35.

    Buonocore MH, Bogren H. Factors influencing the accuracy and precision of velocity-encoded phase imaging. Magn Reson Med. 1992;26(1):141-154.

    Not in our library10.1002/mrm.1910260115 (opens in a new tab)

  36. 36.

    Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics. 2002;22(3):651-671.

    Not in our library10.1148/radiographics.22.3.g02ma11651 (opens in a new tab)

  37. 37.

    Hofman MB, Visser FC, van Rossum AC, Vink QM, Sprenger M, Westerhof N. In vivo validation of magnetic resonance blood volume flow measurements with limited spatial resolution in small vessels. Magn Reson Med. 1995;33(6):778-784.

    Not in our library10.1002/mrm.1910330606 (opens in a new tab)

  38. 38.
  39. 39.

    Jacobs K, Hahn L, Horowitz M, Kligerman S, Vasanawala S, Hsiao A. Structural heart 4D Flow MRI for hemodynamic assessment: how we do it. AJR Am J Roentgenol. 2021.

    Not in our library10.2214/ajr.21.25978 (opens in a new tab)

  40. 40.

    Bailes DR, Gilderdale DJ, Bydder GM, Collins AG, Firmin DN. Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artefacts in MR imaging. J Comput Assist Tomogr. 1985;9(4):835-838.

    Not in our library10.1097/00004728-198507010-00039 (opens in a new tab)

  41. 41.

    Markl M, Harloff A, Bley TA, Zaitsev M, Jung B, Weigang E, et al. Time-resolved 3D MR velocity mapping at 3T: improved navigator-gated assessment of vascular anatomy and blood flow. J Magn Reson Imaging. 2007;25(4):824-831.

    Not in our library10.1002/jmri.20871 (opens in a new tab)

  42. 42.

    van Ooij P, Semaan E, Schnell S, Giri S, Stankovic Z, Carr J, et al. Improved respiratory navigator gating for thoracic 4D flow MRI. Magn Reson Imaging. 2015;33(8):992-999.

    Not in our library10.1016/j.mri.2015.04.008 (opens in a new tab)

  43. 43.

    Spartera M, Pessoa-Amorim G, Stracquadanio A, Von Ende A, Fletcher A, Manley P, et al. Left atrial 4D flow cardiovascular magnetic resonance: a reproducibility study in sinus rhythm and atrial fibrillation. J Cardiovasc Magn Reson. 2021.

    Not in our library10.1186/s12968-021-00729-0 (opens in a new tab)

  44. 44.

    Bock J, Töger J, Bidhult S, Markenroth Bloch K, Arvidsson P, Kanski M, et al. Validation and reproducibility of cardiovascular 4D-flow MRI from two vendors using 2 × 2 parallel imaging acceleration in pulsatile flow phantom and in vivo with and without respiratory gating. Acta Radiol. 2019;60(3):327-337.

    Not in our library10.1177/0284185118784981 (opens in a new tab)

  45. 45.

    Dyverfeldt P, Ebbers T. Comparison of respiratory motion suppression techniques for 4D flow MRI. Magn Reson Med. 2017;78:1877-82.

    Not in our library10.1002/mrm.26574 (opens in a new tab)

  46. 46.

    Dumoulin CL, Souza SP, Walker MF, Wagle W. Three-dimensional phase contrast angiography. Magn Reson Med. 1989;9(1):139-149.

    Not in our library10.1002/mrm.1910090117 (opens in a new tab)

  47. 47.

    François CJ, Lum DP, Johnson KM, Landgraf BR, Bley TA, Reeder SB, et al. Renal arteries: isotropic, high-spatial-resolution, unenhanced MR angiography with three-dimensional radial phase contrast. Radiology. 2011;258(1):254-260.

    Not in our library10.1148/radiol.10100443 (opens in a new tab)

  48. 48.

    Bock J, Frydrychowicz A, Stalder AF, Bley TA, Burkhardt H, Hennig J, et al. 4D phase contrast MRI at 3 T: effect of standard and blood-pool contrast agents on SNR, PC-MRA, and blood flow visualization. Magn Reson Med. 2010;63(2):330-338.

    Not in our library10.1002/mrm.22199 (opens in a new tab)

  49. 49.

    Hess AT, Bissell MM, Ntusi NA, Lewis AJ, Tunnicliffe EM, Greiser A, et al. Aortic 4D flow: quantification of signal-to-noise ratio as a function of field strength and contrast enhancement for 1.5T, 3T, and 7T. Magn Reson Med. 2015;73(5):1864-71.

    Not in our library10.1002/mrm.25317 (opens in a new tab)

  50. 50.

    Vasanawala SS, Nguyen KL, Hope MD, Bridges MD, Hope TA, Reeder SB, et al. Safety and technique of ferumoxytol administration for MRI. Magn Reson Med. 2016;75(5):2107-2111.

    Not in our library10.1002/mrm.26151 (opens in a new tab)

  51. 51.

    Panayiotou HR, Mills LK, Broadbent DA, Shelley D, Scheffczik J, Olaru AM, et al. Comprehensive neonatal cardiac, feed and wrap, non-contrast, non-sedated, free-breathing compressed sensing 4D flow MRI assessment. J Magn Reson Imaging. 2023;57(3):789-799.

    Not in our library10.1002/jmri.28325 (opens in a new tab)

  52. 52.

    Sjöberg P, Hedström E, Fricke K, Frieberg P, Weismann CG, Liuba P, et al. Comparison of 2D and 4D flow MRI in neonates without general anesthesia. J Magn Reson Imaging. 2023;57(1):71-82.

    Not in our library10.1002/jmri.28303 (opens in a new tab)

  53. 53.

    Callaghan FM, Burkhardt B, Valsangiacomo Buechel ER, Kellenberger CJ, Geiger J. Assessment of ventricular flow dynamics by 4D-flow MRI in patients following surgical repair of d-transposition of the great arteries. Eur Radiol. 2021;31(10):7231-7241.

    Not in our library10.1007/s00330-021-07813-0 (opens in a new tab)

  54. 54.

    Lai LM, Cheng JY, Alley MT, Zhang T, Lustig M, Vasanawala SS. Feasibility of ferumoxytol-enhanced neonatal and young infant cardiac MRI without general anesthesia. J Magn Reson Imaging. 2017;45(5):1407-1418.

    Not in our library10.1002/jmri.25482 (opens in a new tab)

  55. 55.

    Cheng JY, Hanneman K, Zhang T, Alley MT, Lai P, Tamir JI, et al. Comprehensive motion-compensated highly accelerated 4D flow MRI with ferumoxytol enhancement for pediatric congenital heart disease. J Magn Reson Imaging. 2016;43(6):1355-1368.

    Not in our library10.1002/jmri.25106 (opens in a new tab)

  56. 56.

    Juffermans JF, Minderhoud SCS, Wittgren J, Kilburg A, Ese A, Fidock B, et al. Multicenter consistency assessment of valvular flow quantification with automated valve tracking in 4D flow CMR. JACC Cardiovasc Imaging. 2021;14(7):1354-1366.

    Not in our library10.1016/j.jcmg.2020.12.014 (opens in a new tab)

  57. 57.

    Chernobelsky A, Shubayev O, Comeau CR, Wolff SD. Baseline correction of phase contrast images improves quantification of blood flow in the great vessels. J Cardiovasc Magn Reson. 2007;9(4):681-685.

    Not in our library10.1080/10976640601187588 (opens in a new tab)

  58. 58.

    Minderhoud SCS, van der Velde N, Wentzel JJ, van der Geest RJ, Attrach M, Wielopolski PA, et al. The clinical impact of phase offset errors and different correction methods in cardiovascular magnetic resonance phase contrast imaging: a multi-scanner study. J Cardiovasc Magn Reson. 2020;22(1):68.

    Not in our library10.1186/s12968-020-00659-3 (opens in a new tab)

  59. 59.
  60. 60.

    Carrillo H, Osses A, Uribe S, Bertoglio C. Optimal dual-VENC unwrapping in phase-contrast MRI. IEEE Trans Med Imaging. 2019;38(5):1263-1270.

    Not in our library10.1109/tmi.2018.2882553 (opens in a new tab)

  61. 61.

    Dymerska B, Eckstein K, Bachrata B, Siow B, Trattnig S, Shmueli K, et al. Phase unwrapping with a rapid opensource minimum spanning tree algorithm (ROMEO). Magn Reson Med. 2021;85(4):2294-2308.

    Not in our library10.1002/mrm.28563 (opens in a new tab)

  62. 62.

    Loecher M, Schrauben E, Johnson KM, Wieben O. Phase unwrapping in 4D MR flow with a 4D single-step Laplacian algorithm. J Magn Reson Imaging. 2016;43(4):833-842.

    Not in our library10.1002/jmri.25045 (opens in a new tab)

  63. 63.

    Ma LE, Markl M, Chow K, Vali A, Wu C, Schnell S. Efficient triple-VENC phase-contrast MRI for improved velocity dynamic range. Magn Reson Med. 2020;83(2):505-520.

    Not in our library10.1002/mrm.27943 (opens in a new tab)

  64. 64.

    Zhang J, Rothenberger SM, Brindise MC, Scott MB, Berhane H, Baraboo JJ, et al. Divergence-free constrained phase unwrapping and denoising for 4D flow MRI using weighted least-squares. IEEE Trans Med Imaging. 2021.

    Not in our library10.1109/tmi.2021.3086331 (opens in a new tab)

  65. 65.

    Wigström L, Ebbers T, Fyrenius A, Karlsson M, Engvall J, Wranne B, et al. Particle trace visualization of intracardiac flow using time-resolved 3D phase contrast MRI. Magn Reson Med. 1999;41(4):793-799.

    Not in our library10.1002/(sici)1522-2594(199904)41:4<793::aid-mrm19>3.0.co;2-2 (opens in a new tab)

  66. 66.

    Casciaro ME, Pascaner AF, Guilenea FN, Alcibar J, Gencer U, Soulat G, et al. 4D flow MRI: impact of region of interest size, angulation and spatial resolution on aortic flow assessment. Physiol Meas. 2021;42(3):035004.

    Not in our library10.1088/1361-6579/abe525 (opens in a new tab)

  67. 67.

    Blanken CPS, Westenberg JJM, Aben JP, Bijvoet GP, Chamuleau SAJ, Boekholdt SM, et al. Quantification of mitral valve regurgitation from 4D flow MRI using semiautomated flow tracking. Radiol Cardiothorac Imaging. 2020;2(5):e200004.

    Not in our library10.1148/ryct.2020200004 (opens in a new tab)

  68. 68.

    Hsiao A, Lustig M, Alley MT, Murphy MJ, Vasanawala SS. Evaluation of valvular insufficiency and shunts with parallel-imaging compressed-sensing 4D phase-contrast MR imaging with stereoscopic 3D velocity-fusion volume-rendered visualization. Radiology. 2012;265(1):87-95.

    Not in our library10.1148/radiol.12120055 (opens in a new tab)

  69. 69.

    Contijoch FJ, Horowitz M, Masutani E, Kligerman S, Hsiao A. 4D flow vorticity visualization predicts regions of quantitative flow inconsistency for optimal blood flow measurement. Radiol Cardiothorac Imaging. 2020;2(1):e190054.

    Not in our library10.1148/ryct.2020190054 (opens in a new tab)

  70. 70.

    Hsiao A, Lustig M, Alley MT, Murphy M, Chan FP, Herfkens RJ, et al. Rapid pediatric cardiac assessment of flow and ventricular volume with compressed sensing parallel imaging volumetric cine phase-contrast MRI. AJR Am J Roentgenol. 2012;198(3):W250-W259.

    Not in our library10.2214/ajr.11.6969 (opens in a new tab)

  71. 71.

    Feneis JF, Kyubwa E, Atianzar K, Cheng JY, Alley MT, Vasanawala SS, et al. 4D flow MRI quantification of mitral and tricuspid regurgitation: reproducibility and consistency relative to conventional MRI. J Magn Reson Imaging. 2018;48(4):1147-1158.

    Not in our library10.1002/jmri.26040 (opens in a new tab)

  72. 72.
  73. 73.

    Calkoen EE, Roest AA, Kroft LJ, van der Geest RJ, Jongbloed MR, van den Boogaard PJ, et al. Characterization and improved quantification of left ventricular inflow using streamline visualization with 4DFlow MRI in healthy controls and patients after atrioventricular septal defect correction. J Magn Reson Imaging. 2015;41(6):1512-1520.

    Not in our library10.1002/jmri.24735 (opens in a new tab)

  74. 74.

    Fidock B, Archer G, Barker N, Elhawaz A, Al-Mohammad A, Rothman A, et al. Standard and emerging CMR methods for mitral regurgitation quantification. Int J Cardiol. 2021;331:316-321.

    Not in our library10.1016/j.ijcard.2021.01.066 (opens in a new tab)

  75. 75.

    Spampinato RA, Jahnke C, Crelier G, Lindemann F, Fahr F, Czaja-Ziolkowska M, et al. Quantification of regurgitation in mitral valve prolapse with four-dimensional flow cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2021;23(1):87.

    Not in our library10.1186/s12968-021-00783-8 (opens in a new tab)

  76. 76.

    Mills MA-O, Grafton-Clarke CA-O, Williams GA-O, Gosling RC, Al Baraikan AA-O, Kyriacou AL, et al. Feasibility and validation of trans-valvular flow derived by four-dimensional flow cardiovascular magnetic resonance imaging in patients with atrial fibrillation. Wellc Open Res. 2021.

    Not in our library10.12688/wellcomeopenres.16655.2 (opens in a new tab)

  77. 77.

    Demir A, Wiesemann S, Erley J, Schmitter S, Trauzeddel RF, Pieske B, et al. Traveling volunteers: a multi-vendor, multi-center study on reproducibility and comparability of 4D flow derived aortic hemodynamics in cardiovascular magnetic resonance. J Magn Reson Imaging. 2021.

    Not in our library10.1002/jmri.27804 (opens in a new tab)

  78. 78.

    Oechtering TH, Nowak A, Sieren MM, Stroth AM, Kirschke N, Wegner F, Balks M, König IR, Jin N, Graessner J, Kooijman-Kurfuerst H, Hennemuth A, Barkhausen J, Frydrychowicz A. Repeatability and reproducibility of various 4D Flow CMR postprocessing software programs in a multi-software and multi-vendor cross-over comparison study. J Cardiovasc Magn Reson. 2023.

    Not in our library10.1186/s12968-023-00921-4 (opens in a new tab)

  79. 79.

    Isorni MA, Moisson L, Moussa NB, Monnot S, Raimondi F, Roussin R, et al. 4D flow cardiac magnetic resonance in children and adults with congenital heart disease: clinical experience in a high volume center. Int J Cardiol. 2020;320:168-177.

    Not in our library10.1016/j.ijcard.2020.07.021 (opens in a new tab)

  80. 80.

    Sierra-Galan LM, François CJ. Clinical applications of MRA 4D-flow. Curr Treat Options Cardiovasc Med. 2019;21(10):58.

    Not in our library10.1007/s11936-019-0758-8 (opens in a new tab)

  81. 81.

    Raimondi F, Martins D, Coenen R, Panaioli E, Khraiche D, Boddaert N, et al. Prevalence of venovenous shunting and high-output state quantified with 4D flow MRI in patients with Fontan circulation. Radiol Cardiothorac Imaging. 2021;3(6):e210161.

    Not in our library10.1148/ryct.210161 (opens in a new tab)

  82. 82.

    Isorni MA, Martins D, Ben Moussa N, Monnot S, Boddaert N, Bonnet D, et al. 4D flow MRI versus conventional 2D for measuring pulmonary flow after tetralogy of Fallot repair. Int J Cardiol. 2020;300:132-136.

    Not in our library10.1016/j.ijcard.2019.10.030 (opens in a new tab)

  83. 83.

    Brix L, Ringgaard S, Rasmusson A, Sørensen TS, Kim WY. Three dimensional three component whole heart cardiovascular magnetic resonance velocity mapping: comparison of flow measurements from 3D and 2D acquisitions. J Cardiovasc Magn Reson. 2009;11(1):3.

    Not in our library10.1186/1532-429x-11-3 (opens in a new tab)

  84. 84.

    Nordmeyer S, Riesenkampff E, Messroghli D, Kropf S, Nordmeyer J, Berger F, et al. Four-dimensional velocity-encoded magnetic resonance imaging improves blood flow quantification in patients with complex accelerated flow. J Magn Reson Imaging. 2013;37(1):208-216.

    Not in our library10.1002/jmri.23793 (opens in a new tab)

  85. 85.

    Uribe S, Beerbaum P, Sørensen TS, Rasmusson A, Razavi R, Schaeffter T. Four-dimensional (4D) flow of the whole heart and great vessels using real-time respiratory self-gating. Magn Reson Med. 2009;62(4):984-992.

    Not in our library10.1002/mrm.22090 (opens in a new tab)

  86. 86.

    Roes SD, Hammer S, van der Geest RJ, Marsan NA, Bax JJ, Lamb HJ, et al. Flow assessment through four heart valves simultaneously using 3-dimensional 3-directional velocity-encoded magnetic resonance imaging with retrospective valve tracking in healthy volunteers and patients with valvular regurgitation. Invest Radiol. 2009;44(10):669-675.

    Not in our library10.1097/rli.0b013e3181ae99b5 (opens in a new tab)

  87. 87.

    Ha H, Kvitting JP, Dyverfeldt P, Ebbers T. Validation of pressure drop assessment using 4D flow MRI-based turbulence production in various shapes of aortic stenoses. Magn Reson Med. 2019;81(2):893-906.

    Not in our library10.1002/mrm.27437 (opens in a new tab)

  88. 88.

    Knobloch V, Binter C, Gülan U, Sigfridsson A, Holzner M, Lüthi B, et al. Mapping mean and fluctuating velocities by Bayesian multipoint MR velocity encoding-validation against 3D particle tracking velocimetry. Magn Reson Med. 2014;71(4):1405-1415.

    Not in our library10.1002/mrm.24785 (opens in a new tab)

  89. 89.

    Töger J, Bidhult S, Revstedt J, Carlsson M, Arheden H, Heiberg E. Independent validation of four-dimensional flow MR velocities and vortex ring volume using particle imaging velocimetry and planar laser-Induced fluorescence. Magn Reson Med. 2016;75(3):1064-1075.

    Not in our library10.1002/mrm.25683 (opens in a new tab)

  90. 90.

    Nordell B, Ståhlberg F, Ericsson A, Ranta C. A rotating phantom for the study of flow effects in MR imaging. Magn Reson Imaging. 1988;6(6):695-705.

    Not in our library10.1016/0730-725x(88)90094-x (opens in a new tab)

  91. 91.

    Vali A, Schmitter S, Ma L, Flassbeck S, Schmidt S, Markl M, et al. Development of a rotation phantom for phase contrast MRI sequence validation and quality control. Magn Reson Med. 2020;84(6):3333-3341.

    Not in our library10.1002/mrm.28343 (opens in a new tab)

  92. 92.

    Lee KL, Doorly DJ, Firmin DN. Numerical simulations of phase contrast velocity mapping of complex flows in an anatomically realistic bypass graft geometry. Med Phys. 2006;33(7):2621-2631.

    Not in our library10.1118/1.2208931 (opens in a new tab)

  93. 93.

    Petersson S, Dyverfeldt P, Gårdhagen R, Karlsson M, Ebbers T. Simulation of phase contrast MRI of turbulent flow. Magn Reson Med. 2010;64(4):1039-1046.

    Not in our library10.1002/mrm.22494 (opens in a new tab)

  94. 94.

    Puiseux T, Sewonu A, Moreno R, Mendez S, Nicoud F. Numerical simulation of time-resolved 3D phase-contrast magnetic resonance imaging. PLoS ONE. 2021;16(3):e0248816.

    Not in our library10.1371/journal.pone.0248816 (opens in a new tab)

  95. 95.
  96. 96.

    Dyvorne H, Knight-Greenfield A, Jajamovich G, Besa C, Cui Y, Stalder A, et al. Abdominal 4D flow MR imaging in a breath hold: combination of spiral sampling and dynamic compressed sensing for highly accelerated acquisition. Radiology. 2015;275(1):245-254.

    Not in our library10.1148/radiol.14140973 (opens in a new tab)

  97. 97.

    Ma LE, Markl M, Chow K, Huh H, Forman C, Vali A, et al. Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction. Magn Reson Med. 2019;81(6):3675-3690.

    Not in our library10.1002/mrm.27684 (opens in a new tab)

  98. 98.

    Garg P, Westenberg JJM, van den Boogaard PJ, Swoboda PP, Aziz R, Foley JRJ, et al. Comparison of fast acquisition strategies in whole-heart four-dimensional flow cardiac MR: two-center, 1.5 Tesla, phantom and in vivo validation study. J Magn Reson Imaging. 2018;47(1):272-81.

    Not in our library10.1002/jmri.25746 (opens in a new tab)

  99. 99.

    Dillinger H, Walheim J, Kozerke S. On the limitations of echo planar 4D flow MRI. Magn Reson Med. 2020;84(4):1806-1816.

    Not in our library10.1002/mrm.28236 (opens in a new tab)

  100. 100.

    Johnson KM, Lum DP, Turski PA, Block WF, Mistretta CA, Wieben O. Improved 3D phase contrast MRI with off-resonance corrected dual echo VIPR. Magn Reson Med. 2008;60(6):1329-1336.

    Not in our library10.1002/mrm.21763 (opens in a new tab)

  101. 101.

    Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47(6):1202-1210.

    Not in our library10.1002/mrm.10171 (opens in a new tab)

  102. 102.

    Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42(5):952-962.

    Not in our library10.1002/(sici)1522-2594(199911)42:5<952::aid-mrm16>3.0.co;2-s (opens in a new tab)

  103. 103.

    Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med. 1997;38(4):591-603.

    Not in our library10.1002/mrm.1910380414 (opens in a new tab)

  104. 104.

    Bammer R, Hope TA, Aksoy M, Alley MT. Time-resolved 3D quantitative flow MRI of the major intracranial vessels: initial experience and comparative evaluation at 1.5T and 3.0T in combination with parallel imaging. Magn Reson Med. 2007;57(1):127-40.

    Not in our library10.1002/mrm.21109 (opens in a new tab)

  105. 105.

    Tsao J, Boesiger P, Pruessmann KP. k-t BLAST and k-t SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med. 2003;50(5):1031-1042.

    Not in our library10.1002/mrm.10611 (opens in a new tab)

  106. 106.

    Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P. Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE. Magn Reson Med. 2005;54(6):1430-1438.

    Not in our library10.1002/mrm.20730 (opens in a new tab)

  107. 107.

    Stadlbauer A, van der Riet W, Crelier G, Salomonowitz E. Accelerated time-resolved three-dimensional MR velocity mapping of blood flow patterns in the aorta using SENSE and k-t BLAST. Eur J Radiol. 2010;75(1):e15-21.

    Not in our library10.1016/j.ejrad.2009.06.009 (opens in a new tab)

  108. 108.

    Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58(6):1182-1195.

    Not in our library10.1002/mrm.21391 (opens in a new tab)

  109. 109.

    Hutter J, Schmitt P, Aandal G, Greiser A, Forman C, Grimm R, et al. Low-rank and sparse matrix decomposition for compressed sensing reconstruction of magnetic resonance 4D phase contrast blood flow imaging (loSDeCoS 4D-PCI). Med Image Comput Comput Assist Interv. 2013;16(Pt 1):558-565.

    Not in our library10.1007/978-3-642-40811-3_70 (opens in a new tab)

  110. 110.

    Nett EJ, Johnson KM, Frydrychowicz A, Del Rio AM, Schrauben E, Francois CJ, et al. Four-dimensional phase contrast MRI with accelerated dual velocity encoding. J Magn Reson Imaging. 2012;35(6):1462-1471.

    Not in our library10.1002/jmri.23588 (opens in a new tab)

  111. 111.

    Schnell S, Ansari SA, Wu C, Garcia J, Murphy IG, Rahman OA, et al. Accelerated dual-venc 4D flow MRI for neurovascular applications. J Magn Reson Imaging. 2017;46(1):102-114.

    Not in our library10.1002/jmri.25595 (opens in a new tab)

  112. 112.

    Zwart NR, Pipe JG. Multidirectional high-moment encoding in phase contrast MRI. Magn Reson Med. 2013;69(6):1553-1564.

    Not in our library10.1002/mrm.24390 (opens in a new tab)

  113. 113.

    Moersdorf R, Treutlein M, Kroeger JR, Ruijsink B, Wong J, Maintz D, et al. Precision, reproducibility and applicability of an undersampled multi-venc 4D flow MRI sequence for the assessment of cardiac hemodynamics. Magn Reson Imaging. 2019;61:73-82.

    Not in our library10.1016/j.mri.2019.05.015 (opens in a new tab)

  114. 114.

    Binter C, Knobloch V, Manka R, Sigfridsson A, Kozerke S. Bayesian multipoint velocity encoding for concurrent flow and turbulence mapping. Magn Reson Med. 2013;69(5):1337-1345.

    Not in our library10.1002/mrm.24370 (opens in a new tab)

  115. 115.

    Johnson KM, Markl M. Improved SNR in phase contrast velocimetry with five-point balanced flow encoding. Magn Reson Med. 2010;63(2):349-355.

    Not in our library10.1002/mrm.22202 (opens in a new tab)

  116. 116.

    Lee AT, Pike GB, Pelc NJ. Three-point phase-contrast velocity measurements with increased velocity-to-noise ratio. Magn Reson Med. 1995;33(1):122-126.

    Not in our library10.1002/mrm.1910330119 (opens in a new tab)

  117. 117.

    Fathi MF, Perez-Raya I, Baghaie A, Berg P, Janiga G, Arzani A, et al. Super-resolution and denoising of 4D-Flow MRI using physics-Informed deep neural nets. Comput Methods Programs Biomed. 2020;197:105729.

    Not in our library10.1016/j.cmpb.2020.105729 (opens in a new tab)

  118. 118.
  119. 119.

    Berhane HA-O, Scott MA-O, Barker AJ, McCarthy P, Avery RA-O, Allen B, et al. Deep learning-based velocity antialiasing of 4D-flow MRI. Magn Reson Med. 2005.

    Not in our library10.1002/mrm.29205 (opens in a new tab)

  120. 120.

    Bastkowski R, Bindermann R, Brockmeier K, Weiss K, Maintz D, Giese D. Respiration dependency of caval blood flow in patients with Fontan circulation: quantification using 5D flow MRI. Radiol Cardiothorac Imaging. 2019;1(4):e190005.

    Not in our library10.1148/ryct.2019190005 (opens in a new tab)

  121. 121.

    Ma L, Yerly J, Di Sopra L, Piccini D, Lee J, DiCarlo A, et al. Using 5D flow MRI to decode the effects of rhythm on left atrial 3D flow dynamics in patients with atrial fibrillation. Magn Reson Med. 2021;85(6):3125-3139.

    Not in our library10.1002/mrm.28642 (opens in a new tab)

  122. 122.

    Ma LE, Yerly J, Piccini D, Di Sopra L, Roy CW, Carr JC, et al. 5D Flow MRI: A fully self-gated, free-running framework for cardiac and respiratory motion-resolved 3D hemodynamics. Radiol Cardiothorac Imaging. 2020;2(6):e200219.

    Not in our library10.1148/ryct.2020200219 (opens in a new tab)

  123. 123.

    Walheim J, Dillinger H, Kozerke S. Multipoint 5D flow cardiovascular magnetic resonance - accelerated cardiac- and respiratory-motion resolved mapping of mean and turbulent velocities. J Cardiovasc Magn Reson. 2019;21(1):42.

    Not in our library10.1186/s12968-019-0549-0 (opens in a new tab)

  124. 124.

    Cheng JY, Zhang T, Alley MT, Uecker M, Lustig M, Pauly JM, et al. Comprehensive multi-dimensional MRI for the simultaneous assessment of cardiopulmonary anatomy and physiology. Sci Rep. 2017;7(1):5330.

    Not in our library10.1038/s41598-017-04676-8 (opens in a new tab)

  125. 125.

    Bastkowski R, Weiss K, Maintz D, Giese D. Self-gated golden-angle spiral 4D flow MRI. Magn Reson Med. 2018;80(3):904-913.

    Not in our library10.1002/mrm.27085 (opens in a new tab)

  126. 126.

    Falcão MA-O, Di Sopra LA-O, Ma LA-O, Bacher MA-O, Yerly JA-O, Speier PA-O, et al. Pilot tone navigation for respiratory and cardiac motion-resolved free-running 5D flow MRI. Magn Reson Med. 2022.

    Not in our library10.1002/mrm.29023 (opens in a new tab)

  127. 127.

    Berhane H, Scott M, Elbaz M, Jarvis K, McCarthy P, Carr J, et al. Fully automated 3D aortic segmentation of 4D flow MRI for hemodynamic analysis using deep learning. Magn Reson Med. 2020;84(4):2204-2218.

    Not in our library10.1002/mrm.28257 (opens in a new tab)

  128. 128.

    Fujiwara T, Berhane H, Scott MB, Englund EK, Schäfer M, Fonseca B, et al. Segmentation of the aorta and pulmonary arteries based on 4D flow MRI in the pediatric setting using fully automated multi-site, multi-vendor, and multi-label dense U-net. J Magn Reson Imaging. 2021.

    Not in our library10.1002/jmri.27995 (opens in a new tab)

  129. 129.

    Bustamante M, Petersson S, Eriksson J, Alehagen U, Dyverfeldt P, Carlhäll CJ, et al. Atlas-based analysis of 4D flow CMR: automated vessel segmentation and flow quantification. J Cardiovasc Magn Reson. 2015;17:87.

    Not in our library10.1186/s12968-015-0190-5 (opens in a new tab)

  130. 130.

    Vishnevskiy V, Walheim J, Kozerke S. Deep variational network for rapid 4D flow MRI reconstruction. Nat Mach Intell. 2020;2:228.

    Not in our library10.1038/s42256-020-0165-6 (opens in a new tab)

  131. 131.

    Pelc NJ, Bernstein MA, Shimakawa A, Glover GH. Encoding strategies for three-direction phase-contrast MR imaging of flow. J Magn Reson Imaging. 1991;1(4):405-413.

    Not in our library10.1002/jmri.1880010404 (opens in a new tab)

  132. 132.

    Dyverfeldt P, Sigfridsson A, Kvitting JP, Ebbers T. Quantification of intravoxel velocity standard deviation and turbulence intensity by generalizing phase-contrast MRI. Magn Reson Med. 2006;56(4):850-858.

    Not in our library10.1002/mrm.21022 (opens in a new tab)

  133. 133.

    Dyverfeldt P, Kvitting JP, Sigfridsson A, Engvall J, Bolger AF, Ebbers T. Assessment of fluctuating velocities in disturbed cardiovascular blood flow: in vivo feasibility of generalized phase-contrast MRI. J Magn Reson Imaging. 2008;28(3):655-663.

    Not in our library10.1002/jmri.21475 (opens in a new tab)

  134. 134.

    Haraldsson H, Kefayati S, Ahn S, Dyverfeldt P, Lantz J, Karlsson M, et al. Assessment of Reynolds stress components and turbulent pressure loss using 4D flow MRI with extended motion encoding. Magn Reson Med. 2018;79(4):1962-1971.

    Not in our library10.1002/mrm.26853 (opens in a new tab)

  135. 135.

    Gu T, Korosec FR, Block WF, Fain SB, Turk Q, Lum D, et al. PC VIPR: a high-speed 3D phase-contrast method for flow quantification and high-resolution angiography. AJNR Am J Neuroradiol. 2005;26(4):743-749.

    Not in our library

  136. 136.

    Schrauben EM, Anderson AG, Johnson KM, Wieben O. Respiratory-induced venous blood flow effects using flexible retrospective double-gating. J Magn Reson Imaging. 2015;42(1):211-216.

    Not in our library10.1002/jmri.24746 (opens in a new tab)

  137. 137.

    Krämer M, Motaal AG, Herrmann KH, Löffler B, Reichenbach JR, Strijkers GJ, et al. Cardiac 4D phase-contrast CMR at 94 T using self-gated ultra-short echo time (UTE) imaging. J Cardiovasc Magn Reson. 2017;19(1):39.

    Not in our library10.1186/s12968-017-0351-9 (opens in a new tab)

  138. 138.

    O'Brien KR, Myerson SG, Cowan BR, Young AA, Robson MD. Phase contrast ultrashort TE: a more reliable technique for measurement of high-velocity turbulent stenotic jets. Magn Reson Med. 2009;62(3):626-636.

    Not in our library10.1002/mrm.22051 (opens in a new tab)

  139. 139.

    Kadbi M, Negahdar M, Cha JW, Traughber M, Martin P, Stoddard MF, et al. 4D UTE flow: a phase-contrast MRI technique for assessment and visualization of stenotic flows. Magn Reson Med. 2015;73(3):939-950.

    Not in our library10.1002/mrm.25188 (opens in a new tab)

  140. 140.

    Sigfridsson A, Petersson S, Carlhäll CJ, Ebbers T. Four-dimensional flow MRI using spiral acquisition. Magn Reson Med. 2012;68(4):1065-1073.

    Not in our library10.1002/mrm.23297 (opens in a new tab)

  141. 141.

    Duerk JL, Simonetti OP. Theoretical aspects of motion sensitivity and compensation in echo-planar imaging. J Magn Reson Imaging. 1991;1(6):643-650.

    Not in our library10.1002/jmri.1880010605 (opens in a new tab)

  142. 142.

    Viola F, Dyverfeldt P, Carlhäll CJ, Ebbers T. Data quality and optimal background correction order of respiratory-gated k-space segmented spoiled gradient echo (SGRE) and echo planar imaging (EPI)-based 4D flow MRI. J Magn Reson Imaging. 2020;51(3):885-896.

    Not in our library10.1002/jmri.26879 (opens in a new tab)

  143. 143.

    Westenberg JJM, van Assen HC, van den Boogaard PJ, Goeman JJ, Saaid H, Voorneveld J, et al. Echo planar imaging-induced errors in intracardiac 4D flow MRI quantification. Magn Reson Med. 2022.

    Not in our library10.1002/mrm.29112 (opens in a new tab)

  144. 144.

    Carlsson M, Töger J, Kanski M, Bloch KM, Ståhlberg F, Heiberg E, et al. Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T. J Cardiovasc Magn Reson. 2011;13(1):55.

    Not in our library10.1186/1532-429x-13-55 (opens in a new tab)

  145. 145.

    Jung B, Honal M, Ullmann P, Hennig J, Markl M. Highly k-t-space-accelerated phase-contrast MRI. Magn Reson Med. 2008;60(5):1169-1177.

    Not in our library10.1002/mrm.21764 (opens in a new tab)

  146. 146.

    Schnell S, Markl M, Entezari P, Mahadewia RJ, Semaan E, Stankovic Z, et al. k-t GRAPPA accelerated four-dimensional flow MRI in the aorta: effect on scan time, image quality, and quantification of flow and wall shear stress. Magn Reson Med. 2014;72(2):522-533.

    Not in our library10.1002/mrm.24925 (opens in a new tab)

  147. 147.

    Giese D, Wong J, Greil GF, Buehrer M, Schaeffter T, Kozerke S. Towards highly accelerated Cartesian time-resolved 3D flow cardiovascular magnetic resonance in the clinical setting. J Cardiovasc Magn Reson. 2014;16(1):42.

    Not in our library10.1186/1532-429x-16-42 (opens in a new tab)

  148. 148.

    Bollache E, Barker AJ, Dolan RS, Carr JC, van Ooij P, Ahmadian R, et al. k-t accelerated aortic 4D flow MRI in under two minutes: feasibility and impact of resolution, k-space sampling patterns, and respiratory navigator gating on hemodynamic measurements. Magn Reson Med. 2018;79(1):195-207.

    Not in our library10.1002/mrm.26661 (opens in a new tab)

  149. 149.

    Neuhaus E, Weiss K, Bastkowski R, Koopmann J, Maintz D, Giese D. Accelerated aortic 4D flow cardiovascular magnetic resonance using compressed sensing: applicability, validation and clinical integration. J Cardiovasc Magn Reson. 2019;21(1):65.

    Not in our library10.1186/s12968-019-0573-0 (opens in a new tab)

  150. 150.

    Braig M, Menza M, Leupold J, LeVan P, Feng L, Ko CW, et al. Analysis of accelerated 4D flow MRI in the murine aorta by radial acquisition and compressed sensing reconstruction. NMR Biomed. 2020;33(11):e4394.

    Not in our library10.1002/nbm.4394 (opens in a new tab)

  151. 151.

    Walheim J, Dillinger H, Gotschy A, Kozerke S. 5D flow tensor MRI to efficiently map Reynolds stresses of aortic blood flow in-vivo. Sci Rep. 2019;9:18794.

    Not in our library10.1038/s41598-019-55353-x (opens in a new tab)

  152. 152.
  153. 153.

    Pathrose A, Ma L, Berhane H, Scott MB, Chow K, Forman C, et al. Highly accelerated aortic 4D flow MRI using compressed sensing: Performance at different acceleration factors in patients with aortic disease. Magn Reson Med. 2021;85(4):2174-2187.

    Not in our library10.1002/mrm.28561 (opens in a new tab)

  154. 154.

    Ehman RL, Felmlee JP. Adaptive technique for high-definition MR imaging of moving structures. Radiology. 1989;173(1):255-263.

    Not in our library10.1148/radiology.173.1.2781017 (opens in a new tab)

  155. 155.

    McConnell MV, Khasgiwala VC, Savord BJ, Chen MH, Chuang ML, Edelman RR, et al. Comparison of respiratory suppression methods and navigator locations for MR coronary angiography. AJR Am J Roentgenol. 1997;168(5):1369-1375.

    Not in our library10.2214/ajr.168.5.9129447 (opens in a new tab)

  156. 156.

    Wang Y, Rossman PJ, Grimm RC, Riederer SJ, Ehman RL. Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography. Radiology. 1996;198(1):55-60.

    Not in our library10.1148/radiology.198.1.8539406 (opens in a new tab)

  157. 157.

    Akçakaya M, Gulaka P, Basha TA, Ngo LH, Manning WJ, Nezafat R. Free-breathing phase contrast MRI with near 100% respiratory navigator efficiency using k-space-dependent respiratory gating. Magn Reson Med. 2014.

    Not in our library10.1002/mrm.24874 (opens in a new tab)

  158. 158.

    Buehrer M, Curcic J, Boesiger P, Kozerke S. Prospective self-gating for simultaneous compensation of cardiac and respiratory motion. Magn Reson Med. 2008;60(3):683-690.

    Not in our library10.1002/mrm.21697 (opens in a new tab)

  159. 159.

    Cheng JY, Alley MT, Cunningham CH, Vasanawala SS, Pauly JM, Lustig M. Nonrigid motion correction in 3D using autofocusing with localized linear translations. Magn Reson Med. 2012;68(6):1785-1797.

    Not in our library10.1002/mrm.24189 (opens in a new tab)

  160. 160.

    Di Sopra L, Piccini D, Coppo S, Stuber M, Yerly J. An automated approach to fully self-gated free-running cardiac and respiratory motion-resolved 5D whole-heart MRI. Magn Reson Med. 2019;82(6):2118-2132.

    Not in our library10.1002/mrm.27898 (opens in a new tab)

  161. 161.

    Feng L, Axel L, Chandarana H, Block KT, Sodickson DK, Otazo R. XD-GRASP: golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing. Magn Reson Med. 2016;75(2):775-788.

    Not in our library10.1002/mrm.25665 (opens in a new tab)

  162. 162.

    Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP. Self-gated cardiac cine MRI. Magn Reson Med. 2004;51(1):93-102.

    Not in our library10.1002/mrm.10664 (opens in a new tab)

  163. 163.

    Uribe S, Muthurangu V, Boubertakh R, Schaeffter T, Razavi R, Hill DL, et al. Whole-heart cine MRI using real-time respiratory self-gating. Magn Reson Med. 2007;57(3):606-613.

    Not in our library10.1002/mrm.21156 (opens in a new tab)

  164. 164.

    Gottwald LM, Blanken CPS, Tourais J, Smink J, Planken RN, Boekholdt SM, et al. Retrospective camera-based respiratory gating in clinical whole-heart 4D flow MRI. J Magn Reson Imaging. 2021;54(2):440-451.

    Not in our library10.1002/jmri.27564 (opens in a new tab)

  165. 165.

    Kanski M, Töger J, Steding-Ehrenborg K, Xanthis C, Bloch KM, Heiberg E, et al. Whole-heart four-dimensional flow can be acquired with preserved quality without respiratory gating, facilitating clinical use: a head-to-head comparison. BMC Med Imaging. 2015;15:20.

    Not in our library10.1186/s12880-015-0061-4 (opens in a new tab)

  166. 166.

    Nordmeyer S, Riesenkampff E, Crelier G, Khasheei A, Schnackenburg B, Berger F, et al. Flow-sensitive four-dimensional cine magnetic resonance imaging for offline blood flow quantification in multiple vessels: a validation study. J Magn Reson Imaging. 2010;32(3):677-683.

    Not in our library10.1002/jmri.22280 (opens in a new tab)

  167. 167.

    Gabbert DD, Trotz P, Kheradvar A, Jerosch-Herold M, Scheewe J, Kramer HH, et al. Abnormal torsion and helical flow patterns of the neo-aorta in hypoplastic left heart syndrome assessed with 4D-flow MRI. Cardiovasc Diagn Ther. 2021;11(6):1379-1388.

    Not in our library10.21037/cdt-20-770 (opens in a new tab)

  168. 168.

    Oechtering TA-O, Sieren M, Schubert K, Schaller T, Scharfschwerdt MA-O, Panagiotopoulos A, et al. In vitro 4D flow MRI evaluation of aortic valve replacements reveals disturbed flow distal to biological but not to mechanical valves. J Card Surg. 2019.

    Not in our library10.1111/jocs.14253 (opens in a new tab)

  169. 169.

    Petersson S, Dyverfeldt P, Ebbers T. Assessment of the accuracy of MRI wall shear stress estimation using numerical simulations. J Magn Reson Imaging. 2012.

    Not in our library10.1002/jmri.23610 (opens in a new tab)

Questions about this paper

We email you if it is retracted, withdrawn, corrected or flagged.

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.