Characterization of magnetic nanoparticle by dynamic light scattering
JitKang Lim, Swee Pin Yeap, Hui Xin Che and 1 more
Nanoscale Research Letters · 2013
Abstract
Here we provide a complete review on the use of dynamic light scattering (DLS) to study the size distribution and colloidal stability of magnetic nanoparticles (MNPs). The mathematical analysis involved in obtaining size information from the correlation function and the calculation of Z-average are introduced. Contributions from various variables, such as surface coating, size differences, and concentration of particles, are elaborated within the context of measurement data. Comparison with other sizing techniques, such as transmission electron microscopy and dark-field microscopy, revealed both the advantages and disadvantages of DLS in measuring the size of magnetic nanoparticles. The self-assembly process of MNP with anisotropic structure can also be monitored effectively by DLS.
Full text
From PubMed Central, U.S. National Library of Medicine, as current on Oct 18, 2026. Paragraph numbers are ours, so an answer can point at one.
Introduction
Magnetic nanoparticles (MNPs) with a diameter between 1 to 100 nm have found uses in many applications. This nanoscale magnetic material has several advantages that provide many exciting opportunities or even a solution to various biomedically and environmentally related problems. Firstly, it is possible to synthesize a wide range of MNPs with well-defined structures and size which can be easily matched with the interest of targeted applications. Secondly, the MNP itself can be manipulated by an externally applied magnetic force. The capability to control the spatial evolution of MNPs within a confined space provides great benefits for the development of sensing and diagnostic system/techniques. Moreover MNPs, such as Fe0 and Fe3O4, that exhibit a strong catalytic function can be employed as an effective nanoagent to remove a number of persistent pollutants from water resources. In addition to all the aforementioned advantages, the recent development of various techniques and procedures for producing highly monodispersed and size-controllable MNPs has played a pivotal role in promoting the active explorations and research of MNPs.
In all of the applications involving the use of MNPs, the particle size remained as the most important parameter as many of the chemical and physical properties associated to MNPs are strongly dependent upon the nanoparticle diameter. In particular, one of the unique features of a MNP is its high-surface-to-volume ratio, and this property is inversely proportional to the diameter of the MNP. The smaller the MNP is, the larger its surface area and, hence, the more loading sites are available for applications such as drug delivery and heavy metal removal. Furthermore, nanoparticle size also determines the magnetophoretic forces (Fmag) experienced by a MNP since Fmag is directly proportional to the volume of the particles. In this regard, having size information is crucial as at nanoregime, the MNP is extremely susceptible to Stoke’s drag and thermal randomization energy. The successful manipulation of MNP can only be achieved if the Fmag introduced is sufficient to overcome both thermal and viscous hindrances. In addition, evidences on the (eco)toxicological impacts of nanomaterials have recently surfaced. The contributing factors of nanotoxicity are still a subject of debate; however, it is very likely due to either (1) the characteristic small dimensional effects of nanomaterials that are not shared by their bulk counterparts with the same chemical composition or (2) biophysicochemical interactions at the nano-bio interface dictated by colloidal forces. For either reason, the MNP’s size is one of the determining factors.
The technique of dynamic light scattering (DLS) has been widely employed for sizing MNPs in liquid phase. However, the precision of the determined particle size is not completely understood due to a number of unevaluated effects, such as concentration of particle suspension, scattering angle, and shape anisotropy of nanoparticles. In this review, the underlying working principle of DLS is first provided to familiarize the readers with the mathematical analysis involved for correct interpretation of DLS data. Later, the contribution from various factors, such as suspension concentration, particle shape, colloidal stability, and surface coating of MNPs, in dictating the sizing of MNPs by DLS is discussed in detail. It is the intention of this review to summarize some of the important considerations in using DLS as an analytical tool for the characterization of MNPs.
Overview of sizing techniques for MNPs
There are numerous analytical techniques, such as DLS, transmission electron miscroscopy (TEM), thermomagnetic measurement, dark-field microscopy, atomic force microscopy (AFM), and acoustic spectrometry measurement, that have been employed to measure the size/size distribution of MNPs (Table 1). TEM is one of the most powerful analytical tools available which can give direct structural and size information of the MNP. Through the use of the short wavelengths achievable with highly accelerated electrons, it is capable to investigate the structure of a MNP down to the atomic level of detail, whereas by performing image analysis on the TEM micrograph obtained, it is possible to give quantitative results on the size distribution of the MNP. This technique, however, suffered from the small sampling size involved. A typical MNP suspension composed of 1010 to 1015 particles/mL and the size analysis by measuring thousands or even tens of thousands of particles still give a relatively small sample pool to draw statistically conclusive remarks.
Thermomagnetic measurement extracts the size distribution of an ensemble of superparamagnetic nanoparticles from zero-field cooling (ZFC) magnetic moment, mZFC(T), data based on the Néel model. This method is an indirect measurement of particle size and relies on the underlying assumption of the mathematical model used to calculate the size distribution. In addition, another limitation of this analytical method includes the magnetic field applied for ZFC measurements which must be small compared to the anisotropy field of the MNPs, and it also neglects particle-particle dipolar interactions which increase the apparent blocking temperature. This technique, however, could give a very reliable magnetic size of the nanoparticle analyzed.
Dark-field microscopy relies on direct visual inspection of the optical signal emitted from the MNP while it undergoes Brownian motion. After the trajectories of each MNP over time t are recorded, the two-dimensional mean-squared displacement <r2> = 4Dt is used to calculate the diffusion coefficient D for each particle. Later on, the hydrodynamic diameters can be estimated via the Stokes-Einstein equation for the diffusion coefficients calculated for individual particles, averaging over multiple time steps. Successful implementation of this technique depends on the ability to trace the particle optically by coating the MNP with a noble metal that exhibits surface Plasmon resonance within a visible wavelength. This extra synthesis step has significantly restricted the use of this technique as a standard route for sizing MNPs. The size of an MNP obtained through dark-field microscopy is normally larger than the TEM and DLS results. It should be noted that dark-field microscopy can also be employed for direct visualization of a particle flocculation event. As for AFM, besides the usual topographic analysis, magnetic imaging of a submicron-sized MNP grown on GaAs substrate has been performed with magnetic force microscopy equipment. Despite all the recent breakthroughs, sample preparation and artifact observation are still the limiting aspect for the wider use of this technology for sizing MNPs.
The particle size and size distribution can also be measured with an acoustic spectrometer which utilizes the sound pulses transmitted through a particle suspension to extract the size-related information. Based on the combined effect of absorption and scattering of acoustic energy, an acoustic sensor measures attenuation frequency spectra in the sample. This attenuation spectrum is used to calculate the particle size distribution. This technique has advantages over the light scattering method in studying samples with high polydispersity as the raw data for calculating particle size depend on only the third power of the particle size. This scenario makes contribution of the small (nano) and larger particles more even and the method potentially more sensitive to the nanoparticle content even in the very broad size distributions.
DLS, also known as photon correlation spectroscopy, is one of the most popular methods used to determine the size of MNPs. During the DLS measurement, the MNP suspension is exposed to a light beam (electromagnetic wave), and as the incident light impinges on the MNP, the direction and intensity of the light beam are both altered due to a process known as scattering. Since the MNPs are in constant random motion due to their kinetic energy, the variation of the intensity with time, therefore, contains information on that random motion and can be used to measure the diffusion coefficient of the particles. Depending on the shape of the MNP, for spherical particles, the hydrodynamic radius of the particle RH can be calculated from its diffusion coefficient by the Stokes-Einstein equation Df= kBT/6πηRH, where kB is the Boltzmann constant, T is the temperature of the suspension, and η is the viscosity of the surrounding media. Image analysis on the TEM micrographs gives the ‘true radius’ of the particles (though determined on a statistically small sample), and DLS provides the hydrodynamic radius on an ensemble average. The hydrodynamic radius is the radius of a sphere that has the same diffusion coefficient within the same viscous environment of the particles being measured. It is directly related to the diffusive motion of the particles.
DLS has several advantages for sizing MNPs and has been widely used to determine the hydrodynamic size of various MNPs as shown in Table 2. First of all, the measuring time for DLS is short, and it is almost all automated, so the entire process is less labor intensive and an extensive experience is not required for routine measurement. Furthermore, this technique is non-invasive, and the sample can be employed for other purposes after the measurement. This feature is especially important for the recycle use of MNP with an expensive surface functional group, such as an enzyme or molecular ligands. In addition, since the scattering intensity is directly proportional to the sixth power of the particle radius, this technique is extremely sensitive towards the presence of small aggregates. Hence, erroneous measurement can be prevented quite effectively even with the occurrences of limited aggregation events. This unique feature makes DLS one of the very powerful techniques in monitoring the colloidal stability of MNP suspension.
The underlying principle of DLS
The interaction of very small particles with light defined the most fundamental observations such as why is the sky blue. From a technological perspective, this interaction also formed the underlying working principle of DLS. It is the purpose of this section to describe the mathematical analysis involved to extract size-related information from light scattering experiments.
The correlation function
DLS measures the scattered intensity over a range of scattering angles θdls for a given time tk in time steps ∆t. The time-dependent intensity I(q, t) fluctuates around the average intensity I(q) due to the Brownian motion of the particles:
where [I(q)] represents the time average of I(q). Here, it is assumed that tk, the total duration of the time step measurements, is sufficiently large such that I(q) represents average of the MNP system. In a scattering experiment, normally, θdls (see Figure 1) is expressed as the magnitude of the scattering wave vector q as
where n is the refractive index of the solution and λ is the wavelength in vacuum of the incident light. Figure 2a illustrates typical intensity fluctuation arising from a dispersion of large particles and a dispersion of small particles. As the small particles are more susceptible to random forces, the small particles cause the intensity to fluctuate more rapidly than the large ones.
The time-dependent intensity fluctuation of the scattered light at a particular angle can then be characterized with the introduction of the autocorrelation function as
where τ = i ∆t is the delay time, which represents the time delay between two signals I(q,i Δt) and I(q,(i + j) Δt). The function C(q,τ) is obtained for a series of τ and represents the correlation between the intensity at t1 (I(q,t1)) and the intensity after a time delay of τ (I(q,t1 + τ)). The last part of the equation shows how the autocorrelation function is calculated experimentally when the intensity is measured in discrete time steps. As for nanoparticle dispersion, the autocorrelation function decays more rapidly for small particles than for the large particles as depicted in Figure 2b. The autocorrelation function has its highest value of [I(q,0)]2 at τ = 0. As τ becomes sufficiently large at long time scales, the fluctuations becomes uncorrelated and C(q,τ) decreases to [I(q)]2. For non-periodic I(q,t), a monotonic decay of C(q,τ) is observed as τ increases from zero to infinity and
where ξ is an instrument constant approximately equal to unity and g(1)(q,τ) is the normalized electric field correlation function. Equation 4 is known as the Siegert relation and is valid except in the case of scattering volume with a very small number of scatterers or when the motion of the scatterers is limited. For monodisperse, spherical particles, g(1)(τ) is given by
Once the value of Df is obtained, the hydrodynamic diameter of a perfectly monodisperse dispersion composed of spherical particles can be inferred from the Stokes-Einstein equation. Practically, the correlation function observed is not a single exponential decay but can be expressed as
where G(Γ) is the distribution of decay rates Γ. For a narrowly distributed decay rate, cumulant method can be used to analyze the correlation function. A properly normalized correlation function can be expressed as
where 〈Γ〉 is the average decay rate and can be defined as
and μ2 = 〈Γ〉2 − 〈Γ〉2 is the variance of the decay rate distribution. Then, the polydispersity index (PI) is defined as PI = μ2/〈Γ〉2. The average hydrodynamic radius is obtained from the average decay rate 〈Γ〉 using the relation
Z-average
In most cases, the DLS results are often expressed in terms of the Z-average. Since the Z-average arises when DLS data are analyzed through the use of the cumulant technique, it is also known as the “cumulant mean.” Under Rayleigh scattering, the amount of light scattered by a single particle is proportional to the sixth power of its radius (volume squared). This scenario causes the averaged hydrodynamic radius determined by DLS to be also weighted by volume squared. Such an averaged property is called the Z-average. For particle suspension with discrete size distribution, the Z-average of some arbitrary property y would be calculated as
where ni is the number of particles of type i having a hydrodynamic radius of RH,i and property y. If we assume that this particle dispersion consists of exactly two sizes of particles 1 and 2, then Equation 10 yields
where RH,i and yi are the volume and arbitrary property for particle 1 (i = 1) and particle 2 (i = 2). Suppose that two particles 1 combined to form one particle 2 and assume that we start with n0 total of particle 1, some of which combined to form n2 number of particle 2. With this assumption, we have n1 = n0– n2 number of particle 1. Moreover, under this assumption RH,2 = 2 RH,1. Substitute these relations into Equation 11; then, the Z-average of property y becomes
where 2n2/n0 is the fraction of total particle 1 existing as particle 2. Solving this fraction, we obtained
However, it should be noted that Z-average should only be employed to provide the characteristic size of the particles if the suspension is monomodal (only one peak), spherical, and monodisperse. As shown in Figure 3, for a mixture of particles with obvious size difference (bimodal distribution), the calculated Z-average carries irrelevant size information.
DLS measurement of MNPs
The underlying challenges of measuring the size of MNPs by DLS lay in the facts that (1) for engineering applications, these particles are typically coated with macromolecules to enhance their colloidal stability (see Figure 4) and (2) there present dipole-dipole magnetic interactions between the none superparamagnetic nanoparticles. Adsorbing macromolecules onto the surface of particles tends to increase the apparent RH of particles. This increase in RH is a convenient measure of the thickness of the adsorbed macromolecules. This section is dedicated to the scrutiny of these two phenomena and also suspension concentration effect in dictating the DLS measurement of MNPs. All DLS measurements were performed with a Malvern Instrument Zetasizer Nano Series (Malvern Instruments, Westborough, MA, USA) equipped with a He-Ne laser (λ = 633 nm, max 5 mW) and operated at a scattering angle of 173°. In all measurements, 1 mL of particle suspensions was employed and placed in a 10 mm × 10 mm quartz cuvette. The iron oxide MNP used in this study was synthesized by a high-temperature decomposition method.
Size dependency of MNP in DLS measurement
In order to demonstrate the sizing capability of DLS, measurements were conducted on three species of Fe3O4 MNPs produced by high-temperature decomposition method which are surface modified with oleic acid/oleylamine in toluene (Figure 5). The TEM image analyses performed on micrographs shown in Figure 5 (from top to bottom) indicate that the diameter of each particle species is 7.2 ± 0.9 nm, 14.5 ± 1.8 nm, and 20.1 ± 4.3 nm, respectively. The diameters of these particles obtained from TEM and DLS are tabulated in Table 3. It is very likely that the main differences between the measured diameters from these two techniques are due to the presence of an adsorbing layer, which is composed of oleic acid (OA) and oleylamine (OY), on the surface of the particle. Small molecular size organic compounds, such as OA and OY, are electron transparent, and therefore, they did not show up in the TEM micrograph (Figure 5). Given that the chain lengths of OA and OY are approximately 2 nm, the best match of DLS and TEM, in terms of measured diameter, can be observed from middle-sized Fe3O4 MNPs.
For small-sized MNPs, the radius of curvature effect is the main contributing factor for the large difference observed on the averaged diameter from DLS and TEM. This observation has at least suggested that for any inference of layer thickness from DLS measurement, the particles with a radius much larger than the layer thickness should be employed. In this measurement, the fractional error in the layer thickness can be much larger than the fractional error in the radius with the measurement standard deviation of only 0.9 nm for TEM but at a relatively high value of 5.2 nm for DLS. At a very large MNP size of around 20 nm (bottom image of Figure 5), the Z-average hydrodynamic diameter is 23 nm larger than the TEM size. Moreover, the standard deviation of the DLS measurement of this particle also increased significantly to 14.9 nm compared to 5.2 and 5.5 nm for small- and middle-sized MNPs, respectively. This trend of increment observed in standard deviation is consistent with TEM measurement. Both the shape irregularity and polydispersity, which are the intrinsic properties that can be found in a MNP with a diameter of 20 nm or above, contribute to this observation. For a particle larger than 100 nm, other factors such as electroviscous and surface roughness effects should be taken into consideration for the interpretation of DLS results.
MNP concentration effects
In DLS, the range of sample concentration for optimal measurements is highly dependent on the sample materials and their size. If the sample is too dilute, there may be not enough scattering events to make a proper measurement. On the other hand, if the sample is too concentrated, then multiple scattering can occur. Moreover, at high concentration, the particle might not be freely mobile with its spatial displacement driven solely by Brownian motion but with the strong influences of particle interactions. This scenario is especially true for the case of MNPs with interparticle magnetic dipole-dipole interactions.
Figure 6 illustrates the particle concentration effects on 6- and 18-nm superparamagnetic iron oxide MNPs, with no surface coating, dispersed in deionized water. Both species of MNPs show strong concentration dependency as their hydrodynamic diameter increases with the concentration increment. The hydrodynamic diameter for small particles increases from 7.1 ± 1.9 nm to 13.2 ± 3.3 nm as the MNP concentration increases from 25 to 50 mg/L. On the other hand, the hydrodynamic diameter of large particles remains to be quite constant until around 100 mg/L and then only experiences a rapid jump of the detected size from 29.3 ± 4.6 nm (at 100 mg/L) to 177.3 ± 15.8 nm (at 250 mg/L). Since the concentration of the MNP is prepared in mass basis, the presence of an absolute number of particles in a given volume of solution is almost two orders of magnitude higher in a small-particle suspension. For example, at 100 mg/L, the concentrations for small and larger particles are calculated as 1.7 × 1020 particles (pts)/m3 and 6.3 × 1018 pts/m3 by assuming that the composition material is magnetite with a density of 5.3 g/cm3. This concentration translated to a collision frequency of 85,608 s−1 and 1,056 s−1. So, at the same mass concentration, it is more likely for small particles to experience the non-self-diffusion motions.
For both species of particles, the upward trends of hydrodynamic diameter, which associates to the decrement of diffusion coefficient, reflect the presence of a strong interaction between the particles as MNP concentration increases. Furthermore, since the aggregation rate has a second-order dependency on particle concentration, the sample with high MNP concentration has higher tendency to aggregate, leading to the formation of large particle clusters. Therefore, the initial efforts for MNP characterization by using DLS should focus on the determination of the optimal working concentration.
Colloidal stability of MNPs
Another important use of DLS in the characterization of MNPs is for monitoring the colloidal stability of the particles. An iron oxide MNP coated with a thin layer of gold with a total diameter of around 50 nm is further subjected for surface functionalization by a variety of macromolecules. The colloidal stability of the MNP coated with all these macromolecules suspended in 154 mM ionic strength phosphate buffer solution (PBS) (physiologically relevant environment for biomedical application) is monitored by DLS over the course of 5 days (Figure 7). The uncoated MNP flocculated immediately after their introduction to PBS and is verified with the detection of micron-sized objects by DLS.
As shown in Figure 7, both polyethylene glycol (PEG) 6k and PEG 10k are capable of tentatively stabilizing the MNPs in PBS for the first 24 and 48 h. Aggregation is observed with the detection of particle clusters with a diameter of more than 500 nm. After this period of relative stability, aggregation accelerated to produce micron-sized aggregates by day 3. Actually, the continuous monitoring of MNP size by DLS after this point is less meaningful as the dominating motion is the sedimentation of large aggregates. For PEG 6k and PEG 10k that have a rather low degree of polymerization, the loss of stability over a day or two could have been due to slow PEG desorption that would not be expected of larger polymers. Nevertheless, PEG 100k-coated MNPs were not as well stabilized as the PEG 6k- or PEG 10k-coated ones, despite the higher degree of polymerization that one might expect to produce greater adsorbed layer thicknesses and therefore longer-ranged steric forces. In addition to the degree of polymerization, as discussed by Golas and coworkers, the colloidal stability of polymeric stabilized MNPs is also dependent on other structural differences of the polymer employed, such as the chain architecture and the identity of the charged functional unit. In their work, DLS was used to confirm the nanoparticle suspensions that displayed the least sedimentation which was indeed stable against aggregation.
In addition to the popular use of DLS in sizing individual MNPs, this analytical technique is also being employed to monitor the aggregation behavior of MNPs and the size of final clusters formed. The study of particle aggregates is important since the magnetic collection is a cooperative phenomenon. Subsequently, it is much easier to harvest submicron-sized MNP clusters than individual particles. Hence, a magnetic nanocluster with loss-packed structure and uniform size and shape has huge potential for various engineering applications in which the real-time separation is the key requirement. Therefore, the use of DLS to monitor the aggregation kinetic of MNPs is important to provide direct feedback about the time scale associated with this process. Figure 8 illustrates the aggregation behavior of three species of 40-nm reactive nanoscale iron particles (RNIP), 27.5-nm magnetite (Fe3O4) MNP, and 40-nm hematite (α-Fe2O3) MNP. Phenrat and coworkers have demonstrated that DLS can be an effective tool to probe the aggregation behavior of MNPs (Figure 8a). The time evolution of the hydrodynamic radius of these particles from monomodal to bimodal distribution revealed the aggregation kinetic of the particles. Together with the in situ optical microscopy observation, the mechanism of aggregation is proposed as the transitions from rapidly moving individual MNPs to the formation of submicron clusters that lead to chain formation and gelation (Figure 8b). By the combination of small-angle neutron scattering and cryo-TEM measurements, DLS can also be used as an effective tool to understand the fractal structure of this aggregate.
DLS measurement of non-spherical MNPs
Even though, under most circumstances, a more specialized analytical technique known as depolarized dynamic light scattering is needed to investigate the structural contribution of anisotropic materials, it is still possible to extract useful information for rod-like MNPs by conventional DLS measurement. For rod-like particles, the decay rate in Equation 6 can be defined as
where in a plot of Γ vs q2, the value of rotational diffusion DR can be obtained directly by an extrapolation of q to zero and the value of translational diffusion DT from the slope of the curve. For rigid non-interacting rods at infinite dilution with an aspect ratio (L/d) greater than 5, DR and DT can be expressed using Broersma’s relations or the stick hydrodynamic theory. By performing angle-dependent DLS analysis on rod-like β-FeOOH nanorods as shown in Figure 9a, we found that the decay rate is linearly proportional to q2 and passes through the origin (Figure 9b), suggesting that the nanorod motion is dominated by translational diffusion. From Figure 9b, the slope of the graph yields the translational diffusion coefficient, DT = 7 × 10−12 m2/s. This value of DT corresponds to an equivalent spherical hydrodynamic diameter of 62.33 nm, suggesting that the DLS results with a single fixed angle of 173° overestimated the true diameter. By taking the length and width of the nanorods as 119.7 and 17.5 nm (approximated from TEM images in Figure 9a), the DT calculated by the stick hydrodynamic theory and Broersma’s relationship is 7.09 × 10−12 m2/s and 6.84 × 10−12 m2/s, respectively, consistent with the DLS results.
Since the β-FeOOH nanorods are self-assembled in a side-by-side fashion to form highly oriented 2-D nanorod arrays and the 2-D nanorod arrays are further stacked in a face-to-face fashion to form the final 3-D layered architectures, DLS can serve as an effective tool to monitor these transient behaviors. Figure 10a depicts the structural changes of self-assembled nanorods over a time course of 7 h. To monitor the in situ real-time behavior of this self-assembly process, DLS was employed to provide the size distribution of the intermediate products that formed in the solution (Figure 10b). The temporal evolution of the detected size from 60 to 70 nm, to dual peaks, to eventually only a single distribution with a peak value of 700 nm indicating that all the building blocks are self-assembled into the large aggregates within the experiment time frame agrees well with the SEM observation (Figure 10a). This kinetic data time scale is involved in the full assembly of anisotropic nanomaterials from single building blocks to 2-D arrays and, eventually, 3-D micron-sized assemblies.
Conclusion
Dynamic light scattering is employed to monitor the hydrodynamic size and colloidal stability of the magnetic nanoparticles with either spherical or anisotropic structures. This analytical method cannot be employed solely to give feedbacks on the structural information; however, by combining with other electron microscopy techniques, DLS provides statistical representative data about the hydrodynamic size of nanomaterials. In situ, real-time monitoring of MNP suspension by DLS provides useful information regarding the kinetics of the aggregation process and, at the same time, gives quantitative measurement on the size of the particle clusters formed. In addition, DLS can be a powerful technique to probe the layer thickness of the macromolecules adsorbed onto the MNP. However, the interpretation of DLS data involves the interplay of a few parameters, such as the size, concentration, shape, polydispersity, and surface properties of the MNPs involved; hence, careful analysis is needed to extract the right information.
Competing interests
The authors declare that they have no competing interests.
Authors’ contributions
JKL synthesized the MNPs, carried out TEM analysis, and drafted the manuscript. SPY carried out DLS measurement and data analysis. HXC carried out DLS measurement and data analysis. SCL participated in the design of the study and drafted the manuscript. All authors read and approved the final manuscript.
References 87
- 1.
Lu AH Salabas EL Schüth F Magnetic nanoparticles: synthesis, protection, functionalization, and applicationAngew Chem Int Ed2007 812221244
Not in our library10.1002/anie.200602866
- 2.
Pankhurst QA Connolly J Jones SK Dobson J Applications of magnetic nanoparticles in biomedicineJ Phys D Appl Phys2003 8R167
Not in our library10.1088/0022-3727/36/13/201
- 3.
Adolphi NL Huber DL Bryant HC Monson TC Fegan DL Lim JK Trujillo JE Tessier TE Lovato DM Butler KS Provencio PP Hathaway HJ Majetich SA Larson RS Flynn ER Characterization of single-core magnetite nanoparticles for magnetic imaging by SQUID relaxometryPhys Med Biol2010 859856003
Not in our library10.1088/0031-9155/55/19/023
- 4.
Gupta AK Gupta M Synthesis and surface engineering of iron oxide nanoparticles for biomedical applicationsBiomaterials2005 839954021
Not in our library10.1016/j.biomaterials.2004.10.012
- 5.
Hao R Xing R Xu Z Hou Y Gao S Sun S Sythesis, functionalization and biomedical applications of multifunctional magnetic nanoparticlesAdv Mater2010 827292742
Not in our library10.1002/adma.201000260
- 6.
Cumbat L Greenleaf J Leun D SenGupta AK Polymer supported inorganic nanoparticles: characterization and environmental applicationsReact Funct Polym2003 8167180
Not in our library10.1016/s1381-5148(02)00192-x
- 7.
Yantasee W Warner CL Sangvanich T Addleman RS Carter TG Wiacek RJ Fryxell GE Timchalk C Warner MG Removal of heavy metals from aqueous systems with thiol functionalized superparamagnetic nanoparticlesEnviron Sci Technol2007 851145119
Not in our library10.1021/es0705238
- 8.
Hu J Lo IMC Chen G Comparative study of various magnetic nanoparticles for Cr(VI) removalSep Purif Technol2007 8249256
Not in our library10.1016/j.seppur.2007.02.009
- 9.
Dobson J Remote control of cellular behavior with magnetic nanoparticlesNat Nanotech2008 8139143
Not in our library10.1038/nnano.2008.39
- 10.
Gao J Zhang W Huang P Zhang B Zhang X Xu B Intracellular spatial control of fluorescent magnetic nanoparticlesJ Am Chem Soc2008 837103711
Not in our library10.1021/ja7103125
- 11.
Fiedor JN Bostick WD Jarabek RJ Farrell J Understanding the mechanism of uranium removal from groundwater by zero-valent iron using X-ray photoelectron spectroscopyEnviron Sci Technol1998 814661473
Not in our library10.1021/es970385u
- 12.
Feng J Hu X Yue PL Zhu HY Lu GQ Degradation of azo-dye orange II by a photoassisted Fenton reaction using a novel composite of iron oxide and silicate nanoparticles as a catalystInd Eng Chem Res2003 820582066
Not in our library10.1021/ie0207010
- 13.
Sun S Recent advances in chemical synthesis, self-assembly, and applications of FePt nanoparticlesAdv Mater2006 8393403
Not in our library10.1002/adma.200501464
- 14.
Park J Joo J Kwon SG Jang Y Hyeon T Synthesis of monodisperse spherical nanocrystalsAngew Chem Int Ed2007 846304660
Not in our library10.1002/anie.200603148
- 15.
Zborowski M Sun L Moore LR Williams PS Chalmers JJ Continuous cell separation using novel magnetic quadrupole flow sorterJ Magn Magn Mater1999 8224230
Not in our library10.1016/s0304-8853(98)00581-2
- 16.
Purcell EM Life at low Reynolds numberAm J Phys1977 8311
Not in our library10.1119/1.10903
- 17.
Lim JK Eggeman A Lanni F Tilton RD Majetich SA Synthesis and single-particle optical detection of low-polydispersity plasmonic-superparamagnetic nanoparticlesAdv Mater2008 817211726
Not in our library10.1002/adma.200702196
- 18.
Lim JK Lanni C Evarts ER Lanni F Tilton RD Majetich SA Magnetophoresis of nanoparticlesACS Nano2011 8217226
Not in our library10.1021/nn102383s
- 19.
Nel A Xia T Mädler L Li N Toxic potential of materials at the nanolevelScience2006 8622627
Not in our library10.1126/science.1114397
- 20.
Auffan M Rose J Bottero JY Lowry GV Jolivet JP Wiesner MR Towards a definition of inorganic nanoparticles from an environmental, health and safety perspectiveNat Nanotech2009 8634641
Not in our library10.1038/nnano.2009.242
- 21.
Nel A Madler T Velegol D Xia T Hoek E Somasundaran P Klaessig F Castranova V Thompson M Understanding biophysicochemical interactions at the nano-bio interfaceNat Mater2009 8543557
Not in our library10.1038/nmat2442
- 22.
Phenrat T Kim HJ Fagerlund F Illangasekare T Tilton RD Lowry GV Particle size distribution, concentration, and magnetic attraction affect transport of polymer-modified Fe0 nanoparticles in sand columnsEnviron Sci Technol2009 850795085
Not in our library10.1021/es900171v
- 23.
Goon IY Lai LMH Lim M Munroe P Gooding JJ Amal R Fabrication and dispersion of gold-shell-protected magnetite nanoparticles: systematic control using polyethyleneimineChem Mater2009 8673681
Not in our library10.1021/cm8025329
- 24.
Takahashi K Kato H Saito T Matsuyama S Kinugasa S Precise measurement of the size of nanoparticles by dynamic light scattering with uncertainty analysisPart Part Syst Charact2008 83138
Not in our library10.1002/ppsc.200700015
- 25.
Goldburg WI Dynamic light scatteringAm J Phys1999 811521160
Not in our library10.1119/1.19101
- 26.
Chatterjee J Haik Y Chen CJ Size dependent magnetic properties of iron oxide nanoparticlesJ Magn Magn Mater2003 8113118
Not in our library10.1016/s0304-8853(02)01066-1
- 27.
DiPietro RS Johnson HG Bennett SP Nummy TJ Lewis LH Determining magnetic nanoparticle size distributions from thermomagnetic measurementsAppl Phys Lett2010 8222506
Not in our library10.1063/1.3441411
- 28.
Silva LP Lacava ZGM Buske N Morais PC Azevedo RB Atomic force microscopy and transmission electron microscopy of biocompatible magnetic fluids: a comparative analysisJ Nanopart Res2004 8209213
Not in our library
- 29.
Dukhin AS Goetz PJ Acoustic and electroacoustic spectroscopyLangmuir1996 843364344
Not in our library10.1021/la951086q
- 30.
Chantrell RW Wohlfarth EP Rate dependent of the field-cooled magnetisation of a fine particle systemPhys Status Solidi A1985 8619626
Not in our library10.1002/pssa.2210910231
- 31.
El-Hilo M O’Grady K Chantrell RW Susceptibility phenomena in a fine particle system: I. Concentration dependence of peakJ Magn Magn Mater1992 8295306
Not in our library10.1016/0304-8853(92)90272-p
- 32.
Jans H Liu X Austin L Maes G Huo Q Dynamic light scattering as a powerful tool for gold nanoparticle bioconjugation and biomolecular binding studiesAnal Chem2009 894259432
Not in our library10.1021/ac901822w
- 33.
Ando K Chiba A Tanoue H Uniaxial magnetic anisotropy of submicron MnAs ferromagnets in GaAs semiconductorsAppl Phys Lett1998 8387
Not in our library10.1063/1.121843
- 34.
Lacava LM Lacava BM Azevedo RB Lacava ZGM Buske N Tronconi AL Morais PC Nanoparticles sizing: a comparative study using atomic force microscopy, transmission electron microscopy, and ferromagnetic resonanceJ Magn Magn Mater2001 87983
Not in our library10.1016/s0304-8853(00)01231-2
- 35.
Dukhin AS Goetz PJ Fang X Somasundaran P Monitoring nanoparticles in the presence of larger particles in liquids using acoustics and electron microscopyJ Colloid Interface Sci2010 81825
Not in our library10.1016/j.jcis.2009.07.001
- 36.
Van de Hulst HC Light Scattering by Small Particles1981 New York: Dover Publications
Not in our library
- 37.
Hiemenz PC Rajagopalan R Principles of Colloid and Surface Chemistry1997 3New York: Marcel Dekker
Not in our library
- 38.
Berne BJ Pecora R Dynamic Light Scattering: With Applications to Chemistry, Biology and Physics2000 New York: Dover Publications
Not in our library
- 39.
He F Zhao D Manipulating the size and dispersibility of zerovalent iron nanoparticles by use of carboxymethyl cellulose stabilizersEnviron Sci Technol2007 862166221
Not in our library10.1021/es0705543
- 40.
Tiraferri A Chen KL Sethi R Elimelech M Reduced aggregation and sedimentation of zero valent iron nanoparticles in the presence of guar gumJ Colloid Interface Sci2008 87179
Not in our library10.1016/j.jcis.2008.04.064
Cited by papers we hold 3
- Exosome Processing and Characterization Approaches for Research and Technology Development · 2022No notice
- Thermosensitive Micellar Hydrogels as Vehicles to Deliver Drugs With Different Wettability · 2020Not checked
- Anti-biofilm and Antibacterial Activities of Silver Nanoparticles Synthesized by the Reducing Activity of Phytoconstituents Present in the Indian Medicinal Plants · 2020Corrected
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.