No retraction, correction or concern found when last checked on Oct 20, 2026.
Validity of activity monitors in health and chronic disease: a systematic review
Hans Van Remoortel, Santiago Giavedoni, Yogini Raste and 14 more
International Journal of Behavioral Nutrition and Physical Activity · 2012 · Meta-analysis
Abstract
The assessment of physical activity in healthy populations and in those with chronic diseases is challenging. The aim of this systematic review was to identify whether available activity monitors (AM) have been appropriately validated for use in assessing physical activity in these groups. Following a systematic literature search we found 134 papers meeting the inclusion criteria; 40 conducted in a field setting (validation against doubly labelled water), 86 in a laboratory setting (validation against a metabolic cart, metabolic chamber) and 8 in a field and laboratory setting. Correlation coefficients between AM outcomes and energy expenditure (EE) by the criterion method (doubly labelled water and metabolic cart/chamber) and percentage mean differences between EE estimation from the monitor and EE measurement by the criterion method were extracted. Random-effects meta-analyses were performed to pool the results across studies where possible. Types of devices were compared using meta-regression analyses. Most validation studies had been performed in healthy adults (n = 118), with few carried out in patients with chronic diseases (n = 16). For total EE, correlation coefficients were statistically significantly lower in uniaxial compared to multisensor devices. For active EE, correlations were slightly but not significantly lower in uniaxial compared to triaxial and multisensor devices. Uniaxial devices tended to underestimate TEE (−12.07 (95%CI; -18.28 to −5.85) %) compared to triaxial (−6.85 (95%CI; -18.20 to 4.49) %, p = 0.37) and were statistically significantly less accurate than multisensor devices (−3.64 (95%CI; -8.97 to 1.70) %, p<0.001). TEE was underestimated during slow walking speeds in 69% of the lab validation studies compared to 37%, 30% and 37% of the studies during intermediate, fast walking speed and running, respectively. The high level of heterogeneity in the validation studies is only partly explained by the type of activity monitor and the activity monitor outcome. Triaxial and multisensor devices tend to be more valid monitors. Since activity monitors are less accurate at slow walking speeds and information about validated activity monitors in chronic disease populations is lacking, proper validation studies in these populations are needed prior to their inclusion in clinical trials.
Full text
From PubMed Central, U.S. National Library of Medicine, as current on Oct 20, 2026. Paragraph numbers are ours, so an answer can point at one.
Introduction
There is evidence that regular physical activity is associated with a reduced risk of mortality and contributes to the primary and secondary prevention of several chronic diseases. For example, a reduced risk of coronary heart disease, cardiovascular disease, stroke and colon cancer has been reported in more active individuals. In patients with chronic obstructive pulmonary disease (COPD), regular physical activity leads to a lower risk of both COPD related hospital admissions and mortality. Physical activity limitation is a major problem in patients with chronic diseases and needs to be accurately measured if therapies aimed at improving this are to be properly evaluated. A range of devices are available for this purpose but most have been validated in young, healthy subjects and their applicability to older or unwell populations, where movements tend to be slower, is not well established.
Physical activity is defined as any bodily movement, produced by skeletal muscles, requiring energy expenditure. Daily physical activity can be considered as “the totality of voluntary movement produced by skeletal muscles during everyday functioning”. Estimates of daily physical activity can be obtained by different approaches; questionnaires, energy expenditure measurements and activity monitors. Questionnaires rely on the subject’s recollection of activities and allow categorization of patients by physical activity (very active, active, sedentary and inactive), but may lack the precision needed to detect changes in physical activity on a day to day basis.
Daily physical activity can be expressed as an overall measure of active energy expenditure, using indirect calorimetry techniques such as doubly labelled water or metabolic carts. Although doubly labelled water is regarded as a criterion method, this technique does not quantify the duration, frequency and intensity of physical activity performed. Metabolic cart systems which measure expired O2 and CO2 however cannot be used over extended periods of time.
Physical activity can also be monitored directly using physical activity monitors. In general, three classes of activity monitors are being used increasingly in chronic disease populations (e.g. COPD): pedometers, accelerometers and integrated multisensor systems. Pedometers are devices which estimate the number of steps taken through mechanical or digital measurements in only the vertical plane. This is a limited measure of physical activity. Accelerometers detect acceleration in one, two or three directions (uni-, bi- or triaxial accelerometers). These devices allow determination of the quantity and intensity of movements. Integrated multisensor systems combine accelerometry with other sensors that capture body responses to exercise (e.g. heart rate or skin temperature) in an attempt to optimise physical activity assessments.
With the advancement of technology, the number of activity monitors available to measure physical activity is growing. However, despite these advances, it remains a challenge to assess physical activity in slowly moving patients (such as those with COPD, chronic heart failure and diabetes type II). In these patients small changes in physical activity are likely to be important effects of interventions aimed at enhancing physical activity. Therefore, in order for investigators to interpret the effect of interventions on physical activity, activity monitors that have been properly validated in these patient groups are needed.
In order to make evidence based statements on the validity of activity monitors, a systematic review was conducted to identify available activity monitors that have been validated in both healthy adults and chronic disease populations.
Inclusion criteria
Studies meeting the following criteria were included: (1) Population: healthy adults and adults with a diagnosis of chronic disease in whom inactivity is a likely contributor to morbidity or a target for treatment, but whose locomotor function is relatively preserved (COPD, heart failure, diabetes type II, frail elderly, primary pulmonary hypertension, chronic low back pain, fibromyalgia syndrome, obesity). (2) Measurement: any commercially available activity monitor for outdoor activity monitoring from uniaxial to triaxial accelerometers and multisensor devices to tools incorporating spatial information (e.g. GPS) or other information on motion. (3) Study design: studies that evaluated the validity of an activity monitor, i.e. testing an activity monitor against a criterion method, such as indirect calorimetry. Two types of validation studies were included; field validation studies (validation of an activity monitor against doubly labelled water) and laboratory validation studies (validation of an activity monitor using a metabolic cart or metabolic chamber and/or manual step-counting or video observation). (4) Clinical trials using activity monitoring as an outcome and which might contain a reference to a validation paper were included for hand-searching. (5) A search window between 1st of January 2000 until 1st of March 2012 was selected in order to capture sensors in contemporary use. This approach still allowed for the identification of older validation studies (published before 2000) of devices in current use in clinical trials. Main exclusion criteria were 1) studies in children (subjects younger than 18 years), 2) studies in subjects with abnormal biomechanical movement patterns (e.g. cerebral palsy, lower limb amputation), 3) studies only investigating the number of steps using pedometers because of the inaccuracy in measurement of total energy expenditure and lack of ability to measure physical activity patterns.
No language restrictions were used; any non-English studies retrieved through the literature search were translated to determine their appropriateness for inclusion.
Search strategy and systematic review
Eligible studies were identified by searching the following databases: MEDLINE, EMBASE and CINAHL. A librarian was consulted prior to initiating the search in order to identify appropriate search terms to describe the population (from healthy adults to patients with chronic disease), physical activity and activity monitoring. A combination of MeSH terms (MEDLINE), Emtree terms (Embase) and Cinahl headings (Cinahl) with free text words (all databases) were used (see Additional file 1 for detailed information). Refworks (www.refworks.com) was used to store and share all papers and to collect all the information of title and abstract screening, full text assessment and the hand-searching process. Each review team consisted of 3 reviewers who independently screened the titles and abstracts of the retrieved articles. Each abstract was labelled as ‘A) excluded papers‘, ‘B) order for full text assessment‘or ‘C) hand-search for references only’, i.e. clinical trials which may have a reference to an older validation study. After independently reviewing the articles for inclusion, the reviewers compared their labels to ensure consensus. Once agreement had been reached, a full text copy of each article that met the inclusion criteria was obtained (Label B). Thereafter, the same review teams looked at the full texts of the potential validation papers in detail and decided in consensus, whether the articles were indeed suitable validation papers for data extraction. Subsequently, hand-searching of the clinical trials using an activity monitor outcome which might contain a reference to a validation paper (Label C), was performed by three independent reviewers. After independently reviewing these full texts, validation papers were identified which met the inclusion criteria for full text assessment. Again, the reviewers compared their decisions to ensure consensus. Data of all included validation papers were extracted into predefined prepared Excel tables.
Data extraction
For the field studies, correlation coefficients between total and active energy expenditure from activity monitor (TEEAM and AEEAM respectively) and total and active energy expenditure measured with doubly labelled water (TEEDLW and AEEDLW respectively) were extracted. The percentage mean differences (ΔTEE and ΔAEE) with 95% confidence intervals were obtained from the reports to assess agreement between energy expenditure estimates from the activity monitor (TEEAM and AEEAM) versus energy expenditure measures from doubly labelled water (TEEDLW and AEEDLW). For the laboratory studies, correlation coefficients between activity monitor outcome and EE measured by metabolic cart/chamber were extracted. A sub-analysis included to compare correlation coefficients derived from walking based protocols to correlation coefficients derived from protocols based on activities of daily living. Agreement between energy expenditure outcomes from the activity monitor versus criterion method (indirect calorimetry) were extracted by the mean difference at different treadmill walking speeds; slow walk (<3.2 km/hr or 1 mph), intermediate speed walk (3.2-6.4 km/hr or 2–4 mph), fast walk (6.5-8.05 km/hr or 4–5 mph) and running (8.06-11 km/hr or 5–7 mph). Accuracy of steps measured by activity monitoring was expressed as the percentage mean difference between steps measured by an activity monitor versus actual steps measured by the criterion method (video observation and/or manual step counting).
Statistical analysis
Descriptive statistics were used to report information about type of activity monitor, activity monitor outcomes and studied population. Papers were separated by type of validation, ‘field validation papers’ (validation of an activity monitor against indirect calorimetry, using the doubly labelled water technique) and ‘lab validation papers’ (validation of an activity monitor against indirect calorimetry, using a metabolic cart, metabolic chamber or direct observation).
We also analysed the results separately per type of device (uni-, bi-, triaxial and multisensor devices). We performed (DerSimonian and Laird) random-effects meta-analyses to pool the correlation coefficients and mean differences across studies and expressed heterogeneity by the I2 statistic, which estimates the percentage of total variation between studies that is due to heterogeneity rather than chance. I2 is calculated from basic results obtained from a typical meta-analysis as I2 = 100% x (Q-df)/Q, where Q is Cochran’s heterogeneity statistic and df the degrees of freedom. Negative values of I2 are put equal to zero so that I2 lies between 0% en 100% with larger values showing larger heterogeneity. We used the Fisher r to z-transformation in order to pool normally distributed data (z scores) rather than the skewed distribution of Pearson correlation coefficients. We back transformed the pooled z-scores to correlation coefficients for easier interpretation.
We used random-effects linear regression models (meta-regression analyses) with the studies’ results as the dependent variable (and considering each studies’ standard error) to compare the type of devices (covariate) and to assess the type of population (covariate) as a potential explanation for heterogeneity. For those few studies where no measures of variability were reported we imputed the median standard deviations of those studies where the standard deviation was available. We did not perform meta-analyses for the laboratory studies where none of the studies provided standard deviations for ΔTEE and ΔAEE but presented the point estimates as graphs. Coefficient of variation for TEEDLW and AEEDLW was calculated per study population to investigate whether the degree of variation in TEEDLW and AEEDLW affected the correlation coefficients and/or mean differences, (i.e. higher correlations/mean differences in populations with larger variation in TEE and/or AEE).
Results
The systematic literature search resulted in a total of 2875 abstracts which were scrutinised by four review teams across Europe. Figure 1 represents the different processes used in the systematic review.
Forty monitors were tested in validation studies; 12 uniaxial, 3 biaxial, 16 triaxial accelerometers and 9 multisensor devices. Fifty-five percent of activity monitors (22/40) were used only in lab validation studies, 10% (4/40) only in field validation studies and 35% (14/40) in both a lab as well as a field validation study. An overview of the different activity monitors is shown in Tables 1, 2, 3 and 4.
The most frequently available outcomes present in validated activity monitors are (total and/or active) energy expenditure (70%, 28/40), steps (38%, 15/40) and different levels of physical activity intensity (38%, 15/40). The majority of the validation studies (118/134, 88%) were performed in healthy adults. Few studies (16/134, 12%) were performed in chronic disease populations; obesity (n = 4), chronic obstructive pulmonary disease (n = 5), chronic heart failure (n = 1), chronic organ failure (n = 1), chronic low back pain (n = 1), fibromyalgia syndrome (n = 1), peripheral arterial disease (n = 1), diabetes mellitus type II (n = 1) and a general chronic disease population (cardiac, obese or knee arthritis, n = 1).
Field validation studies
Individual correlation coefficients, with converted Fisher z-scores, for total energy expenditure (TEE) between TEEAM and TEEDLW are presented in Figure 2.
Variability of study populations’ TEEDLW was relatively small; coefficient of variation (CV) ranged from 0.11 to 0.29. Pooled r in uniaxial devices (r = 0.52 (95%CI, 0.29 to 0.70)) was significantly lower compared to multisensor devices (r = 0.84 (95%CI, 0.78 to 0.88), p<0.001) but not to triaxial devices (r = 0.61 (95%CI, 0.45 to 0.73, p = 0.37)). Because of the relatively large difference in accuracy between the uniaxial, the triaxial and multisensor devices 53% of the between–study heterogeneity was accounted for by type of device in meta-regression analyses.
ΔTEE (TEEAM – TEEDLW) was less accurate in uniaxial compared to triaxial accelerometers and multisensor devices (−12.07 (95%CI, -18.28 to −5.85) % in uniaxial versus −6.85 (95%CI, -18.20 to 4.49) % in triaxial (p = 0.39 for comparison against uniaxial devices) and −3.64 (95%CI, -8.97 to 1.70) % in multisensor devices, p = 0.03 for comparison against uniaxial devices, Figure 3). ΔTEE were smaller in studies with chronic disease populations than in studies with healthy populations (−9% (95%CI −19 to 1)) but the difference did not reach statistical significance (p = 0.09).
Correlations for active energy expenditure (AEE) between AEEAM and AEEDLW were higher in triaxial (0.59 (95%CI, 0.45 to 0.70)) and multisensor devices (0.54 (95%CI, 0.39 to 0.65)) compared to uniaxial (0.39 (95%CI, 0.16 to 0.58)) devices, p = 0.12 for triaxial and p = 0.32 for multisensor against uniaxial devices) (Figure 4) Types of devices accounted for only 12% of the between-study heterogeneity in the meta-regression analysis. All monitors underestimated AEE (ΔAEE (AEEAM – AEEDLW) -24.22 (95%CI, -62.05 to −13.61) % in uniaxial, -21.01 (95%CI, -41.92 to −0.11) % in triaxial and −24.35 (95%CI, -45.28 to −3.42) % in multisensor devices. No significant differences were found between devices (Figure 5). But ΔAEE were statistically significantly smaller in studies with chronic disease populations than in studies with healthy populations (−44%, 95%CI −73 to −13, p = 0.006).
Laboratory validation studies
For correlation analysis, TEE and AEE, as determined from indirect calorimetry, were used as criterion outcomes (in 89% and 11% of the studies, respectively) against different outcomes of the activity monitor (activity counts (37%), vector magnitude units (7%), total energy expenditure (48%), active energy expenditure (2%) or monitor-specific activity scores (6%).
Pooled correlation coefficients between indirect calorimetry (TEEIC) and activity monitor outcome were lower in uniaxial (0.80 (95%CI, 0.75 to 0.84)) compared to multisensor devices (0.85 (95%CI, 0.72 to 0.92)) but the difference did not reach statistical significance (p = 0.43) No differences were found with biaxial (0.73 (95%CI, 0.33 to 0.91), p = 0.50) and triaxial (0.84 (95%CI, 0.78 to 0.89), p = 0.28) devices, either (Figure 6).
Correlation coefficients between TEEIC and activity monitor outcome were higher when tested using laboratory protocols based on walking activities (overall pooled r = 0.84 (95%CI, 0.79 to 0.87), no significant differences between types of devices, Figure 7) compared to protocols using activities of daily living involving the upper and lower limbs (overall pooled r = 0.75 (95%CI, 0.68 to 0.81, no significant differences between types of devices), Figure 8).
There was evidence of heterogeneity of results across all analyses (overall I2 ranged from 84.6% (Figure 7) to 85.9% (Figure 8)). Again, the results did not differ for chronic disease and healthy populations in any of the analyses on laboratory validation studies.
Mean differences between TEEAM and TEEIC at different treadmill walking speeds are presented in Figures 9, 10, 11 and 12. TEE was underestimated during slow walking speed in 69% of studies (n = 16/23), whereas in only 37% of studies (n = 15/40) during intermediate walking speed, 30% of studies (n = 10/33) during fast walking speed and 37% of studies (n = 7/19) during running reported underestimation of TEE. Underestimations in the slow walking group were relatively larger.
All accelerometers underestimate steps during slow walking; from 0.94 to 60% underestimation. One uniaxial device (activPAL), mounted on the thigh, showed a high accuracy in measuring steps during slow walking with only 0.94% overestimation. More accurate estimates of steps were reported at higher speeds; from 13% under to 2% overestimation during intermediate walking speed (except one study with 35% underestimation using SenseWear Armband), and from 0.18 to 4.3% overestimation during fast walking (Figure 13).
Discussion
This systematic review of the literature identified forty activity monitors (12 uniaxial, 3 biaxial, 16 triaxial and 9 multisensor devices) that had been validated against indirect calorimetry (doubly labelled water, metabolic cart and/or metabolic chamber) in healthy adults (88% of studies) or adults with chronic disease (12% of studies).
Field and laboratory validation studies had highly heterogeneous results which could partly be explained by the type of activity monitor and the activity monitor outcome. These factors need consideration when a validation study is evaluated.
First, selecting the type of activity monitor is important. Pedometers are limited in their ability to detect certain physical activity patterns which might occur in chronic disease populations (for example, an unstable gait profile or lack of intensity of physical activity). Accelerometers can overcome this. Multi-axial accelerometers have the ability to measure accelerations in different orientations, which provides information about the total amount, intensity and duration of daily physical activity. Some multisensor devices, which combine physiological parameters with accelerometry, are available to assess both body posture and body movement. An additional promising class of monitors integrate positioning systems (Global Positioning System (GPS) and Bluetooth® systems for outdoor and indoor activities respectively) with accelerometry and other sensors. However, to date, these have been used infrequently in patients with chronic disease. Based on this systematic review, heterogeneity among studies was significantly explained by the types of devices, although no statistical significance was reached between different types of devices.
A second factor to take into consideration is the activity monitor outcome. When measuring TEE in field validation studies (doubly labelled water), high correlations with the TEE estimate of the activity monitor were found in most activity monitors. These correlations are, however, to a large extent driven by patient characteristics (i.e. body weight, age, height) which is an important predictor of TEE. Consequently, the comparison of TEE estimated from activity monitors, with TEE measured with indirect calorimetry or doubly labelled water is not necessarily a proof of validation. In a field setting it has been reported that only 19% of the TEE is accounted for by physical activity in both healthy subjects and in patients with coronary heart disease.
Another factor that needs to be considered is the study population. Most of the study populations (88%) were healthy adults (from young healthy adults to healthy elderly). Only 12% of validation studies were performed in patients with chronic diseases (COPD, chronic heart failure, chronic organ failure, diabetes mellitus type II, obesity, peripheral arterial disease chronic low back pain and fibromyalgia syndrome). These patients walk more slowly than healthy subjects, which is reflected, for example, by a reduced six minute walking distance. This review, as well as original research, suggests that most monitors are less accurate at lower walking speeds. These findings are consistent with a systematic review of pedometers which found evidence of reduced accuracy during slow walking. Hence, there is a need to perform validation studies specifically in chronic disease populations.
When measuring TEE in lab validation studies by assessment of oxygen consumption, higher correlations were reported for walking activities compared to other daily life activities which implies that the walking component of physical activity is better detected than other activities of daily living.
Most activity monitors use prediction equations to calculate energy expenditure from the activity signals. This is helpful to validate monitors against indirect calorimetry, but, given the inherent inaccuracy of these estimates and fundamental differences between the different prediction equations (some of which are proprietary to particular device manufacturers), perhaps greater weight should be given to direct monitor outputs (steps, activity counts, VMU, etc.) and their relation to activity energy expenditure (AEE), rather than the ability of a monitor to estimate energy expenditure precisely. It is very unlikely that an activity monitor will be able to capture accurately all the factors affecting energy expenditure (i.e. movement efficiency, resting metabolism, distribution of fat-free mass and fat mass). In patients with COPD, for example, Baarends et al. showed that non-resting energy expenditure (TEE-REE) was elevated in COPD compared to healthy controls. Since it is generally accepted now that these patients are less active than healthy controls, it is clear that patients expend more energy than controls to achieve the same movements. It would be unrealistic to expect an activity monitor to pick this up. Hence, the lack of accuracy against energy expenditure does not render activity monitors invalid tools to assess physical activity in patients over time (for which precision is more important) or to capture the physical activity level of a patient (for which validity, represented by the correlation with true energy expenditure is more important than absolute accuracy). The acceptable correlations between VO2 and activity monitor outputs in triaxial and multisensor devices are therefore encouraging for the use of monitors to assess physical activity in an adult population. With specific validation studies, these findings can possibly be extrapolated to elderly and patients with chronic diseases.
The current systematic review may also help researchers to decide on appropriate activity monitor outcomes. Combination of the three most frequently available outcomes (TEE/AEE, steps and different levels of physical activity intensity), which is likely to provide a comprehensive insight in overall physical activity of a patient, is available in 3 uniaxial (Actigraph 7164/GT1M, Kenz Lifecorder EX and Polar Activity Watch 200), 1 biaxial (Biotrainer Pro), 3 triaxial (Dynaport Minimod, Actical and Actigraph GT3X) and 2 multisensor activity monitors (SenseWear Armband and multisensor board). Some general considerations can also be taken into account when selecting an activity monitor in clinical trials such as the type of monitoring (e.g. daily physical activity), size and scope of the study, usability of the monitor and cost.
Methodological issues
A point of difficulty in collecting, analysing and interpreting the data was the wide range of statistical approaches used in the original papers. Indeed, we had to compute the standard deviation of the mean difference (between EEAM and EEIC) because some field validation studies didn’t report this.
Correlation analysis but also Bland and Altman analysis were the two main statistical approaches used in validation studies and were used for data extraction. A systematic review of the statistical methods used to validate physical activity questionnaires revealed similar findings, with the majority of the studies using correlation analysis compared to Bland and Altman analysis. Correlation analyses are a common evaluation approach and allow statements on validity, whereas agreement between activity monitor and criterion method (indirect calorimetry) with Bland and Altman plots are preferred when the aim is to identify systematic bias in measures. Since not all activity monitors have the possibility to estimate total and/or active energy expenditure, this type of analysis is not uniformly applicable. Multiple regression analysis with TEE/AEE as the dependent variable is a correct technique to tackle this. Consistent statistical guidelines for reporting the validity of an activity monitor would be helpful.
Conclusion
Validation studies of activity monitors are highly heterogeneous, and this is partly explained by the type of activity monitor and the activity monitor outcome. Since activity monitors are less accurate at slow walking speeds and information about validated activity monitors in chronic disease populations is lacking, proper validation studies in these populations are needed prior to their inclusion in clinical trials.
Abbreviations
AEE, Active energy expenditure; AEEAM, Estimation of active energy expenditure by an activity monitor; AEEDLW, Measurement of active energy expenditure by doubly labelled water; AEEIC, Measurement of active energy expenditure by indirect calorimetry; AM, Activity monitor; COPD, Chronic obstructive pulmonary disease; DLW, Doubly labelled water; GPS, Global positioning system; IC, Indirect calorimetry; TEE, Total energy expenditure; TEEAM, Estimation of total energy expenditure by an activity monitor; TEEDLW, Measurement of total energy expenditure by doubly labelled water; TEEIC, Measurement of total energy expenditure by indirect calorimetry.
Competing interests
The authors declare not having financial or non-financial competing interests.
Authors’ contributions
All authors contributed to the different processes in the systematic review (title and abstract screening, full text assessment and data extraction). HVR worked out an appropriate search term strategy in different databases. HVR and SG have been involved in analysing the data and drafting the manuscript which was revised by all the other authors. MP performed the statistical analysis. All authors read and approved the final manuscript.
Authors’ information
This systematic review is part of the PROactive project (www.proactivecopd.com) which aims to develop a Patient Reported Outcome (PRO) tool capturing physical activity in COPD in close harmony with a valid activity monitor.
References 154
- 1.
Warburton DE Nicol CW Bredin SS Health benefits of physical activity: the evidenceCMAJ2006 174801809
Not in our library10.1503/cmaj.051351
- 2.
Blair SN Cheng Y Holder JS Is physical activity or physical fitness more important in defining health benefits?Med Sci Sports Exerc2001 33S379S399discussion S419-320
Not in our library10.1097/00005768-200106001-00007
- 3.
Garcia-Aymerich J Lange P Benet M Schnohr P Anto JM Regular physical activity reduces hospital admission and mortality in chronic obstructive pulmonary disease: a population based cohort studyThorax2006 61772778
Not in our library10.1136/thx.2006.060145
- 4.
Caspersen CJ Powell KE Christenson GM Physical activity, exercise, and physical fitness: definitions and distinctions for health-related researchPublic Health Rep1985 100126131
Not in our library
- 5.
Steele BG Belza B Cain K Warms C Coppersmith J Howard J Bodies in motion: monitoring daily activity and exercise with motion sensors in people with chronic pulmonary diseaseJ Rehabil Res Dev2003 404558
Not in our library10.1682/jrrd.2003.10.0045
- 6.
Garcia-Aymerich J Serra I Gomez FP Farrero E Balcells E Rodriguez DA de Batlle J Gimeno E Donaire-Gonzalez D Orozco-Levi M Physical activity and clinical and functional status in COPDChest2009 1366270
Not in our library10.1378/chest.08-2532
- 7.
Bassett DR Ainsworth BE Swartz AM Strath SJ O’Brien WL King GA Validity of four motion sensors in measuring moderate intensity physical activityMed Sci Sports Exerc2000 32S471S480
Not in our library10.1097/00005768-200009001-00006
- 8.
Le Masurier GC Tudor-Locke C Comparison of pedometer and accelerometer accuracy under controlled conditionsMed Sci Sports Exerc2003 35867871
Not in our library10.1249/01.mss.0000064996.63632.10
- 9.
Plasqui G Westerterp KR Physical activity assessment with accelerometers: an evaluation against doubly labeled waterObesity (Silver Spring)2007 1523712379
Not in our library10.1038/oby.2007.281
- 10.
Cyarto EV Myers AM Tudor-Locke C Pedometer accuracy in nursing home and community-dwelling older adultsMed Sci Sports Exerc2004 36205209
Not in our library10.1249/01.mss.0000113476.62469.98
- 11.
Giacomozzi C Caselli A Macellari V Giurato L Lardieri L Uccioli L Walking strategy in diabetic patients with peripheral neuropathyDiabetes Care2002 2514511457
Not in our library10.2337/diacare.25.8.1451
- 12.
Guazzi M Dickstein K Vicenzi M Arena R Six-minute walk test and cardiopulmonary exercise testing in patients with chronic heart failure: a comparative analysis on clinical and prognostic insightsCirc Heart Fail2009 2549555
Not in our library10.1161/circheartfailure.109.881326
- 13.
Corey DM Dunlap WP Averaging Correlations: expected Values and Bias in Combined Pearson rs and Fisher’s z TransformationsJ Gen Psychol1998 125245261
Not in our library10.1080/00221309809595548
- 14.
Leenders NY Sherman WM Nagaraja HN Energy expenditure estimated by accelerometry and doubly labeled water: do they agree?Med Sci Sports Exerc2006 3821652172
Not in our library10.1249/01.mss.0000235883.94357.95
- 15.
Adams SA Matthews CE Ebbeling CB Moore CG Cunningham JE Fulton J Hebert JR The effect of social desirability and social approval on self-reports of physical activityAm J Epidemiol2005 161389398
Not in our library10.1093/aje/kwi054
- 16.
Leenders NY Sherman WM Nagaraja HN Kien CL Evaluation of methods to assess physical activity in free-living conditionsMed Sci Sports Exerc2001 3312331240
Not in our library10.1097/00005768-200107000-00024
- 17.
Dalton AF Ni Scanaill C Carew S Lyons D OL G A clinical evaluation of a remote mobility monitoring system based on SMS messagingConf Proc IEEE Eng Med Biol Soc2007 200723272330
Not in our library10.1109/iembs.2007.4352792
- 18.
Kochersberger G McConnell E Kuchibhatla MN Pieper C The reliability, validity, and stability of a measure of physical activity in the elderlyArch Phys Med Rehabil1996 77793795
Not in our library10.1016/s0003-9993(96)90258-0
- 19.
Masse LC Fulton JE Watson KL Heesch KC Kohl HW Blair SN Tortolero SR Detecting bouts of physical activity in a field settingRes Q Exerc Sport1999 70212219
Not in our library10.1080/02701367.1999.10608041
- 20.
McClain JJ Craig CL Sisson SB Tudor-Locke C Comparison of Lifecorder EX and ActiGraph accelerometers under free-living conditionsAppl Physiol Nutr Metab2007 32753761
Not in our library10.1139/h07-060
- 21.
- 22.
Lof M Hannestad U Forsum E Comparison of commonly used procedures, including the doubly-labelled water technique, in the estimation of total energy expenditure of women with special reference to the significance of body fatnessBr J Nutr2003 90961968
Not in our library10.1079/bjn2003975
- 23.
Masse LC Fulton JE Watson KL Mahar MT Meyers MC Wong WW Influence of body composition on physical activity validation studies using doubly labeled waterJ Appl Physiol2004 9613571364
Not in our library10.1152/japplphysiol.00901.2003
- 24.
Patrik Johansson H Rossander-Hulthen L Slinde F Ekblom B Accelerometry combined with heart rate telemetry in the assessment of total energy expenditureBr J Nutr2006 95631639
Not in our library10.1079/bjn20051527
- 25.
Brooks AG Gunn SM Withers RT Gore CJ Plummer JL Predicting walking METs and energy expenditure from speed or accelerometryMed Sci Sports Exerc2005 3712161223
Not in our library10.1249/01.mss.0000170074.19649.0e
- 26.
Crouter SE Churilla JR Bassett DR Estimating energy expenditure using accelerometersEur J Appl Physiol2006 98601612
Not in our library10.1007/s00421-006-0307-5
- 27.
Focht BC Initial Validation of the CSA Activity Monitor During Rehabilitative Exercise Among Older Adults With Chronic DiseaseJournal of Aging and Physical Activity2003 11293304
Not in our library
- 28.
Freedson PS Melanson E Sirard J Calibration of the Computer Science and Applications, Inc. accelerometerMed Sci Sports Exerc1998 30777781
Not in our library10.1097/00005768-199805000-00021
- 29.
Fudge BW Wilson J Easton C Irwin L Clark J Haddow O Kayser B Pitsiladis YP Estimation of oxygen uptake during fast running using accelerometry and heart rateMed Sci Sports Exerc2007 39192198
Not in our library10.1249/01.mss.0000235884.71487.21
- 30.
Heil DP Body size as a determinant of activity moniyor output during overground walkingJEP online2003 6111
Not in our library
- 31.
Hendelman D Miller K Baggett C Debold E Freedson P Validity of accelerometry for the assessment of moderate intensity physical activity in the fieldMed Sci Sports Exerc2000 32S442S449
Not in our library10.1097/00005768-200009001-00002
- 32.
King GA Torres N Potter C Brooks TJ Coleman KJ Comparison of activity monitors to estimate energy cost of treadmill exerciseMed Sci Sports Exerc2004 3612441251
Not in our library10.1249/01.mss.0000132379.09364.f8
- 33.
Leenders NY Nelson TE Sherman WM Ability of different physical activity monitors to detect movement during treadmill walkingInt J Sports Med2003 244350
Not in our library10.1055/s-2003-37196
- 34.
Melanson EL Freedson PS Validity of the Computer Science and Applications, Inc. (CSA) activity monitorMed Sci Sports Exerc1995 27934940
Not in our library
- 35.
Nichols JF Patterson P Early T A validation of a physical activity monitor for young and older adultsCan J Sport Sci1992 17299303
Not in our library
- 36.
Patterson SM Krantz DS Montgomery LC Deuster PA Hedges SM Nebel LE Automated physical activity monitoring: validation and comparison with physiological and self-report measuresPsychophysiology1993 30296305
Not in our library10.1111/j.1469-8986.1993.tb03356.x
- 37.
Slootmaker SM Chin APMJ Schuit AJ van Mechelen W Koppes LL Concurrent validity of the PAM accelerometer relative to the MTI Actigraph using oxygen consumption as a referenceScand J Med Sci Sports2009 193643
Not in our library10.1111/j.1600-0838.2007.00740.x
- 38.
Stone MR Esliger DW Tremblay MS Comparative validity assessment of five activity monitors: does being a child matter?Pediatr Exerc Sci2007 19291309
Not in our library10.1123/pes.19.3.291
- 39.
Strath SJ Bassett DR Thompson DL Swartz AM Validity of the simultaneous heart rate-motion sensor technique for measuring energy expenditureMed Sci Sports Exerc2002 34888894
Not in our library10.1097/00005768-200205000-00025
- 40.
Swartz AM Strath SJ Bassett DR O’Brien WL King GA Ainsworth BE Estimation of energy expenditure using CSA accelerometers at hip and wrist sitesMed Sci Sports Exerc2000 32S450S456
Not in our library10.1097/00005768-200009001-00003
- 41.
Welk GJ Blair SN Wood K Jones S Thompson RW A comparative evaluation of three accelerometry-based physical activity monitorsMed Sci Sports Exerc2000 32S489S497
Not in our library10.1097/00005768-200009001-00008
- 42.
Crouter SE Clowers KG Bassett DR A novel method for using accelerometer data to predict energy expenditureJ Appl Physiol2006 10013241331
Not in our library10.1152/japplphysiol.00818.2005
- 43.
Nichols JF Morgan CG Chabot LE Sallis JF Calfas KJ Assessment of physical activity with the Computer Science and Applications, Inc., accelerometer: laboratory versus field validationRes Q Exerc Sport2000 713643
Not in our library10.1080/02701367.2000.10608878
- 44.
Esliger DW Probert A Gorber SC Bryan S Laviolette M Tremblay MS Validity of the Actical accelerometer step-count functionMed Sci Sports Exerc2007 3912001204
Not in our library10.1249/mss.0b013e3804ec4e9
- 45.
Storti KL Pettee KK Brach JS Talkowski JB Richardson CR Kriska AM Gait speed and step-count monitor accuracy in community-dwelling older adultsMed Sci Sports Exerc2008 405964
Not in our library10.1249/mss.0b013e318158b504
- 46.
Berntsen S Hageberg R Aandstad A Mowinckel P Anderssen SA Carlsen KH Andersen LB Validity of physical activity monitors in adults participating in free-living activitiesBr J Sports Med2010 44657664
Not in our library10.1136/bjsm.2008.048868
- 47.
Miller NE Strath SJ Swartz AM Cashin SE Estimating absolute and relative physical activity intensity across age via accelerometry in adultsJ Aging Phys Act2010 18158170
Not in our library10.1123/japa.18.2.158
- 48.
Assah FK Ekelund U Brage S Corder K Wright A Mbanya JC Wareham NJ Predicting physical activity energy expenditure using accelerometry in adults from sub-Sahara AfricaObesity (Silver Spring)2009 1715881595
Not in our library10.1038/oby.2009.39
- 49.
Rothney MP Brychta RJ Meade NN Chen KY Buchowski MS Validation of the ActiGraph two-regression model for predicting energy expenditureMed Sci Sports Exerc2010 4217851792
Not in our library10.1249/mss.0b013e3181d5a984
- 50.
Colbert LH Matthews CE Havighurst TC Kim K Schoeller DA Comparative validity of physical activity measures in older adultsMed Sci Sports Exerc2011 43867876
Not in our library10.1249/mss.0b013e3181fc7162
- 51.
Abel MG Hannon JC Sell K Lillie T Conlin G Anderson D Validation of the Kenz Lifecorder EX and ActiGraph GT1M accelerometers for walking and running in adultsAppl Physiol Nutr Metab2008 3311551164
Not in our library10.1139/h08-103
- 52.
Edwards AG Hill JO Byrnes WC Browning RC Accuracy of optimized branched algorithms to assess activity-specific physical activity energy expenditureMed Sci Sports Exerc2010 42672682
Not in our library10.1249/mss.0b013e3181bd196d
- 53.
Kozey SL Lyden K Howe CA Staudenmayer JW Freedson PS Accelerometer output and MET values of common physical activitiesMed Sci Sports Exerc2010 4217761784
Not in our library10.1249/mss.0b013e3181d479f2
- 54.
Lyden K Kozey SL Staudenmeyer JW Freedson PS A comprehensive evaluation of commonly used accelerometer energy expenditure and MET prediction equationsEur J Appl Physiol2011 111187201
Not in our library10.1007/s00421-010-1639-8
- 55.
Esliger DW Rowlands AV Hurst TL Catt M Murray P Eston RG Validation of the GENEA AccelerometerMed Sci Sports Exerc2010 4310851093
Not in our library10.1249/mss.0b013e31820513be
- 56.
Kuffel EE Crouter SE Haas JD Frongillo EA Bassett DR Validity of estimating minute-by-minute energy expenditure of continuous walking bouts by accelerometryInt J Behav Nutr Phys Act2011 892
Not in our library10.1186/1479-5868-8-92
- 57.
Harrington DM Welk GJ Donnelly AE Validation of MET estimates and step measurement using the ActivPAL physical activity loggerJ Sports Sci2011 29627633
Not in our library10.1080/02640414.2010.549499
- 58.
Fogelholm M Hiilloskorpi H Laukkanen R Oja P Van Marken Lichtenbelt W Westerterp K Assessment of energy expenditure in overweight womenMed Sci Sports Exerc1998 3011911197
Not in our library10.1097/00005768-199808000-00002
- 59.
Fuller Z Horgan G O’Reilly LM Ritz P Milne E Stubbs RJ Comparing different measures of energy expenditure in human subjects resident in a metabolic facilityEur J Clin Nutr2008 62560569
Not in our library10.1038/sj.ejcn.1602739
- 60.
Gardner AW Womack CJ Sieminski DJ Montgomery PS Killewich LA Fonong T Relationship between free-living daily physical activity and ambulatory measures in older claudicantsAngiology1998 49327337
Not in our library10.1177/000331979804900501
- 61.
Starling RD Matthews DE Ades PA Poehlman ET Assessment of physical activity in older individuals: a doubly labeled water studyJ Appl Physiol1999 8620902096
Not in our library10.1152/jappl.1999.86.6.2090
- 62.
Choquette S Chuin A Lalancette DA Brochu M Dionne IJ Predicting energy expenditure in elders with the metabolic cost of activitiesMed Sci Sports Exerc2009 4119151920
Not in our library10.1249/mss.0b013e3181a6164a
- 63.
Fehling PC Smith DL Warner SE Dalsky GP Comparison of accelerometers with oxygen consumption in older adults during exerciseMed Sci Sports Exerc1999 31171175
Not in our library10.1097/00005768-199901000-00026
- 64.
Hageman PA Comparison of two Physical Activity Monitors During a 1-Mile Walking field testJEP online2004 7102110
Not in our library
- 65.
Pambianco G Wing RR Robertson R Accuracy and reliability of the Caltrac accelerometer for estimating energy expenditureMed Sci Sports Exerc1990 22858862
Not in our library10.1249/00005768-199012000-00020
- 66.
Rafamantanantsoa HH Ebine N Yoshioka M Higuchi H Yoshitake Y Tanaka H Saitoh S Jones PJ Validation of three alternative methods to measure total energy expenditure against the doubly labeled water method for older Japanese menJ Nutr Sci Vitaminol (Tokyo)2002 48517523
Not in our library10.3177/jnsv.48.517
- 67.
Yamada Y Yokoyama K Noriyasu R Osaki T Adachi T Itoi A Naito Y Morimoto T Kimura M Oda S Light-intensity activities are important for estimating physical activity energy expenditure using uniaxial and triaxial accelerometersEur J Appl Physiol2009 105141152
Not in our library10.1007/s00421-008-0883-7
- 68.
Kumahara H Schutz Y Ayabe M Yoshioka M Yoshitake Y Shindo M Ishii K Tanaka H The use of uniaxial accelerometry for the assessment of physical-activity-related energy expenditure: a validation study against whole-body indirect calorimetryBr J Nutr2004 91235243
Not in our library10.1079/bjn20031033
- 69.
Schneider PL Crouter SE Lukajic O Bassett DR Accuracy and reliability of 10 pedometers for measuring steps over a 400-m walkMed Sci Sports Exerc2003 3517791784
Not in our library10.1249/01.mss.0000089342.96098.c4
- 70.
Ohkawara K Ishikawa-Takata K Park JH Tabata I Tanaka S How much locomotive activity is needed for an active physical activity level: analysis of total step countsBMC Res Notes2011 4512
Not in our library10.1186/1756-0500-4-512
- 71.
Yokoyama Y Kawamura T Tamakoshi A Noda A Hirai M Saito H Ohno Y Comparison of accelerometry and oxymetry for measuring daily physical activityCirc J2002 66751754
Not in our library10.1253/circj.66.751
- 72.
Grant PM Dall PM Mitchell SL Granat MH Activity-monitor accuracy in measuring step number and cadence in community-dwelling older adultsJ Aging Phys Act2008 16201214
Not in our library10.1123/japa.16.2.201
- 73.
Ryan CG Grant PM Tigbe WW Granat MH The validity and reliability of a novel activity monitor as a measure of walkingBr J Sports Med2006 40779784
Not in our library10.1136/bjsm.2006.027276
- 74.
Maddocks M Petrou A Skipper L Wilcock A Validity of three accelerometers during treadmill walking and motor vehicle travelBr J Sports Med2010 44606608
Not in our library10.1136/bjsm.2008.051128
- 75.
Chen KY Acra SA Majchrzak K Donahue CL Baker L Clemens L Sun M Buchowski MS Predicting energy expenditure of physical activity using hip- and wrist-worn accelerometersDiabetes Technol Ther2003 510231033
Not in our library10.1089/152091503322641088
- 76.
Welk GJ Almeida J Morss G Laboratory calibration and validation of the Biotrainer and Actitrac activity monitorsMed Sci Sports Exerc2003 3510571064
Not in our library10.1249/01.mss.0000069525.56078.22
- 77.
Kane NA Simmons MC John D Thompson DL Bassett DR Validity of the Nike+ device during walking and runningInt J Sports Med2010 31101105
Not in our library10.1055/s-0029-1242810
- 78.
Brugniaux JV Niva A Pulkkinen I Laukkanen RM Richalet JP Pichon AP Polar Activity Watch 200: a new device to accurately assess energy expenditureBr J Sports Med2010 44245249
Not in our library10.1136/bjsm.2007.045575
- 79.
Connolly CP Coe DP Kendrick JM Bassett DR Thompson DL Accuracy of physical activity monitors in pregnant womenMed Sci Sports Exerc2010 4311001105
Not in our library10.1249/mss.0b013e3182058883
- 80.
Jacobi D Perrin AE Grosman N Dore MF Normand S Oppert JM Simon C Physical activity-related energy expenditure with the RT3 and TriTrac accelerometers in overweight adultsObesity (Silver Spring)2007 15950956
Not in our library10.1038/oby.2007.605
- 81.
Maddison R Jiang Y Hoorn SV Mhurchu CN Lawes CM Rodgers A Rush E Estimating energy expenditure with the RT3 triaxial accelerometerRes Q Exerc Sport2009 80249256
Not in our library10.1080/02701367.2009.10599559
- 82.
DeVoe D Gotshall R McArthur T Comparison of the RT3 Research Tracker and Tritrac R3D accelerometersPercept Mot Skills2003 97510518
Not in our library10.2466/pms.2003.97.2.510
- 83.
Howe CA Staudenmayer JW Freedson PS Accelerometer prediction of energy expenditure: vector magnitude versus vertical axisMed Sci Sports Exerc2009 4121992206
Not in our library10.1249/mss.0b013e3181aa3a0e
- 84.
Campbell KL Crocker PR McKenzie DC Field evaluation of energy expenditure in women using Tritrac accelerometersMed Sci Sports Exerc2002 3416671674
Not in our library10.1097/00005768-200210000-00020
- 85.
Chen KY Sun M Improving energy expenditure estimation by using a triaxial accelerometerJ Appl Physiol1997 8321122122
Not in our library10.1152/jappl.1997.83.6.2112
- 86.
Jakicic JM Winters C Lagally K Ho J Robertson RJ Wing RR The accuracy of the TriTrac-R3D accelerometer to estimate energy expenditureMed Sci Sports Exerc1999 31747754
Not in our library10.1097/00005768-199905000-00020
- 87.
Plasqui G Joosen AM Kester AD Goris AH Westerterp KR Measuring free-living energy expenditure and physical activity with triaxial accelerometryObes Res2005 1313631369
Not in our library10.1038/oby.2005.165
- 88.
Bouten CV Verboeket-van de Venne WP Westerterp KR Verduin M Janssen JD Daily physical activity assessment: comparison between movement registration and doubly labeled waterJ Appl Physiol1996 8110191026
Not in our library10.1152/jappl.1996.81.2.1019
- 89.
Meijer EP Goris AH Wouters L Westerterp KR Physical inactivity as a determinant of the physical activity level in the elderlyInt J Obes Relat Metab Disord2001 25935939
Not in our library10.1038/sj.ijo.0801644
- 90.
Bonomi AG Plasqui G Goris AH Westerterp KR Estimation of free-living energy expenditure using a novel activity monitor designed to minimize obtrusivenessObesity (Silver Spring)2010 1818451851
Not in our library10.1038/oby.2010.34
- 91.
Goris AH Meijer EP Kester A Westerterp KR Use of a triaxial accelerometer to validate reported food intakesAm J Clin Nutr2001 73549553
Not in our library10.1093/ajcn/73.3.549
- 92.
Meijer GA Westerterp KR van Hulsel AM ten Hoor F Physical activity and energy expenditure in lean and obese adult human subjectsEur J Appl Physiol Occup Physiol1992 65525528
Not in our library10.1007/bf00602359
- 93.
Pietilainen KH Kaprio J Borg P Plasqui G Yki-Jarvinen H Kujala UM Rose RJ Westerterp KR Rissanen A Physical inactivity and obesity: a vicious circleObesity (Silver Spring)2008 16409414
Not in our library10.1038/oby.2007.72
- 94.
Verbunt JA Westerterp KR van der Heijden GJ Seelen HA Vlaeyen JW Knottnerus JA Physical activity in daily life in patients with chronic low back painArch Phys Med Rehabil2001 82726730
Not in our library10.1053/apmr.2001.23182
- 95.
Bouten CV Koekkoek KT Verduin M Kodde R Janssen JD A triaxial accelerometer and portable data processing unit for the assessment of daily physical activityIEEE Trans Biomed Eng1997 44136147
Not in our library10.1109/10.554760
- 96.
Bouten CV Westerterp KR Verduin M Janssen JD Assessment of energy expenditure for physical activity using a triaxial accelerometerMed Sci Sports Exerc1994 2615161523
Not in our library
- 97.
Levine JA Baukol PA Westerterp KR Validation of the Tracmor triaxial accelerometer system for walkingMed Sci Sports Exerc2001 3315931597
Not in our library10.1097/00005768-200109000-00024
- 98.
Bonomi AG Plasqui G Goris AH Westerterp KR Estimation of free-living energy expenditure using a novel activity monitor designed to minimize obtrusivenessObesity (Silver Spring)2010 1818451851
Not in our library10.1038/oby.2010.34
- 99.
Brandes M Rosenbaum D Correlations between the step activity monitor and the DynaPort ADL-monitorClin Biomech (Bristol, Avon)2004 199194
Not in our library10.1016/j.clinbiomech.2003.08.001
- 100.
Langer D Gosselink R Sena R Burtin C Decramer M Troosters T Validation of two activity monitors in patients with COPDThorax2009 64641642
Not in our library10.1136/thx.2008.112102
- 101.
van Hees VT van Lummel RC Westerterp KR Estimating activity-related energy expenditure under sedentary conditions using a tri-axial seismic accelerometerObesity (Silver Spring)2009 1712871292
Not in our library10.1038/oby.2009.55
- 102.
Carter J Wilkinson D Blacker S Rayson M Bilzon J Izard R Coward A Wright A Nevill A Rennie K An investigation of a novel three-dimensional activity monitor to predict free-living energy expenditureJ Sports Sci2008 26553561
Not in our library10.1080/02640410701708979
- 103.
Horner FE Rayson MP Bilzon JL Reliability and validity of the 3DNX accelerometer during mechanical and human treadmill exercise testingInt J Obes (Lond)2011 35Suppl 1S88S97
Not in our library10.1038/ijo.2011.39
- 104.
Hikihara Y Tanaka S Ohkawara K Ishikawa-Takata K Tabata I Validation and Comparison of Three Accelerometers for Measuring Physical Activity Intensity During Nonlocomotive Activities and Locomotive MovementsJ Phys Act Health2011
Not in our library10.1123/jpah.9.7.935
- 105.
Dugas LR Carstens MA Ebersole K Schoeller DA Durazo-Arvizu RA Lambert EV Luke A Energy expenditure in young adult urban informal settlement dwellers in South AfricaEur J Clin Nutr2009 63805807
Not in our library10.1038/ejcn.2008.75
- 106.
Heil DP Predicting activity energy expenditure using the Actical activity monitorRes Q Exerc Sport2006 776480
Not in our library10.5641/027013606x13080769703920
- 107.
Crouter SE Schneider PL Karabulut M Bassett DR Validity of 10 electronic pedometers for measuring steps, distance, and energy costMed Sci Sports Exerc2003 3514551460
Not in our library10.1249/01.mss.0000078932.61440.a2
- 108.
Spierer DK Hagins M Rundle A Pappas E A comparison of energy expenditure estimates from the Actiheart and Actical physical activity monitors during low intensity activities, walking, and joggingEur J Appl Physiol2011 111659667
Not in our library10.1007/s00421-010-1672-7
- 109.
Crouter SE Dellavalle DM Horton M Haas JD Frongillo EA Bassett DR Validity of the Actical for estimating free-living physical activityEur J Appl Physiol2011 11113811389
Not in our library10.1007/s00421-010-1758-2
- 110.
Heiermann S Khalaj Hedayati K Muller MJ Dittmar M Accuracy of a portable multisensor body monitor for predicting resting energy expenditure in older people: a comparison with indirect calorimetryGerontology2010 57473479
Not in our library10.1159/000322109
- 111.
Duncan GE Lester J Migotsky S Goh J Higgins L Borriello G Accuracy of a novel multi-sensor board for measuring physical activity and energy expenditureEur J Appl Physiol2011 11120252032
Not in our library10.1007/s00421-011-1834-2
- 112.
Hooker SP Feeney A Hutto B Pfeiffer KA McIver K Heil DP Vena JE Lamonte MJ Blair SN Validation of the actical activity monitor in middle-aged and older adultsJ Phys Act Health2011 8372381
Not in our library10.1123/jpah.8.3.372
- 113.
Atallah L Leong JJ Lo B Yang GZ Energy expenditure prediction using a miniaturized ear-worn sensorMed Sci Sports Exerc2011 4313691377
Not in our library10.1249/mss.0b013e3182093014
- 114.
Manohar C McCrady S Pavlidis IT Levine JA An accelerometer-based earpiece to monitor and quantify physical activityJ Phys Act Health2009 6781789
Not in our library10.1123/jpah.6.6.781
- 115.
- 116.
Ohkawara K Oshima Y Hikihara Y Ishikawa-Takata K Tabata I Tanaka S Real-time estimation of daily physical activity intensity by a triaxial accelerometer and a gravity-removal classification algorithmBr J Nutr2011 10516811691
Not in our library10.1017/s0007114510005441
- 117.
Annegarn J Spruit MA Uszko-Lencer NH Vanbelle S Savelberg HH Schols AM Wouters EF Meijer K Objective physical activity assessment in patients with chronic organ failure: a validation study of a new single-unit activity monitorArch Phys Med Rehabil2011 9218521857e1851
Not in our library10.1016/j.apmr.2011.06.021
- 118.
Harris AM McAlpine DE Shirbhate R Manohar CU Levine JA Measurement of daily activity in restrictive type anorexia nervosaInt J Eat Disord2008 41280283
Not in our library10.1002/eat.20486
- 119.
Arvidsson D Slinde F Nordenson A Larsson S Hulthen L Validity of the ActiReg system in assessing energy requirement in chronic obstructive pulmonary disease patientsClin Nutr2006 256874
Not in our library10.1016/j.clnu.2005.09.001
- 120.
Hustvedt BE Christophersen A Johnsen LR Tomten H McNeill G Haggarty P Lovo A Description and validation of the ActiReg: a novel instrument to measure physical activity and energy expenditureBr J Nutr2004 9210011008
Not in our library10.1079/bjn20041272
- 121.
Bussmann JB Hartgerink I van der Woude LH Stam HJ Measuring physical strain during ambulation with accelerometryMed Sci Sports Exerc2000 3214621471
Not in our library10.1097/00005768-200008000-00015
- 122.
Mignault D St-Onge M Karelis AD Allison DB Rabasa-Lhoret R Evaluation of the Portable HealthWear Armband: a device to measure total daily energy expenditure in free-living type 2 diabetic individualsDiabetes Care2005 28225227
Not in our library10.2337/diacare.28.1.225-a
- 123.
St-Onge M Mignault D Allison DB Rabasa-Lhoret R Evaluation of a portable device to measure daily energy expenditure in free-living adultsAm J Clin Nutr2007 85742749
Not in our library10.1093/ajcn/85.3.742
- 124.
Johannsen DL Calabro MA Stewart J Franke W Rood JC Welk GJ Accuracy of armband monitors for measuring daily energy expenditure in healthy adultsMed Sci Sports Exerc2010 4221342140
Not in our library10.1249/mss.0b013e3181e0b3ff
- 125.
Koehler K Braun H de Marees M Fusch G Fusch C Schaenzer W Assessing energy expenditure in male endurance athletes: validity of the SenseWear ArmbandMed Sci Sports Exerc2011 4313281333
Not in our library10.1249/mss.0b013e31820750f5
- 126.
Mackey DC Manini TM Schoeller DA Koster A Glynn NW Goodpaster BH Satterfield S Newman AB Harris TB Cummings SR Validation of an armband to measure daily energy expenditure in older adultsJ Gerontol A Biol Sci Med Sci2011 6611081113
Not in our library10.1093/gerona/glr101
- 127.
Cole PJ LeMura LM Klinger TA Strohecker K McConnell TR Measuring energy expenditure in cardiac patients using the Body Media Armband versus indirect calorimetry. A validation studyJ Sports Med Phys Fitness2004 44262271
Not in our library
- 128.
Fruin ML Rankin JW Validity of a multi-sensor armband in estimating rest and exercise energy expenditureMed Sci Sports Exerc2004 3610631069
Not in our library10.1249/01.mss.0000128144.91337.38
- 129.
Jakicic JM Marcus M Gallagher KI Randall C Thomas E Goss FL Robertson RJ Evaluation of the SenseWear Pro Armband to assess energy expenditure during exerciseMed Sci Sports Exerc2004 36897904
Not in our library10.1249/01.mss.0000126805.32659.43
- 130.
Patel SA Benzo RP Slivka WA Sciurba FC Activity monitoring and energy expenditure in COPD patients: a validation studyCOPD2007 4107112
Not in our library10.1080/15412550701246658
- 131.
Dwyer TJ Alison JA McKeough ZJ Elkins MR Bye PT Evaluation of the SenseWear activity monitor during exercise in cystic fibrosis and in healthRespir Med2009 10315111517
Not in our library10.1016/j.rmed.2009.04.013
- 132.
Furlanetto KC Bisca GW Oldemberg N Sant’anna TJ Morakami FK Camillo CA Cavalheri V Hernandes NA Probst VS Ramos EM Step counting and energy expenditure estimation in patients with chronic obstructive pulmonary disease and healthy elderly: accuracy of 2 motion sensorsArch Phys Med Rehabil2010 91261267
Not in our library10.1016/j.apmr.2009.10.024
- 133.
Hill K Dolmage TE Woon L Goldstein R Brooks D Measurement properties of the SenseWear armband in adults with chronic obstructive pulmonary diseaseThorax2010 65486491
Not in our library10.1136/thx.2009.128702
- 134.
Drenowatz C Eisenmann JC Validation of the SenseWear Armband at high intensity exerciseEur J Appl Physiol2011 111883887
Not in our library10.1007/s00421-010-1695-0
- 135.
Cavalheri V Donaria L Ferreira T Finatti M Camillo CA Cipulo Ramos EM Pitta F Energy expenditure during daily activities as measured by two motion sensors in patients with COPDRespir Med2011 105922929
Not in our library10.1016/j.rmed.2011.01.004
- 136.
Berntsen S Stafne SN Morkved S Physical activity monitor for recording energy expenditure in pregnancyActa Obstet Gynecol Scand2010 90903907
Not in our library10.1111/j.1600-0412.2011.01172.x
- 137.
Brazeau AS Karelis AD Mignault D Lacroix MJ Prud’homme D Rabasa-Lhoret R Accuracy of the SenseWear Armband during ergocyclingInt J Sports Med2011 32761764
Not in our library10.1055/s-0031-1279768
- 138.
Munguia-Izquierdo D Santalla A Legaz-Arrese A Evaluation of a wearable body monitoring device during treadmill walking and jogging in patients with fibromyalgia syndromeArch Phys Med Rehabil2012 93115122
Not in our library10.1016/j.apmr.2011.08.021
- 139.
Crouter SE Churilla JR Bassett DR Accuracy of the Actiheart for the assessment of energy expenditure in adultsEur J Clin Nutr2008 62704711
Not in our library10.1038/sj.ejcn.1602766
- 140.
Lof M Physical activity pattern and activity energy expenditure in healthy pregnant and non-pregnant Swedish womenEur J Clin Nutr2011 6512951301
Not in our library10.1038/ejcn.2011.129
- 141.
Kwon S Jamal M Zamba GK Stumbo P Samuel I Validation of a novel physical activity assessment device in morbidly obese femalesJ Obes2010 201018
Not in our library10.1155/2010/856376
- 142.
Borrell B Epidemiology: every bite you takeNature2011 470320322
Not in our library10.1038/470320a
- 143.
Schenk AK Witbrodt BC Hoarty CA Carlson RH Goulding EH Potter JF Bonasera SJ Cellular telephones measure activity and lifespace in community-dwelling adults: proof of principleJ Am Geriatr Soc2011 59345352
Not in our library10.1111/j.1532-5415.2010.03267.x
- 144.
Ades PA Savage PD Brochu M Tischler MD Lee NM Poehlman ET Resistance training increases total daily energy expenditure in disabled older women with coronary heart diseaseJ Appl Physiol2005 9812801285
Not in our library10.1152/japplphysiol.00360.2004
- 145.
Gosselink R Troosters T Decramer M Peripheral muscle weakness contributes to exercise limitation in COPDAm J Respir Crit Care Med1996 153976980
Not in our library10.1164/ajrccm.153.3.8630582
- 146.
Balashov K Feldman DE Savard S Houde S Frenette M Ducharme A Giannetti N Michel C Pilote L Percent predicted value for the 6-minute walk test: using norm-referenced equations to characterize severity in persons with CHFJ Card Fail2008 147581
Not in our library10.1016/j.cardfail.2007.09.005
- 147.
Pitta F Troosters T Probst VS Spruit MA Decramer M Gosselink R Quantifying physical activity in daily life with questionnaires and motion sensors in COPDEur Respir J2006 2710401055
Not in our library10.1183/09031936.06.00064105
- 148.
Tudor-Locke C Williams JE Reis JP Pluto D Utility of pedometers for assessing physical activity: convergent validitySports Med2002 32795808
Not in our library10.2165/00007256-200232120-00004
- 149.
Baarends EM Schols AM Pannemans DL Westerterp KR Wouters EF Total free living energy expenditure in patients with severe chronic obstructive pulmonary diseaseAm J Respir Crit Care Med1997 155549554
Not in our library10.1164/ajrccm.155.2.9032193
- 150.
- 151.
Watz H Waschki B Meyer T Magnussen H Physical activity in patients with COPDEur Respir J2009 33262272
Not in our library10.1183/09031936.00024608
- 152.
Murphy SL Review of physical activity measurement using accelerometers in older adults: considerations for research design and conductPrev Med2009 48108114
Not in our library10.1016/j.ypmed.2008.12.001
- 153.
Schmidt ME Steindorf K Statistical methods for the validation of questionnaires–discrepancy between theory and practiceMethods Inf Med2006 45409413
Not in our library
- 154.
Bland JM Altman DG Applying the right statistics: analyses of measurement studiesUltrasound Obstet Gynecol2003 228593
Not in our library10.1002/uog.122
Cited by papers we hold 10
- Online compensation detecting for real-time reduction of compensatory motions during reaching: a pilot study with stroke survivors · 2020Not checked
- Automatic Detection of Compensatory Movement Patterns by a Pressure Distribution Mattress Using Machine Learning Methods: A Pilot Study · 2019Not checked
- Detecting compensatory movements of stroke survivors using pressure distribution data and machine learning algorithms · 2019Not checked
- Counting Steps in Activities of Daily Living in People With a Chronic Disease Using Nine Commercially Available Fitness Trackers: Cross-Sectional Validity Study · 2018Not checked
- A systematic literature review of reviews on techniques for physical activity measurement in adults: a DEDIPAC study · 2018No notice
- Wearable activity monitors to assess performance status and predict clinical outcomes in advanced cancer patients · 2018No notice
- Behavior Change with Fitness Technology in Sedentary Adults: A Review of the Evidence for Increasing Physical Activity · 2017No notice
- Acceptance of Commercially Available Wearable Activity Trackers Among Adults Aged Over 50 and With Chronic Illness: A Mixed-Methods Evaluation · 2016No notice
- Remote Physical Activity Monitoring in Neurological Disease: A Systematic Review · 2016No notice
- Assessment of Physical Activity and Energy Expenditure: An Overview of Objective Measures · 2014No notice
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.