Use of UV-C radiation to disinfect non-critical patient care items: a laboratory assessment of the Nanoclave Cabinet

Moore Ginny, Ali Shanom, Cloutman-Green Elaine A and 5 more

BMC Infectious Diseases · 2012

doi:10.1186/1471-2334-12-174 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

<p>Abstract</p> <p>Background</p> <p>The near-patient environment is often heavily contaminated, yet the decontamination of near-patient surfaces and equipment is often poor. The Nanoclave Cabinet produces large amounts of ultraviolet-C (UV-C) radiation (53 W/m<sup>2</sup>) and is designed to rapidly disinfect individual items of clinical equipment. Controlled laboratory studies were conducted to assess its ability to eradicate a range of potential pathogens including <it>Clostridium difficile</it> spores and Adenovirus from different types of surface.</p> <p>Methods</p> <p>Each test surface was inoculated with known levels of vegetative bacteria (10<sup>6</sup> cfu/cm<sup>2</sup>), <it>C. difficile</it> spores (10<sup>2</sup>-10<sup>6</sup> cfu/cm<sup>2</sup>) or Adenovirus (10<sup>9</sup> viral genomes), placed in the Nanoclave Cabinet and exposed for up to 6 minutes to the UV-C light source. Survival of bacterial contaminants was determined via conventional cultivation techniques. Degradation of viral DNA was determined via PCR. Results were compared to the number of colonies or level of DNA recovered from non-exposed control surfaces. Experiments were repeated to incorporate organic soils and to compare the efficacy of the Nanoclave Cabinet to that of antimicrobial wipes.</p> <p>Results</p> <p>After exposing 8 common non-critical patient care items to two 30-second UV-C irradiation cycles, bacterial numbers on 40 of 51 target sites were consistently reduced to below detectable levels (≥ 4.7 log<sub>10</sub> reduction). Bacterial load was reduced but still persisted on other sites. Objects that proved difficult to disinfect using the Nanoclave Cabinet (e.g. blood pressure cuff) were also difficult to disinfect using antimicrobial wipes. The efficacy of the Nanoclave Cabinet was not affected by the presence of organic soils. <it>Clostridium difficile</it> spores were more resistant to UV-C irradiation than vegetative bacteria. However, two 60-second irradiation cycles were sufficient to reduce the number of surface-associated spores from 10<sup>3</sup> cfu/cm<sup>2</sup> to below detectable levels. A 3 log<sub>10</sub> reduction in detectable Adenovirus DNA was achieved within 3 minutes; after 6 minutes, viral DNA was undetectable.</p> <p>Conclusion</p> <p>The results of this study suggest that the Nanoclave Cabinet can provide rapid and effective disinfection of some patient-related equipment. However, laboratory studies do not necessarily replicate ‘in-use’ conditions and further tests are required to assess the usability, acceptability and relative performance of the Nanoclave Cabinet when used <it>in situ</it>.</p>

Abstract licence: CC BY-NC-ND

Cited by papers we hold1

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.