Biomechanical Evaluation of Extramedullary Versus Intramedullary Reduction in Unstable Femoral Trochanteric Fractures
Tadashi Kawamura MD, Hiroaki Minehara MD, PhD, Ryo Tazawa MD, PhD and 3 more
Geriatric Orthopaedic Surgery & Rehabilitation · 2021
doi:10.1177/2151459321998611 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
Introduction: The failure rate of operations involving the cephalomedullary nail technique for unstable femoral trochanteric fractures is 3-12%. Changing the reduction strategy may improve the stability. This study aimed to confirm whether reducing the proximal fragment with the medial calcar contact, as opposed to utilizing an intramedullary reduction, would improve the stability of such fractures. Materials and Methods: The unstable femoral trochanteric fracture model was created with fixation by cephalomedullary nails in 22 imitation bones. The 2 reduction patterns were as follows: one was with the proximal head-neck fragment external to the distal bone in the frontal plane and anterior in the sagittal plane as “Extramedullary,” while the other was the opposite reduction position, that is, bone in the frontal plane and sagittal plane as “Intramedullary.” We evaluated the tip-apex distance, compression stiffness, change in femoral neck-shaft angle, amount of blade telescoping, and diameter of the distal screw hole after the compression test. Statistical analysis was conducted using the Mann-Whitney U test. Results: No significant differences were seen in compression stiffness ( p = 0.804) and femoral neck-shaft angle change ( p = 0.644). Although the “Extramedullary” tip-apex distance was larger than the “Intramedullary” distance ( p = 0.001), it indicated clinically acceptable lengths. The amount of blade telescoping and the distal screw hole diameter were significantly larger in “Intramedullary” than in “Extramedullary” ( p < 0.001, p = 0.019, respectively). Our results showed that “Intramedullary” had significantly larger blade telescoping and distal screw hole diameters than “Extramedullary,” and contrary to our hypothesis, no significant differences were seen in compression stiffness and femoral neck-shaft angle change. Conclusions: As opposed to the “Intramedullary” reduction pattern, the biomechanical properties of the “Extramedullary” reduction pattern improved stability during testing and decreased sliding.
Abstract licence: CC BY-NC
References27
- 12.
- 16.
- 24.
24 more references, whose wording the paper’s licence keeps private.
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.