Calibration of the Absolute Efficiency of Well-Type NaI(Tl) Scintillation Detector in 0.121–1.408 MeV Energy Range

Dalal Al Oraini

Science and Technology of Nuclear Installations · 2018

doi:10.1155/2018/6432380 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Well-type NaI(Tl) detectors are beneficial for low-level photon activity measurements because of the near 4π solid angle that can be gained with them. The detection efficiency can differ with the source-to-detector system geometries, the absorption of the photon in the detector material, and attenuation layers in front of the detector face. For these purposes, the absolute efficiency and the coincidence corrections of the well-type sodium iodide detector have been measured at 0.121–1.408 MeV energy range (obtained from 152Eu, 137Cs, and 60Co radioactive isotopes). The covenant between the experimental (present work) and the published theoretical values is good, with the high discrepancies being less than 1%.