Computational fluid dynamics modelling in cardiovascular medicine
Paul D Morris, Andrew Narracott, Hendrik von Tengg-Kobligk and 8 more
Heart · 2015 · Review
doi:10.1136/heartjnl-2015-308044
Abstract
This paper reviews the methods, benefits and challenges associated with the adoption and translation of computational fluid dynamics (CFD) modelling within cardiovascular medicine. CFD, a specialist area of mathematics and a branch of fluid mechanics, is used routinely in a diverse range of safety-critical engineering systems, which increasingly is being applied to the cardiovascular system. By facilitating rapid, economical, low-risk prototyping, CFD modelling has already revolutionised research and development of devices such as stents, valve prostheses, and ventricular assist devices. Combined with cardiovascular imaging, CFD simulation enables detailed characterisation of complex physiological pressure and flow fields and the computation of metrics which cannot be directly measured, for example, wall shear stress. CFD models are now being translated into clinical tools for physicians to use across the spectrum of coronary, valvular, congenital, myocardial and peripheral vascular diseases. CFD modelling is apposite for minimally-invasive patient assessment. Patient-specific (incorporating data unique to the individual) and multi-scale (combining models of different length- and time-scales) modelling enables individualised risk prediction and virtual treatment planning. This represents a significant departure from traditional dependence upon registry-based, population-averaged data. Model integration is progressively moving towards ‘digital patient’ or ‘virtual physiological human’ representations. When combined with population-scale numerical models, these models have the potential to reduce the cost, time and risk associated with clinical trials. The adoption of CFD modelling signals a new era in cardiovascular medicine. While potentially highly beneficial, a number of academic and commercial groups are addressing the associated methodological, regulatory, education- and service-related challenges.
Full text
From PubMed Central, U.S. National Library of Medicine, as current on Oct 18, 2026. Paragraph numbers are ours, so an answer can point at one.
Introduction
Computational fluid dynamics (CFD) is a well-established tool used in engineering, in many areas of which it has become the primary method for design and analysis. Bioengineers have adopted CFD to study complex physiological flows and have demonstrated their potential. There is increasing interest in applying these methods in cardiovascular medicine. CFD-based techniques are being used to build complex computer representations (in silico models) of the cardiovascular system in health and disease. CFD modelling is a new field within cardiovascular medicine, enhancing diagnostic assessment, device design and clinical trials. It can predict physiological responses to intervention and compute previously unmeasurable haemodynamic parameters. As CFD modelling continues to translate into clinical tools, it is important that clinicians understand the principles, benefits and limitations of these techniques. This article explores these topics using state-of-the-art examples in key clinical areas, highlighting applications likely to impact clinical practice within the next 5 years (table 1).
What is CFD?
CFD is a specialist area of mathematics and a branch of fluid mechanics. It is used in the design of many safety-critical systems, including aircraft and vehicles, by solving differential equations to simulate fluid flow. A glossary of uesful terms is provided in table 2.
For incompressible flows, almost all CFD analyses solve the Navier-Stokes and continuity equations which govern fluid motion. These equations are non-linear, partial differential equations based upon the principle of conservation of mass and momentum. Simplification of these equations yields familiar formulae (eg, those of Bernoulli and Poiseuille); but for complex geometries analytical solutions are not possible, so specialised software applications (CFD solvers) calculate approximate numerical solutions. Non-linearity, due to convective fluid acceleration, makes this challenging, especially in three dimensional (3D) models; so CFD analyses require significant computational power and time.
CFD model complexity
The applications reviewed in this paper focus on 3D CFD analyses of local regions of the vasculature because this is where promising applications are beginning to translate and impact upon clinical medicine. There is a long history of simplification of the governing equations to lower spatial dimensions. Table 3 summarises the relationship between these approaches and provides clinical examples of their use.
2D analyses typically assume symmetry of the solution about the central axis, 1D models capture variation of the solution along the axial direction only, and 0D representations lump the behaviour of vascular regions into a model with no spatial dimensions, hence the term ‘lumped-parameter models'. Due to the breadth of the literature covering application of these techniques to cardiovascular haemodynamics, the interested reader is referred to recent reviews of the state-of-the-art.
Model construction
CFD model construction and solution can be described in seven stages (figure 1): Clinical imaging
Clinical imaging
A range of medical imaging modalities can be used, including ultrasound, CT, MRI and X-ray angiography. Imaging must provide sufficient anatomical and physiological detail, in an appropriate format and quality, to enable segmentation and data extraction. Segmentation and reconstruction
Segmentation and reconstruction
Segmentation methods convert medical images to in silico geometries which define the physical bounds of the model region of interest. If images are acquired over a cardiac cycle, anatomical motion can be tracked over segmented regions. Discretisation
Discretisation
Spatial discretisation, or ‘meshing’, divides the geometry into a number of discrete volumetric elements or cells. Temporal discretisation divides the solution into discrete time steps. The accuracy and numerical stability of the analysis are influenced by both spatial and temporal refinement. The fabrication of the mesh, and the level of mesh refinement, are influenced by case- and context-specific factors. The mesh and timestep (ie, spatio-temporal discretisation) must be refined enough to capture the important haemodynamic behaviour of the modelled compartment (the final solution should be independent of mesh parameters), but without excessive refinement because this impacts negatively on computational resource and solution time (see online supplementary table S1). Boundary conditions
Boundary conditions
Because it is impossible to discretise the entire cardiovascular system, the region to be analysed will have at least one inlet and one outlet. To enable CFD analysis, the physiological conditions at the wall and inlet/outlet boundaries must be specified. Boundary conditions are a set of applied physiological parameters (which may vary over time) that define the physical conditions at the inlets, outlets and walls. They may be based on patient-specific data, population data, physical models or assumptions. Simulation
Simulation
A computer file defining the physical parameters of the model is written. In addition to the geometric, discretisation and boundary data, this file must define properties including: blood density and viscosity (ie, the fluid model), the initial conditions of the system (eg, whether the fluid is initially static or moving), time discretisation information (time step size and numerical approximation schemes), and the desired output data (eg, number of cardiac cycles to be simulated). This information allows the CFD solver to solve the Navier-Stokes and continuity equations, proceeding incrementally towards a final solution (‘convergence’). A typical 3D cardiovascular simulation involves >1 million elements run over several cardiac cycles, each divided into hundreds or thousands of individual time-steps. Millions of non-linear partial differential equations are solved simultaneously, and repeatedly, over all elements, at all time-steps. 3D CFD modelling is therefore time-consuming and computationally demanding. Post-processing
Post-processing
Typically, the CFD solver produces the pressure and velocity field over all elements at each time-step. Only a small proportion of these data are of interest to the operator, so some post-processing is required to extract and display relevant data. Validation
Validation
It is important that modelled results are validated against an acceptable standard. Commonly, this involves comparison with either values measured within an in vitro phantom or acquired during in vivo assessment. Validation generates confidence in the accuracy and reliability of a CFD model.
The workflow
Collectively, the steps outlined above are known as a workflow or toolchain (see figure 1 and online supplementary video). Although there are many specialised software applications facilitating the construction and operation of CFD-based workflows, considerable skill and experience are required at each stage (especially steps 3–5) to ensure reliability of results.
Advanced boundary conditions
Rather than specifying pressure or flow at a boundary, an additional, lower-order model may be coupled to the 3D solver to generate more realistic conditions proximal and distal to the simulation domain (see table 1 and figure 2). This method of modelling is efficient, because it allows detailed analysis in the 3D region without wasting high temporal and spatial refinement on regions beyond this. In some cases the model representing the distal boundary also provides proximal boundary condition. These closed-loop models, or system models, require very careful tuning.
Assumptions
Many CFD models assume that the segmented region has rigid walls. Although untrue in the cardiovascular system, this approximation is acceptable for some applications. Vessel compliance allows blood to be stored during systole and released during diastole. At the system level this results in a finite speed of pressure wave transmission and tends to reduce the peak pressures associated with the inertial acceleration of the blood. Compliance tends to reduce shear stresses because the vessels are slightly larger when peak flow occurs. It is possible to model deformation of the wall, due to cardiac and respiratory variation, in response to change in pressure using fluid-solid interaction (FSI) models. These are far more complex to solve, boundary conditions are a challenge, and many wall parameters remain unknown, which increases the number of assumptions. Furthermore, it is yet to be established for which applications a full FSI approach improves accuracy. An example of the increased computational cost of FSI is reported by Brown et al where 3D transient (time-varying) analysis of the aorta required 145 h (FSI) compared with 6.6 h (CFD). An alternative is to impose wall movement derived from imaging data (eg, gated MRI). There are exciting developments in the use of data assimilation techniques in which sparse clinical data, for example, from 4D imaging, are integrated with the analysis process so that material properties of tissues in individual patients are recovered as the simulation progresses. In biomedical workflows it is assumed the boundaries of the fluid geometry are smooth, yet medical images may not generate smooth surfaces due to poor resolution or imaging artefacts. Instead, structures may be smoothed in silico after segmentation. Typically, cardiovascular simulations assume blood behaves as an incompressible fluid. Although blood exhibits non-Newtonian behaviour (see glossary in table 2), which must be simulated for flow in small capillaries, in larger vessels these effects are often neglected and a Newtonian fluid model is assumed.
Benefits of cardiovascular CFD modelling
CFD modelling enables investigation of pressure and flow fields at a temporal and spatial resolution unachievable by any clinical methodology. Post-processing provides additional data, generating new insights into physiology and disease processes. For example, it is difficult, and invasive, to measure arterial wall shear stress (WSS), a key factor in the development of atherosclerosis and in-stent restenosis, whereas CFD models can compute WSS and map its spatial distribution. Such work has established the link between haemodynamic disturbance and atherogenesis and has explained the preferential deposition of atherosclerotic plaque at arterial bends and bifurcation regions. CFD modelling has been central to our current understanding of the effects of WSS on endothelial homeostasis: laminar, non-disturbed blood flow is associated with increased WSS which inhibits unnecessary endothelial cell activation; whereas turbulent or disturbed blood flow reduces WSS which stimulates adverse vessel remodelling. A complex series of WSS-related signalling pathways and interactions underlie this phenomenon. Before these pathways can be exploited to generate anti-atherosclerotic therapies their complexity needs to be better understood. Integrated multiscale CFD models provide a powerful tool to combine analysis of fluid mechanics and cellular response. Recently, the effect of disturbed WSS in stented vessels has been modelled to investigate the influence on both endothelial function and neointimal hyperplasia, which preferentially accumulates at regions of low and disturbed WSS (figure 3). Such models can be used to develop stents which minimise the risk of in-stent restenosis and thrombosis.
Device design
In silico methods allow rapid prototyping, with reduced risk to humans, so a priority for the medical devices industry is to replace expensive and time-consuming in vivo and in vitro experimentation with in silico testing. In recognition of this, the US Food and Drug Administration (FDA) issued draft guidance in 2014 on the use of modelling to support regulatory submissions. An example is the significant role which CFD plays in design optimisation of mechanical heart valves. While comparison of major flow features captured in vitro with predictions from CFD shows good agreement, simulation delivers 3D information at much higher resolution within critical regions (eg, the hinges) than flow visualisation, giving invaluable insight into design-related thrombogenic potential. Comprehensive simulation of heart valve mechanics, including the separation of the upstream and downstream fluid regions at closure and the structural instability and snap-through dynamics of tissue valves, remain computationally challenging but achievable.
CFD has been used in the optimisation of several commercial ventricular assist devices (VADs), investigating potential thrombogenicity by highlighting design-related regions of stasis and device characteristics resulting in high shear stress. A tool for optimising thromboresistance has been recently demonstrated in a comparative study of two continuous flow VADs, which combines experimental and numerical simulation. Numerical models can also contribute to the process of VAD implantation, informing the choice of catheter implantation site.
In the context of stent design, the greatest emphasis has been on modelling the mechanical integrity of the stent structure during and after deployment. However, CFD provides a valuable tool to assess the resulting haemodynamics within the stented lesion. This, in turn, has been associated with the biological response of the vessel wall and the development of restenosis.
Diagnostic tools and personalised medicine
Computing intravascular physiology, with the aim of minimising invasive instrumentation, is of major interest. A prime example is fractional flow reserve (FFR), an index of physiological (coronary atherosclerosis) lesion significance, measured with a pressure-sensitive angioplasty guidewire. FFR-guided therapy improves patient outcomes, reduces stent insertions, and reduces costs, yet it is used in <10% of cases due to a host of procedural and operator related factors. Several groups propose models for computing ‘virtual’ FFR from angiography to provide the benefits of physiological assessment without the practical drawbacks which limit the invasive technique (see online supplementary figure S1). The VIRTU-1 trial demonstrated the effectiveness of CFD-derived FFR using invasive angiography (diagnostic accuracy 97% vs invasive FFR). More recently, the HeartFlow-NXT trial demonstrated the performance of virtual FFR using CT coronary angiography (sensitivity 86%, specificity 79% vs invasive FFR). Both models offer a less invasive approach, neither requires hyperaemic flow induction nor the passage of an intracoronary wire, and taken together could offer the benefits of FFR to all patients being assessed for coronary artery disease (CAD). CTFFR (HeartFlow Inc) is now FDA approved for use as a class II Coronary Physiologic Simulation Software Device.
Recent in vitro work combining CFD, colour Doppler and simulated Doppler images demonstrates the potential to reduce Doppler inter-user variability and inform interpretation of complex regurgitant flow fields in valvular heart disease. In silico models of the right heart and pulmonary arteries using phase-contrast MRI capture anatomical and flow velocity data to simulate pulmonary artery physiology. It is hoped that these models will soon deliver a diagnosis of pulmonary hypertension without invasive catheterisation. Added value may come from wave analysis of these models to discriminate between the key aetiological sub-groups of pulmonary hypertension.
Treatment decisions are often based upon the ‘gold standard’ of randomised controlled trials. Everyday clinical practice, however, requires tailored treatment for individual patients. CFD modelling offers a patient-specific approach to management, in which an individual's unique anatomy and physiology are used to define the model. The impact of alternative interventional strategies can be compared and a personalised, optimised strategy selected. Patient-specific modelling will not diminish the need for clinical trials but will allow the delivery of truly objective, personalised management on a wide scale.
This approach is exemplified in aortic aneurysm management, where current simplistic guidelines use aneurysm diameter as the arbiter of treatment. The @neurIST project addresses this, and incorporates patient-specific anatomic, genomic and demographic data, with simulated flow patterns (figure 4), to calculate the risk of rupture. The optimal treatment of aortic type B dissections remains controversial. Thoracic endovascular aortic repair (TEVAR) carries a risk of spinal ischaemia, and multiple communicating channels between true and false lumens confer a risk of proximal rupture if only the primary entry is closed. Therefore, graft length is balanced against the risk of paraplegia. In this context, CFD modelling provides individualised risk stratification and optimised treatment delivery (figure 5). Similar simulations are useful in the context of abdominal aortic aneurysm.
Full system models: the Virtual Physiological Human
There is increasing interest in integrating multiple physiological models into comprehensive system models to simulate the impact of various conditions, pathologies and treatments across multiple organ systems. While this is ambitious, large, international, collaborative research projects under the umbrella of the Virtual Physiological Human (VPH) reflect the seriousness, legitimacy and rationale underlying this long-term vision. Full system models offer the possibility of understanding, holistically, the impact of cardiovascular disease upon individual patients.
In silico trials
In silico techniques can simulate measures of safety, accuracy, and efficacy of interventions, both pharmacological and mechanical, in large cohorts of virtual patient models which represent naturally occurring physiological and pathological variability. This minimises the time, cost, and risk associated with clinical trials. The Avicenna project leverages significant commercial interest and engagement to develop a ‘roadmap’ describing the route by which multi-scale in silico techniques can achieve this.
Challenges and limitations
CFD models in medicine have traditionally been used by two user groups: industrial medical device developers, in rapid, low-cost, device prototyping; and academics, to investigate cardiovascular physiology and compute parameters that cannot otherwise be obtained. Both groups construct models which are typically complex, involve multiple finely-tuned geometric, haemodynamic and material parameters and require long computation times. In contrast, clinicians are a third, emerging user group, who require rapid results with adequate accuracy.
Model accuracy is determined by model design and quality of input data. For CFD applications it is unclear how detailed the clinical data needs to be in terms of geometry (segmented from medical images) and parameterisation (variability described by the model and the tuning of patient-specific boundary conditions). Continuing improvements in imaging, image-registration and segmentation algorithms will augment accuracy. Model parameterisation is more challenging, because it requires detailed knowledge of physiological metrics in the proximal and distal circulations which may be inaccessible and variable in health and in disease, for example, microcirculatory resistance is a major determinant of coronary blood flow. Further understanding of the relative importance of physiological parameters is required to determine those which are most influential, and those which can be assumed or averaged. This allows unnecessary model complexity to be simplified, balancing computing speed against accuracy. A second challenge is the development of relevant industry standards. In the European Union and USA, diagnostic software is regulated via CE marking and FDA directives, respectively; but there are no industry standards governing accuracy, reliability or validation. The FDA is addressing this through benchmarking initiatives, in the same way that aviation authorities adopted computer-aided design over traditional physical testing. Third, large volumes of clinical data, of value for model development and validation, are stored in hospital systems. Access is variable and restrictive; although the VPH-Share project demonstrates how anonymised patient-specific data and in silico models can be shared securely. Access to such data supports model validation against long-term outcomes which may expedite clinical translation. Fourth, CFD modelling can be perceived as a ‘disruptive technology’, and a threat, especially by manufacturers of traditional hardware. Finally, the next generation of doctors will require training in in silico systems, and understanding their principles, methodologies and limitations. Several initiatives are attempting to address this, but in the longer term, in silico medicine may gain prominence in medical curricula.
Future directions
In the context of device development, the major computational modelling and technological challenges have already been addressed and the benefits of simulation are recognised by the regulatory authorities. Sustained investment will enable engineers to continue model refinement and the development of novel applications, specifically targeting increases in precision, personalisation and speed. Beyond technological development, and before these tools become established in routine clinical practice, the most immediate need is to demonstrate equivalence of in silico results relative to invasive measurements through observational trials. Beyond this, efficacy must be demonstrated in large multicentre clinical trials. It is clear that these techniques have the potential to change clinical practice. The beneficiaries will be patients, clinicians and healthcare providers.
References 74
- 1.
Taylor CA, Figueroa CA Patient-specific modeling of cardiovascular mechanics. Annu Rev Biomed Eng 2009;11:109–34.
Not in our library10.1146/annurev.bioeng.10.061807.160521
- 2.
Smith N, de Vecchi A, McCormick M, et al euHeart: personalized and integrated cardiac care using patient-specific cardiovascular modelling. Interface Focus 2011;1:349–64.
Not in our library10.1098/rsfs.2010.0048
- 3.
Gasser TC, Nchimi A, Swedenborg J, et al A novel strategy to translate the biomechanical rupture risk of abdominal aortic aneurysms to their equivalent diameter risk: method and retrospective validation. Eur J Vasc Endovasc Surg 2014;47:288–95.
Not in our library10.1016/j.ejvs.2013.12.018
- 4.
Radaelli AG, Augsburger L, Cebral JR, et al Reproducibility of haemodynamical simulations in a subject-specific stented aneurysm model—a report on the Virtual Intracranial Stenting Challenge 2007. J Biomech 2008;41:2069–81.
Not in our library10.1016/j.jbiomech.2008.04.035
- 5.
Morris PD, Ryan D, Morton AC, et al Virtual fractional flow reserve from coronary angiography: modeling the significance of coronary lesions: results from the VIRTU-1 (VIRTUal Fractional Flow Reserve From Coronary Angiography) Study. JACC Cardiovasc Interv 2013;6:149–57.
Not in our library10.1016/j.jcin.2012.08.024
- 6.
Nørgaard BL, Leipsic J, Gaur S, et al Diagnostic performance of noninvasive fractional flow reserve derived from coronary computed tomography angiography in suspected coronary artery disease: the NXT trial (Analysis of Coronary Blood Flow Using CT Angiography: Next Steps). J Am Coll Cardiol 2014;63:1145–55.
Not in our library10.1016/j.jacc.2013.11.043
- 7.
Tu S, Barbato E, Køszegi Z, et al Fractional flow reserve calculation from 3-dimensional quantitative coronary angiography and TIMI frame count: a fast computer model to quantify the functional significance of moderately obstructed coronary arteries. JACC Cardiovasc Interv 2014;7:768–77.
Not in our library10.1016/j.jcin.2014.03.004
- 8.
Morris PD, van de Vosse FN, Lawford PV, et al “Virtual” (computed) fractional flow reserve: current challenges and limitations. JACC Cardiovasc Interv 2015;8:1009–17.
Not in our library10.1016/j.jcin.2015.04.006
- 9.
Sotiropoulos F, Borazjani I A review of state-of-the-art numerical methods for simulating flow through mechanical heart valves. Med Biol Eng Comput 2009;47:245–56.
Not in our library10.1007/s11517-009-0438-z
- 10.
Simon HA, Ge L, Borazjani I, et al Simulation of the three-dimensional hinge flow fields of a bileaflet mechanical heart valve under aortic conditions. Ann Biomed Eng 2010;38:841–53.
Not in our library10.1007/s10439-009-9857-0
- 11.
Yoganathan AP, Chandran KB, Sotiropoulos F Flow in prosthetic heart valves: state-of-the-art and future directions. Ann Biomed Eng 2005;33:1689–94.
Not in our library10.1007/s10439-005-8759-z
- 12.
Astorino M, Hamers J, Shadden SC, et al A robust and efficient valve model based on resistive immersed surfaces. Int J Numer Method Biomed Eng 2012;28:937–59.
Not in our library10.1002/cnm.2474
- 13.
De Hart J, Peters GW, Schreurs PJ, et al A three-dimensional computational analysis of fluid-structure interaction in the aortic valve. J Biomech 2003;36:103–12.
Not in our library10.1016/s0021-9290(02)00244-0
- 14.
Erhart P, Hyhlik-Dürr A, Geisbüsch P, et al Finite element analysis in asymptomatic, symptomatic, and ruptured abdominal aortic aneurysms: in search of new rupture risk predictors. Eur J Vasc Endovasc Surg 2015;49:239–45.
Not in our library10.1016/j.ejvs.2014.11.010
- 15.
Georgakarakos E, Ioannou CV, Papaharilaou Y, et al Computational evaluation of aortic aneurysm rupture risk: what have we learned so far? J Endovasc Ther 2011;18:214–25.
Not in our library10.1583/10-3244.1
- 16.
Molony DS, Kavanagh EG, Madhavan P, et al A computational study of the magnitude and direction of migration forces in patient-specific abdominal aortic aneurysm stent-grafts. Eur J Vasc Endovasc Surg 2010;40:332–9.
Not in our library10.1016/j.ejvs.2010.06.001
- 17.
Karmonik C, Müller-Eschner M, Partovi S, et al Computational fluid dynamics investigation of chronic aortic dissection hemodynamics versus normal aorta. Vasc Endovascular Surg 2013;47:625–31.
Not in our library10.1177/1538574413503561
- 18.
Chen D, Müller-Eschner M, von Tengg-Kobligk H, et al A patient-specific study of type-B aortic dissection: evaluation of true-false lumen blood exchange. Biomed Eng Online 2013;12:65
Not in our library10.1186/1475-925x-12-65
- 19.
Cheng Z, Juli C, Wood NB, et al Predicting flow in aortic dissection: comparison of computational model with PC-MRI velocity measurements. Med Eng Phys 2014;36:1176–84.
Not in our library10.1016/j.medengphy.2014.07.006
- 20.
Chien S Mechanotransduction and endothelial cell homeostasis: the wisdom of the cell. Am J Physiol Heart Circ Physiol 2007;292:H1209–24.
Not in our library10.1152/ajpheart.01047.2006
- 21.
LaDisa JF Jr., Olson LE, Molthen RC, et al Alterations in wall shear stress predict sites of neointimal hyperplasia after stent implantation in rabbit iliac arteries. Am J Physiol Heart Circ Physiol 2005;288:H2465–75.
Not in our library10.1152/ajpheart.01107.2004
- 22.
Jin S, Yang Y, Oshinski J, et al Flow patterns and wall shear stress distributions at atherosclerotic-prone sites in a human left coronary artery—an exploration using combined methods of CT and computational fluid dynamics. Conf Proc IEEE Eng Med Biol Soc 2004;5:3789–91.
Not in our library10.1109/iembs.2004.1404062
- 23.
Lee J, Smith NP The multi-scale modelling of coronary blood flow. Ann Biomed Eng 2012;40:2399–413.
Not in our library10.1007/s10439-012-0583-7
- 24.
Jiménez JM, Davies PF Hemodynamically driven stent strut design. Ann Biomed Eng 2009;37:1483–94.
Not in our library10.1007/s10439-009-9719-9
- 25.
Seshadhri S, Janiga G, Beuing O, et al Impact of stents and flow diverters on hemodynamics in idealized aneurysm models. J Biomech Eng 2011;133:071005
Not in our library10.1115/1.4004410
- 26.
Morlacchi S, Migliavacca F Modeling stented coronary arteries: where we are, where to go. Ann Biomed Eng 2013;41:1428–44.
Not in our library10.1007/s10439-012-0681-6
- 27.
Wentzel JJ, Krams R, Schuurbiers JC, et al Relationship between neointimal thickness and shear stress after Wallstent implantation in human coronary arteries. Circulation 2001;103:1740–5.
Not in our library10.1161/01.cir.103.13.1740
- 28.
Peach TW, Ngoepe M, Spranger K, et al Personalizing flow-diverter intervention for cerebral aneurysms: from computational hemodynamics to biochemical modeling. Int J Numer Method Biomed Eng 2014;30:1387–407.
Not in our library10.1002/cnm.2663
- 29.
Schneiders JJ, Marquering HA, van Ooij P, et al Additional value of intra-aneurysmal hemodynamics in discriminating ruptured versus unruptured intracranial aneurysms. AJNR Am J Neuroradiol 2015;10:1920–6.
Not in our library10.3174/ajnr.a4397
- 30.
Qureshi MU, Vaughan GA, Sainsbury C, et al Numerical simulation of blood flow and pressure drop in the pulmonary arterial and venous circulation. Biomech Model Mechanobiol 2014;13:1137–54.
Not in our library10.1007/s10237-014-0563-y
- 31.
Lungu A, Wild JM, Capener D, et al MRI model-based non-invasive differential diagnosis in pulmonary hypertension. J Biomech 2014;47:2941–7.
Not in our library10.1016/j.jbiomech.2014.07.024
- 32.
Tang BT, Pickard SS, Chan FP, et al Wall shear stress is decreased in the pulmonary arteries of patients with pulmonary arterial hypertension: an image-based, computational fluid dynamics study. Pulm Circ 2012;2:470–6.
Not in our library10.4103/2045-8932.105035
- 33.
Kheyfets VO, Rios L, Smith T, et al Patient-specific computational modeling of blood flow in the pulmonary arterial circulation. Comput Methods Programs Biomed 2015;120:88–101.
Not in our library10.1016/j.cmpb.2015.04.005
- 34.
Niederer SA, Plank G, Chinchapatnam P, et al Length-dependent tension in the failing heart and the efficacy of cardiac resynchronization therapy. Cardiovasc Res 2011;89:336–43.
Not in our library10.1093/cvr/cvq318
- 35.
Chan BT, Lim E, Chee KH, et al Review on CFD simulation in heart with dilated cardiomyopathy and myocardial infarction. Comput Biol Med 2013;43:377–85.
Not in our library10.1016/j.compbiomed.2013.01.013
- 36.
Aguado-Sierra J, Krishnamurthy A, Villongco C, et al Patient-specific modeling of dyssynchronous heart failure: a case study. Prog Biophys Mol Biol 2011;107:147–55.
Not in our library10.1016/j.pbiomolbio.2011.06.014
- 37.
Sermesant M, Chabiniok R, Chinchapatnam P, et al Patient-specific electromechanical models of the heart for the prediction of pacing acute effects in CRT: a preliminary clinical validation. Med Image Anal 2012;16:201–15.
Not in our library10.1016/j.media.2011.07.003
- 38.
Brown AG Patient-specific local and systemic haemodynamics in the presence of a left ventricular assist device. Department of cardiovascular science. Sheffield: University of Sheffield, 2012.
Not in our library
- 39.
Brown AG, Shi Y, Arndt A, et al Importance of realistic LVAD profiles for assisted aortic simulations: evaluation of optimal outflow anastomosis locations. Comput Methods Biomech Biomed Engin 2012;15:669–80.
Not in our library10.1080/10255842.2011.556628
- 40.
Chiu WC, Girdhar G, Xenos M, et al Thromboresistance comparison of the HeartMate II ventricular assist device with the device thrombogenicity emulation- optimized HeartAssist 5 VAD. J Biomech Eng 2014;136:021014
Not in our library10.1115/1.4026254
- 41.
Farag MB, Karmonik C, Rengier F, et al Review of recent results using computational fluid dynamics simulations in patients receiving mechanical assist devices for end-stage heart failure. Methodist Debakey Cardiovasc J 2014;10:185–9.
Not in our library10.14797/mdcj-10-3-185
- 42.
Ladisa JF Jr, Taylor CA, Feinstein JA Aortic coarctation: recent developments in experimental and computational methods to assess treatments for this simple condition. Prog Pediatr Cardiol 2010;30:45–9.
Not in our library10.1016/j.ppedcard.2010.09.006
- 43.
- 44.
Pittaccio S, Migliavacca F, Dubini G, et al On the use of computational models for the quantitative assessment of surgery in congenital heart disease. Anadolu Kardiyol Derg 2005;5:202–9.
Not in our library
- 45.
van de Vosse FN, Stergiopulos N Pulse wave propagation in the arterial tree. Annu Rev Fluid Mech 2011;43:467–99.
Not in our library10.1146/annurev-fluid-122109-160730
- 46.
- 47.
Kim HJ, Vignon-Clementel IE, Figueroa CA, et al On coupling a lumped parameter heart model and a three-dimensional finite element aorta model. Ann Biomed Eng 2009;37:2153–69.
Not in our library10.1007/s10439-009-9760-8
- 48.
Rodriguez L, Thomas JD, Monterroso V, et al Validation of the proximal flow convergence method. Calculation of orifice area in patients with mitral stenosis. Circulation 1993;88:1157–65.
Not in our library10.1161/01.cir.88.3.1157
- 49.
Weese J, Groth A, Nickisch H, et al Generating anatomical models of the heart and the aorta from medical images for personalized physiological simulations. Med Biol Eng Comput 2013;51:1209–19.
Not in our library10.1007/s11517-012-1027-0
- 50.
Barber DC, Hose DR Automatic segmentation of medical images using image registration: diagnostic and simulation applications. J Med Eng Technol 2005;29:53–63.
Not in our library10.1080/03091900412331289889
- 51.
Zhao F, Xie X An overview of interactive medical image segmentation. Ann BMVA 2013;7:1–22.
Not in our library
- 52.
Kung E, Pennati G, Migliavacca F, et al A simulation protocol for exercise physiology in Fontan patients using a closed loop lumped-parameter model. J Biomech Eng 2014;136:107001
Not in our library10.1115/1.4027271
- 53.
Zeng D, Boutsianis E, Ammann M, et al A study on the compliance of a right coronary artery and its impact on wall shear stress. J Biomech Eng 2008;130:041014
Not in our library10.1115/1.2937744
- 54.
Brown AG, Shi Y, Marzo A, et al Accuracy vs. computational time: translating aortic simulations to the clinic. J Biomech 2012;45:516–23.
Not in our library10.1016/j.jbiomech.2011.11.041
- 55.
Bertoglio C, Barber D, Gaddum N, et al Identification of artery wall stiffness: in vitro validation and in vivo results of a data assimilation procedure applied to a 3D fluid-structure interaction model. J Biomech 2014;47:1027–34.
Not in our library10.1016/j.jbiomech.2013.12.029
- 56.
Reporting of Computational Modeling Studies in Medical Device Submissions. Rockeville, MD: Draft Guidance for Industry and Food and Drug Administration Staff, 2014.
Not in our library
- 57.
Bluestein D, Girdhar G, Einav S, et al Device thrombogenicity emulation: a novel methodology for optimizing the thromboresistance of cardiovascular devices. J Biomech 2013;46:338–44.
Not in our library10.1016/j.jbiomech.2012.11.033
- 58.
Pijls NH, Fearon WF, Tonino PA, et al Fractional flow reserve versus angiography for guiding percutaneous coronary intervention in patients with multivessel coronary artery disease: 2-year follow-up of the FAME (Fractional Flow Reserve Versus Angiography for Multivessel Evaluation) study. J Am Coll Cardiol 2010;56:177–84.
Not in our library10.1016/j.jacc.2010.04.012
- 59.
Papafaklis MI, Muramatsu T, Ishibashi Y, et al Fast virtual functional assessment of intermediate coronary lesions using routine angiographic data and blood flow simulation in humans: comparison with pressure wire—fractional flow reserve. EuroIntervention 2014;10:574–83.
Not in our library10.4244/eijy14m07_01
- 60.
U.S. Food and Drug Administration. De novo classification request for FFRCT V. 1.4 (De Novo Summary (DEN130045)) 2014.
Not in our library
- 61.
Sonntag SJ, Li W, Becker M, et al Combined computational and experimental approach to improve the assessment of mitral regurgitation by echocardiography. Ann Biomed Eng 2014;42:971–85.
Not in our library10.1007/s10439-013-0968-2
- 62.
Swift AJ, Wild JM, Nagle SK, et al Quantitative magnetic resonance imaging of pulmonary hypertension: a practical approach to the current state of the art. J Thorac Imaging 2014;29:68–79.
Not in our library10.1097/rti.0000000000000079
- 63.
Morlacchi S, Colleoni SG, Cárdenes R, et al Patient-specific simulations of stenting procedures in coronary bifurcations: two clinical cases. Med Eng Phys 2013;35:1272–81.
Not in our library10.1016/j.medengphy.2013.01.007
- 64.
Villa-Uriol MC, Berti G, Hose DR, et al @neurIST complex information processing toolchain for the integrated management of cerebral aneurysms. Interface Focus 2011;1:308–19.
Not in our library10.1098/rsfs.2010.0033
- 65.
Chen D, Müller-Eschner M, Kotelis D, et al A longitudinal study of type-B aortic dissection and endovascular repair scenarios: computational analyses. Med Eng Phys 2013;35:1321–30.
Not in our library10.1016/j.medengphy.2013.02.006
- 66.
Venkatasubramaniam AK, Fagan MJ, Mehta T, et al A comparative study of aortic wall stress using finite element analysis for ruptured and non-ruptured abdominal aortic aneurysms. Eur J Vasc Endovasc Surg 2004;28:168–76.
Not in our library10.1016/j.ejvs.2004.03.029
- 67.
Martufi G, Christian Gasser T Review: the role of biomechanical modeling in the rupture risk assessment for abdominal aortic aneurysms. J Biomech Eng 2013;135:021010
Not in our library10.1115/1.4023254
- 68.
Hunter P, Chapman T, Coveney PV, et al A vision and strategy for the virtual physiological human: 2012 update. Interface Focus 2013;3:20130004
Not in our library10.1098/rsfs.2013.0004
- 69.
http://www.avicenna-isct.org (accessed Jul 2015).
Not in our library
- 70.
Donders WP, Huberts W, van de Vosse FN, et al Personalization of models with many model parameters: an efficient sensitivity analysis approach. Int J Numer Method Biomed Eng 2015;31 Published Online First: 27th May 2015. doi:10.1002/cnm.2727. Epub 2015 Jun 15
Not in our library10.1002/cnm.2727;published
- 71.
Stewart S, Hariharan P, Paterson E, et al Results of FDA's first interlaboratory computational study of a nozzle with a sudden contraction and conical diffuser. Cardiovasc Eng Technol 2013;4:374–91.
Not in our library10.1007/s13239-013-0166-2
- 72.
vph-share.eu. http://vph-share.eu (accessed Jul 2015).
Not in our library
- 73.
Cates CU, Gallagher AG The future of simulation technologies for complex cardiovascular procedures. Eur Heart J 2012;33:2127–34.
Not in our library10.1093/eurheartj/ehs155
- 74.
http://vph-portal.eu (accessed Jul 2015).
Not in our library
Cited by papers we hold 2
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.