State-of-the-art pulsed field ablation for cardiac arrhythmias: ongoing evolution and future perspective
Kyoung-Ryul Julian Chun, Damijan Miklavčič, Konstantinos Vlachos and 4 more
Europace · 2024 · Review · CC BY
Abstract
Pulsed field ablation (PFA) is an innovative approach in the field of cardiac electrophysiology aimed at treating cardiac arrhythmias. Unlike traditional catheter ablation energies, which use radiofrequency or cryothermal energy to create lesions in the heart, PFA utilizes pulsed electric fields to induce irreversible electroporation, leading to targeted tissue destruction. This state-of-the-art review summarizes biophysical principles and clinical applications of PFA, highlighting its potential advantages over conventional ablation methods. Clinical data of contemporary PFA devices are discussed, which combine predictable procedural outcomes and a reduced risk of thermal collateral damage. Overall, these technological developments have propelled the rapid evolution of contemporary PFA catheters, with future advancements potentially impacting patient care.
Full text
From PubMed Central, U.S. National Library of Medicine, as current on Oct 20, 2026. Paragraph numbers are ours, so an answer can point at one.
Introduction
Catheter ablation is a well-established therapy option for rhythm control in patients with symptomatic atrial fibrillation (AF). Thermal-based energy sources such as radiofrequency current (RFC) or cryoballoon/laser balloon have shown to be reasonably safe and effective in this indication. However, treatment success remains limited, with single-procedure success rates of up to 80%. Furthermore, rare but significant energy-related complications to surrounding tissues, such as oesophageal injury or phrenic nerve palsy, may occur. Pulsed field ablation (PFA) is a novel, largely non-thermal ablation modality based on the process of irreversible electroporation (IRE). It holds the promise of creating transmural, durable ablation lesions while reducing the risk of collateral damage. Electroporation has a long-ranging history as a treatment modality in medical fields like oncology. In the 1980s, direct current shock was used for ablation of cardiac arrhythmias, but due to significant complications such as arching and barotrauma of these early technologies, as well as the success of the emerging RFC technology, PFA was no longer clinically pursued.
In the last decade, pre-clinical studies have shown that PFA in its modern form has the potential to create tissue-selective ablation lesions with short, high-energy electrical pulses that facilitate cell death while avoiding thermal energy source-related side effects. First in-human PFA trials demonstrated safe and efficient AF ablation procedures with high rates of both immediate pulmonary vein isolation (PVI) and durability on invasive re-mapping, resulting in comparably low rates of arrhythmia recurrence. As a result, various catheter platforms using electroporation have been developed and tested for catheter ablation of cardiac arrhythmias, mainly AF. We review the methodology of PFA and the pre-clinical and clinical data and discuss future directions.
Mechanisms of electroporation
When a cell is exposed to an electric field, e.g. by delivering electric pulses to the tissue through a catheter, a transmembrane voltage is induced across the cell membrane. This is known as pacing in the field of cardiac electrophysiology. Even a relatively small signal will trigger depolarization or action potential. If high-voltage pulses are applied, the cell plasma membrane will undergo electroporation. Electroporation is a phenomenon that describes the transient increase of membrane permeability for ions and molecules otherwise deprived of transmembrane transport mechanisms. Induced transmembrane voltage in the order of several 100 mV is sufficient to cause pore formation in the lipid domain that leads to lipid oxidation but can also cause protein damage. Electroporation results in a transient increase of membrane permeability, which after some time recovers, and the cell can survive. Reversible electroporation as often referred to this phenomenon has been used in gene transfer and drug delivery for decades., Membrane electroporation and consequently membrane permeability (even if only transient) may cause the cell to die. This IRE has gained enormous attention, and it seems it offers an unprecedented speed and safety when it comes to cardiac ablation.,
Even though cell death due to electroporation is mechanistically not fully understood, it was demonstrated that following plasma membrane electroporation, the cell may recover membrane integrity but die within the next few hours (Figure 1). Cell death due to electroporation is most often ascribed to adenosine triphosphate exhaustion, protein damage, e.g. ion channels, and calcium overload and is more generally ascribed to loss of cell homoeostasis. Cardiac ablation by electroporation is often described as non-thermal and cardiac tissue selective. The non-thermal nature of this energy and its selectivity must be taken with a grain of salt though. Namely, if high-voltage pulses are delivered to tissue through catheters in the blood pool (1500–3000 V), a considerable current will be drawn from the generator (10–20 A)—some heating is inevitable. It will be proportional to the product of the local electric field and current density—like RFC ablation: the non-thermal nature comes from the fact that the cells are killed by electroporation rather than by heat. It takes microseconds or even less to cause electroporation. In such a short time, the energy is too low to considerably heat up the tissue. However, if many repetitive pulses are delivered (with high duty cycle), the energy deposited in the tissue may cause a significant temperature increase locally and the ‘non-thermal’ property of PFA will vanish. Of note, cells that may be situated in high electric fields (e.g. erythrocytes) may become electroporated.,
Parameters such as pulse amplitude, pulse width, number of pulses, polarity (biphasic or monophasic), and pulse cycle—altogether ‘waveform,’ but also, catheter shape/geometry and distance of the tissue/cells from the catheter/electrodes since the electric field drops rapidly with the distance from the electrode/catheter—are parameters that affect and determine effective lesion creation.,
Most of pulse parameters were selected by manufacturers during the development, testing, and optimization of the system (pulse generator and catheter) and may differ from PFA system to PFA system (Figure 2).,
The first devices were developed for relatively thin-walled atria and more specifically for PVI where lesion depth of few millimetres suffices, but targeting thicker ventricle may be a greater challenge. It seems that lesion depth, among other parameters, is related to waveform and amplitude largely depends on vectoring; i.e. bipolar PFA may be less efficient to create deep lesions compared with monopolar delivery.
As mentioned above, although PFA is considered non-thermal and selective, none of this is absolute and general. Electroporation is used in bacterial inactivation, and also, bacterial transformation by introducing genes of interest, similar cells, and tissues of all origins can be effectively electroporated—with no exception. Although in vivo tissue selectivity seems convincing, there is little evidence available supporting cell electroporation selectivity in vitro., Energy-specific mechanisms regarding lesion formation are summarized in Table 1. Side effects relevant in thermal energies may not be relevant for PFA (if kept non-thermal), but other unknown side effects may emerge. Although coronary artery spasm came as a surprise,, the effects of electric pulses on blood, perfusion, and vasculature are known. Similar to cardiac pacing, electric pulses depolarize neurons—used for many years in electrical stimulation; it should be expected that the cardiac autonomic nervous system (in particular N. vagus) will be stimulated and may be responsible for bradycardia observed after PFA. Reversible stunning of phrenic nerves was studied in vivo and observed in clinic cases., Of note, known side effects can be worsened with more aggressive waveforms, and others may still need to be discovered.
Pulsed field ablation currently has a major drawback as intracardiac electrograms disappear virtually with the first pulse delivery. The traditional guiding of intracardiac electrocardiogram (ECG) signals in cardiac electrophysiology thus appears less important. It was shown before that electroporation increases membrane permeability for ions (and other molecules), which in other words ‘increases membrane conductivity’. Increased membrane conductivity alone in excitable cells can cause membrane depolarization and spontaneous contractions and trigger action potentials or can render cells unexcitable for a period of time., Upon the resealing of the membrane and regain of membrane selective permeability, the cells again behave ‘normally’. This is a clear consequence and result of reversible electroporation and has been shown theoretically by using a simple Hodgkin–Huxley-type model, which was shown both in vitro on a single cell and monolayer of cells mimicking cardiac tissue. For easier understanding, given a focal catheter, the lesion achieved by PFA will have a central irreversible zone surrounded by a rim of reversibly electroporated cells. The final resulting PFA lesion will be formed with delay (Figure 3). This also means that despite the impressive acute procedural efficacy, i.e. acute PVI reported in clinical studies, using PFA may not necessarily translate into superior long-term effects.
Role of contact force
Compared with thermal energy sources and based on available pre-clinical work, PFA lesion size/depth is much less dependent on contact force. This, together with the ability to create lesions even in the absence of direct contact with tissue (electrically conductive path is provided by relatively high conductivity of blood), provides the basis to create contiguous lesions even in trabeculated areas. In summary, PFA catheter-to-tissue contact is desired but appears to be less important compared with RFC energy.
Technologies
There are numerous PFA catheters, designs, and generators under development and investigated in pre-clinical and clinical studies (Tables 2 and 3 and Figure 4). Due to the clinical need, most catheters so far have been designed to treat atrial arrhythmias but ventricular ablation will be in focus in the near future. In brief, contemporary catheters can be divided into (i) circumferential PVI tools or (ii) point-by-point ablation devices. Within this latter group, novel-specific (large/intermediate footprint) devices or conventional RFC catheters deploying PFA as an alternative energy source may be used. Due to the rapid evolution in the field, not all PFA devices can be discussed but different categories of catheters and stages of evolution are shown (Figure 4).
Categories
Circumferential PVI devices (Table 2)
Focal point-by-point PFA devices (Table 3)
Large/intermediate footprint catheter, 3D electro-anatomical (EA) mapping system integration
Conventional RFC catheter, specific generator offering RFC and/or PFA
Farapulse™: pentaspline device
This ablation device consists of five splines carrying 20 electrodes and is available in two sizes (maximal diameter: 31 and 35 mm), which can be seamlessly changed from the so-called flower to a basket configuration (Farapulse™, Boston Scientific). Pulsed field ablation delivery consists of trains of bipolar, biphasic stimuli using electrical field strengths ranging from 1.8–2.0 kV with fixed 2.5 s duration. The device is able to pace and record cardiac electrograms for simple mapping and electrophysiological manoeuvres. The catheter is navigated through a steerable sheath (Faraflex, ID 13.6F), positioned at the PV antrum and stabilized via a guide wire anchored in a PV branch. After a pair of applications in flower and basket or olive configuration, the catheter is rotated to ensure a circumferential transmural lesion. Navigation and ablation at non-PV sites such as the left atrial (LA) roof or posterior wall are technically feasible. Device navigation is typically guided by fluoroscopy, intracardiac echo, or 3D mapping (Figure 5). Currently, contemporary 3D mapping systems offer only very basic catheter visualization. Sophisticated future 3D map features are under development, which may allow for precise device visualization, spline selection, catheter-to-tissue contact assessment, and PFA lesion simulation (Figure 5D). The Farapulse™ device is Ce marked and has recently received FDA approval.
Varipulse: variable loop circular catheter
This PFA system represents an evolution of the former circular irrigated RFC ablation platform. The VARIPULSE™ catheter (8.5F bidirectional, Biosense Webster) is sizeable from 25–35 mm diameter, fully integrated into a 3D mapping system and incorporates proprietary waveform technology to deliver short-duration, high-voltage bipolar biphasic pulses. The catheter is combined with the Trupulse generator. Each pulse is delivered as a square wave with positive and negative phases. Briefly, PFA is applied in a bipolar configuration with an energy of 1.8 kV. The trains are microsecond-long biphasic pulses resulting in a total application duration of ≈250 ms. The catheter is rotated to ensure overlapping lesions. The generator can also be programmed to deliver energy to specific electrode pairs and to adjust energy delivery based on the clinical needs. The device has obtained Ce mark.
Pulse select: fixed-sized multi-polar circular device
This PVI PFA system represents an evolution of the former circular RFC ablation platform (Medtronic Pulse Select™). The 9F catheter is navigated over the wire through a 10F steerable sheath. It consists of a circular array (25 mm diameter, 20° tilt) including nine gold electrodes. The novel PFA generator delivers biphasic, bipolar waveforms (1.5 kV, on the order of milliseconds per delivery). The catheter is rotated to ensure overlapping lesions. All deliveries are automatically gated to the ventricular refractory period.
In principle, impedance-based visualization of this device within a 3D EA mapping system is feasible, but there is no sophisticated image integration available at this point in time. This device is Ce marked and is the first that has recently received FDA approval.
Globe PF™: spherical mapping and ablation device (Kardium)
The Globe™ Pulsed Field Mapping and Ablation System (Kardium Inc.) is a 30 mm array consisting of 16 ribs with a total of 122 gold-plated electrodes that can map intracardiac electrograms, pace, and measure tissue contact and temperature. The latest generation of this spherical catheter also allows for PFA delivery. Up to 64 electrodes can be selected to customize the PFA ablation path (proprietary waveforms, biphasic, bipolar, single train, biphasic, and bipolar delivery). The catheter (16F) is inserted into the LA through a specifically designed bidirectionally deflectable sheath. In the LA, the coiled ribs fan out to form a spherical array. The pulsed field (PF) generator waveform uses propriety pulse parameters and is not modifiable by the operator (1.7 kV, microsecond PF pulses per application). The system allows to create a 3D electroanatomic reconstruction including voltage, activation, and contact mapping. This device is currently under clinical investigation.
Combined pulsed field ablation and ultralow cooling
This platform has been originally designed for ultralow cooling and enables both single-shot PVI and focal ablation (Adagio iCLAS™). The catheter (8.5F, 20 poles, central lumen, and stylet) requires a steerable sheath. The combination of PFA and cryothermal ablation may lead to a favourable thermal profile, reduced muscle contractions, and microbubbles while extending lesion. A prospective clinical study (parallel trial) is currently randomizing persistent AF patients in Canada and Europe [PFA vs. PFA and cryothermal ablation: PVI + posterior wall isolation and cavo-tricuspid isthmus (CTI) ablation according to the operator]. No clinical data have been published at this point in time.
Focal pulsed field ablation devices
This family of ablation devices combines focal point-by-point large/intermediate footprint PFA integrated into novel or contemporary 3D EA mapping systems to deploy individualized lesions based on strategies beyond empiric ablation. Most of these PFA-only catheters are under development or clinical evaluation lacking large-scale human experience (Table 3).
AFFERA™–Sphere 9™ toggling pulsed field ablation and radiofrequency current ablation integrated into a 3D mapping system
This novel ablation and mapping system uses a 9 mm lattice tip catheter (Sphere 9™) covered with nine mini electrodes, which is fully integrated into a high-resolution 3D mapping system (AFFERA™, Medtronic). All electrodes record local temperature and electrograms allowing for activation mapping. The catheter was originally designed to deploy high-power temperature-controlled radiofrequency ablation resulting in lower current density due to the large footprint lattice design. The generator allows for toggling between monopolar PFA from the catheter tip to indifferent electrodes and irrigated radiofrequency ablation. Lesion depth can be increased with repetitive PFA applications. Therefore, this device allows for point-by-point PVI as well as individualized ablation. Both energy modalities can be adjusted according to the physician's preference. Initial human experience indicates safe and durable ablation on the atrial level. More data regarding workflow considerations such as general anaesthesia or deep sedation are needed. This device is Ce marked but not FDA approved.
Pulsed field ablation generator connected to conventional radiofrequency current catheters
CENTAURI represents the first platform technology (Galvanize Therapeutics, San Carlos, CA, USA) enabling PFA delivery via commercially available established focal RFC ablation catheters. This type of PFA combines contemporary contact force RFC catheters and 3D EA mapping systems with a novel PFA generator. The generator delivers biphasic, monopolar PF energy at three selectable energy settings (19, 22, and 25 A) through the tip electrode of the ablation catheter. The system permits connectivity of compatible focal ablation catheters and their mapping systems that synchronizes PFA delivery to the R wave. Initial clinical data indicate feasibility and safety.,
The device is Ce marked for the treatment of atrial arrhythmias.
In addition, clinical studies are currently investigating the safety and efficacy of conventional contact force RFC catheters using novel dual energy generators toggling between PFA and RFC ablation.
Available pulsed field ablation technology
Numerous different ablation devices for PFA have been designed and are currently under clinical investigation. After recent approval, rapid adoption has been observed in the real world (Figure 4).
Performance
The above-mentioned technologies were primarily designed to achieve durable PVI. Hence, the rate of acute and durable PVI may be considered a marker of performance.
In pre-marketing studies, the Farapulse™ pentaspline catheter was evaluated for performance in protocol-mandated re-mapping studies. Acute PVI was achieved in 475/475 PVs (100%), and PVI durability was noted to be 96 and 84.8% on per PV and a per patient level, respectively.
More re-mapping data stem from clinically indicated repeat procedures in patients with symptomatic atrial tachycardia (AT)/AF recurrences. In a single-centre experience, durable PVI was reported to be 90.9% showing a significant difference in favour of the 31 mm pentaspline catheter as opposed to the 35 mm device. Multi-centre experience from the EU-PORIA registry and the ADVENT study found PVI durability in 72 and 64% of all PVs, respectively., It appears that the anterior aspect of the lateral PVs is prone to recovered LA to PV conduction in particular when the 35 mm pentaspline catheter was used. Future studies may further investigate alternative dosing regimens as well as ways to improve catheter-to-tissue contact.
Accordingly, re-mapping data after PVI with the variable loop PFA catheter showed durable PVI rates between 10 and 30.8% depending on the PFA waveform. In the PULSED-AF study, 41/300 patients underwent a repeat ablation but PVI durability data have not been disclosed.
Protocol-mandated re-mapping was also performed in a study using a PFA system employing commercially available contact force tip catheters. After having optimized the workflow, per PV durable isolation rates ranged from 84 to 92%.
Similarly, a novel lattice tip catheter that allows for toggling between PFA and thermal ablation was evaluated for its performance. During protocol-mandated re-mapping, a 97% durable PVI rate was noticed using the optimized waveform.
Both the focal and the lattice tip catheter enable the operator to apply linear ablation concepts. With the latter, acute bidirectional conduction block was achieved in all patients at the LA roof, the mitral isthmus, and the CTI mostly using both energy modalities. For the linear atrial lesions, the overall durability for the LA roof, mitral isthmus, and CTI lines was 82, 68, and 87%, respectively.
Most recently, it was reported that repeat PVI and LA anterior line ablation can be carried out with a standard tip ablation catheter using PFA. Bidirectional conduction block was achievable in 84% of attempts.
Safety
In first in-human studies, excellent procedural safety was reported. Serious adverse events occurred in 0–2.5% of patients. In particular, no energy-specific complications such as thermal oesophageal injury or phrenic nerve palsy were reported.
A similar favourable safety profile (2.1% procedural complications) was observed in the ADVENT trial. However, it should be noted that one patient died from a pericardial tamponade.
After commercialization, the MANIFEST PF survey was the first study to report complication rates for the pentaspline PFA catheter on a centre level. Implementing workflow changes and after having passed initial learning operator curves, the rate of procedure-related complications decreased (MANIFEST 17k) substantially. Pericardial tamponade and stroke were reported to be as low as 0.36 and 0.12%, respectively.
In an attempt to determine the thromboembolic risk profile of various catheters, cerebral magnetic resonance imaging (MRI) was performed in asymptomatic patients. The rates of silent cerebral lesions (SCLs) varied from 3 to 19% for the pentaspline catheter.,,, In the Pulsed AF trial, 45 patients underwent pre-procedural and post-procedural MRI, and SCLs were detected 9% of patients after the ablation. In the Inspire study, the SCL rate was 67 and 12% before and after workflow optimization.
Similarly, very rarely energy-specific complications such as coronary spasm or ablation-induced haemolysis were described. The latter seems to be correlated to the number of PFA applications, and 70 applications per procedure have been identified as a critical threshold for clinically relevant haemolysis. As a consequence, acute kidney injury was incidentally noted, which can be mitigated by infusion of large amounts of normal saline., Still, in the light of rapid adoption of novel PFA devices, more data are required to understand the role of this complication in the real world. For sure, more PFA dosing data are needed and ablation should be carefully guided by the clinical need.
It seems that the occurrence of coronary spasm is closely related to the distance between the ablation catheter and vessel. In a systematic study, coronary angiograms were performed during PFA applications at various locations. While no coronary spasm was observed during ablation at the PVs, ablation at the CTI led to coronary spasm in all cases. Coronary spasm could be resolved by intracoronary application of nitroglycerin and could also be prevented by intravenous drug application.
Similarly, ablation at the mitral isthmus may be associated with spasm of the circumflex artery in a considerable number of cases and an inferior ablation line may help to avoid this complication. In another observational study, mitral isthmus ablation with the pentaspline PFA catheter led to a high acute success rate and was associated with 4.4% non-fatal coronary spasms. To date, no relevant PV stenosis has been reported. Recently, a sub-analysis of the ADVENT trial was published without any evidence for PV stenosis in PFA-treated patients.
Most importantly, the risk for thermal oesophageal injury seems to be eliminated. In several studies, oesophageal temperature monitoring and post-procedural endoscopy revealed no evidence for significant heating or mucosal injury., Moreover, selective sparing of oesophageal tissue by PFA ablation at the posterior LA wall was elegantly demonstrated using MRI.
Efficacy
Several studies reported freedom from atrial arrhythmia after PFA ablation. When comparing outcomes between different PFA ablation devices, it has to be noted that follow-up strategies (i.e. intensity) may vary substantially. In the ADVENT and Pulsed AF pivotal trials, weekly trans-telephonic ECG submissions and 72 h Holter at varying time points were performed. Freedom from arrhythmia was estimated to be 73.3 and 66.2% in ADVANTAGE and Pulsed AF, after 12 months, respectively.
In real-life clinical registries with less intense rhythm, monitoring rates of freedom from arrhythmia were 80–82 and 66–72% for patients with paroxysmal and persistent AF, respectively, which are in line with thermal ablation. Of note, more real-life long-term outcome data are needed.
Impact on the autonomic nervous system
Due to its relative tissue selectivity, PFA may not permanently affect the cardiac autonomic nervous system. In fact, in an elegant study, extracardiac vagal stimulation was used to compare the effects of PFA and irrigated RFC-guided PVI on the sinus node and atrioventricular node, respectively. It showed that RFC ablation led to a significant attenuation of the response to extracardiac vagal stimulation. Oppositely, this response was much less common after PFA ablation and effects rapidly returned to baseline. The authors concluded that cardiac vagal response is preserved in patients treated with PFA. This observation is supported by further studies analysing serum markers of neuronal damage after PFA and cryothermal ablation., Both studies found a significantly lower release of S100B after PFA ablation.
However, efforts are made to design PFA devices that specifically electroporate cardiac ganglion cells to modify the cardiac autonomic nervous system with minimal harm to cardiac myocytes. First in-human data were recently published providing promising data on the feasibility and safety of ganglionic plexi ablation during cardiac surgery.
Future clinical implications
The integration of PFA catheters with available 3D mapping systems is expected to become reality in the near future. It may allow more precise PFA and thereby reducing the risk of gaps. So far, only AFFERA™ and very recently Varipulse™ have been offering 3D mapping capabilities. The Farapulse™ catheter integration into a specific 3D mapping system is under development and expected to be available soon (Figure 5D). Specific software may provide information on contact, spline labelling, and potentially predicting lesion extent, transmurality, and durability.
AFFERA™ and Sphere9™ use a very modest but detectable increase in electrode temperature during deliveries as a surrogate for tissue contact. Presently, numerous treatment planning software have been developed for IRE and electrochemotherapy (i.e. combination of electroporation and chemotherapy) in oncology. It is currently unclear whether meticulous refinement using advanced mathematical methodologies and computing is needed in catheter ablation of arrhythmias. One could argue that the safety profile of PFA allows for maximizing efficacy without compromising safety. Tailored strategies could be advantageous, particularly if accounting for tissue information such as thickness or fatty/fibrotic infiltration. Fat, by acting as an insulator for electricity, is a major limiting factor for RFC ablation, whether this may be less relevant for PFA needs to be assessed.
Developments in atrial fibrillation ablation
It should be first acknowledged that PVI is likely to be the only needed strategy for the vast majority of paroxysmal AF patients. Initial clinical PFA experience suggests high rates of durable PVI and less operator dependency. This may contribute to future increased and predictable success rates.,, Ablation strategies in persistent AF are much less mature, with no consensus beyond PVI. It remains unclear whether one strategy could fit to all clinical scenarios. Pulsed field ablation may help to understand whether AF recurrences were due to insufficient mapping or ablation. Importantly, as PFA is easy to deploy and powerful, the best ablation strategy is the one restoring sinus rhythm at the least tissue cost.
The Farawave™ catheter has been designed for PVI and is not optimized for linear lesion delivery. Interestingly, it performs technically well in posterior LA wall ablation. Reconnections through that posterior wall area are rare when compared with RFC.
The remarkable absence of oesophageal lesions with this system allows to deliver the lesion in a very low position, where the septo-pulmonary bundle is no longer an issue. It should, however, be noted that focal catheters such as Sphere 9™ may be more suitable for tailored linear lesions. This is probably even more true for the mitral isthmus line. The durability of the Farapulse™ linear block is suboptimal, even if combined with Marshall vein alcohol injection. In addition, post-mortem studies have unveiled diverse configurations of myocardial tissue surrounding the CS and/or Marshall bundle, forming endocardial–epicardial bridges—surrounded by fat—across the mitral isthmus line and connecting with the LA.
Providing safer, faster, and possibly more durable lesions, PFA may offer an opportunity in the future to explore major questions in AF ablation:
Pulmonary vein isolation responder vs. PVI non-responder in AF ablation
What is the real rate of PVI responder in paroxysmal and persistent AF?
Pulmonary vein isolation non-responder
What is the best ablation strategy?
Linear lesions: more safety and efficacy data, particularly for the mitral isthmus is needed (e.g. focal PFA within the CS)
What is the role of tailored ablation in the real world?
What is the role of deep sedation vs. general anaesthesia?
Pulsed field ablation in trabeculated areas and right atrium
In the context of atrial ablation some atypical re-entries or focal targets include trabeculated areas. They are often difficult to ablate using RFC, most likely because of their complex 3D topography. Pulsed field ablation has the advantage of treating similar pouches and crests, as it affects tissue included in the treatment volume. It, therefore, offers a theoretical advantage, but this has to be confirmed in clinical practice. If ablation close to the phrenic nerve was required, PFA may become the preferred energy source. Using PFA for CTI is effective albeit with documented instances of adverse events like coronary artery spasm., Notably, proactive administration of intracoronary/intravenous nitroglycerin has been shown to significantly attenuate vasospasm, emphasizing a potential prophylactic measure. In summary, PFA within the right atrium offers advantages but necessitates a cautious appraisal of potential risks, encompassing coronary artery spasms and disturbances of the sinus node and/or conduction system.
Pulsed field ablation for VT ablation
Although PFA has proceeded to randomized trials in patients with AF, there are still sparse data on ventricular ablation with PFA, especially in infarcted myocardium. Dosing remains unclear and has to be studied for different devices and anatomic locations. At first sight, the benefits may be less for ventricular lesions where collateral damages are not so frequent. Radiofrequency current, particularly with AFFERA™, is producing lesions as deep as 12 mm, which would be critical for septal or mid/epicardial substrate, particularly when wall thickness is preserved or increased. Obviously, it will require more in-depth studies. Another aspect, probably not so obvious but really important, might be the ability of PFA vs. RFC to effectively ablate despite of scar tissue with fibrosis and fat infiltration. Radiofrequency current within heterogeneous scar tissue results in non-uniform tissue injury and unpredictable lesions regarding morphology and dimensions, possibly due to lower tissue temperatures. Demonstrating PFA efficacy in scar tissue will be key, as well as defining and refining dosing possibly based on image integration (Figure 6). Data on target substrate composition including tissue thickness and the presence of fibrosis and fat could impact dosing/number of PF deliveries. Repeated PFA application is the only way to increase both lesion depth and width with current systems. One interesting catheter innovation using PFA platforms represents the concept of focal large footprint/intermediate-sized ablation electrodes. Irrespective of the energy used, those ablation electrodes carry the potential to transect most post-myocardial infarction VT isthmuses with only one location, thereby reducing the risk for non-contiguous lesions and reconnections (Figure 6). Using those platforms could lead to more predictable and simplified post-myocardial infarction VT ablation.
The potential clinical value of reversible electroporation
When low-intensity PF was applied in vivo to porcine atrial tissue, local electrogram amplitude immediately vanished but gradually recovered over several minutes, underscoring the reversible nature of these very short pulses. The application of reversible pulses resulted in transient PR prolongation and dose-dependent atrioventricular block, indicating their capacity to transiently impair electrical conduction. Notably, if low-intensity PFA was deployed in critical sites, reproducible termination of atrioventricular nodal re-entrant tachycardia and atypical atrial flutter was demonstrated. This should also be kept in mind when IRE is desired.
Limitation
Unresolved issues such as optimal energy settings, catheter stability, and the learning curve associated with this novel technology warrant further investigation. Long-term follow-up data and large-scale randomized controlled trials will be crucial for establishing the role of PFA in the real world.
Conclusions
Pulsed field ablation represents a potentially disruptive technology in the field of cardiac electrophysiology, offering a potential paradigm shift in the treatment of arrhythmias. The accumulating body of evidence supports its efficacy and safety, positioning PFA as a more than just a viable alternative to traditional thermal ablation. However, more research and clinical experience is needed.
References 101
- 1.
McDonagh TA, Metra M, Adamo M, Gardner RS, Baumbach A, Böhm M et al 2021 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur Heart J2021;42:3599–726.
Not in our library10.1093/eurheartj/ehab368
- 2.
Kuck K-H, Brugada J, Fürnkranz A, Metzner A, Ouyang F, Chun KRJ et al Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med2016;374:2235–45.
Not in our library10.1056/nejmoa1602014
- 3.
Bordignon S, Chun KJ, Gunawardene M, Fuerkranz A, Urban V, Schulte-Hahn B et al Comparison of balloon catheter ablation technologies for pulmonary vein isolation: the Laser versus cryo study. J Cardiovasc Electrophysiol2013;24:987–94.
Not in our library10.1111/jce.12192
- 4.
Bordignon S, Chen S, Bologna F, Thohoku S, Urbanek L, Willems F et al Optimizing cryoballoon pulmonary vein isolation: lessons from >1000 procedures—the Frankfurt approach. Europace2021;23:868–77.
Not in our library10.1093/europace/euaa406
- 5.
Chun JKR, Bordignon S, Last J, Mayer L, Tohoku S, Zanchi S et al Cryoballoon versus laserballoon: insights from the first prospective randomized balloon trial in catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol2021;14:e009294.
Not in our library10.1161/circep.120.009294
- 6.
Boersma L, Andrade JG, Betts T, Duytschaever M, Pürerfellner H, Santoro F et al Progress in atrial fibrillation ablation during 25 years of Europace journal. Europace2023;25:euad244.
Not in our library10.1093/europace/euad244
- 7.
Bordignon S, My I, Tohoku S, Rillig A, Schaack D, Chen S et al Efficacy and safety in patients treated with a novel radiofrequency balloon: a two centres experience from the AURORA collaboration. Europace2023;25:euad106.
Not in our library10.1093/europace/euad106
- 8.
Furnkranz A, Bordignon S, Schmidt B, Perrotta L, Dugo D, De Lazzari M et al Incidence and characteristics of phrenic nerve palsy following pulmonary vein isolation with the second-generation as compared with the first-generation cryoballoon in 360 consecutive patients. Europace2015;17:574–8.
Not in our library10.1093/europace/euu320
- 9.
Duytschaever M, De Potter T, Grimaldi M, Anic A, Vijgen J, Neuzil P et al Paroxysmal atrial fibrillation ablation using a novel variable-loop biphasic pulsed field ablation catheter integrated with a 3-dimensional mapping system: 1-year outcomes of the multicenter inspIRE study. Circ Arrhythm Electrophysiol2023;16:e011780.
Not in our library10.1161/circep.122.011780
- 10.
Gallagher JJ, Svenson RH, Kasell JH, German LD, Bardy GH, Broughton A et al Catheter technique for closed-chest ablation of the atrioventricular conduction system. N Engl J Med1982;306:194–200.
Not in our library10.1056/nejm198201283060402
- 11.
Wittkampf FHM, van Driel VJ, van Wessel H, Neven KGEJ, Gründeman PF, Vink A et al Myocardial lesion depth with circular electroporation ablation. Circ Arrhythm Electrophysiol2012;5:581–6.
Not in our library10.1161/circep.111.970079
- 12.
van Driel VJHM, Neven K, van Wessel H, Vink A, Doevendans PAFM, Wittkampf FHM. Low vulnerability of the right phrenic nerve to electroporation ablation. Heart Rhythm2015;12:1838–44.
Not in our library10.1016/j.hrthm.2015.05.012
- 13.
van Driel VJHM, Neven KGEJ, van Wessel H, du Pré BC, Vink A, Doevendans PAFM et al Pulmonary vein stenosis after catheter ablation: electroporation versus radiofrequency. Circ Arrhythm Electrophysiol2014;7:734–8.
Not in our library10.1161/circep.113.001111
- 14.
Reddy VY, Gerstenfeld EP, Natale A, Whang W, Cuoco FA, Patel C et al Pulsed field or conventional thermal ablation for paroxysmal atrial fibrillation. N Engl J Med2023;389:1660–71.
Not in our library10.1056/nejmoa2307291
- 15.
Reddy VY, Peichl P, Anter E, Rackauskas G, Petru J, Funasako M et al A focal ablation catheter toggling between radiofrequency and pulsed field energy to treat atrial fibrillation. JACC Clin Electrophysiol2023;9:1786–801.
Not in our library10.1016/j.jacep.2023.04.002
- 16.
Schmidt B, Bordignon S, Tohoku S, Chen S, Bologna F, Urbanek L et al 5S study: safe and simple single shot pulmonary vein isolation with pulsed field ablation using sedation. Circ Arrhythm Electrophysiol2022;15:e010817.
Not in our library10.1161/circep.121.010817
- 17.
Reddy VY, Dukkipati SR, Neuzil P, Anic A, Petru J, Funasako M et al Pulsed field ablation of paroxysmal atrial fibrillation: 1-year outcomes of IMPULSE, PEFCAT, and PEFCAT II. JACC Clin Electrophysiol2021;7:614–27.
Not in our library10.1016/j.jacep.2021.02.014
- 18.
Lee RC. Cell injury by electric forces. Ann N Y Acad Sci2006;1066:85–91.
Not in our library10.1196/annals.1363.007
- 19.
Kotnik T, Rems L, Tarek M, Miklavčič D. Membrane electroporation and electropermeabilization: mechanisms and models. Annu Rev Biophys2019;48:63–91.
Not in our library10.1146/annurev-biophys-052118-115451
- 20.
Rems L, Rainot A, Wiczew D, Szulc N, Tarek M. Cellular excitability and ns-pulsed electric fields: potential involvement of lipid oxidation in the action potential activation. Bioelectrochemistry2024;155:108588.
Not in our library10.1016/j.bioelechem.2023.108588
- 21.
Geboers B, Scheffer HJ, Graybill PM, Ruarus AH, Nieuwenhuizen S, Puijk RS et al High-voltage electrical pulses in oncology: irreversible electroporation, electrochemotherapy, gene electrotransfer, electrofusion, and electroimmunotherapy. Radiology2020;295:254–72.
Not in our library10.1148/radiol.2020192190
- 22.
Yarmush ML, Golberg A, Serša G, Kotnik T, Miklavčič D. Electroporation-based technologies for medicine: principles, applications, and challenges. Annu Rev Biomed Eng2014;16:295–320.
Not in our library10.1146/annurev-bioeng-071813-104622
- 23.
Verma A, Asivatham SJ, Deneke T, Castellvi Q, Neal RE II. Primer on pulsed electrical field ablation: understanding the benefits and limitations. Circ Arrhythm Electrophysiol2021;14:e010086.
Not in our library10.1161/circep.121.010086
- 24.
Verma A, Haines DE, Boersma LV, Sood N, Natale A, Marchlinski FE et al Pulsed field ablation for the treatment of atrial fibrillation: PULSED AF pivotal trial. Circulation2023;147:1422–32.
Not in our library10.1161/circulationaha.123.063988
- 25.
Peng W, Polajžer T, Yao C, Miklavčič D. Dynamics of cell death due to electroporation using different pulse parameters as revealed by different viability assays. Ann Biomed Eng2024;52:22–35.
Not in our library10.1007/s10439-023-03309-8
- 26.
Batista Napotnik T, Polajžer T, Miklavčič D. Cell death due to electroporation—a review. Bioelectrochemistry2021;141:107871.
Not in our library10.1016/j.bioelechem.2021.107871
- 27.
Davalos RV, Mir LM, Rubinsky B. Tissue ablation with irreversible electroporation. Ann Biomed Eng2005;33:223–31.
Not in our library10.1007/s10439-005-8981-8
- 28.
Zimmermann U, Pilwat G, Riemann F. Preparation of erythrocyte ghosts by dielectric breakdown of the cell membrane. Biochim Biophys Acta1975;375:209–19.
Not in our library10.1016/0005-2736(75)90189-3
- 29.
Kinosita K Jr, Tsong TY. Voltage-induced pore formation and hemolysis of human erythrocytes. Biochim Biophys Acta1977;471:227–42.
Not in our library10.1016/0005-2736(77)90252-8
- 30.
Bradley CJ, Haines DE. Pulsed field ablation for pulmonary vein isolation in the treatment of atrial fibrillation. J Cardiovasc Electrophysiol2020;31:2136–47.
Not in our library10.1111/jce.14414
- 31.
Stewart MT, Haines DE, Miklavčič D, Kos B, Kirchhof N, Barka N et al Safety and chronic lesion characterization of pulsed field ablation in a Porcine model. J Cardiovasc Electrophysiol2021;32:958–69.
Not in our library10.1111/jce.14980
- 32.
Tan NY, Ladas TP, Christopoulos G, Sugrue AM, van Zyl M, Ladejobi AO et al Ventricular nanosecond pulsed electric field delivery using active fixation leads: a proof-of-concept preclinical study. J Interv Card Electrophysiol2022
Not in our library10.1007/s10840-022-01268-z
- 33.
Kotnik T, Frey W, Sack M, Haberl Meglič S, Peterka M, Miklavčič D. Electroporation-based applications in biotechnology. Trends Biotechnol2015;33:480–8.
Not in our library10.1016/j.tibtech.2015.06.002
- 34.
Casciola M, Keck D, Feaster TK, Blinova K. Human cardiomyocytes are more susceptible to irreversible electroporation by pulsed electric field than human esophageal cells. Physiol Rep2022;10:e15493.
Not in our library10.14814/phy2.15493
- 35.
Avazzadeh S, O'Brien B, Coffey K, O'Halloran M, Keane D, Quinlan LR. Establishing irreversible electroporation electric field potential threshold in a suspension in vitro model for cardiac and neuronal cells. J Clin Med2021;10:5443.
Not in our library10.3390/jcm10225443
- 36.
Gunawardene MA, Schaeffer BN, Jularic M, Eickholt C, Maurer T, Akbulak RÖ et al Coronary spasm during pulsed field ablation of the mitral isthmus line. JACC Clin Electrophysiol2021;7:1618–20.
Not in our library10.1016/j.jacep.2021.08.016
- 37.
Ekanem E, Reddy VY, Schmidt B, Reichlin T, Neven K, Metzner A et al Multi-national survey on the methods, efficacy, and safety on the post-approval clinical use of pulsed field ablation (MANIFEST-PF). Europace2022;24:1256–66.
Not in our library10.1093/europace/euac050
- 38.
Jarm T, Cemazar M, Miklavcic D, Sersa G. Antivascular effects of electrochemotherapy: implications in treatment of bleeding metastases. Expert Rev Anticancer Ther2010;10:729–46.
Not in our library10.1586/era.10.43
- 39.
Gehl J, Skovsgaard T, Mir LM. Vascular reactions to in vivo electroporation: characterization and consequences for drug and gene delivery. Biochim Biophys Acta2002;1569:51–8.
Not in our library10.1016/s0304-4165(01)00233-1
- 40.
Acerbo E, Safieddine S, Weber P, Botzanowski B, Missey F, Carrère M et al Non-thermal electroporation ablation of epileptogenic zones stops seizures in mice while providing reduced vascular damage and accelerated tissue recovery. Front Behav Neurosci2021;15:774999.
Not in our library10.3389/fnbeh.2021.774999
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.