Plasma lactate measurement as an example of encountered gaps between routine clinical laboratory processes and manufactures' sample-handling instructions

Ibrahim A. Hashim, Mishkat Mohamed, Aileen Cox and 2 more

Practical Laboratory Medicine · 2018

doi:10.1016/j.plabm.2018.e00109 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

OBJECTIVES: Deviation from manufacturers' pre-analytical sample handling recommendations necessitates extensive validation studies. This report uses plasma lactate testing, where a recommended 15 min room temperature sample handling limit cannot be met by the clinical laboratory, as an example for studies to bridge the gap with practice. DESIGN AND METHODS: Triplicate blood samples were collected from patients (n = 51) with lactate requests by clinicians and from normal volunteers (n = 50). One tube was transported on ice (4 °C), the others were maintained at room temperature (23 °C). Tubes stored at 4 °C were processed at 30 min from collection. Tubes stored at 23 °C were processed at 15 and at 30 min from collection. Lactate levels were measured using Roche Diagnostics Cobas 6000® analyzer. RESULTS: Lactate levels in normal subjects ranged from 0.6 to 3.1 mmol/L (median 1.1). Patient lactate levels ranged from 0.8 to 26.3 mmol/L (median 2.2). Bias in lactate levels following extended storage of samples from both normal subjects and patients ranged from - 1.3 to2.2 and from - 1.0-1.0 mmol/L when stored for 30 min at 23 °C or at 4 °C, respectively. The bias between lactate levels at 30 min at 23 °C and 4 °C was - 1.2 to - 0.5 mmol/L for both populations. Although the bias was not statistically significant for all variables, a clinically significant (>0.2 mmol/L) bias was observed in 28% of normal and 7.0% of patient samples. CONCLUSION: Extending the pre-analytical time to 30 min at 23 °C did not significantly impact clinical utility of lactate measurement in our patient population.

Abstract licence: CC BY-NC

References20

    20 more references, whose wording the paper’s licence keeps private.

    Reads

    Read by people on this page

    None yet.

    Checked or read by AI through naulon

    None yet.

    A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

    Are you an author?

    Is this your paper?

    Claim it to add your note and be paid when agents cite it. Free.