Effect of surrounding temperature on long-term drift of quartz Bourdon-type pressure gauges kept at constant pressure
Hiroaki Kajikawa, Hideaki Iizumi, Tokihiko Kobata
Measurement: Sensors · 2021
doi:10.1016/j.measen.2021.100199 (opens in a new tab)
We email you when a new notice or a question about this paper appears.
Abstract
We have been investigating the long-term drift of pressure gauges under constant pressure application with the aim to improve the precision of pressure monitoring at seafloor. This paper focuses on the effect of surrounding temperature on the long-term drift of quartz Bourdon-type pressure gauges. The pressure of 100 MPa was continuously applied to the test gauge for several months at 3 °C and 23 °C. The calibration results at 100 MPa and those at atmospheric pressure were traced. The long-term drift at 3 °C was much smaller than that at 23 °C. In addition, the variations in the calibration results at 100 MPa and 0 MPa were different at 3 °C, whereas the two variations were identical at 23 °C. The effect of temperature on the drift characteristics should be considered for the appropriate selection of gauges and procedure of the actual pressure monitoring at seafloor.
Abstract licence: CC BY-NC
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.