Skip to content
naulon

Not checkedNot checked for notices yet.

Reliability enhancement of non-contact fever screening technology (thermography, etc.) for quarantine inspection

Kuniaki Amemiya, Yuhei Shimizu, Naohiko Sasajima and 2 more

Measurement: Sensors · 2021

doi:10.1016/j.measen.2021.100160 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

We have been developing the technology for reliable non-contact body temperature measurement, which contributes to the prevention of new infectious diseases. As a part of this study, we are developing a highly accurate on-site temperature reference with a planar blackbody (reference infrared radiator) that can correct fever-screening thermography. We have established a process to manufacture a planar blackbody material offering a high emissivity of >0.99, which would be less affected by surroundings (background infrared radiation, etc.) when correcting and evaluating thermographic cameras. The target temperature setting accuracy of the planar blackbody radiator would be < 0.2 °C. Thereby the error factors of fever-screening thermography including device performance (stability, response uniformity, etc.), target parameters (emissivity, distance, etc.), and ambient environment (temperature, etc.) would be inspected and corrected in more detail.

Abstract licence: CC BY-NC-ND

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.