Skip to content
naulon

Not checkedNot checked for notices yet.

Calibrated X-ray computed tomography for testing micro-scale pore defect in metallic powder particles for additive manufacturing

Shihua Wang, Baoxi Xu

Measurement: Sensors · 2021

doi:10.1016/j.measen.2021.100110 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

In this paper, a non-destructive X-ray computed tomography (XCT) was introduced to measure hidden pore defects inside metallic powder used in additive manufacturing. The internally hidden pores were found to have sizes ranging from 10 μm to 30 μm within those powder particles (size: up to 100 μm). Corresponding porosity in four (4) different metallic powders were determined to be about 0.11% for Inconel 718, 0.02% for virgin Inconel 625, 0.31% for Recycled Inconel 625 and 0.03% for Stainless steel 316 powder materials. It demonstrated that the proposed XCT is applicable to determine the porosity of the powder particles.

Abstract licence: CC BY-NC

References8

    8 more references, whose wording the paper’s licence keeps private.

    Reads

    Read by people on this page

    None yet.

    Checked or read by AI through naulon

    None yet.

    A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

    Are you an author?

    Is this your paper?

    Claim it to add your note and be paid when agents cite it. Free.