Sweat conductivity and coulometric quantitative test in neonatal cystic fibrosis screening

Mouseline Torquato Domingos, Neiva Isabel Rodrigues Magdalena, Mônica Nunes Lima Cat and 2 more

Jornal de Pediatria (Versão em Português) · 2015

doi:10.1016/j.jpedp.2015.07.017 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

Objective: To compare the results obtained with the sweat test using the conductivity method and coulometric measurement of sweat chloride in newborns (NBs) with suspected cystic fibrosis (CF) in the neonatal screening program. Methods: The sweat test was performed simultaneously by both methods in children with and without CF. The cutoff values to confirm CF were > 50 mmol/L in the conductivity and > 60 mmol/L in the coulometric test. Results: There were 444 infants without CF (185 males, 234 females, and 24 unreported) submitted to the sweat test through conductivity and coulometric measurement simultaneously, obtaining median results of 32 mmol/L and 12 mmol/L, respectively. For 90 infants with CF, the median values of conductivity and coulometric measurement were 108 mmol/L and 97 mmol/L, respectively. The false positive rate for conductivity was 16.7%, and was higher than 50 mmol/L in all patients with CF, which gives this method a sensitivity of 100% (95% CI: 93.8 to 97.8), specificity of 96.2% (95% CI: 93.8 to 97.8), positive predictive value of 83.3% (95% CI: 74.4 to 91.1), negative predictive value of 100% (95% CI: 90.5 to 109.4), and 9.8% accuracy. The correlation between the methods was r = 0.97 (p > 0.001). The best suggested cutoff value was 69.0 mmol/L, with a kappa coefficient = 0.89. Conclusion: The conductivity test showed excellent correlation with the quantitative coulometric test, high sensitivity and specificity, and can be used in the diagnosis of CF in children detected through newborn screening.

Abstract licence: CC BY-NC-ND

References21

    21 more references, whose wording the paper’s licence keeps private.

    Reads

    Read by people on this page

    None yet.

    Checked or read by AI through naulon

    None yet.

    A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

    Are you an author?

    Is this your paper?

    Claim it to add your note and be paid when agents cite it. Free.