A ‘post-close’ Technique for Femoral Hemostasis After Percutaneous EVAR Using an Ultra-low-profile Endoprosthesis System

A. Chaudhuri

EJVES Short Reports · 2015

doi:10.1016/j.ejvssr.2015.10.001 (opens in a new tab)

We email you when a new notice or a question about this paper appears.

Abstract

: Introduction: We present a ‘post-close’ technique using the Angio-Seal VIP vascular closure device (VCD) after percutaneous endovascular aneurysm repair (p-EVAR) using an ultra-low-profile (ULP) device. Technique: Following percutaneous using an ULP device (here the Ovation Prime system, Trivascular, Eysins, Switzerland), contralateral femoral hemostasis is achieved by using an ‘undersized’ application of an 8F Angio-Seal VIP vascular closure device (VCD) for all punctures wherein upto 12F sheaths have been applied for access. On the ipsilateral side, following ‘double-wire’ preparation, synchronous deployment of an 8F and 6F Angio-Seal VIP VCDs achieves hemostasis. Procedural heparin is reversed as an adjunct, pressure dressings are applied and the patient kept flat for 2 hours. Discussion: Double-wire VCD deployments, and ‘post-close’ techniques have not been described in the context of femoral hemostasis after p-EVAR. This technique is easy to apply and saves time (and potentially cost) used to set up ‘pre-close’ devices, as is typically used these days. Keywords: Percutaneous EVAR, Post-close, Vascular closure devices

Abstract licence: CC BY-NC-ND

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.