Diffusion-weighted imaging of the breast—a consensus and mission statement from the EUSOBI International Breast Diffusion-Weighted Imaging working group
On behalf of the EUSOBI international Breast Diffusion-Weighted Imaging working group, Pascal Baltzer, Ritse M. Mann and 11 more
European Radiology · 2019 · CC BY
doi:10.1007/s00330-019-06510-3
Abstract
Electronic supplementary material. The online version of this article (10.1007/s00330-019-06510-3) contains supplementary material, which is available to authorized users.
Full text
From PubMed Central, U.S. National Library of Medicine, as current on Oct 17, 2026. Paragraph numbers are ours, so an answer can point at one.
Introduction
Diffusion-weighted imaging (DWI) is emerging as a key imaging technique to complement dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) of the breast. DWI can be used to distinguish between benign and malignant breast lesions, stratify in situ from invasive disease, and potentially predict the response to and monitor the effect of neoadjuvant treatment over time. Excellent reviews can be found in the literature; however, there is still no clear consensus within the literature on where and how breast DWI should be applied.
A major strength of DWI is that the apparent diffusion coefficient (ADC) can be derived from it, providing a quantitative measure of observed diffusion restriction. We use the term “restriction” because it is widely used in the breast DWI literature. However, the term refers specifically to water diffusion in enclosed spaces, whereas ADC is sensitive to multiple processes occurring in human tissue (genuine restricted intracellular motion, compartmental exchange due to cell membrane permeability, hindered extracellular motion, water bounding to macromolecules, blood microcirculation, etc.). Thus, the term “hindrance” is also used in literature to describe diffusion effects and might be a more accurate description. Advanced protocols and analyses (as discussed later in this statement) may be used to (partly) disambiguate these effects. However, the focus of this statement is on increasing the translation and standardization of DWI for breast cancer evaluation using the established ADC quantification.
A current challenge is the large variability in results, i.e., specificity, sensitivity and thresholds, reported for ADC in distinguishing between benign and malignant breast lesions. A less evident challenge is the inconsistent image quality due to different MRI system capabilities as well as equipment and imaging sequences/protocols contributing to a perceived limited usefulness of DWI in clinical practice. These challenges, coupled with the lack of prospectively validated thresholds for supporting diagnostic decisions, have prevented DWI from becoming an established measure that can be easily incorporated into the Breast Imaging Reporting and Data System (BI-RADS,). Nevertheless, DWI has been incorporated (albeit in a qualitative way) into the Prostate Imaging Reporting and Data System (PI-RADS).
In spite of the diversity in DWI protocols (e.g., the lack of standardization and the presence of artifacts), ADC estimation and interpretation methods across clinical sites, and composition of breast lesions in the studies, there is common agreement that DWI is sensitive to tissue microstructure and cellularity and provides quantitative information that can be used for lesion characterization. The improved lesion characterization can reduce the number of unnecessary biopsy recommendations, which has been further validated in a multicenter trial. The quantitative nature of ADC measures combined with relatively short acquisition times, typically in the order of 2–4 min but not exceeding 5 min, makes it an ideal imaging biomarker candidate. Consequently, there is great interest to improve the generalizability and reproducibility of breast DWI across institutions and imaging platforms. Moreover, although technically challenging, DWI protocol standardization between different systems and vendors has been achieved in the breast and other organs.
The International Breast DWI Working Group
To address the above-described issues and support standardized implementation of DWI, promote its clinical use, and facilitate its adoption as an integral part of breast MRI interpretation within BI-RADS, an International Breast DWI Working Group was established by the European Society of Breast Radiology (EUSOBI). Members of the working group were invited based upon proven expertise in breast MRI and/or breast DWI, representing 25 sites from 16 countries. Within the working group, a scientific committee was appointed. The full composition of the working group is given in Appendix 1.
The primary goals of the International Breast DWI working group are:To promote the integration of DWI into clinical practice by issuing consensus statements and initiate collaborative research where appropriateTo define standards and provide practical guidance for clinical application of DWITo develop a standardized and translatable multisite multivendor quality assurance protocol, especially for multisite research studiesTo find consensus on optimal methods for image processing/analysis, visualization, and interpretationTo work collaboratively with system vendors to improve breast DWI sequences
To promote the integration of DWI into clinical practice by issuing consensus statements and initiate collaborative research where appropriate
To define standards and provide practical guidance for clinical application of DWI
To develop a standardized and translatable multisite multivendor quality assurance protocol, especially for multisite research studies
To find consensus on optimal methods for image processing/analysis, visualization, and interpretation
To work collaboratively with system vendors to improve breast DWI sequences
Members of the working group were first invited to provide information on their protocols with details of acquisition and processing parameters. This revealed substantial heterogeneity of applied DWI protocols, even among experts. The members of the Scientific Committee were then surveyed with a list of points on basic requirements, acquisition parameters, hardware, quality control, analysis and reporting to define the minimum standards for breast DWI, along with technical recommendations to meet these standards.
In this statement, we report the requirements for breast DWI that obtained general consensual agreement. The items that reached more than 80% consensus are shown in the Supplementary Material. This general guidance should be used to provide consistency across clinical trials and in clinical practice.
The group acknowledges that there are many items that may influence both the acquisition and reporting of breast DWI that are not discussed in detail in this statement. Those items did not reach consensus at this stage, requiring further discussion and/or more extensive research. Some of these issues include field strength effects, diffusion gradient waveform dependence, image registration, eddy current distortion and compensation, and gradient nonlinearity correction. Each of these topics has received attention in the breast DWI literature (see below), providing a roadmap for their incorporation in a consensus process. In that spirit, further work is intended from subgroups of our working group to methodically address such items, especially regarding clinical implementation, standardization and quality control, and advanced diffusion MRI methods. The results of these efforts are to be presented in a second round of topical guidelines.
When to perform breast DWI?
Breast DWI may be part of any multiparametric breast MRI protocol independent of the clinical indication for MRI, as DWI improves the characterization of lesions detected by contrast-enhanced MRI irrespective of the indication. This includes breast MRI performed for pre-operative staging of known breast cancer (ipsi- and contralateral), monitoring neoadjuvant systemic therapy, evaluating carcinomas of unknown primary origin, resolving equivocal findings from other imaging modalities, and solving problem. For all these indications, breast DWI is considered an important addition to DCE-MRI to improve specificity, with the aim of reducing the number of recalls and biopsies of benign breast lesions. A more detailed description of current indications for breast MRI is found in existing guidelines from the EUSOBI, European Society of Breast Cancer Specialists, and American College of Radiology.
The working group highlights a special case: the application of breast DWI for breast cancer screening. There is not yet enough evidence to recommend for or against the inclusion of DWI in screening protocols. While DWI may improve lesion classification in this setting, its use should be balanced against the limited frequency of abnormal findings (because in screening, most exams are normal). For example, rather than performing DWI in all screening examinations, in order to prevent unnecessary biopsies, a multiparametric MRI including DWI could be performed as a secondary evaluation in women with ambiguous findings at the screening test.
Could breast DWI be used as a stand-alone test for breast cancer detection?
In general, even though breast DWI has high specificity for lesion characterization, the sensitivity of DCE-MRI still exceeds that of breast DWI, albeit the sensitivity of DWI alone may be equal to or even higher than that of commonly used screening techniques such as mammography and ultrasound. Thus, currently, its use for cancer detection as part of an unenhanced MRI examination requires further investigation and should be considered only in cases where DCE-MRI is not accessible or not appropriate. In particular, DWI may be a valuable alternative option in patients with contraindications to gadolinium-based contrast agents (such as patients with severe kidney dysfunction at risk of nephrogenic systemic fibrosis and patients with a previous acute reaction to a gadolinium-based contrast agent).
Is breast DWI helpful in the assessment of implants?
Breast DWI is not helpful in the assessment of breast implant integrity, which relies on other dedicated non-contrast techniques such as T2-weighted imaging and silicone specific sequences and thus is not needed for this purpose. It is, however, as useful as in any other indication for lesion classification in patients with implants.
What are the minimal technical requirements for breast DWI?
The working group found consensus on a minimal set of acquisition parameters to be met in clinical practice. Adherence to these minimal requirements should improve the comparison of ADC values from site to site, which is an important step towards the generalizability required to eventually incorporate ADC quantification into standardized guidelines (e.g., BI-RADS).
Which hardware should be used for breast DWI?
Breast DWI should be undertaken in a closed bore magnet with field a strength of 1.5 T or more. The gradient hardware should be capable of reaching a maximum gradient strength of at least 30 mT/m, and the use of a dedicated breast coil with at least four channels is strongly recommended. When possible, DWI should be performed before DCE-MRI, as the presence of contrast agents may physically reduce measured ADC values. It should be mentioned, however, that no significant effect has been shown on overall diagnostic performance when DWI is obtained after contrast administration as long as fat suppression using the STIR technique is avoided.
What sequence type and parameters are recommended for breast DWI?
The recommended minimal requirements for breast DWI are shown in Table 1. The working group acknowledges that for different MRI systems, different parameters might need to be adapted in order to obtain the best possible results. Therefore, Table 1 should be regarded as a guide rather than as a checklist for optimizing a DWI sequence. For instance, the axial orientation was motivated by anatomy and ease of interpretation more than technical optimization, which depends on the coil configuration. Specifying more parameters is beyond the scope of this paper but will be addressed at a later stage.
Single-shot or multi-shot echo-planar imaging (EPI) is regarded as baseline techniques for DWI acquisition. EPI has a wide breadth of applications and a correspondingly deep library of tools to address its shortcomings in image quality. Examples are nonlinearities in diffusion gradients, eddy current or magnetic field–induced image distortions, and motion. Gradient nonlinearities have been identified as a major source of inaccuracy for breast DWI, and prescriptions for correction have been offered. Eddy current and inhomogeneous field distortions are well-known issues for diffusion MRI with evolving solutions. Inhomogeneous field distortion correction and motion correction strategies have been applied to breast DWI. Multi-shot EPI, especially within the RESOLVE sequence framework has the potential to reduce susceptibility-induced geometric distortion and motion artifacts, at the expense of the acquisition time. However, corrective elements for breast DWI acquisitions are currently not sufficiently standardized to recommend a specific approach and will be addressed in future work of our working group. Nonetheless, the group made a recommendation on using an acceleration factor of 2, which reduces eddy current effects (and distortion). Vendor-based standard product pulse sequences with inline workflow elements (eddy current correction, gradient nonlinearity correction) may be used, but the optimal correction strategies are yet unknown.
Which b values should be chosen for breast DWI?
The choice of b values is critical. Due to the non-Gaussian nature of water diffusion in tissues (which results in a curvature of the DWI signal attenuation plot across b values), the ADC value is highly dependent on the choice of b values, with the ADC values getting smaller as larger b values are used (Fig. 1). Higher b values may increase the specificity of DWI and also lead to a decreased signal-to-noise ratio. A high b value of 800 s/mm2 was chosen by the group as a good compromise for standardization. This value can be established theoretically and is backed up by empirical evidence.
In how many directions should breast DWI be acquired?
The number of directions to acquire diffusion measurements from which the final ADC is calculated can vary from one direction to more than 6 directions for diffusion tensor imaging (DTI). Using several different diffusion directions allows calculation of an average ADC and reduces the influence of anisotropy in the tissue, thus improving generalizability. The “3 orthogonal directions” method (commonly available clinically and less time consuming than the 6 directions scheme required for DTI) provides an approximate isotropic weighting (a.k.a. pseudo trace-weighting contrast as an approximation to the true trace which can be obtained from the eigenvalue average derived from the full diffusion tensor). This 3 directions approach may also mitigate residual variations of b values across directions. However, variable eddy current interactions with the EPI readout train may result in different image distortions according to the diffusion-encoding direction, which, when averaged over the 3 directions, can confound image quality and reduce sensitivity to small lesions. Alternatively, acquisitions along only 1 direction but using gradient pulses simultaneously on 2 or 3 axes (sometimes called “diagonal”) allow high b values to be reached using shorter TEs, thus increasing SNR and reducing acquisition time compared with a sequential 3 direction acquisition. Diagonal diffusion gradients, however, would introduce variability in the presence of diffusion anisotropy known to exist in fibroglandular breast tissue. Given these mixed advantages, the compromise recommended is to utilize 3 orthogonal directions. This scheme is universally available and provides approximately isotropic contrast. However, we must emphasize that this approach can introduce non-negligible error and bias in the estimation of diffusion kurtosis imaging-derived indices of the breast and is thus not appropriate for this application.
How can high image quality and SNR be ensured?
The signal-to-noise ratio (SNR) of DWI is also important, especially when a high b value is used as ADC values may be largely underestimated (mimicking malignant tissues) when the SNR is low (Fig. 2). This requires that a sufficiently high SNR is achieved. One qualitative requirement in that vein is that fibroglandular tissue (FGT) (if present) should be well depicted on low b value images. Quantitatively, while the term “sufficient SNR” has yet to be defined in DWI of the breast, acquisition with multiple excitations for averaging, especially at b = 800 s/mm2, will be necessary to reach this goal. Unfortunately, the working group acknowledges that there are currently no universal methods to quantify SNR easily in a clinical setting, especially with the use of multichannel coils. In that setting, prescriptions exist for accurate SNR determination using difference methods for average noise assessment or the Kellman approach for spatially varying SNR, but they require additional acquisitions and advanced reconstruction that are not sufficiently ubiquitous to reach a consensus recommendation. Consequently, before implementing a new DWI sequence in routine clinical care, it is advisable to first test the sequence and assess image quality qualitatively. Ideally, quality control should be performed using dedicated calibrated phantoms with known ADC values, especially for multicenter studies (see below). Such phantoms will also allow the correct application of diffusion gradients (i.e., b values across axes) to be checked. In clinical practice, it remains important to check regularly whether the obtained diffusion images have a sufficient SNR and are reasonably free of structural artefacts that might hamper interpretation. While previous studies have addressed quality controls in DWI in general, one study investigated in detail factors influencing SNR and quantitative ADC measures in breast DWI. The results of this study corroborate the hypothesis that scanners should be adequately characterized in breast DWI. Quality control is also required after the maintenance of the MRI system and after each upgrade of hard- and/or software.
Apart from low SNR, a commonly encountered problem is insufficient fat suppression, which leads to ghosting and potentially to underestimation of ADC values. Spectral attenuated inversion recovery (SPAIR), which employs both T1 contrast and spectral selectivity, is recommended given a moderate preference in literature in comparison with short tau inversion recovery (STIR) and based on the consensus among the panel. However, if unsuccessful, STIR may be an acceptable alternative as this technique is independent from B0 field inhomogeneity.
While not listed in Table 1, diffusion time is a parameter that may significantly impact ADC values. Although it is generally hidden to users (gradient pulse time course is usually not reported), with conventional sequences, diffusion time is approximately equal to TE/2. The degree of diffusion restriction (or hindrance) decreases at short diffusion times (increasing the ADC) and it is important to be aware of this parameter as some vendors are now proposing stronger gradient hardware that allows shorter diffusion times. However, on most clinical MRI scanners, achievable diffusion times cannot be shortened below ~ 25 ms, which leads to comparable results between machines.
How are lesions identified on breast DWI?
In a multiparametric breast MRI protocol, lesion detection should be primarily based on evaluation of the contrast-enhanced sequences, which can be aided by inspection of the b = 800 s/mm2 images. On the b = 800 s/mm2 images, cancers are typically hyperintense given adequate baseline T2 signal. As the basic DWI sequence is T2-weighted, lesions with higher water content (e.g., cysts, myxoid fibroadenoma, highly proliferative cancer) will show a high signal on low b value images and may retain a (relatively) high signal on high b value images. Consequently, a high signal on b = 800 s/mm2 images may be due to a very high T2 signal (commonly referred to as T2 shine-through, although physically misleading) or true diffusion restriction with little signal decrease of a moderately high T2 signal. Cross-correlation of b = 800 s/mm2 image findings with quantitative ADC maps usually allows discrimination between these instances. Therefore, DWI analysis requires the evaluation of raw DWI data and ADC maps together.
On the other hand, tissues and lesions with a very low water content (e.g., fibrotic parenchyma, scars, some invasive lobular cancers) may demonstrate a very low signal at b = 0 s/mm2 (and thus, also on b = 800 s/mm2). Thus, these lesions will be difficult to visualize. In these lesions, the measurement of diffusion-dependent signal loss may not be possible, thus preventing an accurate assessment of diffusion level (Fig. 3). ADC in this condition (that can be referred to as signal blackout) will be low but will not reflect a true diffusion restriction (Fig. 3).
Can lesion location, size, and morphology be assessed on DWI images?
Assessment of location, size, and morphology of lesions is possible on DW images, even though this is limited by a spatial resolution that is inferior to that found in anatomical and contrast-enhanced sequences, and when available, these should be interpreted together. Morphologic assessment on breast DWI may be reported when it is discrepant with other sequences. Lesions can be categorized as foci, masses, or non-mass lesions. For masses, shape (round, oval, irregular) and internal signal pattern (homogeneous, heterogeneous, rim) can be reported while in non-mass lesions, distribution (focal, regional, linear, segmental) and internal signal pattern (homogeneous, heterogeneous) can be reported.
How should the ADC value be measured after DWI acquisition?
The pixel-wise parametric ADC maps enable quantitative assessment of diffusion restriction or hindrance, which should always be measured. The ADC value is obtained by drawing a region of interest (ROI) on the lesion on the ADC map (or the b = 800 s/mm2 image when the workstation allows propagation of the ROI to the ADC map). The ROI should fall completely within the lesion, contain at least 3 voxels and avoid both artifacts and necrotic or hemorrhagic parts of the lesion. This implies that lesions of 6 mm or larger in the axial plane are evaluable with DWI, albeit for small lesions, partial volume effects should be taken into account. While there was no consensus from our group on the size of the ROI to be used (whole lesion or focused ROI), literature suggests that selecting the lowest ADC value within the lesion (potentially reflecting the most active part of the lesion) might provide a more accurate discrimination between malignant and benign breast lesions. Due to the above described signal blackout, the ROI should fall within the enhancing part of the lesion (and hyperintense part on DW images) in order to avoid falsely low ADC values. Using a small ROI in the darkest part of the lesion on ADC maps is analogous to that used for analysis of DCE images. We thus suggest the use of a ROI placed on the darkest part of the ADC map, avoiding necrotic, noisy, or non-enhancing lesion voxels as the preferred method for measuring ADC values in order to reduce inter- and intra-reader variability and improve breast DWI consistency and comparability between sites. For such an ROI, the mean ADC value within the ROI should be reported, and it is suggested that the units should be in 10-3 mm2/s. Volumetric sampling of the whole lesion may be useful when the clinical indication is the evaluation of tumor response. In any case, the type of ROI (whole or focused) used for lesion assessment should be reported.
How is the ADC value interpreted?
Based on ADC measurements and lesion appearance, the proposed classification of diffusion level in lesions is very low, low, intermediate, high, and very high (Table 2 and Fig. 4). Figure 4 shows corresponding ADC ranges based on the most recent meta-analysis of studies evaluating DWI for the differentiation of benign and malignant lesions and considering only those studies that were in line with our consensus b value suggestions. The ADC values provided in Table 2 are solely intended to describe lesions in an objective way according to their diffusion level. Lesion classification should not be based upon diffusion level alone but should always be performed in conjunction with all anatomical and functional information available from all other imaging data.
Of the malignant lesions, invasive ductal and invasive lobular cancers, as well as DCIS with microinvasion, are usually associated with low to very low diffusion levels; pure ductal carcinoma in situ generally shows low or intermediate diffusion (16), whereas particularly invasive mucinous cancer may present with intermediate or even high diffusion levels; and triple-negative cancers with extensive necrosis may also yield high or very high diffusion levels in the necrotic part. Typical examples of benign and malignant lesions are presented in Fig. 5.
ADC measurements are prone to bias from adjacent noise regions (e.g., from voxels containing fat). These can be identified by unrealistically high (> 3 × 10-3 mm2/s), low (< 0.5 × 10-3 mm2/s) or even zero/negative ADC values. Unrealistic ADC values may require repositioning of the ROI.
Why are specific quantitative cut-off values not provided?
Two goals of the working group are (a) to promote more widespread integration of DWI into clinical use with standardized acquisitions, increasing the global evidence base by inclusion of DWI in future clinical trials and (b) to derive quantitative discrimination levels supported by standardized and reproducible acquisitions to be incorporated into clinical diagnostic or prognostic guidelines. A practical reality is that these goals are symbiotic and achieving them is an unavoidably iterative process. Thus, the analysis and interpretation scheme based on both qualitative and preliminary quantitative features (ADC value ranges) presented here are preliminary. It is important to realize that these suggestions are not fully vetted by the procedures of quality control, reproducibility, and multisite concordance that underpin formalized quantitative imaging biomarkers (QIB) and thus may evolve with time. This process will also be amplified by the incorporation of experimental optimization (eddy current distortion correction, gradient nonlinearity correction) into the consensus process.
Consequently, the ADC values in Table 2 may be adjusted by our group in the future once more rigorous standardization guidelines are in place, based upon intended multicenter investigation. Similarly, the minimal consensus recommendations of this document may be further refined to minimize variability from protocol variation.
Which steps are necessary for further implementation of quantitative breast DWI?
The recommendations outlined in this statement aim at improving inter-institutional protocol consistency, increasing homogeneity of reported data, and achieving a standardized assessment of the ADC values. The working group assigns high priority to establishing ADC as a QIB for broad diagnostic (lesion classification/aggressiveness) and prognostic (prediction of treatment response) use. Roadmaps for this process have been laid out by other consortia (e.g., Quantitative Imaging Biomarkers Alliance (QIBA, http://qibawiki.rsna.org/), Quantitative Imaging Network (QIN), and the International Society for Magnetic Resonance in Medicine Ad Hoc Committee on Standards for Quantitative MR). It is expected that the standardized acquisitions and evaluation suggested in this document will positively contribute to this process. Continuing efforts by (subgroups of) the working group to address the many other parameters that may influence acquisition and reporting of DWI will be guided by “metrology principles” of precision, repeatability, and reproducibility, tailored to breast imaging in particular.
As an example, physical phantoms are one key aspect of quality control. The ideal phantom mimics the target tissue, morphologically and, in terms of MRI parameters, provides reproducible results and is either easily moved or reproduced at multiple sites. An existing system for oncologic imaging is based on the ice-water phantom, which provides established ADC values and built-in temperature control. Other phantoms specifically designed for breast MRI contain synthetic material (e.g., alkanes) mimicking the MR properties (T1, T2, ADC) of fibroglandular, adipose, and tumor tissue and built-in housings compatible with breast RF coils. It has also been shown that high ADC phantoms such as solutions have utility in quantifying the difference between desired (nominal) and achieved (effective) b values. In addition to ADC standardization, these systems have the added benefit of allowing quality control of the effectively applied b values and of fat suppression, key elements of breast DWI. While these systems involve a higher level of investment, they may be appropriate for the future stages of inter-site standardization envisioned by our working group.
Another element of quality control is test-retest repeatability. Within breast cancer patients, obtaining this information requires considerable commitment given the practicalities of multiple examinations, especially since normal values may be dependent on the menstrual cycle, and likely requires evaluation in research protocols. However, when available, it allows quantitative distinction of biologic changes from measurement imprecision, which is vital for the use of imaging biomarkers in clinical trials. Thus, research efforts are encouraged that collect repeatability data in patients wherever possible to build the evidence base.
What is the clinical potential of advanced DWI techniques?
While the basic monoexponential diffusion model and resulting ADC values providing a simple and technically reproducible parameter are currently preferable in clinical practice, the working group explicitly acknowledges continuous developments in the field of DWI that hold promise and should be pursued in parallel with ADC standardization. For example, the heterogeneity of ADC values within lesions may be measured using histogram analysis and more advanced artificial intelligence techniques. Furthermore, DWI techniques “beyond the ADC”, such as diffusion tensor imaging (DTI), which allows the analysis of diffusion anisotropy; intravoxel incoherent motion (IVIM) imaging, which distinguishes between intravascular perfusion and extravascular microstructural diffusion components; and non-Gaussian diffusion, which provides enhanced sensitivity to tissue complexity, for instance from the kurtosis model, may enhance the value of DWI. Hybrid combinations of these models have also been tested. However, such advanced methods will add constraints to the DWI acquisition protocol (multiple diffusion-encoding directions, multiple b values in the low and high range, etc.) resulting in longer acquisition times and the need for more sophisticated image processing tools. Currently, there is no reliable evidence regarding the clinical value superiority of advanced DWI techniques over standard ADC assessment. Still, DWI users are explicitly encouraged to investigate more specific (or “tailored”) strategies aimed at screening (with/without the concurrent use of contrast-enhanced sequences), lesion characterization and staging, treatment monitoring, and prognosis.
Conclusion
This statement details the first consensus on breast DWI created by the EUSOBI International Breast DWI working group. The working group considers breast DWI to be an essential part of a multiparametric breast MRI protocol. Basic requirements for routine clinical application of breast DWI are proposed, including recommendations on b values, fat saturation, spatial resolution, TR/TE, and considerations for ROI placement. The working group will focus our future efforts on expanding the technical recommendations of DWI protocols and the development of methods for quality control. Finally, the working group explicitly encourages research into more sophisticated advanced acquisition, modeling, and analysis approaches to further exploit the diagnostic and prognostic potential of diffusion-based breast imaging.
References 94
- 1.
Woodhams R, Matsunaga K, Iwabuchi K, et al. Diffusion-weighted imaging of malignant breast tumors: the usefulness of apparent diffusion coefficient (ADC) value and ADC map for the detection of malignant breast tumors and evaluation of cancer extension. J Comput Assist Tomogr. 2005;29:644-649.
Not in our library10.1097/01.rct.0000171913.74086.1b
- 2.
Woodhams R, Matsunaga K, Kan S, et al. ADC mapping of benign and malignant breast tumors. Magn Reson Med Sci. 2005;4:35-42.
Not in our library10.2463/mrms.4.35
- 3.
Rubesova E, Grell AS, De Maertelaer V, Metens T, Chao SL, Lemort M (2006) Quantitative diffusion imaging in breast cancer: a clinical prospective study. J Magn Reson Imaging 24:319–324. 10.1002/jmri.20643
Not in our library10.1002/jmri.20643
- 4.
Wenkel E, Geppert C, Schulz-Wendtland R, et al. Diffusion weighted imaging in breast MRI: comparison of two different pulse sequences. Acad Radiol. 2007;14:1077-1083.
Not in our library10.1016/j.acra.2007.06.006
- 5.
Baltzer PAT, Renz DM, Herrmann K-H, et al. Diffusion-weighted imaging (DWI) in MR mammography (MRM): clinical comparison of echo planar imaging (EPI) and half-Fourier single-shot turbo spin echo (HASTE) diffusion techniques. Eur Radiol. 2009;19:1612-1620.
Not in our library10.1007/s00330-009-1326-5
- 6.
Partridge SC, DeMartini WB, Kurland BF, Eby PR, White SW, Lehman CD (2009) Quantitative diffusion-weighted imaging as an adjunct to conventional breast MRI for improved positive predictive value. AJR Am J Roentgenol 193:1716–1722. 10.2214/AJR.08.2139
Not in our library10.2214/ajr.08.2139
- 7.
Pinker K, Bickel H, Helbich TH, et al. Combined contrast-enhanced magnetic resonance and diffusion-weighted imaging reading adapted to the “Breast Imaging Reporting and Data System” for multiparametric 3-T imaging of breast lesions. Eur Radiol. 2013;23:1791-1802.
Not in our library10.1007/s00330-013-2771-8
- 8.
- 9.
Dorrius MD, Dijkstra H, Oudkerk M, Sijens PE. Effect of b value and pre-admission of contrast on diagnostic accuracy of 1.5-T breast DWI: a systematic review and meta-analysis. Eur Radiol. 2014;24:2835-2847.
Not in our library10.1007/s00330-014-3338-z
- 10.
Shi R, Yao Q, Wu L, Xu J. Breast lesions: diagnosis using diffusion weighted imaging at 1.5 T and 3.0 T—systematic review and meta-analysis. Clin Breast Cancer. 2018;18:e305-e320.
Not in our library10.1016/j.clbc.2017.06.011
- 11.
Iima M, Kataoka M, Kanao S, et al. Intravoxel incoherent motion and quantitative non-Gaussian diffusion MR imaging: evaluation of the diagnostic and prognostic value of several markers of malignant and benign breast lesions. Radiology. 2018;287:432-441.
Not in our library10.1148/radiol.2017162853
- 12.
Iima M, Yano K, Kataoka M, et al. Quantitative non-Gaussian diffusion and intravoxel incoherent motion magnetic resonance imaging: differentiation of malignant and benign breast lesions. Invest Radiol. 2015;50:205-211.
Not in our library10.1097/rli.0000000000000094
- 13.
Rahbar H, Zhang Z, Chenevert TL, et al. Utility of diffusion-weighted imaging to decrease unnecessary biopsies prompted by breast MRI: a trial of the ECOG-ACRIN cancer research group (A6702). Clin Cancer Res. 2019;25:1756-1765.
Not in our library10.1158/1078-0432.ccr-18-2967
- 14.
Partridge SC, Mullins CD, Kurland BF, et al. Apparent diffusion coefficient values for discriminating benign and malignant breast MRI lesions: effects of lesion type and size. AJR Am J Roentgenol. 2010;194:1664-1673.
Not in our library10.2214/ajr.09.3534
- 15.
Rahbar H, Partridge SC, Eby PR, et al. Characterization of ductal carcinoma in situ on diffusion weighted breast MRI. Eur Radiol. 2011;21:2011-2019.
Not in our library10.1007/s00330-011-2140-4
- 16.
Bickel H, Pinker-Domenig K, Bogner W, et al. Quantitative apparent diffusion coefficient as a noninvasive imaging biomarker for the differentiation of invasive breast cancer and ductal carcinoma in situ. Invest Radiol. 2015;50:95-100.
Not in our library10.1097/rli.0000000000000104
- 17.
Iima M, Le Bihan D, Okumura R, et al. Apparent diffusion coefficient as an MR imaging biomarker of low-risk ductal carcinoma in situ: a pilot study. Radiology. 2011;260:364-372.
Not in our library10.1148/radiol.11101892
- 18.
Ding J-R, Wang D-N, Pan J-L. Apparent diffusion coefficient value of diffusion-weighted imaging for differential diagnosis of ductal carcinoma in situ and infiltrating ductal carcinoma. J Cancer Res Ther. 2016;12:744-750.
Not in our library10.4103/0973-1482.154093
- 19.
Pickles MD, Gibbs P, Lowry M, Turnbull LW. Diffusion changes precede size reduction in neoadjuvant treatment of breast cancer. Magn Reson Imaging. 2006;24:843-847.
Not in our library10.1016/j.mri.2005.11.005
- 20.
Richard R, Thomassin I, Chapellier M, et al. Diffusion-weighted MRI in pretreatment prediction of response to neoadjuvant chemotherapy in patients with breast cancer. Eur Radiol. 2013;23:2420-2431.
Not in our library10.1007/s00330-013-2850-x
- 21.
Galbán CJ, Ma B, Malyarenko D, et al. Multi-site clinical evaluation of DW-MRI as a treatment response metric for breast cancer patients undergoing neoadjuvant chemotherapy. PLoS One. 2015;10:e0122151.
Not in our library10.1371/journal.pone.0122151
- 22.
Li X, Abramson RG, Arlinghaus LR, et al. Multiparametric magnetic resonance imaging for predicting pathological response after the first cycle of neoadjuvant chemotherapy in breast cancer. Invest Radiol. 2015;50:195-204.
Not in our library10.1097/rli.0000000000000100
- 23.
Leong KM, Lau P, Ramadan S. Utilisation of MR spectroscopy and diffusion weighted imaging in predicting and monitoring of breast cancer response to chemotherapy. J Med Imaging Radiat Oncol. 2015;59:268-277.
Not in our library10.1111/1754-9485.12310
- 24.
Partridge SC, Zhang Z, Newitt DC, et al. Diffusion-weighted MRI findings predict pathologic response in neoadjuvant treatment of breast cancer: the ACRIN 6698 multicenter trial. Radiology. 2018;289:618-627.
Not in our library10.1148/radiol.2018180273
- 25.
Newitt DC, Zhang Z, Gibbs JE, et al. Test-retest repeatability and reproducibility of ADC measures by breast DWI: results from the ACRIN 6698 trial. J Magn Reson Imaging. 2019;49:1617-1628.
Not in our library10.1002/jmri.26539
- 26.
Le Bihan D. Apparent diffusion coefficient and beyond: what diffusion MR imaging can tell us about tissue structure. Radiology. 2013;268:318-322.
Not in our library10.1148/radiol.13130420
- 27.
D’Orsi CJ, Sickles EA, Mendelson EB, Morris EA (2013) ACR BI-RADS® Atlas, Breast Imaging Reporting and Data System. Reston, VA, American College of Radiology.
Not in our library
- 28.
Weinreb JC, Barentsz JO, Choyke PL, et al. PI-RADS Prostate Imaging - Reporting and Data System: 2015, Version 2. Eur Urol. 2016;69:16-40.
Not in our library10.1016/j.eururo.2015.08.052
- 29.
Bickel H, Pinker K, Polanec S, et al. Diffusion-weighted imaging of breast lesions: Region-of-interest placement and different ADC parameters influence apparent diffusion coefficient values. Eur Radiol. 2017;27:1883-1892.
Not in our library10.1007/s00330-016-4564-3
- 30.
Giannotti E, Waugh S, Priba L, Davisa Z, Crowe E, Vinnicombe S (2015) Assessment and quantification of sources of variability in breast apparent diffusion coefficient (ADC) measurements at diffusion weighted imaging. Eur J Radiol 84:1729–1736. 10.1016/j.ejrad.2015.05.032
Not in our library10.1016/j.ejrad.2015.05.032
- 31.
O’Flynn EAM, Morgan VA, Giles SL, deSouza NM. Diffusion weighted imaging of the normal breast: reproducibility of apparent diffusion coefficient measurements and variation with menstrual cycle and menopausal status. Eur Radiol. 2012;22:1512-1518.
Not in our library10.1007/s00330-012-2399-0
- 32.
Aliu SO, Jones EF, Azziz A, et al. Repeatability of quantitative MRI measurements in normal breast tissue. Transl Oncol. 2014;7:130-137.
Not in our library10.1593/tlo.13841
- 33.
Spick C, Bickel H, Pinker K, et al. Diffusion-weighted MRI of breast lesions: a prospective clinical investigation of the quantitative imaging biomarker characteristics of reproducibility, repeatability, and diagnostic accuracy. NMR Biomed. 2016;29:1445-1453.
Not in our library10.1002/nbm.3596
- 34.
Spick C, Pinker-Domenig K, Rudas M, Helbich TH, Baltzer PA (2014) MRI-only lesions: application of diffusion-weighted imaging obviates unnecessary MR-guided breast biopsies. Eur Radiol 24:1204–1210. 10.1007/s00330-014-3153-6
Not in our library10.1007/s00330-014-3153-6
- 35.
Baltzer A, Dietzel M, Kaiser CG, Baltzer PA. Combined reading of contrast enhanced and diffusion weighted magnetic resonance imaging by using a simple sum score. Eur Radiol. 2016;26:884-891.
Not in our library10.1007/s00330-015-3886-x
- 36.
Kessler LG, Barnhart HX, Buckler AJ, et al. The emerging science of quantitative imaging biomarkers terminology and definitions for scientific studies and regulatory submissions. Stat Methods Med Res. 2015;24:9-26.
Not in our library10.1177/0962280214537333
- 37.
Winfield JM, Payne GS, Weller A, deSouza NM. DCE-MRI, DW-MRI, and MRS in cancer: challenges and advantages of implementing qualitative and quantitative multi-parametric imaging in the clinic. Top Magn Reson Imaging. 2016;25:245-254.
Not in our library10.1097/rmr.0000000000000103
- 38.
Winfield JM, Tunariu N, Rata M, et al. Extracranial soft-tissue tumors: repeatability of apparent diffusion coefficient estimates from diffusion-weighted MR imaging. Radiology. 2017;284:88-99.
Not in our library10.1148/radiol.2017161965
- 39.
Barnes A, Alonzi R, Blackledge M, et al. UK quantitative WB-DWI technical workgroup: consensus meeting recommendations on optimisation, quality control, processing and analysis of quantitative whole-body diffusion-weighted imaging for cancer. Br J Radiol. 2018;91:20170577.
Not in our library10.1259/bjr.20170577
- 40.
Zhang L, Tang M, Min Z, Lu J, Lei X, Zhang X (2016) Accuracy of combined dynamic contrast-enhanced magnetic resonance imaging and diffusion-weighted imaging for breast cancer detection: a meta-analysis. Acta Radiol 57:651–660. 10.1177/0284185115597265
Not in our library10.1177/0284185115597265
Cited by papers we hold 2
- Current State of Breast Cancer Diagnosis, Treatment, and Theranostics · 2021No notice
- Diffusion-Weighted Imaging in Oncology: An Update · 2020No notice
Reads
Read by people on this page
None yet.
Checked or read by AI through naulon
None yet.
A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.
Are you an author?
Is this your paper?
Claim it to add your note and be paid when agents cite it. Free.