Optimization and validation of methods for mapping of the radiofrequency transmit field at 3T

Antoine Lutti, Chloe Hutton, Jürgen Finsterbusch and 2 more

Magnetic Resonance in Medicine · 2010

doi:10.1002/mrm.22421

Abstract

MRI techniques such as quantitative imaging and parallel transmit require precise knowledge of the radio‐frequency transmit field (B ). Three published methods were optimized for robust B mapping at 3T in the human brain: three‐dimensional (3D) actual flip angle imaging (AFI), 3D echo‐planar imaging (EPI), and two‐dimensional (2D) stimulated echo acquisition mode (STEAM). We performed a comprehensive comparison of the methods, focusing on artifacts, reproducibility, and accuracy compared to a reference 2D double angle method. For the 3D AFI method, the addition of flow‐compensated gradients for diffusion damping reduced the level of physiological artifacts and improved spoiling of transverse coherences. Correction of susceptibility‐induced artifacts alleviated image distortions and improved the accuracy of the 3D EPI imaging method. For the 2D STEAM method, averaging over multiple acquisitions reduced the impact of physiological noise and a new calibration method enhanced the accuracy of the B maps. After optimization, all methods yielded low noise B maps (below 2 percentage units), of the nominal flip angle value (p.u.) with a systematic bias less than 5 p.u. units. Full brain coverage was obtained in less than 5 min. The 3D AFI method required minimal postprocessing and showed little sensitivity to off‐resonance and physiological effects. The 3D EPI method showed the highest level of reproducibility. The 2D STEAM method was the most time‐efficient technique. Magn Reson Med, 2010. © 2010 Wiley‐Liss, Inc.

Shown for reading. Its publisher allows quoting it, not republishing it.

References 40

  1. 1.

    Schepers J, van Osch MJ, Bartels LW, Heukels SN, Viergever MA, Nicolay K. The effect of B1 field inhomogeneity and the nonselective inversion profile on the kinetics of FAIR-based perfusion MRI. Magn Reson Med. 2005;53:1355-1362.

    Not in our library10.1002/mrm.20480

  2. 2.

    Samson RS, Wheeler-Kingshott CA, Symms MR, Tozer DJ, Tofts PS. A simple correction for B1 field errors in magnetization transfer ratio measurements. Magn Reson Imaging. 2006;24:255-263.

    Not in our library10.1016/j.mri.2005.10.025

  3. 3.

    Venkatesan R, Lin WL, Haacke EM. Accurate determination of spin-density and T-1 in the presence of RF-field inhomogeneities and flip-angle miscalibration. Magn Reson Med. 1998;40:592-602.

    Not in our library10.1002/mrm.1910400412

  4. 4.

    Wang D, Heberlein K, LaConte S, Hu X. Inherent insensitivity to RF inhomogeneity in FLASH imaging. Magn Reson Med. 2004;52:927-931.

    Not in our library10.1002/mrm.20217

  5. 5.

    Deichmann R, Good CD, Turner R. RF inhomogeneity compensation in structural brain imaging. Magn Reson Med. 2002;47:398-402.

    Not in our library10.1002/mrm.10050

  6. 6.

    Thomas DL, De Vita E, Deichmann R, Turner R, Ordidge RJ. 3D MDEFT imaging of the human brain at 4.7 T with reduced sensitivity to radiofrequency inhomogeneity. Magn Reson Med. 2005;53:1452-1458.

    Not in our library10.1002/mrm.20482

  7. 7.

    Vovk U, Pernus F, Likar B. A review of methods for correction of intensity inhomogeneity in MRI. IEEE Trans Med Imaging. 2007;26:405-421.

    Not in our library10.1109/tmi.2006.891486

  8. 8.

    Collins CM, Li SZ, Smith MB. SAR and B-1 field distributions in a heterogeneous human head model within a birdcage coil. Magn Reson Med. 1998;40:847-856.

    Not in our library10.1002/mrm.1910400610

  9. 9.

    Van den Berg CA, van den BB, Van de Kamer JB, Raaymakers BW, Kroeze H, Bartels LW, Lagendijk JJ. Simultaneous B1 + homogenization and specific absorption rate hotspot suppression using a magnetic resonance phased array transmit coil. Magn Reson Med. 2007;57:577-586.

    Not in our library10.1002/mrm.21149

  10. 10.

    Zhang Z, Yip CY, Grissom W, Noll DC, Boada FE, Stenger VA. Reduction of transmitter B1 inhomogeneity with transmit SENSE slice-select pulses. Magn Reson Med. 2007;57:842-847.

    Not in our library10.1002/mrm.21221

  11. 11.

    Helms G, Dathe H, Dechent P. Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equation. Magn Reson Med. 2008;59:667-672.

    Not in our library10.1002/mrm.21542

  12. 12.

    Insko EK, Bolinger L. Mapping of the radiofrequency field. J Magn Reson A. 1993;103:82-85.

    Not in our library

  13. 13.

    Wang JH, Qiu ML, Constable RT. In vivo method for correcting transmit/receive nonuniformities with phased array coils. Magn Reson Med. 2005;53:666-674.

    Not in our library10.1002/mrm.20377

  14. 14.

    Stollberger R, Wach P. Imaging of the active B1 field in vivo. Magn Reson Med. 1996;35:246-251.

    Not in our library10.1002/mrm.1910350217

  15. 15.

    Cunningham CH, Pauly JM, Nayak KS. Saturated double-angle method for rapid mapping. Magn Reson Med. 2006;55:1326-1333.

    Not in our library10.1002/mrm.20896

  16. 16.

    Hoult DI. Solution of the Bloch equations in the presence of a varying B1 field: approach to selective pulse analysis. J Magn Reson. 1979;35:69-86.

    Not in our library

  17. 17.

    Wang JH, Mao WH, Qiu ML, Smith MB, Constable RT. Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities. Magn Reson Med. 2006;56:463-468.

    Not in our library10.1002/mrm.20947

  18. 18.

    Helms G, Finsterbusch J, Weiskopf N, Dechent P. Rapid radiofrequency field mapping in vivo using single-shot STEAM MRI. Magn Reson Med. 2008;60:739-743.

    Not in our library10.1002/mrm.21676

  19. 19.

    Haacke EM, Brown RW, Thompson MR, Venkatesan R. Sequence design, artifacts and nomenclature: magnetic resonance imaging. 1999;816.

    Not in our library

  20. 20.

    Duyn JH, Yang Y, Frank JA, Mattay VS, Hou L. Functional magnetic resonance neuroimaging data acquisition techniques. Neuroimage. 1996;4(3 pt 3):S76-S83.

    Not in our library10.1006/nimg.1996.0057

  21. 21.

    Metzger GJ, Snyder C, Akgun C, Vaughan T, Ugurbil K, Van De Moortele PF. Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject-dependent transmit phase measurements. Magn Reson Med. 2008;59:396-409.

    Not in our library10.1002/mrm.21476

  22. 22.

    Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med. 2007;57:192-200.

    Not in our library10.1002/mrm.21120

  23. 23.

    Jiru F, Klose U. Fast 3D radiofrequency field mapping using echo-planar imaging. Magn Reson Med. 2006;56:1375-1379.

    Not in our library10.1002/mrm.21083

  24. 24.

    Sled JG, Pike GB. Correction for B1 and B0 variations in quantitative T2 measurements using MRI. Magn Reson Med. 2000;43:589-593.

    Not in our library10.1002/(sici)1522-2594(200004)43:4<589::aid-mrm14>3.0.co;2-2

  25. 25.

    Nehrke K. On the steady-state properties of actual flip angle imaging (AFI). Magn Reson Med. 2009;61:84-92.

    Not in our library10.1002/mrm.21592

  26. 26.

    Yarnykh VL. 2008;329.

    Not in our library

  27. 27.

    Akoka S, Franconi F, Seguin F, Lepape A. Radiofrequency map of an NMR coil by imaging. Magn Reson Imaging. 1993;11:437-441.

    Not in our library10.1016/0730-725x(93)90078-r

  28. 28.

    Hutton C, Bork A, Josephs O, Deichmann R, Ashburner J, Turner R. Image distortion correction in fMRI: a quantitative evaluation. Neuroimage. 2002;16:217-240.

    Not in our library10.1006/nimg.2001.1054

  29. 29.

    Preibisch C, Deichmann R. Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles. Magn Reson Med. 2009;61:125-135.

    Not in our library10.1002/mrm.21776

  30. 30.

    Bernstein MA, King KF, Zhou XJ. Handbook of MRI pulse sequences. Gradient moment nulling. 2004;331-349.

    Not in our library

  31. 31.

    Le Bihan D. Looking into the functional architecture of the brain with diffusion MRI. Nat Rev Neurosci. 2003;4:469-480.

    Not in our library10.1038/nrn1119

  32. 32.

    Weiskopf N, Hutton C, Josephs O, Deichmann R. Optimal EPI parameters for reduction of susceptibility-induced BOLD sensitivity losses: a whole-brain analysis at 3 T and 1.5 T. Neuroimage. 2006;33:493-504.

    Not in our library10.1016/j.neuroimage.2006.07.029

  33. 33.

    Ashburner J, Friston K. Multimodal image coregistration and partitioning—a unified framework. Neuroimage. 1997;6:209-217.

    Not in our library10.1006/nimg.1997.0290

  34. 34.

    Sled JG, Pike GB. Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRI. IEEE Trans Med Imaging. 1998;17:653-662.

    Not in our library10.1109/42.730409

  35. 35.

    Helms G, Dathe H, Kallenberg K, Dechent P. High-resolution maps of magnetization transfer with inherent correction for RF inhomogeneity and T1 relaxation obtained from 3D FLASH MRI. Magn Reson Med. 2008;60:1396-1407.

    Not in our library10.1002/mrm.21732

  36. 36.

    Ding S, Trillaud H, Yongbi M, Rolett EL, Weaver JB, Dunn JF. High resolution renal diffusion imaging using a modified steady-state free precession sequence. Magn Reson Med. 1995;34:586-595.

    Not in our library10.1002/mrm.1910340415

  37. 37.

    Helms G. A precise and user-independent quantification technique for regional comparison of single volume proton MR spectroscopy of the human brain. NMR Biomed. 2000;13:398-406.

    Not in our library10.1002/1099-1492(200011)13:7<398::aid-nbm660>3.0.co;2-w

  38. 38.

    Price D, Delakis I, Renaud C, Dickinson R. 3T MRI systems: comparative report on four MRI systems. 2007.

    Not in our library

  39. 39.

    Deichmann R, Schwarzbauer C, Turner R. Optimisation of the 3D MDEFT sequence for anatomical brain imaging: technical implications at 1.5 and 3 T. Neuroimage. 2004;21:757-767.

    Not in our library10.1016/j.neuroimage.2003.09.062

  40. 40.

    Howarth C, Hutton C, Deichmann R. Improvement of the image quality of T1-weighted anatomical brain scans. Neuroimage. 2006;29:930-937.

    Not in our library10.1016/j.neuroimage.2005.08.004

Cited by papers we hold 9

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.