Calibration of optical tweezers with positional detection in the back-focal-plane

Simon F. Tolic-Norrelykke, Erik Schaeffer, Jonathon Howard and 3 more

Rev. Sci. Instrum. 77, 103101 (2006) (11 pages) · 2006

We email you when a new notice or a question about this paper appears.

Abstract

We explain and demonstrate a new method of force- and position-calibration for optical tweezers with back-focal-plane photo detection. The method combines power spectral measurements of thermal motion and the response to a sinusoidal motion of a translation stage. It consequently does not use the drag coefficient of the trapped ob ject as an input. Thus, neither the viscosity, nor the size of the trapped ob ject, nor its distance to nearby surfaces need to be known. The method requires only a low level of instrumentation and can be applied in situ in all spatial dimensions. It is both accurate and precise: true values are returned, with small error-bars. We tested this experimentally, near and far from surfaces. Both position- and force-calibration were accurate to within 3%. To calibrate, we moved the sample with a piezo-electric translation stage, but the laser beam could be moved instead, e.g. by acousto-optic deflectors. Near surfaces, this precision requires an improved formula for the hydrodynamical interaction between an infinite plane and a micro-sphere in non-constant motion parallel to it. We give such a formula.

Abstract licence: CC0

Reads

Read by people on this page

None yet.

Checked or read by AI through naulon

None yet.

A person counts once a day per paper. We keep no cookie and no address: a reader is a hash that changes every day. Any month under five reads shows no number.

Are you an author?

Is this your paper?

Claim it to add your note and be paid when agents cite it. Free.